K180802 · Meta Biomed Co., Ltd. · KLE · Aug 29, 2018 · Dental
Device Facts
Record ID
K180802
Device Name
Meta P&Bond
Applicant
Meta Biomed Co., Ltd.
Product Code
KLE · Dental
Decision Date
Aug 29, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3200
Device Class
Class 2
Indications for Use
A dental adhesive formulated to adhesively bond to hard tissues of the oral cavity, enamel and dentin Dentin-enamel primer/bonding agent for direct composite restorations Indirect composite restorative luting system Porcelain veneer luting system Bonding composite to composite Bonding composite to metal/amalgam Adhesive amalgam restorations
Device Story
Meta P&Bond is a one-step, light-cured, wet-bond dental adhesive system. Applied by clinicians to prepared tooth cavities or surfaces; functions as a primer and bonding agent to improve restoration retention. Used for direct composite restorations, indirect luting, porcelain veneers, and bonding composite to metal/amalgam or other composites. Benefits include improved adhesion to hard oral tissues. Packaged with Microbrush accessories.
Clinical Evidence
Bench testing only. Biocompatibility testing performed per ISO 10993-1, -10, -11. Performance testing (visual, capacity, film thickness, solubility, ambient light sensitivity, depth of cure, bond strength) performed per ISO 4049:2009. Shelf life testing conducted.
Technological Characteristics
One-step, light-cured resin tooth bonding agent. Materials and performance tested per ISO 4049:2009. Properties include 2µm film thickness, 1.4µg/mm² solubility, 2.9mm depth of cure, and bond strengths of 17 MPa (dentin) and 31 MPa (enamel).
Indications for Use
Indicated for use as a dental adhesive for bonding to enamel and dentin, as a primer/bonding agent for direct composite restorations, for indirect composite restorative luting, porcelain veneer luting, bonding composite to composite, bonding composite to metal/amalgam, and for adhesive amalgam restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Meta Biomed, Co., Ltd. April Lee Consultant Withus Group Inc 106 Superior Irvine, California 92620
Re: K180802
Trade/Device Name: Meta P&Bond Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: March 21, 2018 Received: March 28, 2018
Dear April Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
August 29, 2018
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K180802
Device Name Meta P&Bond
Indications for Use (Describe)
- · A dental adhesive formulated to adhesively bond to hard tissues of the oral cavity, enamel and dentin
- · Dentin-enamel primer/bonding agent for direct composite restorations
- · Indirect composite restorative luting system
- · Porcelain veneer luting system
- · Bonding composite to composite
- · Bonding composite to metal/amalgam
- · Adhesive amalgam restorations
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### Submitter
META BIOMED CO., LTD. Suk Song Oh 270, Osongsaengmyeong 1-ro, Osong-eup, Heungdeok-gu, Cheongju-si, Chungcheongbuk-do, South Korea Email: ef1459@metabiogw.bizmeka.com Phone: +82-43-230-8841 Fax: +82-43-217-1983
# Device Information
- Trade Name: Meta P&Bond ●
- . Classification Name: agent, tooth bonding, resin
- Product Code: KLE
- Panel: Dental
- Regulation Number: 21 CFR 872.3200
- Device Class: Class II
- Date prepared: 08/22/2018 ●
#### Predicate Devices:
- K081913, Meta P&Bond manufactured by Meta Biomed Co., Ltd.
#### Device Description
Meta P&Bond is a wet bond adhesive system that is activated by visible light curing. This bonding agent is one-step for priming and bonding. It is formulated to adhesively bond to hard surfaces of the oral cavity, namely enamel and dentin, for use with dental cements. It is useful as a primer for use with other methacrylate cements. It is intended to be painted on the interior of a prepared cavity or surface of a tooth to improve retention of a restoration such as a filling or crown. It also is useful as a prime and bonder for porcelain veneer luting, bonding composite, and composite to metal/amalgam. Meta P&Bond is packaged with Microbrush accessories.
# Indication for Use
- . A dental adhesive formulated to adhesively bond to hard tissues of the oral cavity, enamel and dentin
- Dentin-enamel primer/bonding agent for direct composite restorations
- Indirect composite restorative luting system
- Porcelain veneer luting system
- Bonding composite to composite
- Bonding composite to metal/amalgam
- Adhesive amalgam restorations
# Official Correspondent
Withus Group Inc April Lee 106 Superior, Irvine, CA 92620 USA Email: withus6664@gmail.com Phone: 1-909-274-9971 Fax: 1-909-460-8122
{4}------------------------------------------------
# Non-clinical Testing
The subject device was tests in accordance to the following standards:
- Biocompatibility Tests according to ISO 10993-1:2009, ISO 10993-10:2010, ISO 10993-11:2006.
- Performance tests such as visual, capacity, package, film thickness, solubility, sensitivity to ambient light, Depth of cure, Bone strength (Dentin, Enamel) according to ISO 4049:2009.
- Shelf Life tests according to ISO4049:2009.
# Summary of Technological Characteristics:
The subject device and predicate device have same intended use and principle of operation and similar technological characteristics such as capacity, film thickness, sensitivity to ambient light, depth of cure, bond strength and shelf life.
The difference between the subject and predicate device is solubility. As the solubility value of the subject device is within the range that ISO 4049:2009, it doesn't affect safety and effectiveness.
| | Subject Device | Predicate Device |
|---------------------|---------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | META BIOMED CO., LTD. | META BIOMED CO., LTD. |
| Device Name | Meta P&Bond | Meta P&Bond |
| 510(k) Number | K180802 | K081913 |
| Classification Name | Resin tooth bonding agent | Resin tooth bonding agent |
| Product Code | KLE | KLE |
| Regulation Number | 21 CFR 872.3200 | 21 CFR 872.3200 |
| Intended Use | Identical | ● A dental adhesive formulated<br>to adhesively bond to hard<br>tissues of the oral cavity,<br>enamel and dentin<br>● Dentin-enamel primer/bonding<br>agent for direct composite<br>restorations<br>● Indirect composite restorative<br>luting system<br>● Porcelain veneer luting system<br>● Bonding composite to<br>composite |
{5}------------------------------------------------
| | | Bonding composite to metal/amalgam Adhesive amalgam restorations |
|---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Principle of operation | Meta P&Bond adhesive system is<br>wet bond adhesive system that is<br>activated by visible light curing.<br>This bonding agent is one step for<br>priming and bonding. | Meta P&Bond adhesive system is wet<br>bond adhesive system that is activated<br>by visible light curing. This bonding<br>agent is one step for priming and<br>bonding. |
| Capacity | 1.37% | 2.15% |
| Film thickness | $2 \mu m$ | $1 \mu m$ |
| Solubility | $1.4 \mu g/mm^2$ | $7.5 \mu g/mm^2$ |
| Sensitivity to Ambient<br>Light | Material remained physically<br>homogeneous | Material remained physically<br>homogeneous |
| Depth of Cure | 2.9 mm | 3.53 mm |
| Bone Strength, Dentine | 17 MPa | 10 MPa |
| Bone Strength, Enamel | 31 MPa | 30.76 MPa |
| Shelf Life | 2 Years | 2 Years |
# Conclusion:
Metabiomed Co., Ltd believes that Meta P&Bond is substantially equivalent to the currently legally marketed product based on similar intended use, physical, chemical and mechanical properties. Any differences do not raise different questions of safety and effectiveness than the predicate. These differences therefore, do not render the new device NSE in comparison to the predicate.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.