K173684 · G.Tec Medical Engineering GmbH · GYC · May 4, 2018 · Neurology
Device Facts
Record ID
K173684
Device Name
g.Estim PRO
Applicant
G.Tec Medical Engineering GmbH
Product Code
GYC · Neurology
Decision Date
May 4, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1310
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The g.Estim PRO is intended for functional brain mapping via electrical stimulation prior to cortical resections in the vicinity of essential cortex. The device must be used by medically trained and qualified personnel within a medical environment.
Device Story
g.Estim PRO is a battery-powered, constant-current, monophasic/biphasic cortical stimulator; used for functional brain mapping prior to cortical resections. Device inputs include user-defined stimulation parameters (amplitude, frequency, pulse width, polarity) via a host computer GUI; external trigger signals (TTL compatible) via hand/foot switches. Device outputs include electrical stimulation pulses to cortical electrodes; trigger signals to external devices (e.g., EEG acquisition systems) for synchronization. Operated by physicians/trained personnel in medical environments. System includes stimulator unit, USB cable, trigger cables, and software. Impedance measurement performed prior to stimulation; stimulation inhibited if impedance is high. Optical isolation between stimulator and computer ensures patient safety. Output allows clinicians to map essential cortex, aiding surgical planning and potentially improving patient outcomes by minimizing post-surgical deficits.
Clinical Evidence
Bench testing only. Testing verified stimulation pulse characteristics (frequency, amplitude, pulse width), impedance measurement accuracy, and trigger signal functionality (TTL compatibility). Safety verified via optical isolation and battery power. No clinical data presented.
Technological Characteristics
Constant-current stimulator; battery-powered (2 x 9V); optical isolation for USB communication. Delivers bi-phasic/monophasic rectangular pulses. Stimulation range: 0.05-15 mA, 2-100 Hz, 0.1-1.0 ms pulse duration. Connectivity: USB to host PC, TTL-compatible trigger I/O. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-40, ISO 14971, IEC 62304, IEC 62366. Patient isolation: Type BF.
Indications for Use
Indicated for functional brain mapping via electrical stimulation prior to cortical resections in the vicinity of essential cortex. For use by trained/qualified personnel in medical environments.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 4, 2018
g.tec medical engineering GmbH % Olaf Teichert Responsible Third Party Official TUV SUD America Inc. 1775 Old Highway 8 NW New Brighton, Minnesota 55112-1891
Re: K173684
Trade/Device Name: g.Estim PRO Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical electrode Regulatory Class: Class II Product Code: GYC Dated: April 20, 2018 Received: April 25, 2018
Dear Olaf Teichert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{2}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
for
Carlos L. Peña. PhD. MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173684
Device Name g.Estim PRO
Indications for Use (Describe)
The g.Estim PRO is intended for functional brain mapping via electrical stimulation prior to cortical resections in the vicinity of essential cortex. The device must be used by trained and qualified personnel within a medical environment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
g.tec medical engineering GmbH
# 510(k) Summary
The following 510(k) summary has been prepared pursuant to requirements specified in 21CFR 807.92(a).
807.92(a)(1)
#### Submitter Information
| g.tec medical engineering GmbH<br>Sierningstrasse 14<br>4521 Schiedlberg<br>Austria | | |
|-------------------------------------------------------------------------------------|------------------------------------------------------------|---------|
| Phone: | ++43 (7251) 22240-12 | |
| Fax: | ++43 (7251) 22240-39 | |
| Contact Person: | Christoph Guger | |
| Date: | 14th November 2016 | |
| 807.92(a)(2) | | |
| Trade Name: | g.Estim PRO | |
| Common Name: | Cortical Stimulator | |
| Classification Names(s): | Cortical Electrode<br>(per 21 CFR section 21 CFR 882.1310) | |
| Product Code: | GYC | |
| 807.92(a)(3) | | |
| | Predicate Device(s) | |
| Chatten Associates | S12X | K082629 |
g.tec medical engineering GmbH, Sierningstrasse 14, 4521 Schiedlberg, Austria, Europe,
{5}------------------------------------------------
807.92(a)(4)
#### Device Description
The g.Estim PRO is a neural, constant current, monophasic stimulator for electrical stimulation of the brain (cortical stimulator). The device is battery supplied and can be controlled via USB from a computer. It can be trigged via software or with external devices and can provide beside the stimulation pulse also trigger signals for other devices.
The system consists of the stimulator, a USB connector cable to connect the device to a host computer, trigger out cables and trigger in cables, the driver and software package (including a graphical user interface and communication interface). Additionally the device can be triggered with a hand switch and/or foot switch.
The stimulator can be programmed and is a constant-current stimulator that can deliver electrical pulses with alternating polarities, lengths, amplitudes or trains of such pulses. The device is connected via USB to a computer and on the computer stimulator settings can be performed. On the computer a graphical user interface allows to perform all settings required for the stimulator. The stimulator is battery powered and has an optical isolation from the computer for safety reasons. A LED shows the stimulus time point.
The stimulator allows to send trigger pulses to external devices to synchronize the stimulus e.g. with EEG data acquisition devices. Such a trigger pulse can be programmed to occur prior to the stimulus so that e.g. a switching unit can switch before the stimulus occurs and can last longer than the electrical pulse to switch back after the pulse. Furthermore the triggers can be perfectly aligned with the stimulus so that an accurate trigger signal can be sent to the external device. The trigger lines are TTL compatible. Two trigger lines are supported.
The stimulus can be triggered via the software running on the computer or via external inputs to the stimulator. Important is that the stimulus is triggered very accurately for perfect synchronization. A foot switch or hand switch can be used to trigger the stimulator. The trigger input is TTL compatible. One trigger line is supported.
The device can perform an impedance measurement to check the electrodes impedance before the stimulation is done and can measure the actual stimulation current when a stimulation is done. The results are shown in the graphical user interface and can be passed also to other applications via an interface. The impedance measurement is done quickly so that the stimulation can rapidly be performed. If a high impedance is detected the stimulator does not stimulate.
g. Estim PRO works in the same manner as the approved and predicate device.
{6}------------------------------------------------
g.tec medical engineering GmbH
807.92(1)(5)
#### Intended Use(s)
The g.Estim PRO is intended for functional brain mapping via electrical stimulation prior to cortical resections in the vicinity of essential cortex. The device must be used by medically trained and qualified personnel within a medical environment.
> g.tec medical engineering GmbH, Sierningstrasse 14, 4521 SchiedIberg, Austria, Europe, Tel.: +43 7251 22240, Fax: +43 7251 22240-39, office@gtec.at, www.gtec.at
{7}------------------------------------------------
#### g.tec medical engineering GmbH
807.92(a)(6)
| | Item | Chatten Associates<br>S12X<br>K082629 | g.tec medical<br>engineering GmbH<br>g.Estim PRO<br>This Submission | Comments |
|-----|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. | Intended use | The S12X Cortical<br>stimulator is intended for<br>Intraoperative cortical<br>stimulation mapping to aid<br>in cortical resections in the<br>vicinity of essential<br>cortex.<br>The ESAx option is<br>intended for use with the<br>S1 2X for facilitating the<br>switching of the patient<br>electrodes. | The g.Estim PRO is intended<br>for functional brain mapping<br>via electrical stimulation prior<br>to cortical resections in the<br>vicinity of essential cortex. | Reduced intended<br>used, no switching<br>unit option realized,<br>but equivalent in<br>safety and<br>effectiveness |
| 2. | User | The device is intended for<br>use only by medically<br>trained and qualified<br>personnel, with a hospital<br>or medical environment. | The device must be used by<br>medically trained and qualified<br>personnel within a medical<br>environment. | Same as predicate |
| 3. | Constant type | Constant current bi-<br>phasic, rectangular | Constant current, bi-phasic,<br>rectangular | Same as predicate |
| 4. | Maximum stimulation charge | 20 μC | 15 μC | Lower limit for<br>additional safety but<br>equivalent in<br>effectiveness. |
| 5. | Pulse polarity | Selectable, positive,<br>negative or alternating<br>stimulus polarity on<br>successive pulses | Selectable, positive, negative<br>or alternating stimulus polarity<br>on successive pulses | Same as predicate |
| 6. | Train duration | 0.2 to 20 sec in 7 steps | 0.1 to 20 sec. | Shorter minimum<br>train duration but<br>equivalent in safety<br>and effectiveness. |
| 7. | Pulse amplitude | 0.2 - 15 mA (peak),<br>constant current, 17 steps | 0.05 - 15 mA (peak), constant<br>current, 0.01 mA increments | Allows lower start<br>value, lower<br>increments, but<br>equivalent in safety<br>and effectiveness. |
| 8. | Stimulation frequency | 2 - 100 Hz in 7 steps | 2-100 Hz, 0,1 Hz steps | Same frequency<br>range and smaller<br>increments for more<br>flexibility, but<br>equivalent in safety<br>and effectiveness. |
| 9. | Stimulation pulse duration | 0.1 - 1.0 ms per phase (15<br>mA max); up to 2 ms per<br>phase (10 mA max) | 0.1 ms - 1 ms per phase | 1 ms is long enough<br>to achieve the<br>necessary<br>stimulation, but<br>equivalent in safety<br>and effectiveness. |
| 10. | Max. transient voltage | 34V, Zener Diode for<br>hardware safety | 80V, lead off via software for<br>safety | Higher voltage for<br>smaller electrode<br>sizes but equivalent |
| | | | | in safety and<br>effectiveness. |
| 11. | Power supply | 12V DC, External medical<br>power supply | Battery, 2 x 9 V | Battery supply to<br>reduce risk and<br>noise. |
| 12. | Electrode size | Warning that electrodes<br>with less than 4 sq mm<br>can cause tissue damage. | Electrode size defines the<br>safety threshold for stimulation<br>parameters. | The g.Estim PRO is<br>safe also with<br>smaller electrode<br>size, because the<br>stimulator calculates<br>the maximum<br>stimulation current<br>depending on<br>electrode size. So<br>equivalent in safety<br>and effectiveness. |
| 13. | Supported electrodes | Strip, grids, handheld<br>probe | Strip, grids, handheld probe | Same as predicate |
| 14. | Stimulus switching unit | Yes | No | Just a cortical<br>stimulator for 1<br>bipolar channel, but<br>equivalent in safety<br>and effectiveness. |
| 15. | Use standard sensors and<br>electrodes | Electrodes are not<br>included | Electrodes are not included | Same as predicate |
| 16. | User input | Stylus (touch), standard<br>mouse | Standard computer | Standard computer,<br>but equivalent in<br>safety and<br>effectiveness. |
| 17. | Graphical user interface | LCD | LCD display of computer | Same as predicate |
| 18. | Simulator parameters setting | Through displayed<br>controls | Through displayed controls | Same as predicate |
| 19. | Active stimulation LED | Yes | Yes | Same as predicate |
| 20. | Actual current delivered<br>displayed | Yes | Yes | Same as predicate |
| 21. | Dimension | 337 x 165 x 178 mm | 240 x 137 x 80 mm | Smaller and<br>therefore easier to<br>handle, but<br>equivalent in safety<br>and effectiveness. |
| 22. | Weight | 2.25 kg | 0.85 kg | Lighter and therefore<br>easier to carry, but<br>equivalent in safety<br>and effectiveness. |
| 23. | Trigger out for external<br>devices | Yes | Yes | Same as predicate |
| 24. | Digital input | Isolated | Isolated | Same as predicate |
| 25. | Patient isolation | BF | BF | Same as predicate |
| 26. | Safety standards | IEC60601-1<br>IEC60601-1-2 | IEC60601-1<br>IEC60601-1-2<br>IEC60601-2-40<br>ISO14971<br>IEC62304<br>IEC62366 | More standards<br>applied. |
| | | | | |
| 27. | System components | Stimulator<br>Switching array<br>Hand controller<br>Touchscreen Stylus<br>Power Supply<br>Portable carry case<br>Cable to switching array<br>Cable to manual probe | Stimulator with USB cable<br>Hand switch and foot switch<br>Portable carry case<br>1 x Trigger in cable<br>2 x Trigger out cable | g.Estim PRO doesnot provide a<br>switching array,<br>Touchscreen Stylus<br>and power supply,<br>additionally with<br>foot switch and<br>trigger cables.<br>Equivalent in safety<br>and effectiveness. |
| 28. | Firmware / software | Resident firmware | Resident firmware inside<br>g.Estim PRO<br>API/GUI on computer | Additional Front-end<br>driver to operate but<br>equivalent in safety<br>and effectiveness. |
| 29. | Log file | YES on device transferred<br>with USB flash stick | YES on computer | Not saved, on<br>device, but<br>equivalent in safety<br>and effectiveness. |
| 30. | Digital inputs/outputs | 1 external trigger input<br>(disabled)<br>1 train duration out<br>1 trigger out (pulse) | 1 trigger input (disabled)<br>1 pulse output<br>1 pulse output to trigger<br>external devices | Programmable<br>outputs but<br>equivalent in safety<br>and effectiveness. |
| 31. | Connectors, switches | 1 handheld controller<br>connector<br>1 handheld probe<br>connector<br>1 output connector to<br>switching unit<br>2 USB<br>1 external trigger input<br>1 train duration output<br>1 trigger out(pulse)<br>1 DC power in<br>1 Ethernet connector<br>1 on/off switch<br>1 HW all stop switch<br>1 HW stimulation switch | 1 hand/foot switch connector<br>2 safety sockets to connect<br>electrodes<br>1 USB connector<br>DIO - 1 input connector<br>DIO - 1 output connector 1<br>DIO - 1 output connector 2<br>2 battery housings<br>1 on/off switch | Different in output<br>for the switching<br>unit, reduced<br>connectivity with<br>one USB and no<br>Ethernet connector<br>and no HW-switches<br>, but equivalent in<br>safety and<br>effectiveness. |
| 32. | Impedance measurement | Yes | Yes | Same as predicate |
| 33. | Isolated power out | Yes | No | No isolated power<br>out but equal in<br>safety and<br>effectiveness. |
| 34. | Embedded controller | 1 | 2 | One additional to<br>ensure the equal<br>safety and<br>effectiveness. |
g.tec medical engineering GmbH, Siemingstrasse 14, 4521 Schiediberg, Austria, Europe,
{8}------------------------------------------------
# g.tec medical engineering GmbH
g.tec medical engineering GmbH, Sierningstrasse 14, 4521 Schiediberg, Austria, Europe,
{9}------------------------------------------------
# g.tec medical engineering GmbH
g.tec medical engineering GmbH, Sierningstrasse 14, 4521 Schiedlberg, Austria, Europe,
{10}------------------------------------------------
#### 807.92(b)(1)
The stimulator was tested with a data acquisition device and provides rectangular stimulation pulses with the necessary frequency and amplitude range. Furthermore digital outputs were tested to provide trigger signals and digital inputs were tested to trigger the stimulation. The impedance measurement was tested with test impedances. The tests show that the stimulator works like the predicate device.
In g.Estim PRO medical safety is realized by isolating the applied part optically for data transmission via USB and by battery supply. The current for the impedance measurement is limited with resistors.
807.92(b)(2) Not applicable
#### 807.92(b)(3)
The conclusion is that g.Estim PRO and the predicate device provide electrical stimuli with varying pulse width, frequency and amplitude in the same way and that the stimulator is working substantial equivalent and as effective as the marketed device. g.Estim PRO is using battery supply and optical separated USB transmission and is therefore considered to be safe.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.