K181468 · Hermes Medical Solutions AB · KPS · Oct 25, 2018 · Radiology
Device Facts
Record ID
K181468
Device Name
Hybrid3D
Applicant
Hermes Medical Solutions AB
Product Code
KPS · Radiology
Decision Date
Oct 25, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
Hybrid3D that provides software applications used to process, display, analyze and manage nuclear medicine and other medical imaging data transferred from other workstation or acquisition stations. Warning! The results generated by the SIRT function cannot be used to change a treatment plan after the treatment has been delivered, nor can the results be used to impact future treatment planning for any type of radiation therapy or chemotherapy. The SIRT function cannot be used for prospective treatment planning of initial treatment with microspheres in contradiction to the labeling of Class III radiation microsphere devices. Warning! The results from the SIRT function using the partition model for dose planning cannot be used to change a clinical Y90 SIRT treatment plan, either before or after treatment. Nor can the results be used to impact future clinical treatment planning for any type of radiation therapy or chemotherapy. The SIRT function cannot be used for prospective treatment planning, initial or retreatment, with Y90 microspheres in contradiction to the labeling of the Class III radiation microspheres devices.
Device Story
Hybrid3D is a medical imaging software module for multi-modal (PET, CT, MR, SPECT) data processing; enables coregistration, interactive fusion, and 3D rendering. Features include: Lung Lobe Quantification for automated 3D lobar anatomy computation from CT; TumorFinder for automatic lesion segmentation in PET/CT based on background volume criteria; SIRT wizard for processing SIRT planning and verification. Used by clinicians to view, segment, and analyze imaging data; supports VOI definition via cropping, interpolation, and logic operations. Outputs are DICOM-compatible for research or clinical review. Assists in tumor burden evaluation and radionuclide therapy planning. SIRT function is restricted: cannot be used to change treatment plans or for prospective planning of Class III radiation microsphere treatments.
Clinical Evidence
Bench testing only. Validation performed by comparing software-generated calculations against manual spreadsheet calculations based on published formulae for SIRT (Resin and Glass microspheres). Metrics included Lung Shunt, Prescribed Activity, Activity to Implant, and Voxel Dose. Results showed identical values or minor variations (up to 5% in SPECT Dose Map normalization) attributed to methodological differences, confirming no fundamental calculation errors.
Technological Characteristics
Software-based image processing system. Supports multi-modal data (PET, CT, MR, SPECT). Features include 3D segmentation, fusion rendering, and automated quantification algorithms. Connectivity via DICOM import/export. Operates as a workstation module.
Indications for Use
Indicated for processing, displaying, analyzing, and managing nuclear medicine and other medical imaging data transferred from other workstations or acquisition stations.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 25, 2018
HERMES MEDICAL SOLUTIONS AB % Mr. Joakim Arwidson Vice President Quality & Regulatory Affairs Strandbergsgatan 16 STOCKHOLM SWEDEN 11251
Re: K181468
Trade/Device Name: Hybrid3D Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: KPS Dated: August 20, 2018 Received: September 24, 2018
Dear Mr. Arwidson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/ofdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Michael D. | Digitally signed by Michael D. O'hara -S<br>DN: c=US, o=U.S. Government, ou=HHS,<br>ou=FDA, ou=People,<br>0.9.2342.19200300.100.1.1=1300226759,<br>cn=Michael D. O'hara -S<br>Date: 2018.10.25 18:03:53 -04'00' |
|-----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| O'hara -S | For |
| Robert A. Ochs, Ph.D. | |
| Director | |
| Division of Radiological Health | |
| Office of In Vitro Diagnostics<br>and Radiological Health | |
| Center for Devices and Radiological Health | |
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K181468
Device Name Hybrid3D
Indications for Use (Describe)
Hybrid3D that provides software applications used to process, display, and manage nuclear medicine and other medical imaging data transferred from other workstation or acquisition stations.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 5.0 510 (k) SUMMARY
### A. Submitted by:
- Submitters name and address: ● Hermes Medical Solutions AB Strandbergsgatan 16 112 51 Stockholm Sweden
#### . Submitters telephone number
| Phone: | +46 8 19 03 25 |
|---------|-----------------------------------|
| Cell: | +46 708 19 03 08 |
| E-mail: | joakim.arwidson@hermesmedical.com |
#### . Contact person
Joakim Arwidson VP Quality and Regulatory Hermes Medical Solutions AB Strandbergsgatan 16 112 51 Stockholm Sweden
- . Registration number 9710645
- B. Preparation date: 2018-05-30
#### C. Proprietary/Trade name, Common name, Classification name :
- Proprietary/Trade name . Hybrid3D v3.0
- . Common name Image processing systems
- . Classification name Emission Computer Tomography System, Class II, 21CFR892.1200
#### D. Legally marketed device (predicate device):
Following legally marketed device has been used for comparison.
- . Hybrid3D v2.0 (K171719)
#### E. Description of the device that is subject of this premarket notification :
#### Hvbrid3D
HERMES Hybrid3D is a reading and processing module for the advanced needs in medical imaging. It offers multi-modal (PET/CT/MR/SPECT) coregistration and interactive fusion of multiple datasets. HybridViewer 3D handles viewing and fusion of multi-sequence MRI studies with oblique orientation and allows switching between original and standard TCS view orientation as well as defining own slice directions. 3D seqmentation, cropping and interpolation techniques allow
{5}------------------------------------------------
complex tasks in VOI definition and can cover cases like cavities, splitting structures into subsections or logic operations (compute intersections, merge, grow). Results can be imported and exported as DICOM and are therefore available for research in 3rd party tools. Additionally, it provides tools for advanced 3D fusion rendering of studies and VOIs.
#### Lung Lobe Quantification
The Lung Lobe Quantification module in Hybrid3D, introduces an efficient and automated workflow solution to accurately compute 3D lobar anatomy from CT (with or without contrast). The workflow supports the addition of functional images (SPECT V/Q, SUV SPECT, CT iodine maps, hyperpolarized xenon MRI, etc.) to accurately relate lobar anatomy to function. No changes have been made to Lung Lobar Quantification since the previous release.
#### TumorFinder
The Tumor Finder wizard provides automatic segmentation of lesions in a PET study or a combined PET/CT study pair, based on criteria relative to a background volume placed in the liver or mediastinum. This reduces the time required for tumor delineation. It also provides both visual and statistical evaluation of tumor burden, which helps with comparing follow up studies.
#### SIRT
Selective Internal Radionuclide Therapy (SIRT), is currently used in the treatment of liver tumors either from primary liver cancer or metastatic disease (e.g. colorectal primary cancer). The SIRT wizard provides processing for SIRT planning and verification.
#### F. Intended use:
Hybrid3D that provides software applications used to process, display, analyze and manage nuclear medicine and other medical imaging data transferred from other workstation or acquisition stations.
Warning! The results generated by the SIRT function cannot be used to change a treatment plan after the treatment has been delivered, nor can the results be used to impact future treatment planning for any type of radiation therapy or chemotherapy. The SIRT function cannot be used for prospective treatment planning of initial treatment with microspheres in contradiction to the labeling of Class III radiation microsphere devices.
Warning! The results from the SIRT function using the partition model for dose planning cannot be used to change a clinical Y90 SIRT treatment plan, either before or after treatment. Nor can the results be used to impact future clinical treatment planning for any type of radiation therapy or chemotherapy. The SIRT function cannot be used for prospective treatment planning, initial or retreatment, with Y90 microspheres in contradiction to the labeling of the Class III radiation microspheres devices.
#### G. Technological characteristics:
The proposed device Hybrid3D has the same technological characteristics as the predicate device and the same indication for use.
#### H. Testing:
The tests for verification and validation followed Hermes Medical Solutions AB design controlled procedures. The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria.
{6}------------------------------------------------
### I. Substantially Equivalent/Conclusions:
The proposed device Hybrid3Dv3.0and the predicate device Hybrid 3D v2.0 (K171719) have the same indication for use.
The proposed device will use similar technology and fundamental concepts and operation are also the same, as described in the 510(k) submission.
The comparisons between Hybrid 3D v3.0 and Hybrid 3D v2.0 (K171719) were part of the test procedure for V3.0 and showed good results.
Validation of the Hybrid3D SIRT has been made between the calculations performed by Hybrid3D and the results from spreadsheet calculations based on the formulae published by SIRTEX for the Resin Microspheres and BTG for the Glass Theraspheres.
- Lung Shunt was calculated from planar images. The results for Lung Shunt were identical.
- . The Prescribed Activity for Planning using Resin Microspheres using the BSA method were identical to 2 decimal places
- The Activity to Implant for Planning using Glass Microspheres using the PTV method were identical to 2 decimal places
- The Voxel Dose measured from the PETDose Map was identical to 2 decimal places ●
- The Voxel Dose measured from the SPECT Dose Map varied by up to 5% because the . methods used for normalization, by calculating the common field of view by the application and by the test procedure, differ slightly.
- . The values for Maximum Dose and Dose at 90% Volume D90%(Gy) differ because of the inherent inaccuracy of reading values manually from the Dose Volume Histogram. The test was performed only to establish that there were no fundamental errors in the calculations.
In summary, the Hybrid3Dv3.0 described in this submission is in our opinion substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.