K180479 · Sensible Medical Innovations , Ltd. · DSB · Feb 28, 2019 · Cardiovascular
Device Facts
Record ID
K180479
Device Name
ReDS System
Applicant
Sensible Medical Innovations , Ltd.
Product Code
DSB · Cardiovascular
Decision Date
Feb 28, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
The ReDS System is intended for use by qualified healthcare practitioners, under the direction of a physician, in hospitals, hospital-type facilities and home environments, for the non-invasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications. The ReDS System is indicated for patients: With fluid management problems Taking diuretic medication Living with heart failure Recovering from a coronary artery disease-related event
Device Story
ReDS System is a non-invasive thoracic base wave impedance monitor measuring lung fluid content. Device emits low-power electromagnetic (EM) signals (1-2 GHz) into the body; sensors intercept signals to determine tissue conductivity and capacitance. System comprises a bedside console with touch-screen display and a clip-like sensor unit attached to the patient's thorax. Used in hospitals, hospital-type facilities, and home environments by healthcare practitioners. Software manages device, analyzes signals, and provides graphical display/reporting. Cloud application enables remote patient monitoring. Healthcare providers use output to assess fluid status, adjust diuretic therapy, and manage heart failure or post-coronary event recovery. Benefits include non-invasive, objective fluid monitoring to guide clinical decision-making.
Clinical Evidence
Clinical data compared ReDS V2.7 to ReDS V2.6. Results demonstrated equivalent performance between generations, consistent with non-clinical findings. No device-related adverse events reported. Bench testing verified mechanical durability, packaging, cleaning/disinfection, and usability. Biocompatibility confirmed per ISO 10993-5 and 10993-10. Electrical safety and EMC verified per IEC 60601 series.
Technological Characteristics
Non-invasive thoracic impedance monitor. Energy source: 1-2 GHz electromagnetic fields. Components: Bedside console (embedded computer, touch-screen) and clip-like sensor unit. Connectivity: WiFi enabled. Software: Signal analysis and management. Biocompatibility: ISO 10993 compliant. Standards: IEC 60601-1, 60601-1-11, 60601-1-2, 60601-1-6, ISO 14971:2007.
Indications for Use
Indicated for patients with fluid management problems, including those taking diuretic medication, living with heart failure, or recovering from a coronary artery disease-related event.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
February 28, 2019
Sensible Medical Innovations Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1735 Market Street, Floor 23 Philadelphia, Pennsylvania 19103
Re: K180479
Trade/Device Name: ReDS System Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Code: DSB Dated: February 25, 2019 Received: February 25, 2019
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Stephen C. Browning -S5
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K180479
Device Name ReDS System
#### Indications for Use (Describe)
The ReDS System is intended for use by qualified healthcare practitioners, under the direction of a physician, in hospitals, hospital-type facilities and home environments, for the non-invasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications. The ReDS System is indicated for patients:
- With fluid management problems
- Taking diuretic medication
- · Living with heart failure
- · Recovering from a coronary artery disease-related event
Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(K) Summary
#### Update date : February 9, 2019
## ReDS System V2.7
## 510(k) Number K180479
## Applicant's Name:
Sensible Medical Innovations Meir Ariel 6, Netanya, 4059300 POB 8702 Israel Tel: 972-9-8654402, Fax: 972-9-8654472
## Contact Person:
Name: Shlomi Bergida Title: VP Product Tel: Office (+972) 9-8654402, Mobile (+972) 52-6584101 Fax: (+972) 9-8654472 Email: shlomi.b@sensible-medical.com
#### Device Trade Name:
ReDS System
#### Common/Classification Name:
Impedance plethysmograph
#### Classification:
FDA has classified impedance plethysmographs as Class II devices (product code DSB, 21 CFR 870.2770), and they are reviewed by the Cardiovascular panel.
## Predicate Devices:
- > ReDS Wearable System V2.6 (K150095)
{4}------------------------------------------------
## Indications for Use:
The ReDS System is intended for use by qualified health care practitioners, under the direction of a physician, in hospitals, hospital-type facilities and home environments, for the non-invasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
The ReDS System is indicated for patients:
- With fluid management problems ●
- . Taking diuretic medication
- Living with heart failure .
- . Recovering from a coronary artery disease-related event
#### Device Description:
The ReDS System ("ReDS" or "ReDS System") is a non-invasive thoracic base wave impedance monitor that provides measurement of patient lung fluid content. ReDS™ technology measures wave impedance of biological tissues. Low power electromagnetic (EM) signals are emitted into the body, and intercepted by sensors on the body. The wave impedance is a function of the distributed tissue conductivity and capacitance properties, and is representative of tissue fluid content.
The system consists of a Sensor Unit connected to a bedside console, as well as a cloud application. The Sensor Unit encases two sensors and an attachment mechanism. The Bedside Console is an enclosure housing the electronic modules, an embedded computer and a touch-screen display. The device software provides the management of the device as well as analysis of the measured signals, graphic display of readings, reporting, and communications functionality to enable remote patient monitoring.
#### Technological Characteristics:
The ReDS System's technological characteristics are substantially equivalent to those of its predicate device, i.e., the first generation ReDS System V2.6. Both devices are non-invasive, prescription use, transportable bedside devices indicated for use in hospital and hospital-like environments by healthcare providers. Same as the predicate, ReDS V2.7 uses non-invasive measurements of impedance characteristics to assess the amount of fluid in tissue. Both devices use sensors placed on the user's body and measurement initiation through a graphical user interface. ReDS V2.7 sensors are located in a clip-like sensor unit, whereas the predicate device use sensors embedded in the wearable vest. The minor differences noted above between ReDS V2.7 and the ReDs V2.6 predicate device
{5}------------------------------------------------
do not raise new types of safety and effectiveness questions. These differences have been assessed in bench, human factors, and clinical testing. Results established that the ReDS System V2.7 performs as intended and is substantially equivalent to its predicate device.
#### Performance Data:
ReDS System V2.7 was evaluated in non-clinical and clinical testing. Results demonstrated that the device meets specifications and supported substantial equivalence to the predicate device. The device is not provided sterile and does not require end user sterilization.
#### Biocompatibility
ISO 10993 testing demonstrated biocompatibility of the device materials. Cytotoxicity (ISO 10993-5), skin irritation (ISO 10993-10), and sensitization (ISO 10993-10) testing demonstrated all passing results, supporting the biocompatibility of the device for its intended use.
Software
Software validation and verification testing was conducted for the ReDS System software. Results demonstrated that the software was appropriate for release. The software hazard analysis was performed in accordance with ISO 14971:2007.
#### Electrical Safety and Electromagnetic Compatibility
Electrical safety and electromagnetic compatibility testing were conducted. Results demonstrated that the system complies with the applicable testing standards (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-1-6, and IEC 62366).
#### Bench Testing
Bench testing, demonstrating substantial equivalence to the predicate device, provided verification and validation of mechanical durability, packaging and transportation, quality feedback mechanisms, and cleaning and disinfection. Usability testing was also conducted to validate use of the system by the intended professional users. All performance testing passed according to defined acceptance criteria, demonstrating that the device performs as expected.
#### Clinical Data
Clinical data was provided to further assess the ReDS System V2.7 equivalency to ReDS System V2.6. Results showed equivalent performance of the two ReDS System generations. The equivalency presented in the clinical setting was consistent with the findings of the nonclinical studies. No device-related adverse events were reported.
{6}------------------------------------------------
#### Conclusion:
ReDS System V2.7 has the same intended use and similar indications for use, technological characteristics, and principles of operation as the predicate device, ReDS System V2.6. Any minor differences between the two generation of devices do not raise any new questions of safety or effectiveness. Performance tests, confirm that these differences do not adversely impact safety or performance. In summary, the conclusions from the non-clinical tests demonstrate that ReDS System V2.7 is as safe, as effective, and performs as well as or better than the legally marketed predicate device, ReDS System V2.6.
{7}------------------------------------------------
## Substantial Equivalence Comparison Table
| Criteria | ReDS V2.6 (As Cleared In K150095) | ReDS V2.7 (K180479) |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Sensible Medical Innovations Ltd. | Sensible Medical Innovations Ltd. |
| K Number | K150095 | K180479 |
| Product Code | DSB | DSB |
| Classification | Class II | Class II |
| Product<br>Classification/<br>Common<br>Name | Plethysmograph/Fluid Status Monitor | Same |
| Indications for<br>Use | ReDS is intended for use by qualified health care<br>practitioners and by patients, under the direction of a<br>physician, in hospitals, hospital-type facilities and home<br>environment, for the non-invasive monitoring and<br>management of patients with fluid management<br>problems in a variety of medically accepted clinical<br>applications.<br>ReDS is indicated for patients:<br>◇ With fluid management problems<br>◇ Taking diuretic medication<br>◇ Living with Heart Failure<br>◇ Recovering from Coronary Artery Disease related<br>event | The ReDS System is intended for use by qualified<br>health care practitioners, under the direction of a<br>physician, in hospitals, hospital-type facilities and<br>home environments, for the non-invasive monitoring<br>and management of patients with fluid management<br>problems in a variety of medically accepted clinical<br>applications.<br>The ReDS System is indicated for patients:<br>◇ With fluid management problems<br>◇ Taking diuretic medication<br>◇ Living with heart failure<br>◇ Recovering from a coronary artery disease-related<br>event |
| Criteria | ReDS V2.6 (As Cleared In K150095) | ReDS V2.7 (K180479) |
| Range of<br>Patient<br>Anatomies<br>(Height, BMI) | ◊ The system is suitable for both male and female<br>patients with a body mass index of 22 to 36 and a<br>height of between 155 cm (5' 1'') to 190 cm (6' 3''). | ◊ The system is suitable for both male and female<br>patients with a body mass index of 22 to 36 and a<br>height between 155 cm (5' 1'') to 195 cm (6' 5'').<br>◊ Patients within the height range having a BMI of 36<br>to 38 can use the system if their chest size ruler<br>value (as measured by the unit) is 39 or less. |
| Rx Device | Yes | Yes |
| Power Source | AC powered | Same |
| Technology | Induced electromagnetic fields are used to measure<br>wave impedance of the thorax. | Same |
| Energy Type | Electromagnetic fields at 1-2 Ghz | Same |
| Device Display<br>Parameter<br>(Fluid) | Measurement of fluid status by:<br>◊ Base impedance in Ohm (Ω) units (70–150Ω) or<br>◊ Volume percent (%) units (15-60%) | Measurement of fluid status by:<br>◊ Volume percent (%) units (15-60%) (base<br>impedance option no longer presented) |
| Sensor<br>Attachment<br>Configuration | ◊ 2 non-touch sensors embedded in a wearable vest<br>attached to the patient thorax on the front and back<br>of chest<br>◊ Vest is adjustable for patient dimensions | ◊ 2 non-touch sensors embedded in a clip-like sensor<br>unit attached to the patient thorax on the front and<br>back of chest |
| Criteria | ReDS V2.6 (As Cleared In K150095) | ReDS V2.7 (K180479) |
| | ◊ Bedside console connected to the patient sensor vest via a cable | ◊ Unit is adjustable for patient dimensions<br>◊ Bedside console connected to the clip-like sensor unit via a cable |
| Main System Components - Electronics | RF Signal generator and analyzer | Same with added WiFi connectivity |
| Software | V2.6 Software | V2.6 Software with the following main additions:<br>◊ User interface providing Sensor Unit use instructions<br>◊ Multi-patient management with anonymized ID option<br>◊ Historical readings screen |
| Cloud Application | V2.6 SensiCloud application | V2.6 SensiCloud application with the following main additional features:<br>◊ User interface enhancements<br>◊ HCP option to set thresholds range and receive out of range notices.<br>◊ Patient's measurements graph UI improvements (e.g. a green zone was added on the graph marking the center range)<br>◊ Task management indicators for administrative users |
| User Interface | 10" Touch screen graphical user interface and display | Same |
| Algorithm | ◊ Calculation of thorax impedance from measured signals | ◊ Same with a stricter signal quality assessment mechanism |
| Criteria | ReDS V2.6 (As Cleared In K150095) | ReDS V2.7 (K180479) |
| | ◇ Conversion of impedance measurements to present<br>fluid content in volume percentage units<br>◇ Assessment of signal quality | |
| Mode of Use | ◇ Intermittent use | ◇ Intermittent use |
| | ◇ Application on a single patient | ◇ Application on single and multiple patient |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.