MAGNETOM Avanto, MAGNETOM Verio with software syngo MR E11D
K181613 · Siemens Medical Solutions USA, Inc. · LNH · Nov 6, 2018 · Radiology
Device Facts
Record ID
K181613
Device Name
MAGNETOM Avanto, MAGNETOM Verio with software syngo MR E11D
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
LNH · Radiology
Decision Date
Nov 6, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis. Your MAGNETOM MR system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
Device Story
MAGNETOM Avanto and Verio are MR diagnostic systems; input consists of magnetic resonance signals; system transforms signals into cross-sectional images, spectroscopic images, and derived physical parameters. Operated by trained physicians/technicians in clinical settings. Output displayed for physician interpretation to assist in diagnosis. Supports interventional procedures using MR-safe devices. Software syngo MR E11D introduces migrated applications (e.g., LiverLab, MyoMaps, Dot Engines) and sequence optimizations (e.g., HISTO, DSI, BLADE, WARP-SEMAC). System includes MaRS hardware component. Provides diagnostic information to clinicians; aids in clinical decision-making; enables non-invasive visualization of internal structures/functions.
Clinical Evidence
No clinical tests conducted. Substantial equivalence supported by software verification and validation testing, and submission of sample clinical images per FDA guidance for MR diagnostic devices.
Technological Characteristics
MR diagnostic system; software syngo MR E11D; MaRS hardware. Conforms to IEC 60601-1 (safety/EMC), IEC 60601-2-33 (MR safety), IEC 62304 (software lifecycle), and ISO 14971 (risk management). Connectivity includes IPv6 support and syngo.via interface. Features include various pulse sequences (EPI, Dixon, VIBE, etc.) and automated Dot Engines.
Indications for Use
Indicated for use as an MR diagnostic device to produce cross-sectional images and spectroscopic data of the head, body, or extremities for diagnostic purposes. Applicable to patients requiring MR imaging, including during interventional procedures using MR-compatible devices. No specific contraindications listed.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
MAGNETOM Avanto and MAGNETOM Verio with software syngo MR D13A (K121434)
Reference Devices
Software syngo MR E11C for the MAGNETOM System Aera and MAGNETOM System Skyra (K153343)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Siemens Medical Solutions USA, Inc. Cordell Fields Regulatory Specialist 40 Liberty Blvd Mailcode 65-1A Malvern, Pennsylvania 19355
November 6, 2018
Re: K181613
Trade/Device Name: MAGNETOM Avanto, MAGNETOM Verio with software syngo MR E11D Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic Resonance Diagnostic Device Regulatory Class: Class II Product Code: LNH, LNI, AND MOS Dated: October 10, 2018 Received: October 11, 2018
Dear Cordell Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael D. O'Hara
For
Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K181613
Device Name
MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D
#### Indications for Use (Describe)
Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and that displays the internal structure and/or function of the head, body, or extremities.
Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used. These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM MR system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
### MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D
Company: Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355, USA Establishment Registration Number: 2240869
Date Prepared June 14, 2018
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 CFR § 807.92.
#### 1. General Information
#### Importer/Distributor:
Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard Malvern, PA 19355, USA Establishment Registration Number: 2240869
#### Manufacturing Sites:
Siemens Shenzhen Magnetic Resonance Ltd. Siemens MRI Center, Gaoxin C. Ave., 2nd Hi-Tech Industrial Park 518057 Shenzhen China Establishment Registration Number: 3004754211
Siemens Healthcare GmbH Henkestr. 127 91052 Erlangen Germany Establishment Registration Number: 3002808157
#### 2. Contact Information
Cordell L. Fields, Esq. Regulatory Affairs Specialist Siemens Medical Solutions USA, Inc. 40 Liberty Boulevard, 65-1A Malvern, PA 19355, USA
{4}------------------------------------------------
Phone: (610) 448-6545 Fax: (610) 448-6469 E-mail: cordell.fields@siemens-healthineers.com
#### 3. Device Name and Classification
| Trade Name | MAGNETOM Avanto and MAGNETOM Verio with software<br>syngo MR E11D |
|-----------------------|-------------------------------------------------------------------|
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH<br>Secondary: LNI. MOS |
#### 4. Legally Marketed Predicate Device
| Trade Name | MAGNETOM Avanto and MAGNETOM Verio with software |
|-----------------------|--------------------------------------------------|
| | syngo MR D13A |
| 510(k) Number | K121434, Cleared November 05, 2012 |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR § 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH |
| | Secondary: LNL MOS |
### 5. Device Description
The subject device, MAGNETOM Avanto and MAGNETOM Verio D with software syngo MR E11D, is a modification of the previously cleared predicate device, MAGNETOM Avanto and MAGNETOM Verio with software syngo MR D13A (K121434). Software version syngo E11D for MAGNETOM Avanto and MAGNETOM Verio includes software applications migrated from the previously cleared MAGNETOM Aera and Skyra systems with syngo MR E11C (K153343). Only minor adaptations were needed to support the system specific hardware and optimize the sequence/protocols. The following are the software applications migrated from previously cleared software to the subject device:
#### Software
.
- Sequences or Algorithms:
- New sequences for both systems with syngo MR E11D with respect to syngo MR D13A :
- HISTO sequence (as part of the new application LiverLab) o
- O Beat map
- DSI (Diffusion Spectrum Imaging) o
- Improvements in BLADE Imaging (fast TSE) O
{5}------------------------------------------------
# SIEMENS
- Modified sequences for both systems with syngo MR E11D with respect to syngo MR D13A:
- syngo WARP-SEMAC (Advanced WARP) O
- Fast DIXON (TSE DIXON) o
- SPAIR fat sat improvements (applicable for 3T MR Systems addressed by o syngo MR E11D)
- Improvements in Liver Segmentation O
- Improvement in Multi Echo Dixon for Fat Iron Quantification o
- O A flow compensated SE sequence for 3T systems
- Improvements in Inline Ventricular function O
- Improvements in VIBE (TWIST-VIBE) O
- Minor modifications of sequences or algorithms for both systems with syngo MR E11D with respect to syngo MR D13A:
- EPI sequence: B1 control loop for a better DTI signal stability O
- Improved Body DWI imaging using a Flash-based PAT reference scan O
- Improved SPAIR fat saturation in haste acquisitions O
- Improved ACC algorithm to reduce artifacts O
- Improved spoiling for the TurboFLASH sequence, especially for MPRAGE O protocols
- Improved FatSat option for MSK imaging o
- B1-corrected abdominal T1 mapping O
- ICEPAT algorithm change O
- O Improved Dixon sequence
- Prepulese functionality now available for Petra sequence O
#### Applications:
- New applications for MAGNETOM Avanto //erio with syngo MR E11D with respect to syngo D13A:
- o LiverLab
- MyoMaps O
- TWIST VIBE (FREEZEit Body MRI) O
- Advanced WARP (WARP-SEMAC) O
- Dot Cockpit O
- Dynamic Breast Evaluation and Auto Bolus Detection for Breast Dot O Engine
- New Dot Engines for MAGNETOM Avanto /Verio with syngo MR E11D with respect to MAGNETOM Aera/Skyra with software syngo D13A:
- Spine Dot Engine o
- Large Joint Dot Engine (knee, hip, shoulder) O
- Breast Dot Engine o
Minor modifications of applications for MAGNETOM Avanto /Verio with syngo MR E11D with respect to syngo D13A:
{6}------------------------------------------------
# SIEM
- Improvements in Abdomen Dot Engine and Tim CT Onco Dot Engine O
- O Improvement in Auto Coil Selection
- TimCT Onco Improvements O
- o Neuro DTI: Customized diffusion directions
- Minor modifications to support the usability of the Avanto /Verio with syngo MR E11D with respect to syngo D13A:
- Improvement based on stimulation monitoring O
- AutoAlign Spine is integrated within Spine Dot Engine O
- Support for Internet protocol v6 support O
- O Economic mode intended for energy saving
- Inclusion of: O
- ഗ്ന Local AIF within Neuro Perfusion
- syngo BreVis
- Ability to interface with syngo.via post-processing O
- SEEit (Marketing bundle being made available for all subject device O systems)
### Hardware
.
- New HW-components for MAGNETOM Avanto /Verio:
- MaRS (Measurement and reconstruction system)
The MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D will be offered as in-field upgrades for the currently installed MAGNETOM Avanto and Verio systems.
### 6. Indication for Use
The indications for use for the subject device is the same as the predicate device and is as follows:
Your MAGNETOM MR system is indicated for use as a magnetic resonance diagnostic device (MRDD) that produces transverse, sagittal, coronal and oblique cross sectional images, spectroscopic images and/or spectra, and that displays the internal structure and/or function of the head, body, or extremities. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, contrast agents may be used.
These images and/or spectra and the physical parameters derived from the images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.
Your MAGNETOM MR system may also be used for imaging during interventional procedures when performed with MR compatible devices such as in-room displays and MR Safe biopsy needles.
{7}------------------------------------------------
### 7. Substantial Equivalence
The MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D is substantially equivalent to the following device:
| Predicate Device | FDA Clearance<br>Number | FDA Clearance<br>Date | Product Code |
|-------------------------------------------------------------------|-------------------------|-----------------------|------------------|
| MAGNETOM Avanto and MAGNETOM<br>Verio with software syngo MR D13A | K121434 | November 05, 2012 | LNH, LNI,<br>MOS |
As described above, the MAGNETOM Avanto Dot and MAGNETOM Verio Dot with software synqo MR E11D includes features already cleared on the following devices:
| Reference Device | FDA Clearance Number | FDA Clearance Date | Product Code |
|-------------------------------------------------------------------------------------|----------------------|--------------------|------------------|
| Software syngo MR E11C for the<br>MAGNETOM System Aera and<br>MAGNETOM System Skyra | K153343 | April 15, 2016 | LNH, LNI,<br>MOS |
#### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device
The subject device MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D is substantially equivalent to the predicate device, MAGNETOM Avanto and MAGNETOM Verio with software syngo MR D13A, with regard to the operational environment, programming language, operating system, and performance.
The subject device, MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D, conforms to the IEC 62304, Edition 1.1, 2015-06, standard for software medical devices and other relevant IEC and NEMA standards.
While there are some differences in technological characteristics between the subject device and predicate device including new and modified software applications and hardware additions, these differences have been tested and the conclusions from the non-clinical data suggests that the features bear an equivalent safety and performance profile as that of the predicate device.
### 9. Nonclinical Performance Testing
The following performance testing was conducted on the subject device:
- Software verification and validation testing was completed in accordance with the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005.
- Performance testing was completed in accordance with the FDA guidance । document, "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices", dated November 18, 2016
{8}------------------------------------------------
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate device to which it has been compared.
MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D conforms to the following FDA recognized and international IEC and ISO standards:
| Recogniton<br>Number | Product<br>Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
|----------------------|--------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|------------------------------------------|
| 19-4 | General II<br>(ES/EMC) | C1:2009/(R)2012 and<br>A2:2010/(R)2012<br>(Consolidated Text) Medical<br>electrical equipment - Part 1:<br>General requirements for basic<br>safety and essential<br>performance (IEC 60601-<br>1:2005, MOD) | ES60601-<br>1:2005/(R)20<br>12 and<br>A1:2012, | AAMI ANSI |
| 19-1 | General II<br>(ES/EMC) | Medical electrical equipment -<br>Part 1-2: General<br>requirements for basic safety<br>and essential performance -<br>Collateral standard:<br>Electromagnetic compatibility -<br>Requirements and tests | 60601-1-2<br>Edition 3:<br>2007-03 | IEC |
| 12-271 | Radiology | Medical electrical equipment -<br>Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis | 60601-2-33<br>Ed. 3.1:2013 | IEC |
| 5-40 | General I<br>(QS/RM) | Medical devices - Application<br>of risk management to medical<br>devices | 14971<br>Second<br>Edition 2007-<br>03-01 | ISO |
| 5-89 | General I<br>(QS/RM) | Medical electrical equipment -<br>Part 1-6: General<br>requirements for basic safety<br>and essential performance -<br>Collateral standard: Usability | 60601-1-6<br>Edition 3.1<br>2013-10 | IEC |
| 13-32 | Software/<br>Informatics | Medical device software -<br>Software life cycle processes | IEC 62304<br>Edition 1.1<br>2015-06 | AAMI ANSI<br>IEC |
{9}------------------------------------------------
# SIEMENS
No clinical tests were conducted to support the claim of substantial equivalence between the subject and predicate device. Sample clinical images have been provided in accordance with the FDA guidance document, "Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices," dated November 18, 2016.
### 10. General Safety and Effectiveness Concerns
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk management is ensured via compliance with ISO 14971:2007 to identify and provide mitigation of potential hazards in a risk analysis early in the design phase and continuously throughout the development of the product. These risks are controlled via measures realized in hardware and software development, testing, and product labeling. To minimize risks, Siemens adheres to recognized and established industry practices and standards, such as the IEC 60601-1 series, to minimize electrical and mechanical risk. Furthermore, the device is intended for healthcare professionals familiar with and responsible for the acquisition and-post processing of magnetic resonance images.
### 11.Conclusion as to Substantial Equivalence
There are no changes to the indications for use for the subject device as compared to that of the legally marketed predicate device. MAGNETOM Avanto and MAGNETOM Verio with software syngo MR D13A (K121434).
While the new and modified software and hardware features provide additional capabilities compared to the predicate device, the additional capabilities are currently cleared features of the reference devices MAGNETOM Aera and MAGNETOM Skyra with software syngo MR E11C (K153343) and do not raise new questions of safety and effectiveness. All features have been verified and validated to support the claim of substantial equivalence to the predicate device.
Siemens believes that the subject device, MAGNETOM Avanto and MAGNETOM Verio with software syngo MR E11D, is substantially equivalent to the predicate device, MAGNETOM Avanto and MAGNETOM Verio with software syngo MR D13A.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.