Electrical Stimulator - Model R-C1 TENS and EMS stimulator, Model R-E1 EMS stimulator, Model R-T1 TENS Stimulator
Applicant
Shenzhen Roundwhale Technology Co. , Ltd.
Product Code
NUH · Neurology
Decision Date
Jun 7, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5890
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
R-C1 TENS and EMS Stimulator TENS: This mode is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities. EMS: This mode is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance. R-E1 EMS Stimulator This device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance. R-T1 TENS Stimulator This device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Device Story
RW Series A Electrical Stimulators (R-C1, R-E1, R-T1) are portable, battery-powered (6.0V DC) devices for TENS and/or EMS. Devices deliver gentle electrical current to nerves and muscles via skin-applied silica gel electrodes. User controls intensity via press buttons; LCD displays status. R-C1 combines TENS (18 programs) and EMS (15 programs); R-E1 provides EMS (15 programs); R-T1 provides TENS (18 programs). Used in home healthcare environment by patients for pain relief or muscle conditioning. Output is constant current, biphasic square waveform. Safety features include automatic overload, over-current, no-load trips, and automatic shut-off. Healthcare providers or patients use device to manage muscle pain or facilitate muscle performance, potentially reducing pain and improving muscle function.
Clinical Evidence
No clinical test data was used to support the decision of substantial equivalence. Evidence consists of bench testing, including shelf life, software validation, electromagnetic compatibility, electrical safety, and functional performance testing.
Technological Characteristics
Portable, battery-powered (4xAAA) stimulators. ABS plastic enclosure. Silica gel electrodes (50x50mm). Constant current, biphasic square waveform output. Two channels. Microprocessor-controlled. Compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, IEC 60601-2-10, and ISO 14971. Type BF applied part.
Indications for Use
Indicated for temporary relief of pain associated with sore/aching muscles in shoulder, waist, back, neck, upper/lower extremities due to strain from exercise or household work (TENS mode); indicated for stimulation of healthy muscles to improve/facilitate muscle performance (EMS mode).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services logo. The logo on the right is the FDA logo, which is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
June 7, 2018
Shenzhen Roundwhale Technology Co., Ltd. Kevin Zhang General Manager No. 615, Building C of Sanlian Industrial Zone, Shiyan, Baoan District Shenzhen, GuangDong, CN 518108
Re: K180956
Trade/Device Name: Electrical Stimulator - Model R-C1 TENS and EMS stimulator, Model R-E1 EMS stimulator, Model R-T1 TENS Stimulator Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous Electrical Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH, NGX Dated: February 9, 2018 Received: April 12, 2018
Dear Kevin Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Vivek J. Pinto -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
#### 510(k) Number (if known) K180956
#### Device Name
Electrical Stimulator - Model R-C1 TENS and EMS stimulator, Model R-E1 EMS stimulator, Model R-T1 TENS Stimulator
#### Indications for Use (Describe)
R-C1 TENS and EMS Stimulator
TENS: This mode is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
EMS: This mode is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
#### R-E1 EMS Stimulator
This device is designed to be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
#### R-T1 TENS Stimulator
This device is designed to be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
Type of Use (Select one or both, as applicable)
| <input type="checkbox"/> <span>Research Use Only (Part 301 CFR 301.3: Select One)</span> | <input checked="checked" type="checkbox"/> <span>For In-Vitro Use Only (Part 301 CFR 301.3: Select One)</span> |
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for ROOVJOY. The text "ROOVJOY" is in white, block letters against a black background. To the right of the text is a red, crescent-shaped design that partially encircles the "J" and "O" in "ROOVJOY."
Sponsor: Shenzhen Roundwhale Technology Co., Ltd File No: RW-Stimulator A-FDA-8 Date: Jun. 6, 2018
## 510(k) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92.
The assigned 510(k) number is: K180956
#### 1. Information of Submitter and Correspondent
#### Submitter's information:
| Company Name: | Shenzhen Roundwhale Technology Co., Ltd |
|------------------|------------------------------------------------------------------------|
| Street Address: | No. 615, Building C of Sanlian Industrial Zone, Shiyan, Baoan District |
| City: | Shenzhen |
| State/ Province: | Guangdong |
| Country: | China |
| Telephone: | +86(755) 23212776 |
| Fax: | +86(755) 23212776 |
| Contact Person: | Kevin Zhang |
| Contact Title: | General Manager |
| Contact Email: | info@yjing.net |
#### Submission correspondent's information:
Shenzhen Reanny Medical Devices Management Consulting Co., Ltd
Address: Room 2012#, Gebu commercial building, Hongxing community, Songgang street, Baoan district, Shenzhen 518000, China
Contact Person: Reanny Wang; E-mail: reanny@reanny.com
#### 2. Device Information
- a) Trade Name: R-C1, TENS and EMS Stimulator Common Name: Electro-Stimulator or Electrical Stimulator
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "ROOVJOY" in white letters on a black background. The letters are in a bold, sans-serif font. A red crescent shape is behind the letters "VJ" of the word. The logo is simple and modern.
| Classification Name: | Stimulator, Muscle, Powered, for muscle conditioning<br>per 21 CFR § 890.5850;<br>Transcutaneous Electrical Nerve Stimulator for Pain Relief;<br>Stimulator, Nerve, Transcutaneous, Over-the-Counter<br>per 21 CFR § 882.5890 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Class: | Class II |
| Product Code: | NUH, NGX |
| b) Trade Name: | R-E1, EMS Stimulator |
| Common Name: | Powered Muscle Stimulator, OTC |
| Classification Name: | Stimulator, Muscle, Powered, for muscle conditioning<br>per 21 CFR § 890.5850; |
| Device Class: | Class II |
| Product Code: | NGX |
| c) Trade Name: | R-T1, TENS Stimulator |
| Common Name: | TENS or TENS Device |
| Classification Name: | Transcutaneous Electrical Nerve Stimulator for Pain Relief;<br>Stimulator, Nerve, Transcutaneous, Over-the-Counter<br>per 21 CFR § 882.5890 |
| Device Class: | Class II |
| Product Code: | NUH |
## 3. Identification of Predicate Device(s)
| Manufacturer | Shenzhen Dongdixin Technology Co., Ltd |
|-------------------------|----------------------------------------|
| Legally Marketed Device | MT9001, LT3060 |
| 510 (K) Number | K130802 |
## 4. Description of Device
The RW Series A Electrical Stimulators, which includes models R-C1 TENS and EMS stimulator, R-E1 EMS stimulator and R-T1 TENS stimulator, are Transcutaneous Electrical Nerve Stimulator and muscle stimulator for pain relief and/or Electrical Muscle Stimulator. The stimulator sends gentle electrical current to underlying nerves and muscle group via electrodes applied on the skin. The parameters of units are controlled
{5}------------------------------------------------
by the press buttons. Its intensity level is adjustable according to the needs of patients.
The device unit of R-C1 was portable device, battery powered (6.0V DC) multi-function device offering both Transcutaneous Electrical Nerve Stimulator(TENS) and Powered Muscle Stimulator (EMS) qualities in one device.
The electrode size is 50*50mm, the manufacturer is Top-Rank Health care Equipment Co., Ltd. The electrode is OTC use and cleared with 510(K), the cleared number is K132588. It has been evaluated and tested according to ISO 10993-1/-5 and -10 standards.
The three models of R-C1, R-E1 and R-T1 have the same appearance, structure, display, electrical principle and CDF, the main differences are software and color. The R-C1 have TENS mode (18 programs), EMS mode (15 programs). The R-E1 only have EMS mode (15 programs) and R-T1 only have TENS mode (18 programs). The R-C1 was combination the function of R-E1 and R-T1.
## 5. Intended Use
## R-C1, TENS and EMS Stimulator
## For TENS mode
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
#### For EMS mode
To be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
## R-E1, EMS Stimulator
To be used for stimulate healthy muscles in order to improve and facilitate muscle performance.
## R-T1, TENS Stimulator
To be used for temporary relief of pain associated with sore and aching muscles in the shoulder, waist, back, neck, upper extremities (arm), and lower extremities (leg) due to strain from exercise or normal household work activities.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image is a logo for ROOVJOY. The text "ROOVJOY" is in white, sans-serif font on a black background. To the right of the text is a red, semi-circular shape that resembles a crescent moon or a partial circle.
## 6. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows:
## 6.1 Non-clinical testing
A series of safety and performance tests were conducted on the subject device, RW series A Electrical stimulators (Model R-C1 TENS and EMS stimulator, Model R-E1 EMS stimulator and model R-T1 TENS Stimulator).
- Shelf life
- Software validation
- Electromagnetic compatibility and electrical safety
- Function test
All the test results demonstrate RW series A Electrical stimulators (Model R-C1 TENS and EMS stimulator, Model R-E1 EMS stimulator and model R-T1 TENS Stimulator) meets the requirements of its pre-defined acceptance criteria and intended uses, and is substantially equivalent to the predicate devices.
## 6.2 Clinical Testing
No clinical test data was used to support the decision of substantial equivalence.
## 7. Performance Summary
| Item | Description | FDA recognized<br>consensus<br>standards | Justification |
|--------------------------------|---------------------------------|------------------------------------------|---------------|
| Safety | IEC 60601-1:2005+<br>A1:2012 | Yes | Conform |
| EMC | IEC 60601-1-2:2014 | Yes | Conform |
| Home healthcare<br>environment | IEC 60601-1-11:2015 | Yes | Conform |
| Performance | IEC 60601-2-10:2012<br>+A1:2016 | Yes | Conform |
| Software | IEC 62304:2006 | Yes | Conform |
| Usability | IEC 62366-1:2015 | | |
| Risk management | ISO 14971:2007 | Yes | Conform |
The devices conform to applicable standards as follow table:
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "ROOVJOY" in white letters on a black background. There is a red semi-circle that wraps around the letters "OVJ". The font is sans-serif and appears to be bolded. The logo is simple and modern.
## 8. Comparison for Predicate Device & Subject Device
The RW series A Electrical stimulators (Model R-C1 TENS and EMS stimulator, Model R-E1 EMS stimulator and model R-T1 TENS Stimulator) submitted in this 510(k) file is substantially equivalent in intended use, design, technology/ principles of operation, materials and performance to the cleared OTC Electrical stimulator models MT9001, LT3060 (K130802). Differences between the devices cited in this section do not raise any new issue of substantial equivalence.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "ROOVJOY" in white letters on a black background. To the right of the word is a red crescent shape. The font is sans-serif and the letters are all capitalized.
| Comparison<br>item | New device | | | Predicate device | | Substantial<br>equivalence<br>determination | |
|------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------|
| | R-C1TENS and EMS<br>Stimulator | R-E1 EMS<br>Stimulator | R-T1 TENS<br>Stimulator | MT9001 | LT3060 | | |
| 510K# | | Pending | | K130802 | | N/A | |
| Manufacturer | | Shenzhen Roundwhale Technology Co., Ltd. | | | Shenzhen Dongdixin Technology Co., Ltd | | N/A |
| Intended use | For TENS: This mode is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>neck, upper extremities<br>(arm), and lower<br>extremities (leg) due to<br>strain from exercise or<br>normal household work<br>activities.<br>For EMS: This mode is<br>designed to be used for<br>stimulate healthy<br>muscles in order to<br>improve and facilitate<br>muscle performance. | This device is<br>designed to be<br>used for<br>stimulate<br>healthy muscles<br>in order to<br>improve and<br>facilitate muscle<br>performance. | This device is<br>designed to be<br>used for<br>temporary relief<br>of pain<br>associated with<br>sore and<br>aching muscles<br>in the shoulder,<br>waist, back,<br>neck, upper<br>extremities<br>(arm), and lower<br>extremities (leg)<br>due to strain<br>from exercise or<br>normal<br>household work<br>activities. | For TENS: This mode is<br>designed to be used for<br>temporary relief of pain<br>associated with sore and<br>aching muscles in the<br>shoulder, waist, back,<br>neck, upper extremities<br>(arm), and lower<br>extremities (leg) due to<br>strain from exercise or<br>normal household work<br>activities.<br>For EMS: This mode is<br>designed to be used for<br>stimulate healthy muscles<br>in order to improve and<br>facilitate muscle<br>performance. | The device is<br>designed to be<br>used for<br>temporary relief<br>of pain<br>associated with<br>sore and aching<br>muscles in the<br>shoulder, waist,<br>back, neck,<br>upper<br>extremities<br>(arm), and lower<br>extremities (leg)<br>due to strain<br>from exercise or<br>normal<br>household work<br>activities. | Same | |
| Comparison item | | New device | | | Predicate device | | Substantial<br>equivalence<br>determination |
| | | R-C1TENS and<br>EMS Stimulator | R-E1 EMS<br>Stimulator | R-T1 TENS<br>Stimulator | MT9001 | LT3060 | |
| Prescription or OTC | | OTC | OTC | OTC | OTC | OTC | Same |
| FDA product code | | NUH, NGX | NGX | NUH | NUH, NGX | NUH | Same |
| Basic technological characteristics | | | | | | | |
| Power source | | Battery powered, d.c. 6.0V, 4 X AAA batteries | | | Battery powered, d.c.<br>9.0V, one 6F22 battery | | Different but does not<br>adversely impact<br>safety and<br>effectiveness of<br>subject device |
| User interface | | By LCD display | | | By LCD display | By LCD display | Same |
| Output channel | | Two channels | | | Two channels | | Same |
| Number of output models | | TENS and EMS | EMS | TENS | TENS and<br>EMS | TENS | Same |
| Number of treatment programs | | For TENS: 18<br>For EMS: 15 | 15 | 18 | For TENS: 4;<br>For EMS: 3 | ? | N/A |
| Number of<br>output channels | Synchronous<br>or<br>Alternating? | | Alternating | | Alternating | | Same |
| Number of output channels | | 2 | | | 2 | | Same |
| Method of channel isolation | | By electrical circuit and software | | | By electrical circuit and<br>software | | Same |
| Constant Current or Constant<br>Voltage? | | Constant current | | | Constant current | | Same |
| | New device | | | Predicate device | | Substantial<br>equivalence<br>determination | |
| Comparison item | R-C1TENS and<br>EMS Stimulator | R-E1 EMS<br>Stimulator | R-T1 TENS<br>Stimulator | MT9001 | LT3060 | | |
| Constant Current or Constant<br>Voltage? | Constant current | | | Constant current | | Same | |
| Waveform | Biphasic square | | | Biphasic square | | Same | |
| Software/Firmware/<br>Microprocessor Control? | Yes | | | Yes | | Same | |
| Automatic overload trip? | Yes | | | Yes | | Same | |
| Automatic Over Current Trip? | Yes | | | Yes | | Same | |
| Automatic No Load Trip? | Yes | | | Yes | | Same | |
| Automatic shut off? | Yes | | | Yes | | Same | |
| Patient Override Control? | Yes | | | Yes | | Same | |
| Indication<br>function | On/off status? | Yes | | Yes | | Same | |
| | Low battery? | Yes | | Yes…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.