K111381 · Acacia, Inc. · LZH · Jan 13, 2012 · General Hospital
Device Facts
Record ID
K111381
Device Name
MEDI-SIS SYRINGE INFUSION SYSTEM
Applicant
Acacia, Inc.
Product Code
LZH · General Hospital
Decision Date
Jan 13, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
1.0 The Medi-SIS Syringe Infusion System is indicated for the delivery of enteral fluids through clinically accepted routes of administration. 2.0 The Medi-SIS Administration Extension Set is intended for single patient use only. 3.0 The Medi-SIS Syringe Driver can be reused up to 5,000 times
Device Story
Medi-SIS Syringe Infusion System is a mechanical syringe pump for enteral fluid delivery. System components: reusable mechanical Syringe Driver (plastic housing, steel spring), proprietary administration extension set (non-DEHP PVC tubing, flow restrictor, connectors), and compatible 30mL/60mL syringes. Principle of operation: mechanical spring force presses against syringe plunger to deliver fluids at a continuously decreasing rate; flow restrictor in administration set controls infusion time. Operated by clinicians in hospital settings (e.g., NICU). Clinicians calculate average flow rate based on volume and time. System benefits include controlled enteral fluid delivery with specialized connectors to prevent IV/enteral cross-connection errors.
Clinical Evidence
Human factors study conducted to assess clinical risks for enteral feeding of neonatal patients by NICU staff; risks mitigated. Non-clinical bench testing verified enteral fluid delivery within specified time and maximum residual volume met predicate specifications.
Technological Characteristics
Mechanical syringe pump; plastic housing and steel spring actuation. Non-DEHP PVC tubing. Flow restrictor for rate control. Reusable driver; single-use sterile administration sets. Connectors designed to prevent IV/enteral cross-connection. No software or electronic components.
Indications for Use
Indicated for delivery of enteral fluids via clinically accepted routes. Applicable to patients requiring enteral feeding, including neonates in NICU settings.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
B. Braun Perfusor Space Infusion Syringe Pump System (K062699)
Submission Summary (Full Text)
{0}------------------------------------------------
K111381
JAN 1 3 2012
# Section 5 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
#### 510(k) Summary of safety and effectiveness
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT: | Acacia, Inc. |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| APPLICANT ADDRESS | 785 Challenger Street<br>Brea, CA 92821 |
| APPLICANT PHONE | (714) 257-0470 |
| CONTACT PERSON | Fergie Ferguson |
| PREPARATION DATE | 01/11/12 |
| TRADE NAME: | Medi-SIS Syringe Infusion System |
| COMMON NAME: | Infusion Pump |
| CLASSIFICATION NAME: | Infusion Pump |
| DEVICE CLASSIFICATION: | Class II |
| PRODUCT CODE | FRN, LZH |
| PREDICATE DEVICES: | Medi-SIS Syringe Infusion System (K953028, K954059, K972173)<br>B. Braun Perfusor Space Infusion Syringe Pump System<br>(K062699) |
#### Purpose
The purpose of this submission is to add the delivery of enteral fluids through clinically accepted routed of administration to the intended use. No changes to the design of the device or performance features were made.
#### Substantially Equivalent To:
The Acacia, Inc. Medi-SIS Syringe Infusion System is substantially equivalent in intended use. principle of operation and technological characteristics to the Medi-SIS Syringe Infusion System (K953028, K954059, K972173), and is substantially equivalent in intended use with the B. Braun Perfusor Space Infusion Syringe Pump System (K062699) for the delivery of enteral fluids.
## Description of the Device Subject to Premarket Notification:
The Medi-SIS Syringe Infusion System consists of a Syringe Driver (mechanical syringe pump), proprietary administration extension set, and compatible syringe.
The Medi-SIS Syringe Driver consists of a mechanical syringe pump assembly that is manufactured from plastic housings and end caps, and steel spring (mechanical force from the spring presses against the syringe plunger), and holds a 30mL or 60mL Becton Dickinson syringe (preamendment device) or a 30mL or 60mL Acacia syringe (K092986). The Medi-SIS Syringe Driver must be used in conjunction with a Medi-SIS Administration Extension Set, which controls the time the infusion will occur (by flow restrictor) for the entire volume of the filled syringe. The
Page 1 of 3 Premarket Notification
{1}------------------------------------------------
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS Section 5
Medi-SIS Syringe Infusion System infuses fluids in a continuously decreasing rate and the user can calculate the average flow rate based from volume and time infused.
The Medi-SIS Syringe Driver is reusable and provided non-sterile.
The Medi-SIS Administration Set consists of various lengths of flexible non-DEHP PVC tubing. on/off clamp, flow restrictor, and male and female connectors. The male and female connectors meet Acacia's internal specifications for enteral applications to ensure cross connections do not occur between IV and Enteral infusions. Additional componentry may be added and configurations modified based on the needs of clinicians.
The following Administration Extension Sets will be available:
60mL Syringe Size, 1.0mL per minute delivery rate 60mL Syringe Size, 2.0mL per minute delivery rate 30mL Syringe Size, 0.5mL per minute delivery rate 30mL Syringe Size, 1.0mL per minute delivery rate 30mL Syringe Size. 2.0mL per minute delivery rate
The approximate residual volume for the syringe/administration set combination is approximately 2.20mL for all configurations. The Medi-SIS Administration Extension Set is provided sterile, single use, and non-pyrogenic.
### Indications for Use:
- 1. The Medi-SIS Syringe Infusion System is indicated for the delivery of enteral fluids through clinically accepted routes of administration.
- 2. The Medi-SIS Administration Extension Set is intended for single patient use only.
- 3. The Medi-SIS Syringe Driver can be reused up to 5,000 times
## Technical Characteristics:
Other than differences in connector sizes between the Medi-SIS Administration Sets for IV and Enteral applications (to ensure cross connections do not occur between IV and Enteral infusions), the Medi-SIS Syringe Infusion System is identical to the predicate device (the same Syringe Driver, tubing, and flow restrictor will be used), along with the same compatible syringes. There are no differences in design, material, chemical composition, or any other physical (other than connector sizes) and technical characteristics to the predicate device.
### Performance Data:
A human factors study was conducted to assess the clinical risks related to the Medi-SIS Syringe Infusion System for enteral feeding of neonatal patients by the hospital NICU staff. All identified risks have been identified and adequately miligated to ensure the system is safe for use.
The following non-clinical tests were performed to ensure the device is capable of meeting the predicate device's specifications when tested for the addition of enteral fluids to the intended use:
Acacia, Inc. MEDI-SIS Syringe Infusion System
Page 2 of 3 Premarket Notification
{2}------------------------------------------------
#### Section 5 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
- . Enteral fluid delivery within the specified time
- . Maximum residual volume
Since the addition of enteral fluid delivery was the only difference between the proposed and the predicate device, the significant test to ensure equivalency was to test the delivery time of enteral fluids and to verify they met the predicate device's specifications. The testing concluded that the device met all of the specifications.
Since the proposed and predicate devices are the same, it is concluded that the proposed device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate device.
### Basis for Determination of Substantial Equivalence:
Since the only change to the predicate device was to add the delivery of enteral fluids through clinically accepted routes of administration to the intended use, and to differentiate the connectors between the IV and Enteral Medi-SIS Administration Sets, it was concluded that the differences do not affect the performance specifications of the predicate device and do not add new hazards that would pose a safety risk to the patient or clinician. Verification was performed to ensure that the testing with an equivalent enteral fluid standard met all of the predicate device specifications.
Upon reviewing the safety and efficacy information provided in this submission and comparing the intended use, principle of operation and overall technological characteristics, the Medi-SIS Syringe Infusion System is determined by Acacia, Inc., to be substantially equivalent to the existing legally marketed device.
Acacia, Inc. MEDI-SIS Syringe Infusion System
Page 3 of 3 Premarket Notification
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird-like figure, composed of three curved, parallel lines.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Ms. Fergie Ferguson Director of Operations Acacia. Incorporated 785 Challenger Street Brea. California 92821
JAN 1 3 2012
Re: K111381
Trade/Device Name: Medi-SIS Syringe Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: LZH Dated: January 6, 2012 Received: January 9, 2012
Dear Ms. Ferguson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Ferguson
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Anton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
| 510(k) Number (if known): | K111381 |
|---------------------------|---------|
|---------------------------|---------|
Device Name: Medi-SIS Syringe Infusion System
Indications for Use:
1.0 The Medi-SIS Syringe Infusion System is indicated for the delivery of enteral fluids through clinically accepted routes of administration.
2.0 The Medi-SIS Administration Extension Set is intended for single patient use only.
3.0 The Medi-SIS Syringe Driver can be reused up to 5,000 times
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Ri C. Ch 1/13/12
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K11138
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.