K181767 · Siemens Medical Solutions USA, Inc. · OWB · Aug 17, 2018 · Radiology
Device Facts
Record ID
K181767
Device Name
Cios Select
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OWB · Radiology
Decision Date
Aug 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1650
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Cios Select is a mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. The Cios Select can operate in three different modes, Digital Radiography, Fluoroscopy, and Pulsed Fluoroscopy, which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of a intra-medullary nail implants in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes.
Device Story
Mobile fluoroscopy C-arm system; consists of mobile acquisition unit (X-ray generator, tube, flat panel detector) and monitor cart (display, image processing, storage). Inputs: X-ray radiation; transforms via PaxScan 2121DXV detector into digital images. Used in OR, trauma, ICU, etc., by healthcare professionals. Output: real-time fluoroscopic or radiographic images on TFT displays. Supports clinical decision-making for surgical guidance, implant positioning, and interventional procedures. Benefits: enables intraoperative visualization, reducing need for invasive exploration; supports low-dose pediatric imaging.
Clinical Evidence
Bench testing only. Includes unit, subsystem, and system integration testing; software verification and regression testing; electrical safety (IEC 60601-1) and EMC (IEC 60601-1-2) testing; and imaging performance evaluation per SSXI guidance. Human factors usability validation performed. No clinical data presented.
Technological Characteristics
Mobile C-arm X-ray system. Detector: PaxScan 2121DXV flat panel. Software: VA20, Windows 10 OS. Connectivity: Wired/wireless DICOM, optional WLAN. User Interface: Touchscreen panel. Energy: High-frequency generator. Standards: IEC 60601-1, 60601-1-2, 60601-1-3, 60601-2-28, 60601-2-43, 60601-2-54, ISO 14971, IEC 62304.
Indications for Use
Indicated for use in operating rooms, traumatology, endoscopy, intensive care, pediatrics, ambulatory care, and veterinary medicine. Supports digital radiography, fluoroscopy, and pulsed fluoroscopy for procedures including bile duct display, intra-medullary nail positioning, pediatric low-dose imaging, pain therapy, and catheter/probe placement.
Regulatory Classification
Identification
An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
August 17, 2018
Siemens Medical Solutions USA, Inc. % Ms. Patricia Jones Senior Regulatory Affairs Technical Specialist 40 Liberty Boulevard, 65-1A MALVERN PA 19355
Re: K181767
Trade/Device Name: Cios Select Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO Dated: July 02, 2018 Received: July 03, 2018
Dear Ms. Jones:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Hol. 2. Nils
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the text is a graphic of orange dots.
| | DEPARTMENT OF HEALTH AND HUMAN SERVICES | Form Approved: OMB No. 0910-0120 |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------|
| | Food and Drug Administration | Expiration Date: 06/30/2020 |
| | Indications for Use | See PRA Statement below. |
| 510(k) Number (if known) | K181767 | |
| Device Name | Cios Select | |
| Indications for Use (Describe) | The Cios Select is a mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. | |
| | The Cios Select can operate in three different modes, Digital Radiography, Fluoroscopy, and Pulsed Fluoroscopy, which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of a intra-medullary nail implants in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes. | |
| Type of Use (Select one or both, as applicable) | | |
| | <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> | |
| | CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. | |
| | <b>*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*</b> | |
| | The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: | |
| | Department of Health and Human Services | |
| | Food and Drug Administration | |
| | Office of Chief Information Officer | |
| | Paperwork Reduction Act (PRA) Staff | |
| | PRAStaff@fda.hhs.gov | |
| | "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." | |
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange. To the right of the words is a cluster of orange dots of varying sizes.
# Section 7 510(k) Summary
The 510(k) Summary is provided on the next page and is suitable for publication on the FDA website.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logos for Siemens and Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots.
# 510(k) Summary: Cios Select
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355
Date Prepared: August 16, 2018
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
#### General Information: 1.
### lmporter / Distributor:
Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Reqistration Number: 2. 2240869
### Manufacturing Site:
Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Establishment Registration Number: 3003202425
#### 2. Contact Person:
Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 448-6474 Email: patricia.d.jones@siemens-Healthineers.com
### 3. Device Name and Classification:
| Trade Name: | Cios Select |
|--------------------------|---------------------------------------------|
| Classification Name: | Image-intensified fluoroscopic x-ray system |
| Classification Panel: | Radiology |
| CFR Section: | 21 CFR §892.1650 |
| Submission Type: | Traditional 510(k) |
| Device Class: | Class II |
| Product Code: | OWB |
| Subsequent Product Code: | OXO |
{5}------------------------------------------------
# SIEMENS Healthineers
- 4. Legally Marketed Predicate Device: Primarv Predicate: Cios Select Trade Name: K153232 510(k) #: Clearance Date: February 10, 2016 Classification Name: Image-intensified fluoroscopic x-ray system Classification Panel: Radiology Classification Regulation: 21 CFR §892.1650 Device Class: Class II Product Code: OWB Subsequent Product Code: JAA, OXO Recall Information: No Recalls Secondary Predicate: Trade Name: Cios Fusion 510(k) #: K153244 March 07, 2016 Clearance Date: Classification Name: Image-intensified fluoroscopic x-ray svstem Classification Panel: Radiology Classification Regulation: 21 CFR §892.1650 Device Class: Class II Product Code: OWB Subsequent Product Code: ОХО No Recalls Recall Information:
# 5. Device Description:
The Siemens Healthineers Cios Select mobile fluoroscopy C-arm system is an X-ray imaging system consisting of two mobile units: a mobile acquisition unit and a monitor cart as the image display station. The mobile acquisition unit is comprised of the X-ray control, the C-arm which supports the single-tank high-frequency generator/X-ray tube assembly, the flat panel detector, and user controls. The monitor cart connects to the acquisition unit by a cable. It integrates the TFT flat panel displays, Digital Imaging Processing System, user controls and image storage devices (DVD, USB).
The new detector PaxScan 2121DXV is a new component to the subject device Cios Select. This is a hardware change that impacts the system software VA20. The essential performance of the detector is to generate (a sequence of) digital images in which the digital pixel values are representative of the intensity of the spatially distributed X-Ray radiation at the position of the detector. All the non-clinical performance data demonstrate that the detector PaxScan 2121DXV for Cios Select VA20 does not raise any new issues of safety or effectiveness and is as safe, and effective when compared to the secondary predicate devices Cios Fusion (K153244) that is currently marketed for the detector technology.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logos for Siemens and Healthineers. The Siemens logo is in teal, and the Healthineers logo is in orange. To the right of the Healthineers logo is a graphic of orange dots.
Interfaces are provided for optional devices such as external monitors, thermal printers, as well as wired and wireless DICOM network interfaces.
The following modifications are incorporated to create the Subject Device the Cios Select (VA20):
- 1. Updated the Indications for Use Statement with minor changes
- 2. Upgraded software, from Version VA10 to Version VA20:
- a. New User Interface touch panel (software)
- b. New flat detector (PaxScan 2121DXV) (software)
- (OTS) software from Windows 7 to Windows 10. (software) C.
- d. Updated Image storage to a maximum of 300000 frames (software)
- e. Optional Metal Artifact Reduction Software (cleared in Secondary Predicate Cios Fusion K153244)
- 3. New user interface in-touch panel for navigating and operating the system. (Hardware)
- 4. New optional wireless LAN for network connectivity (Hardware)
- 5. New optional wireless footswitch in addition to the wired footswitch. (Hardware)
- 6. New flat detector (PaxScan 2121DXV) to replace image intensifier for image receptor.(Hardware)
# 6. Indication for Use:
The Cios Select is a mobile X-ray system intended for use in Operating room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. The Cios Select can operate in three different modes, Digital Radiography, Fluoroscopy, and Pulsed Fluoroscopy which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of an intra-medullary nail implants in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes.
# 7. Substantial Equivalence:
The Cios Select (VA20) is substantially equivalent to the commercially available primary predicate device Siemens Cios Select (VA10), cleared February 10, 2016, with K153232.
The Indications for Use was slightly modified, however, the change does not impact the Intended Use of the device. Table 1 provides primary and secondary Predicate Devices comparable information:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots that are arranged in a pattern.
# The Cios Select is substantially equivalent to the following predicate devices:
## Table 1: Predicate Device Comparable Properties:
| Predicate Device(s) Name and<br>Manufacturer | 510(k)<br>Number | Clearance<br>Date | Comparable Properties |
|--------------------------------------------------------------------------------------------------------------------------------|------------------|-------------------|--------------------------------------------------------------------------------------------------|
| Primary Predicate Device<br>Cios Select (VA10)<br>Product Code: OWB, OXO,<br>JAA<br>Siemens Shanghai Medical<br>Equipment Ltd. | K153232 | 2/10/2016 | ●<br>ndications for use<br>●<br>-ray technology<br>●<br>mage processing<br>●<br>echanical design |
| Secondary Predicate Device<br>Cios Fusion (VA20)<br>Product Code:<br>OWB, OXO<br>Siemens Healthcare GmbH | K153244 | 3/07/2016 | ●<br>etector Technology |
# 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device:
The Subject device Cios Select design is based on the Siemens Cios Select(K153232) including the system control, mechanical design and image processing functions. The flat panel detector technology is based on the Cios Fusion (K153244).
The subject device has the same fundamental scientific technologies as the
predicate devices. The modifications do not affect the intended use of the device. Table 2 has Subject Device characteristics compared to the Predicate Device.
| Modifications | Subject Device<br>Cios Select<br>(VA20) | Primary<br>Predicate Device<br>Cios Select<br>(VA10) K153232 | Secondary<br>Predicate Device<br>Cios Fusion<br>(VA20) K153244 | Comparison Results |
|---------------------------|----------------------------------------------------------------------------------|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| System<br>Software | Upgraded<br>software, from<br>Version VA10 to<br>Version VA20. | Software Version<br>VA10 | N/A | System Software<br>upgraded to VA20 to<br>support new software /<br>hardware modifications<br>and tested per software<br>and guidance and SSXI<br>Guidance. All test<br>results passed. |
| User<br>Interface | New user<br>interface in a<br>touch panel for<br>navigating and<br>operating the | Keyboard button<br>based User<br>interface | N/A | The functionality of the<br>device remains the<br>same as the predicate<br>device except for the<br>new User Interface with |
| Modifications | Subject Device<br>Cios Select<br>(VA20) | Primary<br>Predicate Device<br>Cios Select<br>(VA10) K153232 | Secondary<br>Predicate Device<br>Cios Fusion<br>(VA20) K153244 | Comparison Results |
| | system. | | | Touchscreen panel. |
| WLAN | New optional<br>wireless LAN for<br>network<br>connectivity. | N/A | N/A | The WLAN network<br>connectivity feature is in<br>compliance with the<br>Wireless Guidance<br>document. |
| Footswitch | Wired<br>Optional<br>Wireless<br>footswitch | Wired Only | N/A | Optional wireless<br>footswitch with the same<br>functionality as a wired<br>footswitch. Wireless<br>version does not raise<br>any new safety or<br>effectiveness issue.<br>This feature is in<br>compliance with the<br>Wireless Guidance<br>document. |
| Image<br>Receptor | New flat<br>detector<br>PaxScan<br>(2121DXV) | Image Intensifier | Detector<br>PaxScanTM 2020X<br>and PaxScanTM<br>3030X Detector<br>was cleared in the<br>predicate | Detector technology is<br>comparable to<br>secondary predicate<br>device Cios Fusion<br>(K153244) per the SSXI<br>Guidance document and<br>does not raise any new<br>safety or effectiveness<br>issues. |
| Off The Shelf<br>Software | Upgraded Off-<br>the-Shelf (OTS)<br>software from<br>Windows<br>version 10. | Windows 7 | N/A | Tested per Software<br>Guidance and Off the<br>Shelf Software<br>Guidance. |
| Image<br>Storage | Updated Image<br>storage to a<br>maximum of<br>300000 frames. | 150000 frames | N/A | Increased system<br>capability for image<br>storage. This<br>modification does not<br>raise new issues of<br>safety or effectiveness. |
### Table 2: Comparison of Technological Characteristics
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots.
# 9. Non-Clinical Testing
The Siemens Cios Select has been tested to meet the requirements for conformity to multiple industry standards. Performance testing confirmed, that the Siemens Cios Select complies with the following 21 CFR Federal Performance Standards:
1020.30 Diagnostic X-Ray Systems and their major components 1020.32 Fluoroscopic equipment
1040.10Laser products
and with the following relevant voluntary FDA Recognized Consensus Standards as listed in the table below:
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
# Table 3: Voluntary Conformance Standards:
| Recog-<br>nition<br>Number | Product<br>Area | Title of Standard | Reference Number<br>and Date | Standards<br>Development<br>Organization |
|----------------------------|----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------|------------------------------------------|
| 19-4 | General | Medical Electrical Equipment - Part<br>1: General Requirements for<br>Safety, 2005 | 60601-1:2012, ed. 3.1 | IEC |
| 19-8 | General | Medical electrical equipment - Part<br>1-2: General requirements for basic<br>safety and essential performance -<br>Collateral Standard:<br>Electromagnetic disturbances -<br>Requirements and tests | 60601-1-2 Edition 4.0<br>2014-02 | IEC |
| 12-269 | Radiology | Medical electrical equipment - Part<br>1-3: General requirements for basic<br>safety and essential performance -<br>Collateral Standard: Radiation<br>protection in diagnostic X-ray<br>equipment | 60601-1-3 Edition 2.1<br>2013-04 | IEC |
| 5-114 | General | Medical devices - Part 1:<br>Application of usability engineering<br>to medical devices [Including<br>CORRIGENDUM 1 (2016)] | 62366-1 Edition 1.0<br>2015-02 | IEC |
| 5-70 | General I<br>(QS/RM) | Medical devices – application of<br>risk management to medical<br>devices | 14971:2012 | ISO |
| 13-32 | Software | Medical device software - Software<br>life cycle processes | 62304 Ed. 1.0<br>2006 | IEC |
| 12-204 | Radiology | Medical electrical equipment - Part<br>2-28: Particular requirements for<br>the basic safety and essential<br>performance of X-ray tube<br>assemblies for medical diagnosis | 60601-2-28 Edition<br>2.0 2010-03 | IEC |
| 12-202 | Radiology | Medical electrical equipment- Part<br>2-43: Particular requirements for<br>the basic safety and essential<br>performance of X-ray equipment for<br>interventional procedures | 60601-2-43<br>2010 | IEC |
| 12-274 | Radiology | Medical electrical equipment - Part<br>2- | 60601-2-54: | IEC |
| Recog-<br>nition<br>Number | Product<br>Area | Title of Standard | Reference Number<br>and Date | Standards<br>Development<br>Organization |
| | | 54: Particular requirements for the<br>basic safety and essential<br>performance of X-ray equipment for<br>radiography and radioscopy | 2009 | |
| 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set | PS 3.1 - 3.20 (2011) | NEMA |
| 12-273 | Radiology | Safety of laser products - Part 1:<br>Equipment classification, and<br>requirements | 60825-1 Edition 2.0<br>2007-03 | IEC |
| 12-290 | Radiology | Medical electrical equipment -<br>Radiation dose documentation -<br>Part<br>1: Equipment for radiography and<br>radioscopy | 61910-1 Edition 1.0<br>2014-09 | IEC |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
# Table 4: FDA Guidance Documents
| FDA Guidance Documents and Effective Date | |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Guidance for Industry and FDA Staff – User Fees and Refunds for Premarket<br>Notification Submissions 510(k)<br>Document issued on October 2, 2017 |
| 2. | Guidance for Industry and Food and Drug Administration Staff: Refuse to Accept Policy<br>for 510(k)s<br>Document issued on January 30, 2018 |
| 3. | Guidance for Industry and FDA Staff: Format for Traditional and Abbreviated 510(k)s -<br>Guidance for Industry and FDA Staff<br>Document issued on August 12, 2005 |
| 4. | Guidance for Industry and FDA Staff: Deciding when to submit a 510(k) for a change<br>to an existing device.<br>Document issued on October 25, 2017 |
| 5. | Guidance for Industry and Food and Drug Administration Staff: The 510(k) Program:<br>Evaluating Substantial Equivalence in Premarket Notifications [510(k)]<br>Document Issued on July 28, 2014 |
| 6 | Guidance for Industry and FDA Staff: Guidance for the Submission Of 510(k)'s for<br>Solid State X-ray Imaging Devices<br>Document issued on September 1, 2016 |
| 7. | Guidance for Industry and FDA Staff: Guidance for the Content of Premarket<br>Submission for Software in Medical Devices<br>Document issued on May 11, 2005 |
| 8. | Guidance for Industry and FDA Staff: Guidance for Off-The-Shelf Software Use in<br>Medical Devices<br>Document issued on September 9, 1999 |
| 9. | Guidance for Industry and FDA Staff: Applying Human Factors and Usability<br>Engineering to Medical Devices.<br>Document issued February 3, 2016 |
| 10 | Guidance for Industry and FDA Staff: Pediatric Information for X-ray Imaging Device<br>Premarket Notifications.<br>Document issued on November 28, 2017 |
| 11. | Guidance for Industry and FDA Staff: Content of Premarket Submissions for |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is in teal, and the word "Healthineers" is in orange. To the right of the words is a graphic of orange dots in a circular pattern.
| FDA Guidance Documents and Effective Date | |
|-------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Management of Cybersecurity in Medical devices.<br>Document issued on October 2, 2014 |
| 12. | Guidance for Industry and FDA Staff: Radio Frequency Wireless Technology in<br>Medical Device<br>Document issued on August 14, 2007. |
| 13. | Guidance for Industry and FDA Staff: Information to Support a Claim of<br>Electromagnetic Compatibility (EMC) of Electrically-Powered Medical Devices<br>Document issued on July 11, 2016 |
## Verification and Validation:
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005, and "Off-The-Shelf Software Use in Medical Devices" is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. Non-clinical tests were conducted on the Subject Device Cios Select software version VA20 during product development.
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results support that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence.
Bench testing in the form of Unit, Subsystem and System Integration testing were performed to evaluate the performance and functionality of the new features, hardware, and software updates. All testable requirements in the Engineering Requirements Specifications keys, Subsystem Requirements Specifications keys, and the Risk Management Hazard keys have been successfully verified and traced in accordance with the Siemens product development (lifecycle) process. The software verification and regression testing have been performed successfully to meet their previously determined acceptance criteria as stated in the test plans.
Electrical safety and EMC testing were conducted on the Cios Select, consisting of the acquisition unit (C-arm system) and the image processing and display station. The system complies with the IEC 60601-1, IEC 60601- 2-43 and IEC 60601-2-54 standards for safety and the IEC 60601-1-2 the standard for EMC.
The Cios Select software (VA20) was tested and found to be safe and effective for intended users, uses and use environments through the design control verification and validation process. All new software functions present in the Subject Device (New User Interface touch panel, New flat detector (PaxScan 2121DXV), (OTS) software from Windows 7 to Windows 10 and Updated Image storage to a maximum of 300000 frames) have been validated through detailed software testing and it was
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image contains the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots in a circular pattern.
founded they worked as intended. The Human Factor Usability Validation showed that Human factors are addressed in the system test according to the operator's manual and in clinical use tests with customer report and feedback form. Customer employees are adequately trained in the use of this equipment.
Siemens conforms to the cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed or transferred from a medical device to an external recipient. Provided in this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 80001-1-2010 is the hospital. Provided in the Software Section, is the required cybersecurity information.
Additional engineering bench testing was performed including the nonclinical testing identified in the guidance for submission of 510(k)s for Solid State X-Ray Imaging Devices (SSXI); demonstration of system performance; and an imaging performance evaluation. X-ray Imaging Devices- Laboratory Image Quality and Dose Assessment, Tests and Standards" was performed with acceptable results.
### Summary:
Performance tests were conducted to test the functionality of the Cios Select (VA20). These tests have been performed to assess the functionality of the Subject Device. Results of all conducted testing and clinical assessments were found acceptable and do not raise any new issues of safety or effectiveness.
### 10. General Safety and Effectiveness Concerns:
Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical and mechanical hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore, the operators are health care professionals familiar with and responsible for the evaluating and postprocessing of X-ray images.
### 11. Conclusion as to Substantial Equivalence:
The Cios Select has the same indications for use, operating environment, and mechanical design as the Primary and Secondary predicates. Siemens concludes via the documentation provided in this 510(k) submission that the Cios Select does not introduce any new potential safety risks and is substantially equivalent to, and performs as well as the
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the logo for Siemens Healthineers. The word "SIEMENS" is written in teal, and the word "Healthineers" is written in orange below it. To the right of the words is a graphic of orange dots arranged in a circular pattern.
predicate devices even though the Subject Device is using a different detector.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.