Caretaker Remote Display App and Caretaker Software Library
Applicant
Caretaker Medical, LLC
Product Code
DRG · Cardiovascular
Decision Date
Oct 3, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Caretaker Remote Display App and Caretaker Software Library are indicated for use when remote display and/or forwarding of securely transmitted wireless data from the Caretaker Physiological Monitor device (K163255) is required. The Remote Display App and the Software Library provides bi-directional data transfer capabilities with the Caretaker Physiological Monitor device (K163255). To facilitate communications between the Caretaker Physiological Monitor device and other FDA Cleared patient monitors and Remote Data Displays (RDDS), the Caretaker Software Library provides a well-defined interface for 3rd party data integrations. In addition to providing data display and forwarding, the Remote Display App can operate as an MDDS for 3rd party wireless medical devices. The alerts in the subject device are only to be used for historical review and are not intended for time-critical intervention. The Software Library is intended only for use by licensed partners of the company, and the Caretaker Remote Display App is intended for use by clinicians or other properly trained medical personnel. Both the Remote Display App and Software Library are for prescriptive use only.
Device Story
Software application and library running on COTS hardware (tablets, phones, computers); provides remote display, storage, and forwarding of physiological data/waveforms from Caretaker4 Physiological Monitor (K163255). Enables bi-directional control of the Caretaker4 monitor via a more convenient interface than the device's native 1-button control. Functions as an MDDS for 3rd party wireless medical devices. Data forwarded via WiFi/USB to external systems or EMRs. Used by clinicians for remote patient consultation. Alerts are software-generated based on operator-set limits; intended for historical review, not time-critical intervention. Benefits include improved clinical workflow and remote access to patient data.
Clinical Evidence
Bench testing only. Testing confirmed functionality by connecting to the Caretaker4 Physiological Monitor via the Software Library and to SpO2/temperature sensors using BLE protocols. Data forwarding verified via WiFi and USB to external RDDS. Usability and risk assessments performed.
Technological Characteristics
Software-based application and library; runs on COTS hardware (computers, tablets, cell phones). Connectivity via WiFi, USB, and Bluetooth (BLE). Provides bi-directional data transfer and control interface for Caretaker4 monitor. No physiological data processing performed; alarms are threshold-based (operator-set).
Indications for Use
Indicated for clinicians or trained medical personnel requiring remote display or forwarding of wireless physiological data from the Caretaker Physiological Monitor (K163255). Also functions as an MDDS for 3rd party wireless medical devices. Alerts are for historical review, not time-critical intervention. Prescriptive use only.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 3, 2018
Caretaker Medical, LLC Jeff Pompeo President 941 Glenwood Station Ln, Suite 301 Charlottesville, Virginia 22901
Re: K181196
Trade/Device Name: Caretaker Remote Display App and Caretaker Software Library Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II Product Code: DXN, DRG Dated: September 5, 2018 Received: September 5, 2018
Dear Jeff Pompeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{2}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
| Sincerely, | |
|--------------------------------------------|---------------------------------------------|
| Shawn W. | |
| for Forrest -A | |
| | Digitally signed by Shawn W. Forrest -A |
| | DN: c=US, o=U.S. Government, ou=HHS, ou=FDA |
| | ou=People, |
| | 0.9.2342.19200300.100.1.1=1300403341, |
| | cn=Shawn W. Forrest -A |
| | Date: 2018.10.03 22:18:36 -04'00' |
| Bram D. Zuckerman, M.D. | |
| Director | |
| Division of Cardiovascular Devices | |
| Office of Device Evaluation | |
| Center for Devices and Radiological Health | |
Enclosure
{3}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K181196
Device Name
Caretaker Remote Display App and Caretaker Software Library
### Indications for Use (Describe)
The Caretaker Remote Display App and Caretaker Software Library are indicated for use when remote display and/or forwarding of securely transmitted wireless data from the Caretaker Physiological Monitor device (K163255) is required. The Remote Display App and the Software Library provides bi-directional data transfer capabilities with the Caretaker Physiological Monitor device (K163255). To facilitate communications between the Caretaker Physiological Monitor device and other FDA Cleared patient monitors and Remote Data Displays (RDDS), the Caretaker Software Library provides a well-defined interface for 3rd party data integrations.
In addition to providing data display and forwarding, the Remote Display App can operate as an MDDS for 3rd party wireless medical devices. The alerts in the subject device are only to be used for historical review and are not intended for time-critical intervention.
The Software Library is intended only for use by licensed partners of the Caretaker Remote Display App is intended for use by clinicians or other properly trained medical personnel. Both the Remote Display App and Software Library are for prescriptive use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{4}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
- 1) Preparation Date: Aug 31, 2018 2) Submitted by: CareTaker Medical, LLC 941 Glenwood Station Ln, Suite 301 Charlottesville, Virginia 2290 User Fee Organization Number 397095 Owner/Operator #: 10054848 3) Contact Person/Prepared by: Jeff Pompeo President & CEO CareTaker Medical 941 Glenwood Station Ln. Suite 301 Charlottesville. Virginia 22901-1455 Phone: 434-409-1945 Email: Jeff@caretakermedical.net 4) Device Identification: Trade Name: CareTaker Remote Display App and CareTaker Software Library Common Name: CareTaker App Classification: 21 CFR 870.1130, Product Codes: DXN -System, Measurement, Blood-Pressure, Non-Invasive DRG - Transmitters And Receivers, Physiological Signal, Radiofrequency Device Class: II 5) Predicate Devices: VIOS Monitoring System, K150992 Airstrip ONE Web Client, K160862 6)
- Device Description: CareTaker Medical's App ("CareTaker Remote Display App") is a software App that runs on Commercial, Off the Shelf (CoTS) hardware to provide additional controls and features to the CareTaker4 Physiological Monitor, K163255, as well as an interface to other devices.
The App is a software platform that displays and stores physiological parameters and waveforms from patient monitoring medical devices. CareTaker Remote Display App can also forward this data to third party systems, including Electronic Medical Records systems.
The integrated CareTaker Software library provides new control and configuration functions to a connected CareTaker4 Physiological Monitor. This library can also be used by 3rd parties to develop their own Apps that are capable of controlling a CareTaker4 as well. The library does not and cannot control any other medical devices or systems it is connected with.
CareTaker Remote Display App can function without the CareTaker Software Library, in which case it is the same as the RDDS in K163255.
{5}------------------------------------------------
### K181196
The relationships between the various items as well as between K163255 and K181196 are shown conceptually below.
Image /page/4/Figure/2 description: The image shows a diagram of two systems, K163255 and K181196, with various components and connections. System K163255 includes a CT4 Physiological Monitor connected to an RDDS. System K181196 contains a Library, an App, an RDDS, and Other Devices, all interconnected. Arrows indicate the flow of data or interaction between these components within and between the systems.
#### 7) Intended Use:
The Caretaker Remote Display App and Caretaker Software Library are indicated for use when remote display and/or forwarding of securely transmitted wireless data from the Caretaker Physiological Monitor device (K163255) is required. The Remote Display App and the Software Library provides bi-directional data transfer capabilities with the Caretaker Physiological Monitor device (K163255). To facilitate communications between the Caretaker Physiological Monitor device and other FDA Cleared patient monitors and Remote Data Displays (RDDS), the Caretaker Software Library provides a well-defined interface for 3rd party data integrations.
In addition to providing data display and forwarding, the Remote Display App can operate as an MDDS for 3rd party wireless medical devices. The alerts in the subject device are only to be used for historical review and are not intended for time-critical intervention.
The Software Library is intended only for use by licensed partners of the company, and the Caretaker Remote Display App is intended for use by clinicians or other properly trained medical personnel. Both the Remote Display App and Software Library are for prescriptive use only.
| Feature | VIOS Monitoring System | Airstrip ONE Web Client | CareTaker App |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications<br>for Use | The Vios Monitoring System<br>(VMS) is intended for use by<br>medically qualified personnel for<br>physiological and vital signs<br>monitoring of adult (18+) patients<br>in healthcare facilities. It is<br>indicated for use in monitoring of<br>ECG, heart rate, pulse rate,<br>functional oxygen saturation of<br>arterial hemoglobin, and axillary | Information is generated by other<br>medical devices and patient<br>information system, and not by<br>this device. This device<br>captures this information from<br>these other systems and displays it<br>for clinicians.<br>This device is intended to be used<br>by clinicians for the following<br>purposes: | The Caretaker Remote Display<br>App and Caretaker Software<br>Library are indicated for use<br>when remote display and/or<br>forwarding of securely<br>transmitted wireless data from<br>the Caretaker Physiological<br>Monitor device (K163255) is<br>required. The Remote Display |
#### Comparison to Predicate: 8)
{6}------------------------------------------------
| K181196 |
|---------|
|---------|
| | temperature. VMS allows for the<br>input of non-invasive blood<br>pressure and can display data<br>from peripheral devices. VMS can<br>generate alerts when rate-based<br>cardiac arrhythmias are detected<br>and when physiological vital signs<br>fall outside of selected<br>parameters. | K181196<br>• To view the near real-time<br>waveforms remotely<br>• To remotely review other<br>standard or critical near real-time<br>patient data from the monitored<br>system<br>• To provide a request for remote<br>consultation regarding a patient's<br>waveform or other data<br>This device software can display<br>the following physiologic data<br>captured by other medical<br>devices:<br>• ECG Waveform<br>• Heart Rate Monitored<br>• Respiratory Rate<br>• Oxygen Saturation<br>• Intracranial Pressure<br>• Central Venous Pressure<br>• Pulmonary Capillary Wedge<br>Pressure<br>• Cardiac Index<br>• Cardiac Output<br>• Cerebral Perfusion Pressure<br>• Systolic Blood Pressure Invasive<br>• Mean Arterial Pressure Invasive<br>• Diastolic Blood Pressure<br>Invasive<br>• Systolic Blood Pressure Cuff<br>• Mean Arterial Pressure Cuff<br>• Diastolic Blood Pressure Cuff<br><b>Contraindications</b><br>This device is intended for use by<br>clinicians when they cannot be at<br>the hospital. This device is<br>intended for use by clinicians as a<br>diagnostic aid, and not as a<br>replacement for direct viewing of<br>any of the monitoring devices<br>from which it obtains its data. | App and the Software Library<br>provides bi-directional data<br>transfer capabilities with the<br>Caretaker Physiological Monitor<br>device (K163255). To facilitate<br>communications between the<br>Caretaker Physiological Monitor<br>device and other FDA Cleared<br>patient monitors and <b>Remote</b><br>Data Displays (RDDS), the<br>Caretaker Software Library<br>provides a well-defined interface<br>for 3rd party data integrations.<br>In addition to providing data<br>display and forwarding, the<br><b>Remote</b> Display App can operate<br>as an MDDS for 3rd party<br>wireless medical devices. The<br>alerts in the subject device are<br>only to be used for historical<br>review and are not intended for<br>time-critical intervention.<br>The Software Library is intended<br>only for use by licensed partners<br>of the company, and the<br>Caretaker <b>Remote</b> Display App<br>is intended for use by clinicians<br>or other properly trained<br>medical personnel. Both the<br><b>Remote</b> Display App and<br>Software Library are for<br>prescriptive use only. |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Regulatory<br>Classification | - 21 CFR 870.2300; Cardiac<br>monitor (including<br>cardiotachometer and rate<br>alarm)<br>- 21 CFR 890.2910<br>Transmitters And Receivers,<br>Physiological<br>Signal, Radiofrequency (Class<br>II, Cardiovascular)<br>- 21 CFR 880.2910 Clinical<br>Electronic Thermometer (Class | 21 CFR 870.2300; Cardiac<br>monitor (including<br>cardiotachometer and rate<br>alarm) | - 21 CFR 870.1130; System,<br>Measurement, Blood-<br>Pressure, Non-Invasive<br>- 21 CFR 890.2910<br>Transmitters And Receivers,<br>Physiological<br>Signal, Radiofrequency (Class<br>II, Cardiovascular) |
{7}------------------------------------------------
| | II exempt, General Hospital) | | K181196 |
|----------------------------------------------|--------------------------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| Product Code | DRT, DRG, FLL | MSX | DXN, DRG |
| Device Class | II | II | II |
| 510(k) | K150992 | K160862 | K181196 |
| Other<br>Technologies<br>Used | Standard computers, network,<br>and WIFI | Standard computers, network,<br>and WIFI | Standard computers, tablets,<br>cell phones, network, WIFI,<br>and Bluetooth |
| Where Used | Anywhere clinicians remotely<br>consult on patients. | Used by clinicians when they<br>cannot be at the hospital. | Anywhere clinicians consult<br>on patients (local or remote). |
| Wireless | Yes, handheld display devices | Yes, multiple devices and<br>technologies | Yes, handheld display devices |
| Ability to<br>view near<br>real-time<br>data | Yes | Yes | Yes |
| Alerts | Software generated | No | Software generated |
| Synopsis of<br>Functionality | Store, route and display patient<br>data, generate and display<br>alarms | Remotely view near real-time<br>waveforms and patient data | Store and display patient data<br>and optionally route to EMR,<br>provide alarms set by operator |
#### 9) Conclusions:
The safety and effectiveness of the CareTaker Remote Display App have been confirmed through nonclinical testing and conformance to vital signs monitoring, usability, and risk assessment. The App was tested by connecting to a CareTaker4 Physiological Monitor Device using the CareTaker Software Library and to both a SpO2 monitor and a temperature patch using industry standard BLE Protocols to receive data. Data forwarding was tested using both WiFi and USB protocols to an external RDDS.
# Storing, Displaying, and Forwarding Patient Data
The Caretaker Remote Display App and CareTaker Software Library are safe to use as they are substantially equivalent to the predicate devices. There are no significant differences in technologies, operating principles, or intended use.
# Alarm Functions
The Caretaker Remote Display App and CareTaker Software Library are safe to use as the visual and audio alarms only indicate when a received parameter exceeds a limit set by the operator. No processing of physiological data is done by the App or library. The alarms latch to ensure they are viewed by the operator, but are not intended for time-critical intervention. The alarms are retained for review of historical data.
# Controlling the CareTaker4 Physiological Monitor
The Caretaker Remote Display App and CareTaker Software Library are safe to use as the control functions are the same as are available on the previously cleared CareTaker4 Physiological Monitor Device, K163255, and do not alter the safety or efficacy of the CareTaker4 Physiological Monitor itself. The App provides a more convenient method to control the device rather than using the 1-button interface on the device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.