1) Direct bonding for light-cured composites to tooth surface 2) Bonding of dual-cured core build up composites to tooth surface 3) Intraoral repair of composite, PFM and ceramic restoration using cements 4) Sealing of tooth preparation for indirect restoration
Device Story
Bright Bond Universal is a light-cured dental adhesive; combines etching, priming, and bonding functions in a single-bottle liquid solution. Used by dental professionals in clinical settings to facilitate bonding between dentin/enamel and restorative materials. Applied directly to tooth surfaces; light-cured to achieve polymerization. Provides strong, durable bonds for direct and indirect restorations; seals tooth preparations. Benefits include simplified clinical workflow and improved restoration retention.
Clinical Evidence
Bench testing only. No clinical data provided. Performance evaluated via ISO 10993 biocompatibility testing (cytotoxicity, irritation, sensitization, systemic toxicity, genotoxicity, implantation, chronic toxicity) and ISO 4049/ISO 29022 bench testing (film thickness, depth of cure, bonding strength). Results: film thickness 2.0 µm (≤50 µm limit), depth of cure 3.0 mm (≥1.5 mm limit), dentin bonding strength 16 MPa (≥8 MPa limit), enamel bonding strength 14 MPa (≥8 MPa limit).
Technological Characteristics
Single-bottle liquid dental adhesive. Composition: Multifunctional methacrylate resins (HEMA, 10-MDP). Non-sterile. Shelf-life: 2 years. Performance specs: pH 2.1, film thickness 2.0 µm, depth of cure 3.0 mm. Complies with ISO 4049:2009 and ISO 29022:2013 standards.
Indications for Use
Indicated for dental patients requiring direct bonding of light-cured composites, bonding of dual-cured core build-up composites, intraoral repair of composite, PFM, and ceramic restorations, and sealing of tooth preparations for indirect restorations.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
October 28, 2020
GENOSS Co., Ltd. Hong Jeon Manager 1F, Gyeonggi R&DB Center, 105 Gwanggyo-ro, Yeongtong-gu, Suwon-si Suwon-si. Gyeonggi-do 16229 SOUTH KOREA
Re: K200153
Trade/Device Name: Bright Bond Universal Regulation Number: 21 CFR 872.3200 Regulation Name: Resin tooth bonding agent Regulatory Class: Class II Product Code: KLE Dated: August 3, 2020 Received: August 3, 2020
Dear Hong Jeon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{2}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for GENOSS. The word "GENOSS" is written in a sans-serif font, with the first three letters in a dark green color and the last three letters in a lighter green color. The letters are connected to each other, and the logo has a modern and clean look.
## Indication for use
K200153 510(k) Number:
Device Name: Bright Bond Universal
#### Indication for use:
- 1) Direct bonding for light-cured composites to tooth surface
- 2) Bonding of dual-cured core build up composites to tooth surface
- 3) Intraoral repair of composite, PFM and ceramic restoration using cements
- 4) Sealing of tooth preparation for indirect restoration
Prescription Use __ √________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for GENOSS. The word "GENOSS" is written in a sans-serif font. The letters "GEN" are in a dark green color, while the letters "OSS" are in a lighter green color. The logo has a modern and clean design.
## 510(k) Summary K200153
12/20/2019
## 1. Company
| | Submitter |
|----------------|----------------------------------------------------------------------------------------------------------|
| Name | GENOSS Co., Ltd. |
| Address | 1F, Gyeonggi R&DB Center / 226, 2F, GSBC, 105<br>Gwanggyo-ro, Yeongtong-gu, Suwon-si, Gyeonggi-do, Korea |
| Phone/Fax | +82-70-7098-7541/ +82-31-888-5105 |
| Contact person | HongJun Jeon / RA<br>hjjeon@genoss.com |
| Summary Date | 12/20/2019 |
## 2. Device Name
Proprietary name: Bright Bond Universal Regulation description : Resin tooth bonding agent Classification name: Agent, Tooth Bonding, Resin
#### 3. Predicate Device
K110302 Scotchbond Universal
## 4. Description
Bright Bond Universal is a light-cured dental adhesive that combines etching, priming, and bonding functions in a single bottle solution. It provides strong and durable bonds between dentin/enamel and light-curable direct/indirect restorative materials.
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first four letters, "GENO", are in a dark green color, while the last three letters, "OSS", are in a lighter green color. The letters are connected to each other, and the overall design is modern and sleek.
#### 5. Indication for use
- 1) Direct bonding for light-cured composites to tooth surface
- 2) Bonding of dual-cured core build up composites to tooth surface
- 3) Intraoral repair of composite, PFM and ceramic restoration using cements
- 4) Sealing of tooth preparation for indirect restoration
#### 6. Technological Characteristics
Bright Bond Universal has the similar technological characteristics as the predicate device; main material, indication for use and design. Technological characteristics of Bright Bond Universal, Scotchbond Universal are as following
| Device name | | Bright Bond Universal | Scotchbond Universal |
|----------------------|------------|-------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | | Genoss Co., Ltd. | 3M ESPE |
| 510(k) Number | | New Device | K110302 |
| Target population | | Human tooth<br>(Dentin, Enamel) | Human tooth<br>(Dentin, Enamel) |
| Purpose | Clinical | Restoration | Restoration |
| Site of application | | Dentin, Enamel | Dentin, Enamel |
| Clinical Performance | | Strong bonding with composite resin | Strong bonding with composite resin |
| Materials | Biological | Multifunctional methacrylate resins<br>(HEMA, 10-MDP) | Multifunctional methacrylate resins<br>(HEMA, 10-MDP) |
| Chemical Safety | | Biocompatible | Biocompatible |
| Sterilization | | Non-sterile | Non-sterile |
| Shelf-Life | | 2 years | 2 years |
| Form | Technical | Liquid | Liquid |
| Indication for use | | 1) Direct bonding for light-cured<br>composites to tooth surface<br>2) Bonding of dual-cured core build<br>up composites to tooth surface | 1) Bonding light-cured composite or<br>compomer for all classes of direct<br>restorations<br>2) Bonding of dual-cure cements and<br>core build-up materials<br>3) Bonding of self-cure composites |
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for GENOSS. The logo is written in a sans-serif font, with the letters "GENO" in a dark green color and the letters "SS" in a lighter green color. The logo is simple and modern, and the colors are eye-catching.
| | and ceramic restoration using cements | 5) Sealing of dentin prior to amalgam<br>restorations | |
|------------------|-------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|--|
| | 4) Sealing of tooth preparation for<br>indirect restoration | 6) Protective varnish for glass ionomer<br>restorative materials<br>7) Repair of composite or compomer<br>restorations<br>8) Bonding sealants | |
| | | 9) Bonding veneers in combination<br>with 3MTM RelyXTM Veneer Cement | |
| Use | Prescription | Prescription | |
| pH | 2.1 | 2.7 | |
| Film thickness | 2.0 µm | 10.6 µm | |
| Depth of cure | 3.0 mm | 2.8 mm | |
| Bonding Strength | 24.5 Mpa | 22.0 MPa | |
| | | | |
## 7. Performance Data
Biocompatibility testing on the proposed Bright Bond Universal has been completed. Requirements for biological evaluation of the proposed device were based on FDA recognized concensus standard of ISO10993, "Biological Evaluation of Medical Devices, Part 1: Evaluation and Testing." The biocompatibility test results show that the materials used in the design and manufacture of the components of the proposed device are non-toxic and nonsensitizing to biological bone and tissues with its intended use. The following biocompatibility tests were completed:
(P: Pass, F: Fail)
| No. | Test | Method | Acceptance criteria | P/F | Report No. |
|-----|-------------------------|--------------|------------------------|-----|---------------------|
| 1 | Cytotoxicity | ISO 10993-5 | None cytotoxicity | P | CDM-18-0161-11 |
| 2 | Irritation | ISO 10993-10 | None oral irritation | P | CDM-18-0161-12 |
| 3 | Sensitization | ISO 10993-10 | None sensitization | P | MTK-2019-<br>000014 |
| 4 | Acute systemic toxicity | ISO 10993-11 | None systemic toxicity | P | CDM-18-0161-13 |
#### 510(k) Summary
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color, while the last three letters, "OSS", are in a lighter green color. The font is sans-serif and has a modern, slightly rounded appearance. The letters are connected to each other, creating a sense of unity and flow.
| 5 | Genotoxocity | ISO 10993-3 | None genotoxicity | P | BER-18-056 |
|---|------------------|--------------|---------------------|---|------------|
| 6 | Implantation | ISO 10993-6 | Biocompatible | P | BER-18-056 |
| 7 | Chronic toxicity | ISO 10993-11 | No chronic toxicity | P | BER-18-056 |
The proposed Bright Bond Universal was evaluated using the following performance bench testing to confirm the performance characteristics:
| No. | Items | Method | Acceptance Criteria | Result | File No. |
|-----|---------------------------------|-----------------------|---------------------------------------------------------|------------------------------------------|--------------------|
| 1 | Visual test | Bare eyes | No impurities and<br>No specific changes | No impurities and<br>No specific changes | CDM-18-0161-<br>01 |
| 2 | Capacity test | Weight difference | Capacity error of;<br>Standard Capacity<br>$\u00b1 5 %$ | 2.75 % | CDM-18-0161-<br>02 |
| 3 | Package test | Bare eyes | No damage | No damages, cracks | CDM-18-0161-<br>03 |
| 4 | Film Thickness | ISO 4049:2009<br>7.5 | $\u2264 50 \u00b5m$ | $2 \u00b5m$ | CDM-18-0161-<br>04 |
| 5 | Sensitivity to<br>Ambient Light | ISO 4049:2009<br>7.9 | Must be physically<br>uniform | Homogeneous | CDM-18-0161-<br>05 |
| 6 | Depth of Cure | ISO 4049:2009<br>7.10 | $\u2265 1.5 mm$ | 3.0 mm | CDM-18-0161-<br>06 |
| 7 | Bonding<br>Strength<br>(Dentin) | ISO 29022:2013<br>7 | $\u2265 8 MPa$ | 16 MPa | CDM-18-0161-<br>09 |
| 8 | Bonding<br>Strength<br>(Enamel) | ISO 29022:2013<br>7 | $\u2265 8 MPa$ | 14 MPa | CDM-18-0161-<br>10 |
All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Bright Bond Universal met the established specifications necessary for consistent performance according to its intended use.
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "GENOSS" in a stylized font. The first three letters, "GEN", are in a dark green color, while the last three letters, "OSS", are in a lighter, lime green color. The letters are bold and slightly slanted, giving the word a dynamic appearance. The logo appears to be for a company or organization named "GENOSS".
## 8. Conclusion
Based on the information provided in this premarket notification of GENOSS Co., Ltd. Concludes that Bright Bond Universal is acceptable and safe, substantially equivalent to predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.