Cios Select

K153232 · Siemens Medical Solutions USA, Inc. · OWB · Feb 10, 2016 · Radiology

Device Facts

Record IDK153232
Device NameCios Select
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeOWB · Radiology
Decision DateFeb 10, 2016
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 892.1650
Device ClassClass 2
AttributesPediatric

Intended Use

The Cios Select is a mobile x-ray system intended for use in Operation room, Traumatology, Endoscopy, Intensive Care Station, Paediatrics, Ambulatory patient care and in Veterinary Medicine. The Clos Select can operate in four different modes, Digital Radiography, Fluoroscopy and Pulsed Fluoroscopy and Cassette exposure which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of intramedullary nail implant in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes.

Device Story

Mobile fluoroscopic C-arm X-ray system; consists of acquisition unit (C-arm, generator, tube housing, image intensifier) and image display station (trolley, processing/storage system, display). Used in OR, trauma, endoscopy, ICU, pediatrics, ambulatory care, and veterinary settings by healthcare professionals. Inputs: X-ray radiation; transforms inputs via Fluorospot Compact (SW VA10) platform; outputs: digital X-ray images for clinical visualization. Supports intraoperative procedures, implant positioning, and catheter guidance. Enhances clinical decision-making by providing real-time imaging for surgical and diagnostic guidance. Benefits patients through precise, low-dose intraoperative visualization.

Clinical Evidence

Bench testing only. Verification and validation testing performed to assess functionality and performance. Conformance to IEC 60601-1, 60601-1-2, 60601-1-3, 62366, 14971, 62304, 60601-2-28, 60601-2-43, 60601-2-54, and 60825-1 standards. Electrical safety and EMC testing passed.

Technological Characteristics

Mobile C-arm X-ray system. Components: X-ray tube, generator, image intensifier, display station. Connectivity: DVI/VGA, data network. Software: Fluorospot Compact (SW VA10) on Windows 7. Standards: IEC 60601-1:2012, 60601-1-2, 60601-1-3, 62366, 14971, 62304, 60601-2-28, 60601-2-43, 60601-2-54, PS 3.1-3.20 (DICOM), 60825-1.

Indications for Use

Indicated for use in operating rooms, traumatology, endoscopy, intensive care, pediatrics, ambulatory care, and veterinary medicine for digital radiography, fluoroscopy, pulsed fluoroscopy, and cassette exposure procedures, including bile duct display, intramedullary nail positioning, pediatric low-dose imaging, pain therapy, and catheter/probe placement.

Regulatory Classification

Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features the department's name arranged in a circular fashion around a central emblem. The emblem consists of a stylized design with three overlapping profiles, possibly representing people, beneath three curved lines. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 10, 2016 Siemens Medical Solutions USA, Inc. % Ms. Patricia Jones Regulatory Affairs Technical Specialist 40 Liberty Boulevard, 65-1A MALVERN PA 19355 Re: K153232 Trade/Device Name: Cios Select Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, OXO, JAA Dated: January 18, 2016 Received: January 19, 2016 Dear Ms. Jones: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Michael D.'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ Image /page/2/Picture/0 description: The image features the word "SIEMENS" in a bold, sans-serif font. The letters are a light blue color, and there is a thin black line underlining the entire word. The background is plain and white, which makes the word stand out. DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. 510(k) Number (if known) K153232 Device Name Cios Select ## Indications for Use (Describe) The Cios Select is a mobile x-ray system intended for use in Operation room, Traumatology, Endoscopy, Intensive Care Station, Paediatrics, Ambulatory patient care and in Veterinary Medicine. The Clos Select can operate in four different modes, Digital Radiography, Fluoroscopy and Pulsed Fluoroscopy and Cassette exposure which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of intramedullary nail implant in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes. Type of Use (Select one or both, as applicable) > Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!" The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (8/14) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The word is in all capital letters and is a light blue color. There is a black line underneath the word. # 510(k) Summary: Cios Select Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 ### Date Prepared: February 9, 2016 This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92. ### 1. General Information: # Importer / Distributor: Siemens Medical Systems USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Establishment Registration Number: 2240869 # Manufacturing Site: Siemens Shanghai Medical Equipment Ltd. 278 Zhou Zhu Road, Shanghai 201318, China Establishment Registration Number: 3003202425 ## 2. Contact Person: Ms. Patricia D Jones Technical Specialist, Regulatory Submissions Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355 Phone: (610) 448-6474 Fax: (610) 640-4481 Email: patricia.d.jones@siemens.com ## 3. Device Name and Classification: Trade Name: Cios Select Classification Name: Image-Intensified Fluoroscopic X-Ray System. Mobile Classification Panel: Radiology CFR Section: 21 CFR §892.1650 {4}------------------------------------------------ # SIEMENS | Submission Type: | Special | |-------------------------|----------| | Device Class: | Class II | | Product Codes: | OXO, OWB | | Secondary Product Code: | JAA | - 4. Legally Marketed Predicate Device Trade Name: Siremobil Compact 510(k) #: K963093 Clearance Date: September 3, 1996 Classification Name: Image-Intensified Fluoroscopic X-Ray System, Mobile Classification Panel: Radiology Classification Regulation: 21 CFR §892.1650 Device Class: Product Code: Recall Information: Class II ОХО 1 Recall (See Appendix H) ## 5. Device Description: This 510(k) submission, Cios Select is a Mobile Fluoroscopic C-arm X-ray The Cios Select is a modification of the Siremobil Compact System. originally cleared under Premarket Notification K963093 on September 3rd, 1996. The Cios Select consists of two major units: One is the acquisition unit with the C-arm and movable base containing the generator, power unit, system control, and tube housing assembly on one side of the C-arm and the image intensifier on the opposite side. The second unit is the image display station with a moveable trolley for the image processing and storage system, image display and documentation. Both units are connected to each other with a cable. The main unit is connected to the main power outlet and the trolley is connected to a data network. The following modifications were made to the predicate device the Siremobil Compact Mobile X-ray System cleared under Premarket Notification K963093 on September 03, 1996. Siemens Medical Solutions USA, Inc. submits this Special 510(k) to request clearance for the Subject Device the Cios Select. The following minor modifications are incorporated to create the Subject Device and have not received prior 510(k) clearance for the Subject Device: - 1. Upgraded software platform from Memoskop to Fluorospot Compact (SW VA10) Modification Affects: Hardware & Software {5}------------------------------------------------ # SIEMENS - 2. Fluorospot Compact SW VA10 contains the following software functional changes: - New user interface in navigating and operating the system in > terms of interactive elements on screen and on control kevboard. - Modification Affects: Hardware, Software & Labeling - 3. A new single tank with a new integrated X-ray tube and a new qenerator. Modification Affects: Hardware & Software - 4. Updated Image storage to maximum of 150000 frames Modification Affects: Hardware & Software - 5. New optional detachable grid. Modification Affects: Hardware & Software - 6. Upgraded Off-the-Shelf (OTS) software from Windows XP to Windows 7 . Modification Affects: Software - 7. Optional upgrade of external monitor connectivity from VGA to DVI port. Modification Affects: Hardware - 8. Added integrated power supply (UPS) Modification Affects: Hardware - Product Claims List (See Appendix F) 9. - Provided Update 510(k) Information (See Appendix G) 10. The modified Subject Device is within the same classification regulation with the same indication for use as the predicate device (Siremobil Compact). Provided in Section 14 of this 510(k) submission, is a completed "510(k) Decision Making FlowChart". We believe these modifications are eligible for the Special 510(k) process since the Subject Device has the same fundamental scientific technology, Intended Use and Indications for Use as the predicate device system. Documentation is provided to support a claim of substantial equivalence to Siemens' predicate device the Siremobil Compact K963093. ## 6. Indications for Use: The Cios Select is a mobile x-ray system intended for use in Operation room, Traumatology, Endoscopy, Intensive Care Station, Pediatrics, Ambulatory patient care and in Veterinary Medicine. The Cios Select can operate in four different modes, Digital Radiography, Fluoroscopy and Pulsed Fluoroscopy and Cassette exposure which are necessary in performing wide variety of clinical procedures, such as intraoperative bile duct display, fluoroscopic display of intra-medullary nail implant in various positions, low dose fluoroscopy in pediatrics, fluoroscopic techniques utilized in pain therapy and positioning of catheters and probes. {6}------------------------------------------------ Mechanical design # SIEMENS ## 7. Substantial Equivalence: The Subiect Device (Cios Select) is within the same classification requlation with the same indication for use as the predicate device Siremobil Compact (K963093 on September 3, 1996) (see Table 1). Documentation is provided to support a claim of substantial equivalence to the Siemens' predicated device. | Predicate Device(s) Name and<br>Manufacturer: Siemens AG | 510(k)<br>Number | Clearance<br>Date | Comparable Properties | |-----------------------------------------------------------|------------------|-------------------|------------------------------------------------------------| | Predicate Device : Siremobil Compact<br>Product Code: OXO | K963093 | 09/03/1996 | Indication for use<br>X-ray technology<br>Image processing | # Table 1: Predicate Device Comparable Properties to Subject Device ## 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device: The Cios Select with system software VA10 version contains the following minor modifications that were made to the predicate device: a different software platform software, new user interface features, several components contained within a single tank (new x-ray tube, new x-ray generator, and a collimator), an increase in storage space by 150,000 frames, an optional detachable grid. update OTS software from Windows XP to Windows 7. an optional upgrade of external monitor connectivity from VGA to DVI port and added an integrated power supply (UPS). Explanation of these minor modifications is supported in the submission with non-clinical testing. These minor modifications difference between the new device and the predicate device do not raise different questions of safety and effectiveness. The Subject Device has the same comparable properties (see Table 1 Comparable properties) as the predicate device. ## 9. Non-Clinical Testing: Non-clinical tests were conducted for the Cios during product development. The modifications described in this Premarket Notification were supported with verification and validation testing. Siemens claims conformance to the following performance standards: 60601-1:2012, ed. 3.1, 60601-1-2 300 2007; 60601-1-3: 2008+A1:2013; 62366: 2007; 14971:2010; 62304 Ed. 1.0 2006; 60601-2-28 Edition 2.0 2010-03; 60601-2-43:2010; 60601-2-54 2009; PS 3.1 - 3.20 (2011) and 60825-1:2007. Electrical safety and Electromagnetic Compatibility (EMC) testing were conducted on the Subject Device to comply with the IEC 60601-1 30, and the IEC 60601-1-2 3rd standard for EMC. All tests passed. {7}------------------------------------------------ Image /page/7/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are a light blue color. There is a black line underneath the word. # Verification and Validation: Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission. The performance data demonstrates continued conformance with special controls for medical devices containing software. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation for the device was found acceptable to support the claims of substantial equivalence. Siemens conforms to the Cybersecurity requirements by implementing a process of preventing unauthorized access, modifications, misuse or denial of use, or the unauthorized use of information that is stored, accessed, or transferred from a medical device to an external recipient. Contained in Attachment 7 of this submission is a cybersecurity statement that considers IEC 80001-1:2010. The responsibility for compliance with IEC 8001-1-2010 is the hospital. Provided in the Software section, (Section 18) and the Operator's Manual (Appendix B) is the required cybersecurity information. # Summary: Performance tests were conducted to test the functionality of the Subject Device system. These tests have been performed to assess the functionality of the Subject Device. Results of all conducted testing were found acceptable in supporting the claim of substantial equivalence. #### 10. General Safety and Effectiveness Concerns: Instructions for use are included within the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical and mechanical hazards. Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing. Furthermore the operators are health care professionals familiar with and responsible for the evaluating and post processing of X-ray images. #### 11. Conclusion as to Substantial Equivalence: The predicate device was cleared based on non-clinical supportive information. The results of these tests also demonstrate that the Cios Select acceptance criteria are adequate for this intended use. The comparison of technological characteristic, non-clinical performance data and software validation data demonstrates that the Subject Device is as safe, and effective {8}------------------------------------------------ Image /page/8/Picture/1 description: The image contains the word "SIEMENS" in a sans-serif font. The word is written in a light blue color. There is a black line underneath the word. when compared to the predicate device that is currently marketed for the same intended use and indication for use.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...