K180345 · Urogen Pharma, Ltd. · QBL · Oct 17, 2018 · General Hospital
Device Facts
Record ID
K180345
Device Name
Uroject12 Syringe Lever
Applicant
Urogen Pharma, Ltd.
Product Code
QBL · General Hospital
Decision Date
Oct 17, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Uroject12 Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.
Device Story
Uroject12 Syringe Lever is a reusable, manually operated device for controlled administration of sterile materials from a standard 20 mL syringe. Device consists of a syringe housing and a rotary dispensing mechanism; user rotates a knob to advance a clutch-equipped plunger, applying pressure to the syringe piston. Device is used in clinical urology settings by clinicians. It provides visual markings for volume monitoring during administration. Device is non-sterile and requires cleaning and sterilization before each use. Benefits include controlled, accurate fluid delivery during urological procedures.
Clinical Evidence
Bench testing only. Testing included instillation pressure and force, extreme dispensing resistance, residual volume, clutch mechanism functionality, and lifetime functionality. Cleaning validated per AAMI TIR30:2011; sterilization validated to SAL <10^-6.
Technological Characteristics
Reusable syringe housing and rotary driver mechanism. Compatible with 20 mL Medallion and Medallion COP syringes. Manual operation via rotary knob and clutch. Sterilization required before each use; cleaning validated per AAMI TIR30:2011.
Indications for Use
Indicated for clinicians in a urology setting for the administration of sterile materials using a compatible 20 mL syringe.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health & Human Services - USA logo. The logo on the right is the FDA U.S. Food & Drug Administration logo. The FDA logo is in blue.
October 17, 2018
UroGen Pharma Ltd. James G. Ottinger, R.Ph. Vice President, Regulatory Affairs 499 Park Avenue, 12th Floor, Suite 1200 New York, NY 10022
Re: K180345 Trade/Device Name: Uroject12 Syringe Lever Regulation Number: 21 CFR§ 880.5860 Regulation Name: Piston syringe Regulatory Class: II Product Code: QBL Dated: September 14, 2018 Received: September 18, 2018
Dear James G. Ottinger:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You mav, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be avare that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.html; good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark R. Kreitz -S
for
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
### K180345
Device Name Uroject12 Syringe Lever
Indications for Use (Describe)
The Uroject 12 Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M5.41666 12.5L15 2.91669L13.0833 1.00002L3.58333 10.5L5.41666 12.5Z" fill="black" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="2"></path> <path d="M1.33333 6.75L5.41667 12.5L7.25 10.5L3.16667 4.75L1.33333 6.75Z" fill="black" stroke="black" stroke-linecap="round" stroke-linejoin="round" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <rect height="14" stroke="black" stroke-width="2" width="14" x="1" y="1"></rect> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
9 Ha'Ta'asiya St. POB 2397 Ra'anana 4365405, Israel Tel.: +972 - 9 - 770 - 7600 Fax:+972 - 77 - 417 - 1410 e-mail: info@urogen.com
# 510(K) SUMMARY
# Uroject12 Svringe Lever
# 510(k) Number K180345
Date Prepared: September 17, 2018
#### I. SUBMITTER
## Applicant's Name:
UroGen Pharma Ltd. 9 Ha`Ta`asiya Street Ra'anana, Israel, 4365405 Tel: +972-9-770-7600 Fax: +972-77-4171410
# Primary Contact:
James G. Ottinger, R.Ph. Vice President, Regulatory Affairs UroGen Pharma Ltd. Telephone: 610-745-1230 Fax: (646) 918-1262 e-mail: jim.ottinger(@urogen.com
#### II. DEVICE
Trade Name: Uroject12 Syringe Lever Common or Usual Name: Syringe Lever Classification: Product Code: QBL (Piston Syringe Lever) Regulation: 21 CFR 880.5860 (Piston Syringe) Class: II Panel: General Hospital
#### III. PREDICATE DEVICES
Predicate device: iSecure Syringe Cartridge Holder, by Hospira, Product code FMF 21 CFR 880.5860; K063180
Reference device: Medi-SIS Syringe Infusion System; by: I-Flow Corp.; Product code: MEB, 21 CFR 880.5725 (K953028, K954059, K972173); FRN, LZH, 21 CFR 880.5725 (K111381)
#### IV. DEVICE DESCRIPTION
007 510k summary
Image /page/3/Picture/18 description: The image contains the logo for UroGen Pharma. The logo features a stylized kidney shape in blue, with a lighter blue splash-like design inside. Below the kidney graphic, the word "UroGen" is written in a bold, light blue font. Underneath "UroGen", the word "Pharma" is written in a smaller, less bold, light blue font.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for UroGen Pharma. The logo features a stylized kidney shape in blue, with a lighter blue abstract design inside. Below the kidney graphic, the word "UroGen" is written in a bold, blue sans-serif font. Underneath "UroGen", the word "Pharma" is written in a smaller, lighter blue font.
9 Ha'Ta'asiya St. POB 2397 Ra'anana 4365405. Israel Tel.: +972 - 9 - 770 - 7600 Fax:+972 - 77 - 417 - 1410 e-mail: info@urogen.com
The Uroject12 Syringe Lever body contains syringe housing for holding a standard 20 mL syringe and a dispensing mechanism that is manually operated by rotating a knob that presses the syringe piston. A clutch mechanism facilitates fast adaptation of the injection mechanism to the syringe piston position. The Uroject12 Syringe Lever is a reusable, reprocessed device to be sterilized before each use and cleaned after use.
The compatible syringes to be used with the Uroject12 are: 20 mL MEDALLION® syringe (Catalog Number MSS121, Merit Medical, Salt Lake City, UT, USA) and 20 mL MEDALLION® COP syringe (Catalog Number COP121PC, Merit Medical, Salt Lake City, UT, USA).
#### V. INDICATIONS FOR USE
The Uroject12 Syringe Lever Device is intended for use in the administration of sterile materials under aseptic conditions, in a clinical urology setting, by a clinician and in accordance with the best judgment of a physician.
#### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The principle of operation of the Uroject12 Syringe Lever device is similar to the predicate device, the iSecure Syringe Cartridge Holder, in that:
- Both devices encase a syringe to actuate an injection a.
- b. Both devices are used with a container closure system containing a drug solution with no direct contact to the drug.
The design and operating principles of the Uroject12 Syringe Lever Device are similar to those of the Medi-SIS Syringe Infusion System reference device. Both devices:
- a. Consist of an outer syringe holder and an inner syringe driver;
- b. Are used for injection of substances from a standard syringe;
- Are operated by the user rotating a rotary driver mechanism in order to apply C. pressure on the housed syringe plunger, thereby pushing out the fluid in the syringe in a controlled and accurate manner; and
- d. Are designed to provide visual markings on the syringe and allows visualization of the drug volume during administration
# E. SUBSTANTIAL EQUIVALENCE
The Uroject12 Syringe Lever is substantially equivalent to the predicate device based on the following:
# Intended Use
007 510k summary
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for UroGen Pharma. The logo features a stylized kidney shape in blue, with a lighter blue abstract design inside. Below the kidney graphic, the text "UroGen" is written in a bold, sans-serif font, also in blue. Underneath "UroGen", the word "Pharma" is written in a smaller, less bold font.
9 Ha'Ta'asiya St. POB 2397 Ra'anana 4365405. Israel Tel.: +972 - 9 - 770 - 7600 Fax:+972 - 77 - 417 - 1410 e-mail: info@urogen.com
The intended use of the proposed Uroject 12 Syringe Lever is similar to the predicate device, the iSecure Syringe Cartridge Holder, in which both devices are intended to administer sterile materials under aseptic conditions in accordance with the best judgement of the physician. While not identical to the predicate device, the intended use of the Uroject12 is limited to administration of fluids compared to the predicate device, which includes both administration and withdrawal. Performance of the Uroject12 in a urological setting has been demonstrated through bench testing and supports that these differences do not raise new questions of safety and effectiveness of the Uroject12 device.
# Technology
The proposed Uroject12 Syringe Lever is similar in technology to the predicate device, the iSecure Syringe Cartridge Holder, on the basis that the devices are used in conjunction with a pre-loaded syringe to perform administration of fluids operated by the user. The technological characteristics of the Uroject12 Syringe Lever is supported by the reference device, the Medi-Sis Syringe Infusion System, which uses the same principle of operation (rotary mechanism). The technological characteristics of the Uroject12 device have also been demonstrated through performance testing.
# Discussion
A direct comparison of characteristics demonstrates that the Uroject12 Syringe Lever is substantially equivalent to the iSecure Syringe Cartridge Holder by intended use and the principle of operation. The reference device, the Medi-SIS Syringe Infusion System, further supports the technological characteristics of the Uroject12 device. Performance characteristics demonstrate that the proposed Uroject12 Syringe Lever is as safe and effective as the predicate device for use in a urological setting.
#### F. PERFORMANCE DATA
# Bench Testing
Performance of the Uroject12 Syringe Lever has been demonstrated through the following:
- Instillation Pressure and Force Testing ●
- Extreme Dispensing Resistance Testing ●
- Residual Volume Testing ●
- Clutch Mechanism Functionality Testing
- Lifetime Functionality Testing
# Cleaning and Sterilization
The Uroject12 Syringe Lever is provided non-sterile and is a reusable, reprocessed device. The Uroject12 device is intended to be cleaned and sterilized before each
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for UroGen Pharma. The logo features a stylized kidney shape in blue, with a lighter blue splash-like design inside. Below the kidney graphic, the word "UroGen" is written in a bold, sans-serif font, also in blue. Underneath "UroGen," the word "Pharma" is written in a smaller, lighter blue font.
9 Ha'Ta'asiya St. POB 2397
Ra'anana 4365405, Israel Tel.: +972 - 9 - 770 - 7600 Fax:+972 - 77 - 417 – 1410 e-mail: info@urogen.com
use. The cleaning instructions were validated in accordance with AAMI TIR30:2011. The sterilization instructions were validated to demonstrate a sterility assurance level (SAL) of <10-6.
The results of performance testing demonstrate that the Uroject12 Syringe Lever is considered safe and effective for its intended use.
#### G. CONCLUSION
UroGen has demonstrated that the Uroject12 Syringe Lever is substantially equivalent to the predicate device, the iSecure Syringe Cartridge Holder. Differences between the proposed Uroject12 Syringe Lever and the predicate device do not raise new questions of safety or efficacy.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.