The iCertainty System is intended for non-invasive two-dimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.
Device Story
iCertainty is a mobile, cart-based imaging system for real-time visualization of blood flow and tissue perfusion in an operating room. It uses a dual-laser system (450nm and 785nm) and a digital camera to illuminate and record target tissues. The device employs optical reflectance and laser speckle contrast analysis (LSCA) to process reflected light patterns, generating perfusion images up to 4-5mm deep. Operated by surgeons or trained medical professionals, the system displays perfusion images alongside visible light anatomical images on a central monitor. A sterile drape protects the sterile field. By providing real-time, non-invasive perfusion data without requiring injected dyes, the device assists surgeons in evaluating tissue viability and blood flow distribution during procedures, potentially improving clinical decision-making and patient outcomes.
Clinical Evidence
Bench testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), sterile drape compatibility, and cleaning validation. In vivo animal studies evaluated skin, solid organs, and vascular structures in normal and diseased states. Clinical evaluation of human anatomy (hand, forearm, ankle) in 40 patients demonstrated the device's ability to image blood flow distribution and perfusion.
Technological Characteristics
Mobile cart-based system with adjustable extension arm and optical head unit. Dual-laser illumination (450nm, 785nm). Imaging principle: optical reflectance and laser speckle contrast analysis. Resolution: 112µm/pixel; 1600x800 pixels. Imaging distance: 32±3 cm. Field of view: 9cm x 9cm. Connectivity: integrated computer/display. Power: standard US mains. Sterile barrier: Premier Guard drape.
Indications for Use
Indicated for non-invasive 2D imaging as an adjunctive method for evaluating blood flow distribution (flow and perfusion) in tissues up to 4-5mm depth in patients requiring surgical visualization of tissue perfusion.
Regulatory Classification
Identification
A cardiovascular blood flowmeter is a device that is connected to a flow transducer that energizes the transducer and processes and displays the blood flow signal.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health & Human Services and the U.S. Food & Drug Administration (FDA). The Department of Health & Human Services logo is on the left, featuring a stylized depiction of a human figure. To the right of it is the FDA logo, with the acronym "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in bold blue letters, and "ADMINISTRATION" in smaller blue letters below.
Rfpi % Jerzy Wojcik Sr Director RA/QA EdgeOne Medical 455 N Campbell Ave Chicago, Illinois 60612
December 14, 2018
Re: K181269
Trade/Device Name: iCertainty Regulation Number: 21 CFR 870.2100 Regulation Name: Cardiovascular Blood Flowmeter Regulatory Class: Class II Product Code: DPW Dated: November 20, 2018 Received: November 26, 2018
Dear Jerzy Wojcik:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Digitally signed by Neil R Neil R Ogden -S Date: 2018.12.14 Ogden -S Date: 2018.12.1-For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K181269
Device Name iCertainty
Indications for Use (Describe)
The iCertainty System is intended for non-invasive two-dimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| Date Summary Prepared: | December 12, 2018 |
|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner: | RFPi |
| Contact Person: | Jeffery Basham |
| | CEO, RFPi |
| | 1800 N Greene Street, Suite K |
| | Greenville, NC 27834 |
| | 919-280-2953 |
| | Jeff.basham@rfpi-co.com |
| 510(k) Consultant Contact: | |
| | Jerzy Wojcik |
| | Sr. Director RA/QA, EdgeOne Medical |
| | 455 N Campbell Ave, Suite 2N |
| | Chicago, IL 60612 |
| | 312-300-6643 |
| | Jerzy.wojcik@edgeonemedical.com |
| Device Name: | iCertainty |
| Trade Name: | iCertainty |
| Common Name: | Flowmeter, Blood,<br>Cardiovascular |
| Regulation: | 870.2100 |
| Class: | II |
| Product Code: | DPW |
| Predicate Device(s): | |
| | Primary Predicate K120884 PeriCam PSI |
| | Reference Device K063345 Novadaq SPY |
| Device Description: | The iCertainty System is a device used to image blood<br>flow distribution (flow in vessels and perfusion in<br>tissues) in real time. The device employs a mobile cart so<br>that it can be configured in the appropriate orientation to<br>properly image the tissue in an operating room<br>environment without restrictions. When a practitioner<br>elects to use imaging as an adjunct to evaluate blood flow<br>distribution the device is moved into the appropriate |
## 510(k) Summary
{5}------------------------------------------------
position and an extension arm with an optical head unit (OHU, containing the lasers/camera equipment) is positioned over the patient as described in the instructions for use. The OHU is moved to a specified height (focused) over the target image area and the imaging is performed. The imaging can be performed as often as deemed necessary by the surgeon during the procedure to visualize blood flow distribution (blood flow in vessels and perfusion in tissues) in the target tissues being operated upon. The device employs a dual laser-based system with a camera imaging system. The two lasers (450nm and 785nm) illuminate the surface and blood flow distribution up to a depth of 4-5mm of the target tissue. A digital camera records the image in real time. All real-time blood flow distribution analyses are performed on a central computer within the device. Realtime images are displayed to the surgeon on an imaging display. Data entry is performed on a separate display and data entry keyboard. A sterile drape is used as an accessory to protect the sterile operating field from contamination during use.
- Statement of Intended Use: The iCertainty System is intended for non-invasive twodimensional imaging as an adjunctive method for evaluation of blood flow distribution (flow and perfusion) in tissues, up to a depth of 4-5mm.
- Accessory Device: Premier Guard sterile drape (K041501)
{6}------------------------------------------------
| Device Comparisons | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Feature | iCertainty Proposed<br>Device | Primary Predicate<br>PeriCam PSI (K120884) | Reference Device<br>Novadaq Spy (K063345) |
| Intended Use /<br>Indications | ...for non-invasive two-<br>dimensional imaging as<br>an adjunctive method for<br>evaluation of blood flow<br>distribution (flow and<br>perfusion) in tissues, up to<br>a depth of 4-5mm. | ...for non-invasive two-<br>dimensional imaging of<br>peripheral tissue blood<br>perfusion. | ... for capturing and<br>viewing fluorescent<br>images for the visual<br>assessment of blood flow<br>as an adjunctive method<br>for the evaluation of<br>tissue perfusion, and<br>related tissue transfer<br>circulation in tissue and<br>free flaps used in plastic,<br>micro- and reconstructive<br>surgical procedures. |
| Conclusion – Both the subject and predicate devices have the same intended use as<br>an adjunctive method for evaluation of blood flow and perfusion. | | | |
| Target Users | Surgeons, physicians,<br>nurses and other trained<br>medical professionals | Surgeons, physicians,<br>nurses and other trained<br>medical professionals | Surgeons, physicians,<br>nurses and other trained<br>medical professionals |
| Conclusion – Substantially equivalent. All three devices have the same target users. | | | |
| Targeted<br>anatomical sites | Targeted surgical tissues<br>exposed to the device field<br>of view, including both<br>central organ and<br>peripheral tissues | Targeted peripheral<br>tissues exposed to the<br>device field of view | Targeted surgical tissues<br>exposed to the device field<br>of view, including both<br>central organ and<br>peripheral tissues |
| Conclusion - Substantially equivalent. All three devices image tissue that are in the<br>field of view of the device, including central organ and peripheral tissues in the subject<br>and primary predicate devices. | | | |
| Mechanism of<br>Action /<br>Technology | Laser light illuminates the<br>target surface tissue.<br>Using both optical<br>reflectance and laser<br>speckle contrast analysis,<br>the pattern reflected from<br>the blood in the tissue is<br>processed and imaged to<br>generate images of blood<br>perfusion in target tissues<br>up to a depth of 4-5 mm. | Laser light illuminates the<br>target surface tissue.<br>Using laser speckle<br>contrast analysis, the<br>pattern reflected from the<br>blood in the tissue is<br>processed and imaged to<br>generate images of blood<br>perfusion (depth not<br>determined, should be<br>similar to iCertainty by<br>optical physics<br>parameters). | Laser light illuminates the<br>target surface tissue<br>which has a fluorescent<br>dye coursing through<br>blood vessels introduced<br>through an intravenous<br>injection. The laser<br>excites the dye, which<br>emits an infrared light.<br>The infrared pattern<br>reflected is processed and<br>imaged to generate<br>images of blood perfusion |
| Device Comparisons | | | |
| Feature | iCertainty Proposed<br>Device | Primary Predicate<br>PeriCam PSI (K120884) | Reference Device<br>Novadaq Spy (K063345) |
| | Conclusion - Substantially equivalent to the predicate, PeriCam PSI and similar to<br>the reference, Novadaq SPY. All three devices: | | |
| | • | use laser light illumination of target tissues, then analyze the reflected light<br>to generate images of blood perfusion; | |
| | • | have a laser illumination system, camera, computer software for analysis and<br>display capabilities when in use; | |
| | • | have the capability to view, record and replay images of blood flow and<br>perfusion in target tissues. | |
| | Differences between the subject and predicate and reference devices are: | | |
| | • | the reference device requires an invasive injection of a fluorophobe ICG dye; | |
| | • | the subject and predicate device do not require an invasive injection; | |
| | • | the subject and predicate device use laser speckle contrast imaging analysis; | |
| | • | the reference device relatively quantifies the intensity of fluorescence as an<br>indicator of flow and perfusion; | |
| | • | the subject device displays both the visible light and perfusion image side by<br>side for user reference and orientation; | |
| | • | the predicates and reference present only the perfusion image. | |
| | The iCertainty technology used to achieve the intended use as compared to the<br>predicate does not raise new questions of safety or effectiveness. Refer to Section 12.2<br>for further substantial equivalence discussion. | | |
| Method of Use | After preparing the device<br>for use, the iCertainty is<br>positioned above the<br>targeted imaging site. The<br>device is activated, and<br>images are collected and<br>viewed by the operator. If<br>desired, the images can be<br>viewed again at a later<br>time. | After preparing the device<br>for use, the PeriCam PSI is<br>positioned above the<br>targeted imaging site. The<br>device is activated, and<br>images are collected and<br>viewed by the operator. If<br>desired, the images can be<br>viewed again at a later<br>time. | After preparing the device<br>for use, the patient is<br>injected with a fluorescent<br>dye. Novadaq SPY is<br>positioned above the<br>targeted imaging site. The<br>dye must be delivered to<br>the target tissue by the<br>circulatory system. The<br>device is activated, and<br>images are collected and<br>viewed by the operator. If<br>desired, the images can be<br>viewed again at a later<br>time. |
| | Conclusion - Substantially equivalent. All three devices utilize the same method of<br>use, in that the device is positioned over the target area and activated for collection of<br>images for immediate viewing and/or again at a later time. Refer to Section 12.3 for<br>further substantial equivalence discussion. | | |
| Electrical Safety<br>and<br>Electromagnetic<br>Compatibility | IEC 60601-1 and<br>60601-1-2 compliant | IEC 60601-1 and<br>60601-1-2 compliant | IEC 60601-1 and<br>60601-1-2 compliant |
| | Conclusion - Substantially equivalent. All three devices are compliant with<br>standards for electrical safety and compatibility. | | |
| Imaging<br>Distance (Optical<br>Head to Target<br>Tissue) | $32 +/- 3$ cm | 30 cm | 10 cm |
| | Conclusion – substantially equivalent. The subject device and SPY predicate device<br>have comparable imaging distances. |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.