K150095 · Sensible Medical Innovations , Ltd. · DSB · Aug 6, 2015 · Cardiovascular
Device Facts
Record ID
K150095
Device Name
ReDs Wearable System
Applicant
Sensible Medical Innovations , Ltd.
Product Code
DSB · Cardiovascular
Decision Date
Aug 6, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2770
Device Class
Class 2
Indications for Use
ReDS is intended for use by qualified health care practitioners and by patients, under the direction of a physician, in hospitals, hospital-type facilities and home environment, for the non-invasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
Device Story
ReDS Wearable System is a non-invasive thoracic impedance monitor measuring lung fluid content. System comprises a wearable vest with embedded sensors, a bedside console with touch-screen display, and a cloud application. Principle of operation: emission of low-power electromagnetic (EM) signals into the body; sensors intercept signals; wave impedance calculated as function of tissue conductivity and capacitance, representing fluid content. Used in hospitals, hospital-type facilities, and home environments by healthcare providers or patients (self-use). Console processes signals, displays readings, and enables remote monitoring. Output assists clinicians in managing fluid status, potentially improving outcomes for heart failure and CAD patients.
Clinical Evidence
Evaluated via non-clinical bench testing, porcine animal model, and clinical studies. Clinical studies assessed pulmonary fluid quantification across a range of congestion levels (normal to highly congested). Results were consistent with non-clinical findings; no device-related adverse events reported. Bench testing verified accuracy, reproducibility, mechanical durability, and EM signal specifications. Biocompatibility confirmed per ISO 10993-5 and ISO 10993-10.
Technological Characteristics
Non-invasive thoracic impedance plethysmograph. Components: wearable vest with embedded sensors, bedside console with embedded computer. Energy source: low-power electromagnetic signals. Connectivity: bedside console with cloud application for remote monitoring. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-1-6, IEC 62366, ISO 10993, ISO 14971.
Indications for Use
Indicated for patients with fluid management problems, those taking diuretic medication, patients living with Heart Failure, and patients recovering from a Coronary Artery Disease related event.
Regulatory Classification
Identification
An impedance plethysmograph is a device used to estimate peripheral blood flow by measuring electrical impedance changes in a region of the body such as the arms and legs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a body composition analyzer which is not intended to diagnose or treat any medical condition, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird in flight, composed of three overlapping profiles facing to the right. The profiles are arranged in a way that creates the impression of movement or flight. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2015
Sensible Medical Innovations Ltd. % Janice Hogan Partner Hogan Lovells US LLP 1835 Market St Suite 2920 Philadelphia, Pennsylvania 19103
Re: K150095
Trade/Device Name: ReDSTM Wearable System Regulation Number: 21 CFR 870.2770 Regulation Name: Impedance Plethysmograph Regulatory Class: Class II Product Code: DSB Dated: July 8, 2015 Received: July 8, 2015
Dear Janice Hogan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Bram D. Zuckerman -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K150095
Device Name
ReDS™ Wearable System
Indications for Use (Describe)
ReDS is intended for use by qualified health care practitioners and by patients, under the direction of a physician, in hospitals, hospital-type facilities and home environment, for the non-invasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
ReDS is indicated for patients:
- · With fluid management problems
- · Taking diuretic medication
- · Living with Heart Failure
- Recovering from Coronary Artery Disease related event .
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
{3}------------------------------------------------
# 510(K) Summary DATE PREPARED: July 24, 2015 ReDS Wearable System 510(k) Number: K150095
# Applicant's Name:
Sensible Medical Innovations HaAlon 4 Kfar Neter, 4059300 POB 8702, Netanya Israel Tel: 972-9-8654402, Fax: 972-9-8654472
# Contact Person:
Name: Inbal Ben-Tzvi Title: VP Requlatory Affairs Tel: Office (+972) 9-8654402 (x 103), Mobile (+972) 54-2445822 Fax: (+972) 9-8654472 Email: Inbal.b@sensible-medical.com
#### Device Trade Name:
ReDS Wearable System
# Common/Classification Name:
Impedance plethysmograph
## Classification:
FDA has classified impedance plethysmographs as Class II devices (product code DSB, 21 CFR 870.2770), and they are reviewed by the Cardiovascular panel.
{4}------------------------------------------------
# Predicate Devices:
- > ZOE Fluid Status Monitor (NMT Medical), product code DSB, cleared for marketing under K131509, K112830, K042113, K133301 (primary predicate)
- > Aesculon (Osypka Medical), product code DSB, cleared for marketing under K070985, K081035
#### Indications for Use:
ReDS is intended for use by qualified health care practitioners and by patients, under the direction of a physician, in hospital-type facilities and home environment, for the non-invasive monitoring and management of patients with fluid management problems in a variety of medically accepted clinical applications.
ReDS is indicated for patients:
- . With fluid management problems
- Taking diuretic medication ●
- Living with Heart Failure
- Recovering from Coronary Artery Disease related event .
# Device Description:
The ReDS Wearable System ("ReDS") is a non-invasive thoracic base wave impedance monitor that provides measurement of patient lung fluid content. ReDS™ technology measures wave impedance of biological tissues. Low power electromagnetic (EM) signals are emitted into the body, and intercepted by sensors on the body. The wave impedance is a function of the distributed tissue conductivity and capacitance properties, and is representative of tissue fluid content.
The system consists of a Wearable Vest connected to a bedside console, as well as a cloud application. The Wearable Vest encases two sensors and an attachment mechanism. The Bedside Console is an enclosure housing the electronic modules, an embedded computer and a touch-screen display. The device software provides the management of the device as well as analysis of the measured signals, graphic display of readings, reporting, and communications functionality to enable remote patient monitoring.
# Technoloqical Characteristics:
The ReDS Wearable System's technological characteristics are substantially equivalent to those of its predicate devices. All of the devices are non-invasive, prescription use, transportable bedside devices indicated for use in hospital and hospital-like environments by healthcare providers, and both the ReDS and the Aesculon are also indicated for use in the home environment by patients for selfmeasuring. Same as both predicates, ReDS uses non-invasive measurements of impedance characteristics to assess the amount of fluid in tissue. All of the devices use sensing elements placed on the user's body and measurement
{5}------------------------------------------------
initiation through a graphical user interface. ReDS sensors are embedded in the wearable vest, whereas the predicate devices use sensors adhered to the body. The minor differences between the ReDS and the predicate devices, e.g., range of parameters, number of sensors, and the sensor placement methods, do not raise new types of safety and effectiveness questions. These differences have been assessed in bench, preclinical, and clinical testing. Results established that the ReDS System performs as intended and is substantially equivalent to its predicate devices.
#### Performance Data:
The ReDS Wearable System was evaluated in non-clinical, and clinical testing. Results demonstrated that the device meets specifications and supported substantial equivalence to the predicate devices. The device is not provided sterile and does not require end user sterilization.
#### Biocompatibility
ISO 10993 testing demonstrated biocompatibility of the device materials. Cytotoxicity (ISO 10993-5), intracutaneous (ISO 10993-10), and sensitization (ISO 10993-10) testing demonstrated all passing results, supporting the biocompatibility of the device for its intended use.
#### Software
Software validation and verification testing was conducted for the ReDS System software. Results demonstrated that the software was appropriate for release. The software hazard analysis was performed in accordance with ISO 14971:2007.
#### Electrical Safety and Electromagnetic Compatibility
Electrical safety and electromagnetic compatibility testing were conducted. Results demonstrated that the system complies with the applicable testing standards (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-1-6, and IEC 62366).
#### Bench Testing
Bench testing, demonstrating substantial equivalence to the predicate display provided devices, reproducibility and accuracy), mechanical durability and EM signal specifications verification. Usability testing was also conducted to validate use of the system by the intended professional and lay users. All performance testing passed according to defined acceptance criteria, demonstrating that the device performs as expected.
#### Animal Study
The ReDS System was also evaluated in a porcine animal model to demonstrate the ReDS accuracy in quantifying lung fluid content.
{6}------------------------------------------------
#### Clinical Data
Clinical studies were conducted to provide preliminary clinical performance data of the ReDS technology for the quantification of pulmonary fluid level in patients presenting various fluid content levels across the operational range (including normal to highly congested patients). No device-related adverse events were reported. The performance of ReDS in the clinical setting was consistent with the findings of the nonclinical studies.
# Conclusion:
The ReDS Wearable System has the same intended use and similar indications for use, technological characteristics and principles of operation as the predicate devices. Any minor differences in the ReDS System compared to the predicate devices do not raise any new questions of safety or effectiveness. Performance tests, including bench studies, animal testing and clinical studies, have been conducted to confirm that these differences do not adversely impact safety or performance. In sum, the conclusions from the non-clinical and clinical tests demonstrate that the ReDS Wearable System performs similarly to the leqally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.