Glidesheath Slender Tibial Pedal Kit

K181237 · Terumo Medical Corporation · DYB · Aug 3, 2018 · Cardiovascular

Device Facts

Record IDK181237
Device NameGlidesheath Slender Tibial Pedal Kit
ApplicantTerumo Medical Corporation
Product CodeDYB · Cardiovascular
Decision DateAug 3, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1340
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K181237 · Aug 3, 2018Glidesheath Slender Tibial Pedal KitTerumo Medical CorporationPRIME Registry; Clinical literature datasetThe PRIME Registry data was used to evaluate the clinical performance of the Glidesheath Slender in patients undergoing endovascular procedures via tibial access, comparing outcomes against literature-based radial and femoral access procedures.Registry; Tibial access; Endovascular procedure; Comparative analysis

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
PRIME Registry; Registry-based observational cohort studyPatients undergoing index endovascular procedures using a tibial access point; Sample Size: 71 patientsLiterature-based dataset of procedures using radial and femoral accessClinical performance/feasibility of tibial access

Indications for Use

The Glidesheath Slender Tibial Pedal Kit is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee.

Device Story

The Glidesheath Slender Tibial Pedal Kit is a disposable, single-use introducer kit used in cath labs by physicians to gain arterial or venous access in the lower extremity below the knee. The kit includes a stainless steel entry needle, a nitinol mini guide wire, a guide inserter, a dilator, and a hydrophilic-coated sheath. The physician uses the needle to access the vessel, inserts the guide wire, and then advances the sheath and dilator over the wire. The dilator facilitates vessel entry and maintains sheath integrity. Once the sheath is positioned, the dilator and wire are removed, allowing for the insertion of diagnostic or interventional catheters. The sheath features a 1-way valve and 3-way stopcock for fluid management. The device is visible under fluoroscopy due to bismuth content. The hydrophilic coating reduces frictional resistance during insertion. The kit provides a standardized, sterile set of components to facilitate peripheral vascular procedures, potentially improving access success in challenging below-the-knee anatomy.

Clinical Evidence

Clinical evidence included a summary of the PRIME Registry data comparing the first 71 patients treated with the Glidesheath Slender via tibial access to a literature-based dataset of radial and femoral access procedures. Additionally, a simulated-use feasibility study was conducted using cadaver lower leg models to confirm compatibility with ultrasound-guided below-the-knee (BTK) access protocols.

Technological Characteristics

Sheath assembly: ETFE copolymer, bismuth trioxide, silicone oil, hydrophilic coating (dimethyl acrylamide-glycidyl methacrylate copolymer), polypropylene housing/cap, silicone rubber valve. Dilator: polypropylene, bismuth subcarbonate. Guide wire: nitinol core, palladium coil. Sterilization: Ethylene Oxide (ISO 11135, SAL 10^-6). Biocompatibility: ISO 10993-1. Packaging standards: ASTM D4169, ASTM F1929, ASTM F88, ASTM F2825.

Indications for Use

Indicated for patients requiring catheter placement into the lower extremity peripheral vasculature below the knee.

Regulatory Classification

Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. August 3, 2018 Terumo Medical Corporation Liang Lu Senior Regulatory Affairs Specialist 950 Elkton Blvd. Elkton, Maryland 21921 Re: K181237 Trade/Device Name: Glidesheath Slender Tibial Pedal Kit Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Regulatory Class: Class II Product Code: DYB Dated: April 24, 2018 Received: May 10, 2018 Dear Liang Lu: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Finn E. Donaldson -S Digitally signed by Finn E. Donaldson - S DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=20009796 73, cn=Finn E. Donaldson -S Date: 2018.08.03 15:29:23 -04'00' Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K181237 Device Name Glidesheath Slender Tibial Pedal Kit Indications for Use (Describe) The Glidesheath Slender Tibial Pedal Kit is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee. | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:12pt">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:12pt">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(K) SUMMARY #### A. SUBMITTER INFORMATION (807.92(a)(1)) #### Prepared bv: Liang Lu Senior Regulatory Affairs Specialist Terumo Medical Corporation Tel. (410) 392-7321 Fax (410) 398-6079 Prepared for: Owner/Operator Terumo Medical Corporation 2101 Cottontail Lane Somerset, NJ 08873, USA Registration Number: 2243441 #### Manufacturer (510(k) Applicant) Terumo Medical Corporation 950 Elkton Boulevard, Elkton, MD 21921 Registration Number: 1118880 #### Sterilization Facility Steris Isomedix Services Inc. 3459 South Clinton Ave South Plainfield, NJ 07080 Registration Number: 2246552 #### Contact Person: Liang Lu Senior Regulatory Affairs Specialist Terumo Medical Corporation 950 Elkton Boulevard, Elkton, MD 21921 Tel. (410) 392-7321 Fax (410) 398-6079 E-mail: liang.lu@terumomedical.com Date prepared: August 3, 2018 {5}------------------------------------------------ ### B. DEVICE NAME (807.92(a)(2)) | Proprietary Name: | Glidesheath Slender Tibial Pedal Kit | |-----------------------|--------------------------------------| | Common Name: | Introducer Sheath | | Classification Name: | Catheter Introducer | | Classification Panel: | Cardiovascular | | Regulation: | 21 CFR 870.1340 | | Product Code: | DYB | | Classification: | Class II | # C. PREDICATE DEVICE (807.92(a)(3)) The legally marketed device(s) to which substantial equivalence is claimed are: - Predicate Device: K142183 Glidesheath Slender, manufactured by Terumo . Corporation - Reference Device 1: K152173 Glidesheath, manufactured by Terumo ● Medical Corporation - Reference Device 2: K111606 - Pinnacle Precision Access System, manufactured by Terumo Medical Corporation # D. REASON FOR 510(k) SUBMISSION This premarket notification (510(k)) is being submitted for the Glidesheath Slender Tibial Pedal Kit, manufactured by Terumo Medical Corporation, for the purposes of establishing substantial equivalence to a legally marketed predicate device. ### E. DEVICE DESCRIPTION (807.92(a)(4)) The Glidesheath Slender Tibial Pedal Kit is used to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee. It consists of an introducer (sheath and dilator), which are packaged together with a mini guide wire, an entry needle and a guide inserter. During a diagnostic or interventional procedure in a cath lab, the stainless steel entry needle (cannula) is used to gain access to the vein or artery for placement of the mini guide wire. The nitinol mini guide wire is inserted through the cannula into the patient's blood vessel. The wire is used for placement of the sheath and dilator into the vein or artery. The Guide Inserter which is attached to the Mini Guide Wire holder is used to assist the placement of the wire into the needle. Following guide wire insertion, the cannula is removed and the sheath and dilator are then inserted over the mini guide wire and into the blood vessel. The mini guide wire is then withdrawn from the vessel. The dilator maintains the integrity of the sheath and dilates the blood vessel during insertion. Once the sheath is situated in the vessel, the dilator is removed and an appropriate catheter can then be inserted through the sheath. The sheath incorporates a 1-way valve and a 3-way stopcock connected by a side tube. The sheath is coated with hydrophilic coating to minimize frictional resistance when inserting or removing the sheath from the patient's blood vessel. In addition, the sheath and dilator contain bismuth, making these devices visible under fluoroscopy. {6}------------------------------------------------ The sheath, dilator, entry needle, mini guide wire and guide inserter are provided in a single package and sterilized together. The Glidesheath Slender Tibial Pedal Kit is a disposable, ethylene oxide gas sterilized device intended for single use only. # F. INDICATIONS FOR USE (807.92(a)(5)) The Glidesheath Slender Tibial Pedal Kit is indicated to facilitate placing a catheter through the skin into the lower extremity peripheral vasculature below the knee. ### G. SUBSTANTIAL EQUIVALENCE COMPARISON (807.92(a)(6)) The Glidesheath Slender Tibial Pedal Kit, subject of this 510(k), is substantially equivalent in its intended use/indications for use, technology/principal of operation, materials, and performance to the predicate device, manufactured by Terumo Medical Corporation. A comparison of the technological characteristics is summarized on the table below. {7}------------------------------------------------ Table 1: Summary of Substantial Equivalence | Device<br>Characteristic | New Device:<br>Glidesheath Slender Tibial Pedal Kit | Predicate:<br>Glidesheath Slender (K142183) | |------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Terumo Medical Corporation | Terumo Corporation | | Intended Use /<br>Indications for<br>Use | The Glidesheath Slender Tibial<br>Pedal Kit is indicated to facilitate<br>placing a catheter through the skin<br>into the lower extremity peripheral<br>vasculature below the knee. | The Glidesheath Slender is used to<br>facilitate placing a catheter through<br>the skin into the radial artery. | | Operation<br>Principle | Operated manually or by a manual<br>process | Same | | Design /<br>Construction | Sheath, Dilator,<br>Guide Wire, Guide Inserter, Entry<br>Needle | Sheath, Dilator,<br>Guide Wire, Guide Inserter, Entry<br>Needle,<br>Surflo IV catheter | | Materials | <b>Sheath assembly</b><br>Tube:<br>Ethylene-Tetrafluoroethylene<br>(ETFE) copolymer,<br>Bismuth trioxide, Colorant, Silicone<br>oil<br><br>Hydrophilic Coating: Dimethyl<br>acrylamide-glycidyl methacrylate<br>copolymer<br><br>Housing: Polypropylene<br><br>Cap: Polypropylene<br><br>Valve: Silicone Rubber<br><br>Caulking Pin: Stainless Steel<br><br>Sheath support:<br>Styrene-ethylene-butylene-styrene<br>block copolymer, Colorant<br><br>Side tube:<br>Polybutadiene<br><br>Three-way stopcock:<br>Holder: Polycarbonate;<br>Cock: Polyethylene;<br>Fastener pin: Polyethylene<br><br><b>Dilator assembly</b><br>Tube: Polypropylene, Bismuth<br>subcarbonate, Colorant, Silicone oil<br><br>Hub: Polypropylene, Colorant | Note: Sheath and Dilator are<br>Identical to the subject device<br><br><b>Sheath Assembly</b><br>Identical<br><br>Identical<br><br>Identical<br><br>Identical<br><br>Identical<br><br>Identical<br><br>Identical<br><br>Identical<br><br><b>Dilator Assembly</b><br>Identical<br><br>Identical | | | Mini guide wire | Mini Guide Wire | | | Core: Nitinol (Nickel Titanium<br>Alloy) | Nitinol (Nickel Titanium Alloy),<br>Tungsten,<br>Polyurethane | | | Coil: Palladium | | | | Adhesive: Epoxy Adhesive | | | | Guide inserter | Guide inserter | | | High-density polyethylene (HDPE),<br>Colorant | Polyethylene | | | Stainless steel entry needle | Stainless steel entry needle | | | Cannula:<br>Stainless Steel, Silicone oil | Cannula:<br>Stainless Steel | | | Hub: Styrene-Butadiene copolymer | Hub: Polycarbonate | | | Protective Sleeve: Polypropylene | Surflo IV Catheter:<br>Ethylene-Tetrafluoroethylene<br>(ETFE) copolymer, Barium sulfate;<br>Polypropylene; Stainless Steel;<br>Polystyrene, Polyester-Chlorinated<br>polyvinyl chloride; Polycarbonate | | Package | Unit Pouch<br>Shelf Box<br>Shipping Carton | Unit Pouch<br>Shelf Box<br>Shipping Carton | | Specifications | Sheath Size: 5 Fr.<br>Sheath Length: 10 cm<br>Hydrophilic Coating: full effective<br>length (10 cm) | Sheath Size: 5,6,7 Fr.;<br>Sheath Length: 10,16 cm<br>Hydrophilic Coating: full effective<br>length (10, 16 cm) | | | Dilator applicable to Guide Wire<br>OD: 0.021"<br>Dilator Length: 15.5 cm | Dilator applicable to Guide Wire<br>OD: 0.018,0.021,0.025,0.035"<br>Dilator Length: 15.5 and 21.5 cm | | | Guide Wire OD: 0.021"<br>Guide Wire Length: 43 cm | Guide Wire OD:<br>0.018,0.021,0.025,0.035"<br>Guide Wire Length:<br>43,80 cm | | | Entry Needle Type:<br>21/19 (G)<br>Entry Needle Length: 70 mm | Entry Needle Type:<br>21/20 (G)<br>Entry Needle Length: 35 mm | | | | Surflo IV Catheter Type:<br>18,20,22(G);<br>Length: 25,32,51,64(mm) | | Sterilization | Ethylene Oxide (validated in<br>accordance with ISO 11135 to<br>achieve SAL 10-6) | Same | | Shelf life | 30 months | Same | | Disposable Single<br>Use | Yes | Same | {8}------------------------------------------------ {9}------------------------------------------------ ### H. NON-CLINICAL TESTS (807.92(b)(1)) Performance All components of the Glidesheath Slender Tibial Pedal Kit come from the predicate device and reference devices. They are identical in materials, formulation, geometry, source (manufacturer), processing, chemicals, and intended patient exposure and utilize the same sterilization method (Ethylene Oxide) as those in the predicate devices and reference devices. All components are packaged together in a pouch, labeled, and sterilized as a finished good for the Glidesheath Slender Tibial Pedal kit. Therefore, no design verification testing was performed on the individual components to demonstrate the substantial equivalence of the proposed Glidesheath Slender Tibial Pedal kit. Since the packaging configurations of the needle in the Glidesheath Slender Tibial Pedal Kit were different from the reference device (K152173), packaging testing was re-evaluated. The following packaging verification testing was performed to ensure the durability of the proposed Glidesheath Slender Tibial Pedal kit packaging throughout distribution: - . Visual Inspection (Post Environmental Conditioning and Distribution Simulation) ### Cadaver Testing To evaluate the clinical feasibility of the Glidesheath Slender Tibial Pedal kit, devices were tested in a simulated use situation using cadaver lower leg models to test arterial access at specific BTK (below-the-knee) arterial access sites. It was demonstrated that the Glidesheath Slender sheath is compatible with the insertion angles used in ultrasound guided BTK access protocols in the four BTK vessels tested. The study also demonstrated that the kit and its components are acceptable for use in BTK access. The study along with a follow-up survey confirmed that the protocols used by physicians in standard practice are equivalent to the access protocol used in the simulated use study. This provides confidence that the results from the testing are applicable to the broader BTK operator community. Therefore, the results of the feasibility study ensure the substantial equivalence of the proposed Glidesheath Slender Tibial Pedal kit. Standards referenced in the testing of the proposed Glidesheath Slender Tibial Pedal kit by Terumo Medical Corporation are provided in Table 2 below. {10}------------------------------------------------ | Standard Designation | Standard Name | FDA<br>Recognition #<br>(if applicable) | |------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------| | ISO 10993-1: 2009<br>Cor. 1:2010 | Biological Evaluation of Medical Devices - Part<br>1: Evaluation and Testing Within a Risk<br>Management Process | 2-220 | | ISO 10993-7: 2008<br>Cor.1:2009 | Biological Evaluation of Medical Devices - Part<br>7: Ethylene Oxide Sterilization Residuals | 14-408 | | USP 38 <85> | Bacterial Endotoxins Test (Sterility) | NA | | ISO 11135: 2014 | Sterilization Of Health-Care Products - Ethylene<br>Oxide - Requirements For The Development,<br>Validation And Routine Control Of A<br>Sterilization Process For Medical Devices<br>(Sterility) | 14-452 | | ASTM D4169-14 | Standard Practice for Performance Testing of<br>Shipping Containers and Systems | NA | | ASTM F1929-15 | Standard Test Method for Detecting Seal Leaks<br>in Porous Medical Packaging by Dye Penetration | 14-484 | | ASTM F88/F88M-15 | Standard Test Method for Seal Strength of<br>Flexible Barrier Materials | 14-482 | | ASTM F2825-10<br>(Reapproved 2015) | Standard Practice for Climatic Stressing of<br>Packaging Systems for Single Parcel Delivery | 14-344 | | | | | Table 2: Standards Referenced in Testing of the Glidesheath Slender Tibial Pedal Kit | |--|--|--|--------------------------------------------------------------------------------------| | | | | | # Biocompatibility Biocompatibility classification is based on the FDA Guidance Use of International Standard ISO 10993-1, "Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing within a Risk Management Process." Glidesheath Slender Tibial Pedal kit is categorized as an external communicating device, circulating blood with limited contact duration (up to 24 hours). The biological evaluation of Glidesheath Slender Tibial Pedal kit was performed per EN ISO 10993-1 and FDA Guidance on Use of International Standard ISO 10993-1. All components of the Glidesheath Slender Tibial Pedal kit are identical in materials, formulation, geometry, source (manufacturer), processing, chemicals, and intended patient exposure and utilize the same sterilization method (Ethylene Oxide) as those in the predicate device and reference devices. Therefore, no additional biocompatibility testing was performed, and the Glidesheath Slender Tibial Pedal kit is considered to have substantially equivalent biocompatibility for the indicated use. #### Sterilization Glidesheath Slender Tibial Pedal kit was adopted into an existing ethylene oxide sterilization process validated via the overkill half cycle approach in accordance with ISO 11135:2014, Sterilization of health-care products - Ethylene oxide -Requirements for the development, validation and routine control of a sterilization {11}------------------------------------------------ process for medical devices. The microbiological and performance qualifications for the existing sterilization process used for the reference devices were leveraged to provide a Sterility Assurance Level (SAL) of 10-6. Glidesheath Slender Tibial Pedal kit is a limited exposure device. Sterilant residual testing was performed to demonstrate that after 24 hours of heated aeration, the level of residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) do not exceed an average daily dose of 4 mg and 9 mg respectively per EN ISO 10993-7:2008. #### CLINICAL TESTS (807.92(b)(2)) I. This 510(k) included a summary of the PRIME Registry data. The first 71 patients treated with the Terumo Glidesheath Slender when undergoing an index endovascular procedure using a tibial access point was compared to a literature based dataset of procedures using radial and femoral access. No issues that may affect the determination of substantial equivalence were noted in the comparison. # J. CONCLUSION (807.92(b)(3)) In summary, the Glidesheath Slender Tibial Pedal kit, subject of this 510(k), is substantially equivalent in its intended use, technology/principal of operation, materials, and performance to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...