K192761 · Onlume, Inc. · IZI · Jan 17, 2020 · Radiology
Device Facts
Record ID
K192761
Device Name
Asimov-MKS Imaging System
Applicant
Onlume, Inc.
Product Code
IZI · Radiology
Decision Date
Jan 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1600
Device Class
Class 2
Indications for Use
The Asimov-MKS Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after vascular, gastrointestinal, organ transplant, plastic, reconstructive, and micro surgeries.
Device Story
Asimov-MKS Imaging System is a mobile fluorescence angiography system for surgical suites. Device uses near-infrared laser (805 nm) to excite indocyanine green (ICG) injected into patient blood; CMOS cameras capture fluorescence and white-light images. System displays real-time fluorescence overlaid on white-light color images, allowing surgeons to assess tissue perfusion and blood flow relative to anatomy without darkening the room. Imager is mounted on an articulated arm attached to a mobile cart containing processing hardware. Surgeon operates device; sterile drapes cover imager and arm. Output includes real-time video and recorded images for clinical review. Benefits include improved visualization of perfusion during surgery, aiding intraoperative decision-making.
Clinical Evidence
Bench testing only. Verification and validation testing confirmed conformance to design specifications. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60825-1, and IEC 62471 standards demonstrated via independent laboratory testing.
Technological Characteristics
Mobile cart-based system with articulated arm. Fluorescence excitation: 805 nm infrared laser diode (55mW). Detectors: CMOS (12-bit fluorescence, 8-bit white light). Illumination: Integral white LEDs and ambient light. Connectivity: WiFi for data export. Software: Linux-based. Sterile drape used for field maintenance. Complies with IEC 60601-1, 60601-1-2, 60601-1-6, 60825-1, and 62471.
Indications for Use
Indicated for fluorescence imaging of blood flow and tissue perfusion in patients undergoing vascular, gastrointestinal, organ transplant, plastic, reconstructive, and micro surgeries.
Regulatory Classification
Identification
An angiographic x-ray system is a device intended for radiologic visualization of the heart, blood vessels, or lymphatic system during or after injection of a contrast medium. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
SPY Elite® Intraoperative Perfusion Assessment System (K182907)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services USA logo. To the right of that is the FDA logo, with the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below that.
OnLume Inc. Mr. Greg Bange Senior Manager - Regulatory Affairs and Quality Assurance 3300 Commercial Ave Madison, Wisconsin 53714
January 17, 2020
## Re: K192761
Trade/Device Name: Asimov-MKS Imaging System Regulation Number: 21 CFR 892.1600 Regulation Name: Angiographic X-Ray System Regulatory Class: Class II Product Code: IZI Dated: December 18, 2019 Received: December 20, 2019
#### Dear Greg Bange:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Neil R.P. Ogden Assistant Director, THT4A4 DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192761
Device Name Asimov-MKS Imaging System
#### Indications for Use (Describe)
The Asimov-MKS Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after vascular, gastrointestinal, organ transplant, plastic, reconstructive, and micro surgeries.
Type of Use (*Select one or both, as applicable*)
| <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
|--------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo and contact information for Onlume. The logo features the word "Onlume" in a blue sans-serif font, with a stylized lightbulb above the "l". Below the logo is the address "3300 Commercial Ave, Madison, WI 53714" and the website "www.onlume.com".
#### Section 8 510(k) Summary
## Submitter
OnLume Incorporated 3300 Commercial Avenue Madison, WI 53714 Phone: 888-751-5131 Contact: Greg Bange Date Prepared: December 18, 2019
#### Device Identification
| Device Name: | Asimov-MKS Imaging System |
|-------------------------|---------------------------------|
| Trade Name: | Asimov-MKS Imaging System |
| Common Name: | Fluorescence Angiography System |
| Regulation Number: | 21 CFR 892.1600 |
| Regulation Description: | Angiographic x-ray system |
| Regulation Name: | System, X-Ray, Angiographic |
| Regulatory Class: | Class II |
| Product Code: | IZI |
## Predicate Device
SPY Elite® Intraoperative Perfusion Assessment System - K182907
#### Reference Devices
No reference devices were used in this submission.
#### Device Description
The Asimov-MKS Imaging System is a fluorescence angiography imaging system. It is a prescription device intended to be used in hospital surgical suites, ambulatory surgery centers, and similarly equipped healthcare facilities. The Asimov-MKS Imaging System consists of an imager mounted to the distal end of an articulated arm that is affixed to a mobile cart.
Under the supervision of the surgeon, the imager illuminates a region of interested with near infrared and/or white light. Cameras, integral to the imager, acquire fluorescence and/or white light color images and direct the acquired images to processing hardware within the cart.
The articulated arm allows the imager to be directed at the region of interest. The arm affords unencumbered access to the patient and permits the cart to be positioned outside of the sterile field.
The cart allows the Asimov-MKS Imaging System to be positioned as needed for a surgical procedure and to be relocated within a facility for additional procedures or storage. The cart houses the processing hardware that display and record the acquired images. Software executed on the
{4}------------------------------------------------
processing hardware provides secondary functions such as patient biographical data entry, image review, and data archive.
To maintain the sterile surgical field, a sterile drape covers the imager and portions of the arm during clinical use. The single use drape streamlines device cleaning after use.
The fluorescent imaging agent utilized is indocyanine green (ICG). ICG is injected into the patient's blood stream, rapidly binds to the blood plasma, and is transported on the blood from the injection point to the region of interest. The ICG fluoresces when illuminated with the near infrared light emitted by the imager. Images of the fluorescence allow the clinician to visually assess circulation and related tissue perfusion.
#### Indications for Use / Intended Use
The Asimov-MKS Imaging System is indicated for fluorescence imaging of blood flow and tissue perfusion before, during, and after vascular, gastrointestinal, organ transplant, plastic, reconstructive, and micro surgeries.
#### Technological Characteristics
The subject and predicate devices employ the same fundamental scientific principles and have substantially equivalent technological characteristics. The predicate and subject devices have functionally equivalent subsystems for the imager, articulated arm, and mobile cart. The subject device illuminates the surgical field with infrared light of the same wavelength as that of the predicate device. The predicate device uses ambient room light when capturing a color image, whereas, the subject device utilizes both ambient light and the integral white LEDs during color imaging. Both the subject and predicate devices are draped during surgical use and utilize ICG as the fluorescent agent. Neither device is intended to contact the patient during use.
Two operational differences between the predicate and the Asimov-MKS Imaging System are:
- . The Asimov-MKS Imaging System supports the real-time display of combined white-light color and fluorescence images (i.e. fluorescence overlaid on color images). Thus, the user may assess blood flow and tissue perfusion as indicated in the fluorescence images referenced directly to the anatomy as indicated in the white-light color images during the surgical procedure. The predicate device does not support the real-time display of overlaid images.
- The Asimov-MKS Imaging System is able to acquire fluorescence images with no changes in the ambient lighting of a typical suite. The predict device requires that the surgical suite be darkened during fluorescence image acquisition.
The following table compares key attributes of the Asimov-MKS Imaging System to those of the predicate device.
{5}------------------------------------------------
| Attribute | Subject Device | Predicate Device |
|------------------------------------------------------------------------|-------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Trade Name | Asimov-MKS Imaging System<br>(subject to change when placed<br>on the market) | SPY Elite® Intraoperative<br>Perfusion Assessment System -<br>K182907 |
| Common Name | fluorescence angiography<br>imaging system | fluorescence angiography<br>imaging system |
| Product Code | IZI (21 CFR 892.1600) | IZI (21 CFR 892.1600) |
| Regulation Device Name | System, X-Ray, Angiographic | System, X-Ray, Angiographic |
| Combination Product | No (used with equivalent third-<br>party items) | Yes |
| Fluorescence agent | indocyanine green (ICG) | indocyanine green (ICG) |
| Fluorescence Excitation<br>Source | Infrared laser diode<br>805 nm<br>55mW (at laser aperture) | Infrared laser diode<br>805 nm<br>119mW (at laser aperture) |
| Fluorescence Detector | CMOS | CCD |
| Fluorescence Detector<br>Output | 12 bits/pixel<br>1024 x 768 (pixels/frame) | 8 bits/pixel<br>1024 x 768 (pixels/frame) |
| Fluorescence Frame rates | Display rate:<br>30 fps<br>Capture/Record rate:<br>30 fps | Display rate:<br>30 fps<br>Capture/Record rate:<br>30, 15, 7.5, or 3.75 fps (user<br>selected) |
| Fluorescence record<br>capacity (continuous real-<br>time image record | Up to 160,000 frames | Up to 1024 frames |
| White Light Excitation<br>Source (for color images) | Integral LED and Ambient | Ambient |
| White Light Detector | CMOS | CCD |
| White Light Detector<br>Output | 8 bits/pixel<br>1024 x 768 (pixels/frame) | 8 bits/pixel<br>1024 x 768 (pixels/frame) |
| White Light Frame rates | Display:<br>30 fps<br>Record:<br>30 fps and single snapshot<br>image | Display:<br>30 fps<br>Record:<br>Single snapshot image |
| White Light record capacity<br>(continuous real-time image<br>record) | Up to 160,000 frames | 1 frame per snapshot |
| Working distance (cm) | 30 | 30 |
| Field of View (cm2) | 266 (19 x 14) | 266 (19 x 14) |
| Output dimensions (in<br>pixels) | 1024 x 768 | 1024 x 768 |
| Attribute | Subject Device | Predicate Device |
| Display Bit Depth | 8 | 8 |
| Real-time<br>Color image with<br>Fluorescence Overlay | Yes | No |
| Ambient Light Level | High (room lights on) | Low (room lights off) |
| Mobile | Yes | Yes |
| Power Source | Wall | Wall1 |
| Patient contact<br>(w/ imaging device) | None | None |
| Consumables<br>(single use) | Sterile Drape, ICG | Sterile Drape, ICG |
| Operating System | Linux | Windows |
| Report Generation | Various formats and data | Various formats and data |
| Data Archive | WiFi export | USB export |
{6}------------------------------------------------
1 The predicate device has battery support for completion of image data record in the event of a wall power failure. Battery support is unnecessary in the subject device as images are recorded in real-time.
## Performance Data
Results of verification and validation testing performed by OnLume Inc. confirm the Asimov-MKS Imaging System conforms to design specifications and meets the needs of the intended users. Additionally, testing performed by an independent certified testing laboratory demonstrates the device complies with the requirements of applicable FDA recognized consensus safety standards for medical devices as listed in the following table.
| Applicable Standards | | |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| Standard Number<br>(Edition and Publication<br>Date) | Title | FDA<br>Recognition<br>number |
| IEC 60601-1:2005 +<br>A1:2012<br><br>(ed 3.1 2012-08) | Medical electrical equipment Part 1:<br>General requirements for basic safety and essential<br>performance | 19-4 |
| IEC 60601-1-2:2014<br><br>(ed 4.0 2014-02) | Medical electrical equipment Part 1-2:<br>General requirements for basic safety and essential<br>performance – Collateral Standard:<br>Electromagnetic disturbances – Requirements and<br>test | 19-8 |
| IEC 60601-1-6:2010 +<br>A1:2013<br><br>(ed 3.1 2013-10) | Medical electrical equipment Part 1-6:<br>General requirements for basic safety and essential<br>performance - Collateral Standard: Usability | 5-89 |
{7}------------------------------------------------
| Applicable Standards | | |
|---------------------------------------|--------------------------------------------------------------------------------|--------|
| IEC 60825-1:2007<br>(ed 2.0, 2007-03) | Safety of laser products –Part 1:<br>Equipment classification and requirements | 12-273 |
| IEC 62471<br>(ed. 1.0, 2006-07) | Photobiological safety of lamps and lamp systems | 12-249 |
In aggregate the test results and associated performance data demonstrate that the Asimov-MKS Imaging System is equivalent and performs at least as well as the predicate device.
## Conclusion
The Asimov-MKS Imaging System is substantially equivalent to the predicate device. The devices have the same intended use and the same technological characteristics. The technological differences that exist do not raise new types of safety or effectiveness questions. Further, results of performance testing demonstrate the Asimov-MKS Imaging System is equivalent to the predicate.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.