K181667 · Centese, Inc. · BTA · Nov 29, 2018 · General, Plastic Surgery
Device Facts
Record ID
K181667
Device Name
Thoraguard System
Applicant
Centese, Inc.
Product Code
BTA · General, Plastic Surgery
Decision Date
Nov 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Thoraguard System is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. The Thoraguard System is indicated for all situations where chest drains are applied - especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural empyema or other related conditions. The Thoraguard System is intended for use on patients in appropriate care settings.
Device Story
Thoraguard System is an AC/DC powered suction pump for aspiration of surgical fluids, tissue, gases, and infectious materials. System comprises a reusable electronic control module, disposable drainage canister/tubing, and a dual-lumen chest tube kit. Control module regulates suction and provides automated drainage; when used with the proprietary chest tube kit, it performs automated clog clearance. Device features a touchscreen GUI for displaying drainage information and safety alarms. Used in clinical settings for patients requiring chest drainage; operated by healthcare professionals. Output allows clinicians to monitor patient drainage status and manage thoracic conditions; automated clog clearance aims to improve drainage efficacy and patient outcomes.
Clinical Evidence
Bench testing only. Performance verified via suction testing (ISO 10079-1:2015), catheter testing (EN 1617:1997), system-level verification, biocompatibility (ISO 10993-1:2009), electrical safety, EMC, software V&V, and usability validation.
Technological Characteristics
AC/DC powered suction pump. Components: reusable control module, disposable drainage kit, dual-lumen chest tube. Features touchscreen GUI, automated drainage, and automated clog clearance. Biocompatibility per ISO 10993-1:2009. Electrical safety/EMC compliant. Suction performance per ISO 10079-1:2015.
Indications for Use
Indicated for patients requiring thoracic drainage in pleural and mediastinal cavities, including pneumothorax, post-operative cardiac/thoracic surgery, thorax injury, pleural effusion, and pleural empyema.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 29, 2018
Centese, Inc. Evan Luxon President & CEO 3929 Harney St Suite 3008 Omaha, Nebraska 68131
Re: K181667
Trade/Device Name: Thoraguard System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: November 9, 2018 Received: November 13, 2018
Dear Evan Luxon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/5 description: The image shows a digital signature from Long H. Chen -S. The signature indicates that the document was digitally signed on November 29, 2018, at 15:54:34. The time zone is -05'00'. The signature is for Long H. Chen -S.
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K181667
Device Name Thoraguard System
#### Indications for Use (Describe)
The Thoraguard System is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. The Thoraguard System is indicated for all situations where chest drains are applied - especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural empyema or other related conditions. The Thoraguard System is intended for use on patients in appropriate care settings.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### K181667
This summary of 510(k) safety and effectiveness information for the Centese Thoraguard System is being submitted in accordance with the requirements of 21 CFR 807.92.
#### 1. Submitter
Centese, Inc. 3929 Harney St Suite 3008 Omaha. NE 68131
Phone: (402) 417-0740 Email: eluxon@centese.com
Contact Person: Evan Luxon, President & CEO Date Prepared: June 22, 2018
#### 2. Device
Name of Device: Thoraguard System Common or Usual Name: Powered Suction Pump Classification Name: Pump, Portable, Aspiration (Manual or Powered) [21 CFR 878.4780] Regulatory Class: II Product Code: BTA
#### 3. Predicate Device
Thopaz+ Suction Pump by Medela AG (K141553)
# 4. Device Description
The Thoraguard Surgical Drainage System is an AC/DC powered suction pump designed for the aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. The Thoraguard System offers automated drainage functionality, acquisition and display of drainage information, and safety alarms associated with changes in patient output.
The Thoraguard System consists of the three sub-systems: the Thoraguard Control Module, a reusable electronic device for generating and regulating suction; the Thoraguard Drainage Kit, a disposable drainage canister and dual lumen drainage tubing set; and the Thoraguard Chest Tube Kit, a dual lumen drainage catheter with a passive valve and filter.
Similar to traditional chest drainage systems, the Thoraguard System drains accumulated blood and fluid through the chest tube using regulated suction. The control module facilitates this drainage automatically and works in conjunction with the drainage kit to function as a powered suction pump. The system may be used with any drainage catheter; when used with the Thoraguard Chest Tube Kit, the system provides automated clearance of clogs within the chest tube itself. During normal operation, the control module displays relevant system information via the touchscreen graphical user interface.
{4}------------------------------------------------
#### 5. Indications for Use
The Thoraguard System is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. The Thoraguard System is indicated for all situations where chest drains are applied - especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural effusion, pleural empyema or other related conditions. The Thoraguard System is intended for use on patients in appropriate care settings.
# 6. Comparison of Technological Characteristics with Predicate Device
The Thoraguard system has substantially equivalent technological characteristics as the predicate device. Similarities to the predicate device include the following:
- Pump Type
- · Suction capacity
- Max Vacuum
- Vacuum regulator type
- Vacuum gauge type
Differences in the technologic characteristics between the Thoraguard System and the predicate device do not raise different questions of safety and effectiveness. Centese has confirmed equivalent performance of the Thoraguard System to the predicate device using accepted scientific methods for assessing the effect of these different characteristics, such as bench testing.
# 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
# Performance Bench Testing
The Thoraquard System has been verified for its intended use by conducting a series of bench tests that qualified its functionality, integrity and performance. The testing performed included suction performance tests per ISO 10079-1:2015, catheter performance test per EN 1617:1997, and systemlevel verification testing.
# Biocompatibility Testing
Biocompatibility for the Thoraguard chest tube has been assessed in accordance with ISO 10993-1:2009 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management system.
# Electrical Safety and Electromagnetic Compatibility (EMC)
The Thoraguard Surgical Drainage System has been evaluated for electrical safety and electromagnetic compatibility (emissions and immunity) and has passed all electrical safety and EMC testing requirements.
{5}------------------------------------------------
# Software Verification and Validation Testing
Software verification and validation has been performed and the Thoraguard System has passed all testing requirements.
#### Usability Testing
Centese has performed usability validation of the Thoraguard System and has found the it to be safe and effective for the intended users, uses, and use environments.
#### 8. Conclusions
Results from bench testing for design verification and validation have demonstrated acceptable performance on all tests to indicate that the Thoraguard System is suitable for its intended use and is as safe and effective as the predicate device to support a determination of substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.