K141553 · Medela AG · BTA · Aug 25, 2014 · General, Plastic Surgery
Device Facts
Record ID
K141553
Device Name
MEDELA THOPAZ+
Applicant
Medela AG
Product Code
BTA · General, Plastic Surgery
Decision Date
Aug 25, 2014
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Thopaz+ is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. Thopaz+ is indicated for all situations where chest drains are applied - especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural effusion, pleural empyema or other related conditions. Thopaz+ is intended for use on patients in appropriate care settings.
Device Story
Medela THOPAZ+ is a portable, AC/DC powered suction pump for thoracic drainage. Device utilizes a membrane system to provide quiet, continuous suction (up to 5 L/min, -10 kPa). It features an integrated electronic monitoring system for air and fluid drainage, displaying data digitally and graphically on a color screen. A capacitive sensor measures fluid levels through the device wall. Used in clinical settings by healthcare providers to manage pleural/mediastinal drainage; supports patient mobility. Data can be transferred to a PC post-therapy. Device includes alarms for status monitoring. Benefits include improved patient mobilization, user-friendly interface, and reliable, maintenance-free operation.
Clinical Evidence
Bench testing only. No clinical data presented. Verification and validation activities were performed to support the modifications to the predicate device, including risk management evaluations.
Technological Characteristics
Portable, AC/DC powered suction pump. Membrane-based suction mechanism. Features capacitive fluid level sensing, color display with ambient light sensor, and digital communication interface. Dry system (no fluids required for operation). IP-rated housing sealed against disinfection agents. Connectivity via PC data transfer. Software-controlled via firmware.
Indications for Use
Indicated for patients requiring thoracic drainage in pleural and mediastinal cavities, including pneumothorax, post-operative cardiac/thoracic surgery, thorax injury, pleural effusion, and pleural empyema. Intended for use in appropriate care settings.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human figure in profile, with three overlapping faces suggesting community and connection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 25, 2014
Medela AG Mr. Orlando Antunes VP Regulatory Affairs Laettichstrasse 4b CH-6341 Baar, Switzerland
Re: K141553
Trade/Device Name: Medala ® THOPAZ+ Regulation Number: 21 CFR 878.4780 Regulation Name: Pump, Portable, Aspiration (manual or powered) Regulatory Class: II Product Code: BTA Dated: July 28, 2014 Received: July 31, 2014
Dear Mr. Antunes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number K141553
Device Name Medela® THOPAZ+
Indications for Use
Thopaz+ is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. Thopaz+ is indicated for all situations where chest drains are applied - especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural effusion, pleural empyema or other related conditions. Thopaz+ is intended for use on patients in appropriate care settings.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "medela" in black font, followed by a logo of three orange teardrop shapes arranged in a triangular pattern. The word "medela" is written in a simple, sans-serif font. The logo is positioned to the right of the word.
Medela AG, Laettichstrasse 4b, CH-6341 Baar, Switzerland t Person: Orlando Antunes +41 (41) 561 66 71 / +41 (41) 769 51 00 orlando.antunes@medela.ch cial 510(k) Submission for Medela® THOPAZ* Suction Pump
# K141553
# Section 7 - 510(k) Summary
This 510(k) summary for the Medela * THOPAZ* Suction Pump meets the requirements of 21 CFR 807.92.
#### Sponsor's Name, Address and Contact Person 1
- Sponsor: Contact Person Medela AG Orlando Antunes Medical Equipment VP Regulatory Affairs Laettichstrasse 4b 6341 Baar Switzerland Phone: +41 41 5626671 Fax: +41 41 769 5100
Summary prepared 02 June 2014
#### 2 Name of Device
| Trade Name: | Medela® THOPAZ+<br>Secretion & Surgical Aspirator |
|----------------------|--------------------------------------------------------------------------------------------|
| Common Name: | Powered Suction Pump |
| Classification Name: | PUMP, PORTABLE, ASPIRATION (MANUAL OR POWERED)<br>Classified Class II, per 21 CFR 878.4780 |
| Product Code: | BTA |
#### 3 Name of the unmodified legally marketed device
- Medela® THOPAZ, K130210
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "medela" in black, bold font. To the right of the word is an orange graphic of three teardrop shapes clustered together. The teardrop shapes are arranged in a triangular pattern, with one at the top and two at the bottom.
Medela AG, Laettichstrasse 4b, CH-6341 Baar, Switzerland son: Orlando Antunes +41 (41) 561 66 71 / +41 (41) 769 51 00 orlando.antunes@medela.ch cial 510(k) Submission for Medela® THOPAZ+ Suction Pump
#### 4 Device Description
The Medela® THOPAZ* suction pump is an innovative secretion aspirator intended to be used for thoracic drainage. Its well-proven membrane system guarantees maximum suction performance and quiet, dependable operation. Additional advantages of the Medela® THOPAZ* are: User friendliness, patient mobility, simple cleaning and integrated safety features. A comprehensive range of accessories makes the Medela® THOPAZ* suction pump ideally suited for thoracic drainage while mobilizing the patient.
The Medela® THOPAZ* suction pump is an AC/DC powered, maintenance-free aspirator which incorporates a DC-motor with membrane aggregate power actuation in its housing. A user friendly MMI (man machine interface) guides the user through first installation, change of settings, use, data transfer and alarm handling.
The Medela® THOPAZ* suction pump has an electronic measuring and monitoring system for air and fluid being drained with optical and acoustic status display. The device is a dry system, which means that no fluids are necessary for operation. Important information about the course of therapy is displayed digitally and as graphics in the display. These data can be transferred to a PC upon completion of the therapy.
The Medela® THOPAZ* suction pump has a suction capacity of 5 liters per minute and a maximum vacuum up to -10 kPa (-75 mmHg). The pump is marked "low flow – low vacuum".
A variety of reusable and disposable accessories for pleural and mediastinal drainage are available.
#### 5 Indications for use
Thopaz* is intended to be used for aspiration and removal of surgical fluids, tissue, gases, bodily fluids or infectious materials. Thopaz* is indicated for all situations where chest drains are applied – especially for thoracic drainage in the pleural and mediastinal cavity in situations such as pneumothorax, after cardiac or thoracic surgery (post-operative), thorax injury, pleural effusion, pleural empyema or other related conditions. Thopaz* is intended for use on patients in appropriate care settings.
#### 6 Summary of Technological Characteristics
The Medela THOPAZ* has the same performance characteristics as the unmodified legally marketed device Medela® THOPAZ. Modifications to the unmodified legally marketed device are outlined in section 12 – Outline of modifications – Justification for Special 510k. There are no differences in performance or technology which significantly affect the safety and effectiveness of the device or the application (pleural and mediastinal drainage). The outline of changes in-
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Medela logo. The word "medela" is written in black, lowercase letters. To the right of the word is a symbol of three orange teardrop shapes arranged in a triangular pattern.
Medela AG, Laettichstrasse 4b, CH-6341 Baar, Switzerland : Person: Orlando Antunes +41 (41) 561 66 71 / +41 (41) 769 51 00 orlando.antunes@medela.ch ecial 510(k) Submission for Medela® THOPAZ* Suction Pump
cludes the identification of the changes that indicate submission of a Special 510(k) as required per FDA memorandum #K97-1 "Deciding When to Submit a 510(k) for a Change to an Existing Device". Subsequent verification and validation activities for the applied changes are referenced and evidence is contained within this submission. All modifications to the unmodified legally marketed device are evaluated as part of the risk management as outlined in attachment 16 Risk Management. Documented evidence is contained in this submission.
#### 7 Conclusion
With regards to the legally marketed device Medela® THOPAZ, K130210 several minor changes have been implemented and are documented in this Special 510(k) device modification as shortly summarized below.
| 1. | Fixation of bed holder | The new bed holder is optimized to allow fixation of the pump on the hospital bed. To reflect the increasing variety of bed types, the new optimized version of the bed holder is implemented to allow continuous proper fixation of the THOPAZ+ to hospital beds. |
|------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 2. | Sealing of housing | The device is now sealed to prevent from any disinfection agents going into the housing. The additional sealing is only an improve- ment and the IP classification remains the same as before. Never- theless it was introduced to enhance resistance against fluids, e.g. disinfectant fluids used. |
| 3. | Inclusion of new level<br>PCB in housing | The new sensor is included in the housing which is used for fluid level detection that is now introduced (refer also to change 09). |
| 4. | Seal / O-ring (color and<br>shape) | The color of the seal which seals the connection of the canister to the device has been changed to orange and has been slightly adapted in order to prevent undesired removal of the seal.<br>In order to improve the detection of a missing seal the color of the seal was changed to orange and the new seal is sticking better to the housing with a slightly adapted geometry. |
| 5. | New tubing accesso-<br>ries | New tubing sizes are introduced to the subject device. These sizes extend the range of sizes that has existed for the legally marketed device in order to improve connection to large catheter sizes. |
| 6. | Optimization of carry- | As a result of the optimized handle the carrying strap was slightly |
| ing strap | adapted in diameter and wideness. | |
| 7. IFU revised to include<br>information on all<br>changes | Because of all changes applied to the legally marketed device it<br>was necessary to adapt the instructions for use. The new instruc-<br>tions for use generally compromises the same information.<br>Though, it has been complemented with the information for the<br>new level measurement and all other changes discussed in this<br>document. | |
| 8. Adapted device PCB | Because of the usage of a new level sensor, some minor changes<br>of the assembly were implemented:<br>- New level sensor power supply<br>- Digital communication interface<br>- Optimization of resistors and capacitors to enhance com-<br>munication speed | |
| 9. Fluid level measure-<br>ment | In order to measure also the fluid levels resulting from the drain-<br>age a new PCB was added. The PCB can measure fluids capacitive<br>through the device wall providing a reliable fluid measurement<br>independent from the fluids consistency (e.g. blood, water, fatty<br>liquids etc.). | |
| 10. Device Firmware | The device FW from Thopaz+ is based on the firmware of the le-<br>gally marketed device THOPAZ, K130210. However in order to<br>implement the fluid measurement the user interface had to be<br>revised. The device firmware controls, checks and regulates all<br>main device functions.<br>Generally, except for the new fluid measurement function, the<br>user interface remains unchanged. All functions regarding regula-<br>tion of the negative pressure work for both devices.<br>For the new Thopaz+ the firmware had to be adapted mainly in<br>order to show the new fluid amount parameters. None of these<br>changes impact the normal therapy function and the generation<br>of the negative pressure. | |
| 11. Color display and key- | In order to include the fluid measurement the device was | |
| | | |
| pad | equipped with a new slightly bigger color display. The active area | |
| | is now 73.4X49mm. | |
| | Because of the bigger size of the display the keypad needed to be | |
| | adapted. However, the keypad still provides the same amount of | |
| | buttons to approve a good transition from the legally marketed | |
| | device to Thopaz+. Because of the higher brightness of the new | |
| | color display an environment sensor was included to the new | |
| | keypad, which allows to adapt the display brightness according to | |
| | the environmental brightness. | |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "medela" in black font. To the right of the word is a logo consisting of three orange teardrop shapes arranged in a triangular pattern. The logo is slightly smaller than the word "medela".
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the word "medela" in bold, black font. To the right of the word is an orange graphic of three teardrop shapes. The teardrop shapes are arranged in a triangular pattern.
Medela AG, Laettichstrasse 4b, CH-6341 Baar, Switzerland t Person: Orlando Antunes +41 (41) 561 66 71 / +41 (41) 769 51 00 orlando.antunes@medela.ch Special 510(k) Submission for Medela® THOPAZ* Suction Pump
None of the modifications alters the Indications for Use in a significant way, nor the fundamental scientific technology, and do not introduce a fundamentally new scientific technology.
We therefore believe that the information presented in this Special 510(k) demonstrates that the product is safe for the patient, user, and bystander and does not raise any new questions regarding safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.