Arthrex Corkscrew FT

K173788 · Arthrex, Inc. · MAI · Mar 1, 2018 · Orthopedic

Device Facts

Record IDK173788
Device NameArthrex Corkscrew FT
ApplicantArthrex, Inc.
Product CodeMAI · Orthopedic
Decision DateMar 1, 2018
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3030
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Arthrex Corkscrew FT is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, hip, knee, hand/wrist, and elbow in the following procedures: • Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. • Foot/Ankle:Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy. • Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. • Hand/Wrist: Scapholunate Ligament Reconstruction, Unar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. • Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction, Lateral Epicondylitis Repair. • Hip: Capsular Repair, acetabular labral repair.

Device Story

Arthrex Corkscrew FT is a fully threaded, vented suture anchor; pre-loaded with suture on a disposable inserter. Used by surgeons in clinical settings (OR) for soft tissue-to-bone fixation in various orthopedic procedures (shoulder, foot/ankle, hip, knee, hand/wrist, elbow). Device provides mechanical anchor point for sutures to secure tissue to bone. Benefits include stable fixation during healing. Device is sterile, single-use.

Clinical Evidence

Bench testing only. Tensile testing performed after 26 weeks of degradation to confirm pull-out force equivalence to predicate. Bacterial endotoxin testing conducted per EP 2.6.14/USP <85> to verify compliance with pyrogen limit specifications.

Technological Characteristics

Fully threaded, vented suture anchor. Material: PLLA/BTCP (Poly-L-lactic acid/Beta-tricalcium phosphate). Sterile, single-use. Delivered via disposable inserter. Mechanical fixation principle.

Indications for Use

Indicated for soft tissue-to-bone fixation in shoulder, foot/ankle, hip, knee, hand/wrist, and elbow orthopedic procedures. Applicable to patients requiring ligament, tendon, or capsular repair/reconstruction.

Regulatory Classification

Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. March 1, 2018 Arthrex Inc. David L. Rogers Project Manager, Regulatory Affairs 1370 Creekside Boulevard Naples, Florida 34108-1945 Re: K173788 Trade/Device Name: Arthrex Corkscrew FT Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: MAI Dated: January 29, 2018 Received: January 30, 2018 Dear Mr. Rogers: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {2}------------------------------------------------ 1000-1050. and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, ## Vincent J. Devlin -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ ### Indications for Use 510(k) Number (if known) K173788 Device Name Arthrex Corkscrew FT #### Indications for Use (Describe) The Arthrex Corkscrew FT is intended for fixation of suture (soft tissue) to bone in the shoulder, foot/ankle, hip, knee, hand/wrist, and elbow in the following procedures: - . Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction. - . Foot/Ankle:Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair, Hallux Valgus Reconstruction, Midfoot reconstruction, Metatarsal Ligament Repair/Tendon Repair, Bunionectomy. - . Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Iliotibial Band Tenodesis. - . Hand/Wrist: Scapholunate Ligament Reconstruction, Unar or Radial Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction. - . Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial collateral Ligament Reconstruction, Lateral Epicondylitis Repair. - . Hip: Capsular Repair, acetabular labral repair. Type of Use (Select one or both, as applicable) | <span style="font-family: Arial, sans-serif;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |-----------------------------------------------------------------------------------------------------| | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(k) Summary | Date Prepared | February 27, 2018 | |-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 | | Contact Person | David L Rogers<br>Project Manager, Regulatory Affairs<br>1-239-643-5553, ext. 71924<br>david.rogers@arthrex.com | | Name of Device | Arthrex Corkscrew FT | | Common Name | Suture Anchor | | Product Code | MAI | | Classification Name | 21 CFR 888.3030: Fastener, Fixation, Biodegradable, Soft Tissue | | Regulatory Class | II | | Predicate Device | K082810: Arthrex BioComposite Suture Anchors<br>K061665: Arthrex Corkscrew FT | | Purpose of Submission | This Special 510(k) premarket notification is submitted to add a line extension to<br>the Arthrex BioComposite Anchors cleared under predicate K082810. | | Device Description | The Arthrex Corkscrew FT is a fully threaded, vented suture anchor pre-loaded<br>with Arthrex Suture on a disposable inserter. The anchor is manufactured from<br>PLLA/BTCP and is offered sterile. | | Indications for Use | The Arthrex Corkscrew FT is intended for fixation of suture (soft tissue) to bone in<br>the shoulder, foot/ankle, hip, knee, hand/wrist, and elbow in the following<br>procedures:<br>• Shoulder: Rotator Cuff Repairs, Bankart Repair, SLAP Lesion Repair, Biceps<br>Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular<br>Shift or Capsulolabral Reconstruction.<br>• Foot/Ankle:Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair,<br>Hallux Valgus Reconstruction, Mid-foot reconstruction, Metatarsal Ligament<br>Repair/Tendon Repair, Bunionectomy.<br>• Knee: Anterior Cruciate Ligament Repair, Medial Collateral Ligament Repair,<br>Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique<br>Ligament Repair, Iliotibial Band Tenodesis.<br>• Hand/Wrist: Scapholunate Ligament Reconstruction, Ulnar or Radial<br>Collateral Ligament Reconstruction, Radial Collateral Ligament<br>Reconstruction.<br>• Elbow: Biceps Tendon Reattachment, Tennis Elbow Repair, Ulnar or Radial<br>collateral Ligament Reconstruction, Lateral Epicondylitis Repair.<br>• Hip: Capsular Repair, acetabular labral repair. | | Performance Data | Tensile testing was conducted after 26 weeks degradation to demonstrate that<br>the pull-out force of the device is equivalent to the predicate.<br><br>Bacterial endotoxin per EP 2.6.14/USP <85> was conducted to demonstrate that<br>the device meets pyrogen limit specifications. | | Conclusion | Arthrex Corkscrew FT is substantially equivalent to the predicate device in which<br>the basic design features and intended uses are the same. Any differences<br>between the proposed device and the predicate device are considered minor and<br>do not raise questions concerning safety or effectiveness. | | | Based on the indications for use, technological characteristics, and the summary<br>of data submitted, Arthrex Inc. has determined that the Arthrex Corkscrew FT is<br>substantially equivalent to the currently marketed predicate device. | {5}------------------------------------------------
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...