Viz CTP

K180161 · Viz. Ai, Inc. · LLZ · Apr 20, 2018 · Radiology

Device Facts

Record IDK180161
Device NameViz CTP
ApplicantViz. Ai, Inc.
Product CodeLLZ · Radiology
Decision DateApr 20, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesAI/ML, Software as a Medical Device

Indications for Use

Viz CTP is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VMware, and can be used to perform image processing, analysis, and communication of computed tomography (CT) perfusion scans of the brain. Data and images are acquired through DICOM-compliant imaging devices. Viz CTP provides both analysis and communication capabilities for dynamic imaging datasets that are acquired with CT Perfusion imaging protocols. Analysis includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Results of image processing which include CT perfusion parameter maps generated from a raw CTP scan are exported in the standard DICOM format and may be viewed on existing radiological imaging viewers.

Device Story

Viz CTP is a standalone software package for processing dynamic CT perfusion (CTP) brain scans. It receives DICOM-compliant imaging data from scanners or PACS via local network. The software automatically identifies, extracts, and analyzes CTP studies to calculate perfusion parameters: Cerebral Blood Flow (CBF), Cerebral Blood Volume (CBV), Mean Transit Time (MTT), Residue function time-to-peak (TMax), and Arterial Input Function (AIF). It generates parametric maps, writes them back to DICOM format, and automatically transmits results to preconfigured destinations. Used by medical imaging professionals in clinical settings; output is viewed on diagnostic radiological viewers to assist in clinical decision-making regarding tissue perfusion. Benefits include automated, accurate computation of perfusion parameters to support diagnostic workflows.

Clinical Evidence

Bench testing only. No clinical data. Performance validated using a digital phantom dataset simulating tracer kinetic theory across a range of clinically relevant perfusion values. Correlations between Viz CTP outputs and ground truth were calculated and compared against seven other commercially available and academic CTP software packages. Results confirmed the device met pre-established performance goals for AIF detection, soft matter extraction, and calculation of CBF, CBV, MTT, and TMax.

Technological Characteristics

Standalone software; runs on standard off-the-shelf PC or virtual platform (e.g., VMware). DICOM-compliant (NEMA PS 3.1-3.20). Performs automated image analysis and parametric mapping. No hardware components. Software-based processing of dynamic CT perfusion datasets.

Indications for Use

Indicated for trained professionals, including physicians and medical technicians, to perform image processing, analysis, and communication of brain CT perfusion scans. No specific patient population age or gender restrictions stated.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text. Viz.ai, Inc. % Mr. Gregory Ramina Director of Regulatory Affairs 855 El Camino Real Suit 13A-252 PALO ALTO CA 94301 April 20, 2018 Re: K180161 Trade/Device Name: Viz CTP Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 5, 2018 Received: April 5, 2018 Dear Mr. Ramina: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ 510(k) Number (if known) K180161 Device Name #### Viz CTP Indications for Use (Describe) Viz CTP is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelt" computer or a virtual plafform, such as VMware, and can be used to perform image processing, and communication of computed tomography (CT) perfusion scans of the brain. Data and images are acquired through DICOM-compliant imaging devices. Viz CTP provides both analysis and communication capabilities for dynamic imaging datasets that are acquired with CT Perfusion imaging protocols. Analysis includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Results of image processing which include CT perfusion parameter maps generated from a raw CTP scan are exported in the standard DICOM format and may be viewed on existing radiological imaging viewers. | Type of Use (Select one or both, as applicable) | | |---------------------------------------------------------------|--------------------------------------------------------------| | <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ## 510(k) SUMMARY #### Viz.ai, Inc.'s Viz CTP Applicant Name: Viz.ai, Inc. 855 El Camino Real Suite 13A-252 Palo Alto. CA 94301 Contact Person: Gregory Ramina Director of Regulatory Affairs 343 Texas St San Francisco, CA 94107 Tel. (415) 663-6130 Greg@viz.ai Date Prepared: Name of Device: Viz CTP Common or Usual Name: PACS - Picture Archiving and Communications System Classification: Classification Panel: Radiology Regulation No: 21 C.F.R. § 892.2050 Regulatory Class: Class II Product Code: LLZ #### Predicate Devices iSchemaView Inc.'s RAPID (K121447) #### Device Description Viz CTP is a standalone software package that is comprised of several modules including DICOM receiving and sending modules, a study processor, image analysis algorithm, as well as software system components including a DICOM storage database and system health-monitoring. Viz CTP allows for bi-directional communication of data and may be implemented to allow a DICOM-compliant device to send files directly from the imaging modality, through a node on a local network, or from a PACS server. The device is designed to automatically receive, identify, extract, and analyze a CTP study of the head embedded in DICOM image data. The software outputs parametric maps related to tissue blood flow (perfusion) and tissue blood volume that are written back to the source DICOM. Following such analysis, the software automatically sends the results of analysis to a preconfigured destination point. The software allows for repeated use and continuous processing of data and {4}------------------------------------------------ can be deployed on a supportive infrastructure that meets the minimum system requirements. Viz CTP image analysis includes calculation of the following perfusion related parameters: - . Cerebral Blood Flow (CBF) - Cerebral Blood Volume (CBV) . - . Mean Transit Time (MTT) - . Residue function time-to-peak (TMax) - . Arterial Input Function (AIF) The primary users of Viz CTP are medical imaging professionals who analyze dynamic CT perfusion studies. The results of image analysis produced by Viz CTP should be viewed through appropriate diagnostic viewers when used in clinical decision making. ## Intended Use / Indications for Use Viz CTP is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VMware, and can be used to perform image processing, analysis, and communication of computed tomography (CT) perfusion scans of the brain. Data and imaqes are acquired through DICOM-compliant imaging devices. Viz CTP provides both analysis and communication capabilities for dynamic imaging datasets that are acquired with CT Perfusion imaging protocols. Analysis includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. Results of image processing which include CT perfusion parameter maps generated from a raw CTP scan are exported in the standard DICOM format and may be viewed on existing radiological imaging viewers. ## Difference in Indications for Use with Respect to the Predicate Device: Viz CTP differs from it's predicate, IschemaView's RAPID (K121447), in that the Viz CTP is limited to analysis of CT Perfusion imaging data sets. Viz CTP does not include the functionality, and is therefore not indicated for processing or analyzing images acquired through Magnetic Resonance Imaging (MRI) protocols. As such, Viz CTP is indicated for a subset of functionality when compared with the predicate device, however Viz CTP is indicated for the same functionality as the predicate device with respect to CT Perfusion datasets, i.e. for performing analysis which includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume. As Viz CTP is only indicated to be used with CT Perfusion imaging datasets, the difference in indications between Viz CTP and the predicate device limit the Viz CTP's functional or computational capabilities with respect to the device's indicated clinical use to a subset of functionalities provided by the predicate device. These differences, {5}------------------------------------------------ however, between the indications for use of the predicate and Viz CTP devices do not affect the Viz CTP device's diagnostic effect and do not raise any new issues of safety or efficacy. | | Viz CTP | iSchemaView RAPID | |------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | LLZ | LLZ | | Regulation | 21 C.F.R. § 892.2050 | 21 C.F.R. § 892.2050 | | Intended Use /<br>Indications for<br>Use | Viz CTP is an image processing<br>software package to be used by trained<br>professionals, including but not limited<br>to physicians and medical technicians.<br>The software runs on a standard "off-<br>the-shelf" computer or a virtual<br>platform, such as VMware, and can be<br>used to perform image processing,<br>analysis, and communication of<br>computed tomography (CT) perfusion<br>scans of the brain. Data and images<br>are acquired through DICOM-compliant<br>imaging devices.<br>Viz CTP provides both analysis and<br>communication capabilities for dynamic<br>imaging datasets that are acquired with<br>CT Perfusion imaging protocols.<br>Analysis includes calculation of<br>parameters related to tissue flow<br>(perfusion) and tissue blood volume.<br>Results of image processing which<br>include CT perfusion parameter maps<br>generated from a raw CTP scan are<br>exported in the standard DICOM format<br>and may be viewed on existing<br>radiological imaging viewers. | iSchemaView's RAPID is an image<br>processing software package to be used<br>by trained professionals, including but<br>not limited to physicians and medical<br>technicians. The software runs on a<br>standard "off-the-shelf" computer or a virtual<br>platform, such as VMware, and<br>can be used to perform image viewing,<br>processing and analysis of brain<br>images. Data and images are acquired<br>through DICOM compliant imaging<br>devices.<br>iSchemaView's RAPID provides both<br>viewing and analysis capabilities for<br>functional and dynamic imaging<br>datasets acquired with CT Perfusion<br>and MRI including a Diffusion Weighted<br>MRI (DWI)Module and a Dynamic<br>Analysis Module (dynamic contrast<br>enhanced imaging data for MRI and<br>CT).<br>The DWI Module is used to visualize<br>local water diffusion properties from the<br>analysis of diffusion weighted MRI data.<br>The Dynamic Analysis Module is used<br>for visualization and analysis of dynamic<br>imaging data, showing properties of<br>changes in contrast over time. This<br>functionality includes calculation of<br>parameters related to tissue flow<br>(perfusion) and tissue blood volume. | | PACS Functionality | | | | Basic PACS<br>Functions | Yes | Same | | Computer | Standard "Off-the-Shelf" PC | Same | | Platform | Workstation or VMWare | | | DICOM Compliance | Yes | Yes | | Functional Overview | Viz CTP is a software package that provides for the visualization and study of changes of tissue perfusion in digital images captured by CT. Viz CTP allows viewing and quantification. | Same | | Data Acquisition | Acquires medical image data from DICOM compliant imaging devices and modalities. | Same | | Data/Image Types | Computed Tomography (CT) | Same | | Acquisition and Modalities Features | | | | CT | CT Perfusion (CTP) | Yes | | Computed Parameter Maps | | | | Perfusion CT | Cerebral Blood Flow (CBF) | Yes | | | Cerebral Blood Volume (CBV) | Yes | | | Mean Transit Time (MTT) | Yes | | | Tissue residue function time to peak (TMax) | Yes | | Measurements/Tools | | | | CT Tools | Arterial Input Function (AIF) | Yes | | | Brain mask | Yes | | | Export perfusion files to PACS and DICOM file systems | Yes | | | Acquire, transmit, process, and store medical images | Yes | # Substantial Equivalence {6}------------------------------------------------ # Performance Data Viz CTP complies with DICOM (Digital Imaging and Communications in Medicine), developed by the American College of Radiology and the National Electrical Manufacturers Association. NEM PS 3.1-3.20 (2016). Viz.ai Inc. performed software verification and validation testing of the device and additional performance testing on a commercially available simulated dataset (digital phantom) generated by simulating tracer kinetic theory, and includes a wide range of clinically relevant values of perfusion parameters as ground truth. Correlations between the output of the Viz CTP device and the ground truth values were calculated, and compared to published correlations between the ground truth and the outputs of 7 other commercially available and academic CTP post-processing software. The results of performance testing showed that the Viz CTP device achieved the preestablished performance goals for AIF detection, soft matter extraction, and each {7}------------------------------------------------ perfusion parameter: cerebral blood flow (CBF), cerebral blood volume (CBV), mean transit time (MTT), and time to maximum residue (TMax). Thus, the performance testing demonstrated that the Viz CTP device provides accurate computation of perfusion parameters, similar to other, commercially available or academic software aimed for the computation of the same perfusion parameters. Together with software verification and validation, this performance evaluation demonstrates that the device satisfies all design requirements and device specifications. # Conclusions The Viz CTP device is as safe and effective as the RAPID (K121447). The Viz CTP has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate device with respect to CTP functionalities. The minor differences in indications do not alter the intended diagnostic use of the device and do not affect its safety and effectiveness when used as labeled. In addition, the minor technological differences between the Viz CTP device and its predicate device raise no new issues of safety or effectiveness. Performance data demonstrate that the Viz CTP is as safe and effective as the predicate, iSchemaView's RAPID for performing CTP analysis. Thus, the Viz CTP device is substantially equivalent.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...