K121447 · Ischemaview, Inc. · LLZ · Oct 4, 2013 · Radiology
Device Facts
Record ID
K121447
Device Name
ISCHEMA VIEW RAPID
Applicant
Ischemaview, Inc.
Product Code
LLZ · Radiology
Decision Date
Oct 4, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
iSchemaView's RAPID is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices. iSchemaView's RAPID provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast enhanced imaging data for MRI and CT). The DWI Module is used to visualize local water diffusion properties from the analysis of diffusionweighted MRI data. The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data. showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
Device Story
RAPID is a software-only PACS-integrated system for visualization and analysis of brain CT/MRI perfusion and diffusion data. It ingests DICOM images from scanners or PACS via hospital LAN. The system comprises RAPID Servers (physical or virtual) for image processing and a centralized iSchemaView Server for data repository and user interface. It performs volumetry of threshold maps, time-intensity plots, and mismatch measurement between co-registered volumes. Operators (physicians/technicians) use the software to calculate parameters like CBF, CBV, MTT, Tmax, and ADC. Output is delivered back to PACS/workstations for clinical review. The device aids in assessing tissue perfusion/diffusion, supporting diagnostic decision-making in clinical settings. Secure connectivity is maintained via VPN/SSH.
Clinical Evidence
Bench testing only. Performance validation testing demonstrated that the RAPID system provides accurate representation of key diffusion and perfusion processing parameters under a range of clinically relevant conditions. Software verification and validation testing confirmed the system met all design requirements and specifications.
Technological Characteristics
Software-only PACS-integrated system; runs on standard Linux-based PCs or virtual platforms (VMware). Connectivity via DICOM protocol over hospital LAN; supports secure VPN (RFC2401/RFC4301) and SSH. Data storage on encrypted hard disks. Features include DWI module (water diffusion properties) and Dynamic Analysis module (contrast-enhanced perfusion/blood volume).
Indications for Use
Indicated for trained professionals (physicians, medical technicians) to perform viewing, processing, and analysis of brain images acquired via CT or MRI. Used for functional and dynamic imaging datasets, including diffusion-weighted MRI and dynamic contrast-enhanced CT/MRI, to visualize tissue perfusion and diffusion properties.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
## 510(k) SUMMARY iSchemaView, Inc.'s RAPID
| Applicant's name: | iSchemaView, Inc.<br>323 Olmsted Road<br>Stanford, CA 94305 |
|-------------------|-------------------------------------------------------------------------------------------------------------------|
| Contact Person: | John J. Smith, M.D., J.D.<br>Partner<br>Hogan Lovells US LLP<br>555 Thirteenth Street, NW<br>Washington, DC 20004 |
**OCT 04 2013**
Date Prepared: September 13, 2013
Name of Device: RAPID
PACS - Picture Archiving Communications System Common or Usual Name:
Classification: Product Code: LLZ Regulation No: 21 C.F.R. §892.2050 Class: Il Classification Panel: Radiology Devices
Tel. (202) 637-3638 Fax (202) 637-5910
#### Predicate Devices
OLEA MEDICAL's Olea Sphere (K120196)
### Device Description
RAPID is a software package that provides for the visualization and study of changes of tissue perfusion and diffusion in digital images captured by CT (Computed Tomography) and MRI (Magnetic Image Resonance). RAPID can be installed on a customer PC or it can be accessed online as virtual system. It provides viewing, quantification, analysis and reporting capabilities.
RAPID works with the following types of (DICOM compliant) medical image data:
- · CT (Computed Tomography)
- · MRI (Magnetic Image Resonance)
RAPID acquires (DICOM compliant) medical image data from the following sources:
- · DICOM file
- · DICOM CD-R
- · Network using DICOM protocol
{2}------------------------------------------------
RAPID provides tools for performing the following types of analysis:
- · volumetry of threshold maps
- · time intensity plots for dynamic time courses
- measurement of mismatch between labeled volumes on co-registered image volumes
RAPID is a software-only device consisting of one or more RAPID Servers (dedicated or virtual and an iSchemaView Server). The RAPID Server is an image processing engine that connects to a hospital LAN, inside the Hospital Firewall. It can be a dedicated RAPID Server or a vmRAPID appliance, which is a virtualized RAPID Server that runs on a dedicated hospital server. Where available, the RAPID Server is placed logically in the demilitarized zone (DMZ) of the hospital's network to facilitate bidirectional secure connection between the (local) RAPID Server and the centralized iSchemaView Server.
The RAPID Server is configured to connect to applicable DICOM devices (PACS, Imaging Modalities, Research Workstations) via the hospital LAN and to receive diffusion and perfusion (DICOM) data directly and automatically from imaging modalities as they become available. It processes acquired data and delivers postprocessed images directly back to imaging modalities, local PACS and/or workstations, again using DICOM communication. It also transmits data to the iSchemaView Server for storage, retrieval and viewing.
The iSchemaView Server is a dedicated server that provides a central repository for RAPID data. All iSchemaView Server data is stored on encrypted hard disks. It also provides a user interface for accessing RAPID data. It connects to a firewalled Data Center Network and has its own firewall for additional data security. The iSchemaView Server connects to one or more RAPID Servers via WAN. Available types of connection include VPN (Virtual Private Network - RFC2401 and RFC4301 Standards) Tunnel and SSH (Secure Shell).
## Intended Use / Indications for Use
iSchemaView's RAPID is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices.
iSchemaView's RAPID provides both viewing and analysis capabilities for functional and dynamic imaging datasets acquired with CT Perfusion and MRI including a Diffusion Weighted MRI (DWI) Module and a Dynamic Analysis Module (dynamic contrast enhanced imaging data for MRI and CT).
The DWI Module is used to visualize local water diffusion properties from the analysis of diffusionweighted MRI data.
The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data. showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
{3}------------------------------------------------
## Technological Characteristics
RAPID performs the following functions:
- processes DICOM images from multiple sources to provide visualization of changes of . tissue perfusion and diffusion
- receives DICOM images from external DICOM image providers (modalities (CT/MRI . Scanners), PACS and Workstations) and sends DICOM images to external image consumers
- processes requests, statuses and results, and references therein, which are stored in a . queryable database
- processing status is available through a web browser using HTTP, HTML and PHP. .
- can send summary results to the user over email. For this, RAPID generally connects to the . infrastructure of the medical partner (e.g., the hospital). In particular, RAPID uses a SMTP protocol with security extensions to provide secure emailing.
RAPID is available in the following configurations:
- Standard RAPID, which is installed directly on a customer's Linux-based PC and integrated . with medical image processing software such as commercial PACS.
- . Virtual RAPID, wherein the user accesses RAPID online and uses it to process DICOM images otherwise available on his/her computer.
RAPID is a DICOM-compliant PACS software that provides comprehensive functionality to transfer, process, and display diffusion-weighted MRI (DWI) and dynamically acquired CT and MR imaging data (following the administration of a bolus of contrast media). RAPID runs on standard "off-theshelf" computer and networking hardware. RAPID is entirely independent from CT, MRI, or PACS platforms. It supports secure VPN (Virtual Private Network) networking or encapsulated Secure Shell (SSH), and seamlessly integrates into an existing radiological data network.
The primary users of RAPID PACS software are medical imaging professionals who analyze diffusion MRI and/or dynamic CT or MRI images. The images generated by RAPID provide additional diagnostic information, which is derived from the temporal/diffusion features of the native CT or MRI images.
### Performance Data
RAPID complies with DICOM (Digital Imaging and Communications in Medicine) - Developed by the American College of Radiology and the National Electrical Manufacturers Association. NEMA PS 3.1 - 3.20 (2011).
Additionally. iSchemaView conducted extensive performance validation testing and software verification and validation testing of the RAPID system. This performance validation testing demonstrated that the RAPID system provides accurate representation of key diffusion and perfusion processing parameters under a range of clinically relevant parameters and perturbations associated with the intended use of the software. Software validation and verification testing demonstrated that the RAPID system met all design requirements and specifications.
{4}------------------------------------------------
## Substantial Equivalence
RAPID is as safe and effective as the Olea Sphere. RAPID has the same intended use and similar indications, technological characteristics and principles of operation as its predicate device. The minor technological differences between the RAPID and its predicate raise no new issues of safety or effectiveness as demonstrated by the testing conducted with RAPID that confirms that the software reliably processes CT diffusion and MRI diffusion and perfusion medical images. Thus, the RAPID software is substantially equivalent.
{5}------------------------------------------------
| Parameter | RAPID | Olea Sphere (K120196) |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | LLZ | LLZ |
| Regulation | 21 CFR §892.2050 | 21 CFR §892.2050 |
| Intended Use/ Indications for use | iSchemaView's RAPID is an image processing<br>software package to be used by trained<br>professionals, including but not limited to physicians<br>and medical technicians. The software runs on a<br>standard "off-the-shelf" computer or a virtual<br>platform, such as VMware, and can be used to<br>perform image viewing, processing and analysis of<br>brain images. Data and images are acquired<br>through DICOM compliant imaging devices. | Olea Sphere is an image processing software package to be used by<br>trained professionals including but not limited to physicians and<br>medical technicians. The software runs on a standard "off-the-shelf"<br>workstation and can be used to perform image viewing, processing<br>and analysis of medical images. Data and images are acquired<br>through DICOM compliant imaging devices and modalities. |
| | The iSchemaView RAPID provides both viewing and<br>analysis capabilities for functional and dynamic<br>imaging datasets acquired with CT Perfusion and<br>MRI including a Diffusion Weighted MRI (DWI)<br>Module and a Dynamic Analysis Module (dynamic<br>contrast-enhanced imaging data for MRI and CT). | Olea Sphere provides both viewing and analysis capabilities of<br>functional and dynamic imaging datasets acquired with MRI or other<br>relevant modalities, including a Diffusion Weighted MRI (DWI) / Fiber<br>Tracking Module and a Dynamic Analysis Module (dynamic contrast<br>enhanced imaging data for MRI and CT). |
| | The DWI Module is used to visualize local water<br>diffusion properties from the analysis of diffusion-<br>weighted MRI data. | The DWI Module is used to visualize local water diffusion properties<br>from the analysis of diffusion-weighted MRI data. The Fiber Tracking<br>feature utilizes the directional dependency of the diffusion to display<br>the white matter Structure in the brain or more generally the central<br>nervous system. |
| | The Dynamic Analysis Module is used for<br>visualization and analysis of dynamic imaging data,<br>showing properties of changes in contrast over time.<br>This functionality includes calculation of parameters<br>related to tissue flow (perfusion) and tissue blood<br>volume. | The Dynamic Analysis Module is used for visualization and analysis<br>of dynamic imaging data, showing properties of changes in contrast<br>over time where such techniques are useful or necessary. This<br>functionality is referred to as:<br>Perfusion Module - the calculation of parameters related to<br>tissue flow (perfusion) and tissue blood volume.<br>Permeability Module - the calculation of parameters related<br>to leakage of injected contrast material from intravascular to<br>extracellular space. |
| PACS Functionality<br>Basic PACS Functions | View, process and analyze medical images.<br>Performs standard PACS functions with respect to<br>querying and listing. | Same |
| Computer Platform | Standard "off-the-shelf" PC workstation | Same |
| DICOM Compliance | | Yes |
| Funcional Overview | RAPID is a software package that provides for the<br>visualization and study of changes of tissue<br>perfusion and diffusion in digital images captured by<br>CT and MRI. RAPID provides viewing and<br>quantification. | Same |
| Data Acquisition | Acquires medical image data from DICOM compliant<br>imaging devices and modalities | Same |
| Data/Image Types | Computed Tomography (CT)<br>Magnetic Image Resonance (MRI) | Same |
| Acquisition and Modalities Features | | |
| MRI | Diffusion Weighted Image (DWI)<br>Perfusion Weighted Image (PWI) | Yes |
| CT | CT Perfusion (CTP) | Yes |
| Computed Parameter Maps | Isotropic DWI (isoDWI) | Yes |
| | Exponential apparent diffusion coefficient (eADC) | Yes |
| | ADC | Yes |
| Diffusion MRI | Trace of diffusion tensor (Trace) | Yes |
| | Fractional Anisotropy (FA) and color FA | Yes |
| | Eigenvector | Yes |
| | Eigenvalue | Yes |
| Perfusion MRI and Perfusion<br>CT | Cerebral blood flow (CBF) | Yes |
| | Cerebral blood volume (CBV) | Yes |
| | Mean transit time (MTT) | Yes |
| | Tissue residue function time to peak (Tmax) | Yes |
| Measurements/Tools | Arterial input function (AIF)/Venous output function<br>(VOF) | Yes |
| | Time-course | Yes |
| | Brain mask | Yes |
| MRI and CT Tools | Region of interest (ROI) and Volumetry | Yes |
| | Volumetric comparison between 2 ROIs | Yes |
| | Motion correction | Yes |
| | Export perfusion and diffusion files to PACS and<br>DICOM file systems | Yes |
| | Acquire, transmit, process, and store medical<br>images | Yes |
Substantial Equivalence
Note: RAPID also may be used with a virtual platform such as VMware
-
NDC - 038926/00001 - 4886242 v2
{6}------------------------------------------------
NDC: 03892600001-4886242 V2
K121442
page 6 of 6
0137
{7}------------------------------------------------
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-0609 Silver Spring, MD 20993-0002
October 4, 2013
iSchemaView, Inc. % John J. Smith, M.D., J.D. Pariner Hogan Lovells US L.LP Columbia Square 555 Thirtcenth Street. NW WASHINGTON DC 20004
Re: K121447
Trade/Device Name: RAPID Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ. Dated: September 16, 2013 Received: September 16, 2013
Dear Dr. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for too May 28, 1976. the enactment date of the Medical Device Amendments, or 10 devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic North the device, subject to the general controls provisions of the Act. The rou may, merely provisions of the Act include requirements for annual registration. Ilisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
ff your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act mat I DA has made a devel regulations administered by other Federal agencies. You must or any roothall the Act s requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{8}------------------------------------------------
Page 2-Dr. Smith
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.hum for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{9}------------------------------------------------
# Indications for Use
510(k) Number (if known): K121447
Device Name: iSchemaView's RAPID
Indications for Use:
iSchemaView's RAPID is an image processing software package to be used by trained professionals, including but not limited to physicians and medical technicians. The software runs on a standard "off-the-shelf" computer or a virtual platform, such as VMware, and can be used to perform image viewing, processing and analysis of brain images. Data and images are acquired through DICOM compliant imaging devices.
iSchemaView's RAPID provides both viewing and analysis capabilities for functional and dynamic Imaging datasets acquired with CT Perfusion and MRI including a Diffusion Weighted MRI (DVM) Module and a Dynamic Analysis Module (dynamic contrast enhanced imaging data for MRI and CT).
The DWI Module is used to visualize local water diffusion properties from the analysis of diffusionweighted MRI data.
The Dynamic Analysis Module is used for visualization and analysis of dynamic imaging data, showing properties of changes in contrast over time. This functionality includes calculation of parameters related to tissue flow (perfusion) and tissue blood volume.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$\sin(7)\$
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> 长121447 510(k)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.