The S12X Cortical Stimulator is intended for Intraoperative cortical stimulation mapping to aid in cortical resections in the vicinity of essential cortex. The device is intended for use only by medically trained and qualified personnel, within a hospital or medical environment. The ESAx option is intended for use with the S12X for facilitating the switching of the patient electrodes.
Device Story
S12X Cortical Stimulator provides bi-phasic, constant current electrical pulses for cortical mapping; used during surgery to identify essential cortex. System includes S12X unit, external medical-grade power supply, and optional ESAx electrode selector. ESAx uses optically coupled relays to switch between electrode grid/strip pairs; connects to stimulator and EEG amplifiers. Operated by physician/trained personnel via touch screen GUI or mouse. Hardware buttons provide 'Stimulate' and 'All-Stop' safety functions. Device logs stimulation parameters and timing to internal memory; transferable via USB. Transformer-coupled output prevents DC current; Zener diode limits transient voltage to 34V; pulse train terminates if voltage exceeds 30V. Benefits include rapid functional brain mapping and improved electrode switching efficiency. Alternating stimulus polarity feature minimizes electrode polarization for faster EEG recovery.
Clinical Evidence
Bench testing only. Testing confirmed S12X meets design and performance requirements, including stimulus output ranges and safety limits (voltage/current/power). No clinical data provided.
Technological Characteristics
Constant current, bi-phasic pulse generator. Transformer-coupled output. Medical-grade 12V DC power supply (120V/250V AC input). Touch screen GUI with hardware-based safety controls. ESAx uses optically coupled relays. Connectivity: USB port for log transfer. Standards: IEC 60601-1-1, IEC 60601-1-2. Embedded controller with crystal-controlled timing.
Indications for Use
Indicated for intraoperative cortical stimulation mapping to assist in cortical resections near essential cortex. For use by medically trained/qualified personnel in hospital/medical environments only.
Regulatory Classification
Identification
A cortical electrode is an electrode which is temporarily placed on the surface of the brain for stimulating the brain or recording the brain's electrical activity.
{1}------------------------------------------------
### Chatten Associates
1094 New Dehaven Avenue West Conshohocken, PA 19428
www.chattenassociates.com 610 940-6040 fax 610 940-6068 JUI - 1 2009
# 5, 510(k) Summary
Submission Date: September 5, 2008
Submitted by: Chatten Associates, Inc. 1094 New Dehaven Avenue, Suite 200 West Conshohocken, PA 19428 U.S.A Phone: 610-940-6040 FAX: 610-940-6068
Contact Person: John B. Chatten, President Tel: (610) 940-6040 E-mail: jchatten@chattenassociates.com
# Proprietary Name: S12X with ESAx Option
# Common/Usual Name: Cortical Stimulator with Switching Unit
#### Product Classification:
(based on Predicate Device Ojemann Cortical Stimulator #K924226) Category: Electrode, Cortical (based on Predicate Device Ojemann Cortical Stimulator #K924226) Regulation Number: 882.1310 (based on Predicate Device Ojemann Cortical Stimulator #K924226) Product Code: GYC Product Class: Class II
Substantial Equivalence to:
- · Ojemann Cortical Stimulator (K924226), Radionics/Integra Lifesciences Corporation, Plainsboro, NJ
- · Olemann Cortical Guindiator (1024229), Gallerino mografia (XLTEK), Oakville, ONT, Canada; now a division of Natus Medical, Inc., San Carlos, CA
- EEG-1100A Switch Box (K073491), Nihon Kohden America, Inc., Foothill Ranch, CA .
- \$12 Cortical Stimulator, marketed prior to May 28, 1976 and sold as a research device, Grass . Instruments Inc., now Grass-Technologies, an Astro-Med, Inc. Product Group, West Warwick, Ri
#### Intended Use:
The S12X Cortical Stimulator is intended for Intraoperative cortical stimulation mapping to aid in cortical The STZX Gortical Stimulator is intended for The device is intended for use only by medically trained and resections in the vicinly of essential or medical environment. The ESAx option is intended for use with the S12X for facilitating the switching of the patient electrodes.
### Device Description:
### S12X Cortical Stimulator:
The S12X Stimulator supplies a bi-phasic, constant current electrical pulse sequence that is suitable for The ST2X Summator Supplied to the surface of the surface of the cortex either with a Manual Probe or contical sumulation. Currents can be appliod to the ouries. ESAx) option. Electrodes, either the Manual through a Sct of Blootrouss are not considered part of the S12X.
2
{2}------------------------------------------------
The S12X is intended to replace and discontinued S12 Stimulator in the product line of Grass Instruments, Inc. (now Grass-Technologies, an Astro-Med, Inc. Product Group), marketed prior to May 28. 1976 and sold as a research device thereafter. The essential functions of the S12 have been retained, but appropriate isolation and insulation features have been modern requlatory requirements for use with human patients. In order to provide for applications and expansions of function compatible with modern medical data systems, hardware timing circuits have been replaced with an embedded controller whose crystal controlled counters replace the hardware timing functions, and the panel controls have been largely replaced with a touch screen graphical user interface (GUI).
The S12X device consists of an external medical grade power supply, the S12X Stimulator Unit, and the optional ESAx electrode selector adapter. The S12X front panel consists of hardware switches for controlling the "Stimulate" and "All Stop" functions, and a color LCD touch screen display with instrumentstyle controls for setting up the stimulation parameters, calibrating the stimulation log file contents, and for selecting electrodes for stimulation. User input is via a stylus (for touch panel use) or via a standard mouse. The optional ESAx unit connects to a multi-pin connector on the side panel of the S12X.
The S12X accumulates a log of stimulation events, their time-of-day, and their parameters. The log file can be easily transferred to a USB flash drive or equivalent for electronically filing or printing of the stimulation log.
No transient voltage can exceed 34 volts on the patient electrodes (limited by a Zener diode across the patient leads). Further, if after two pulses the internal voltage monitor detects pulse voltages greater than 30 volts, the pulse train will terminate.
No sustained direct current can pass through the patient leads since the output stimulus is entirely transformer coupled. The unit is powered by an external regulated 12 volt DC medical-grade supply powered by standard mains voltage of 120V or 250V AC, and meets U.S. and European requlatory standards: IEC 60601-1-1, and IEC 60601-1-2.
### ESAx Electrode Switch Array:
The ESAx is an optional accessory to the S12X which facilitates rapid functional brain mapping during presurgical workups and during surgery. The ESAx is an array of optically coupled relays that can select electrodes from an electrode grid array when more than one electrode pair is in contact with the cerebral cortex, usually in the form of strips or grids. Note that the cortical electrode strips or grids are not part of the device, and they are purchased by the user from other vendors. The ESAx connects the desired electrode pair to the stimulator and simultaneously disconnects the corresponding EEG amplifier leads from the amplifier. The choice of electrodes to be stimulated is controlled from a touch screen display on the S12X,
The ESAx has no buttons or switches, and consists of multi pin connectors for connecting to the S12X and for connecting EEG electrode iack boxes and EEG amplifier unit(s). Note that the EEG jack boxes and the multi-channel EEG amplifiers which can connect to the ESAx are not part of the S12X device, and therefore are not part of this 510(k) submission.
#### Summary of Comparison to Predicate Devices:
The basic performance parameters and intended use of the S12X are essentially the same as the OCS-1 Ojemann Cortical Stimulator currently marketed and sold by Integra Lifesciences Corporation and authorized by 510/k) #K924226. The S12X is also equivalent to the predicate device in terms of safety, effectiveness, and performance.
The differences between the S12X and OCS-1 are the following:
- · The Ojemann OCS-1 is battery powered with four 9V alkaline batteries, where the S12X uses a medical grade isolating power supply with a +12V DC output.
- The maximum available current of 10 milliamperes peak in the OCS-1 Ojemann Stimulator has been increased to 15 milliamperes in the S12X, for all pulse widths up to and including 1000 microseconds. At the largest pulse width of 2000 microseconds, the available current is still limited to 10 milliamperes in
{3}------------------------------------------------
order to stay within the 20 microcoulomb charge limit of the OCS-1 Ojemann stimulator. The current increase at shorter pulse widths was made available in the S12X because common stimulation practice when used at pulse widths of 300 microseconds often requires a maximum current of 15 milliamperes. The maximum available power output to the patient leads in the OCS-1 is 80 milliwatts. The maximum currents, voltages, pulse widths and frequencies of the S12X have been specified to limit the maximum power to the patient to 60 milliwatts.
- · The OCS-1 Ojemann front panel has knobs and switches for controls, whereas the S12X uses a touchscreen for setting parameters. However, controls essential to safety, such as "Stimulate" and "All-Stop", are still controlled by hardware push-buttons and hard-wired logic.
- · The S12X has an additional pulse shape feature, Alternating Stimulus Polarity, made available in order to minimize polarization of the stimulus electrodes. Intended to provide quicker recovery of the EEGamplifiers after stimulus, this option allows the first pulse of each pulse-pair to be made alternately positive and negative.
- The S12X keeps a running log of stimulus time and parameters in internal memory, in a format easily transferred to a USB flash drive or equivalent via the slave USB port on the S12X,
The ESAx electrode switch array has no function other than speeding up the process of functional brain mapping during presurgical and surgical workups. The optical relays are powered by the same isolated power supply in the S12X, and do not affect the stimulus waveform or isolation except to add approximately 100 ohms resistance to the stimulation circuit. The ESAx is essentially the same in functionality, effectiveness, and performance as the EMU-128S MODEL EX-NW-128S, marketed by XLTEK a division of Natus Medical, Inc., and authorized by 510(k) #K040360, and the EEG-1100A Switch Box, marketed by Nihon Kohden America, Inc., and authorized by 510(k) #K073491. All three are intended to switch external signals, as those of an electrical stimulation unit, to the appropriate electrodes. The only significant difference is that the XLTEK EM128S MODEL EX-NW128S switch matrix is incorporated into the XLTEK 128 channel EEG head box, whereas the Nihon Kohden EEG-1100A and the ESAx are external switches to the EEG amplifiers.
#### Conclusion from test data demonstrating Safety. Effectiveness, and Performance:
In house testing shows that the S12X meets design and performance functional requirements. The test data also demonstrates substantial equivalence with the predicate devices is so far as range of available stimulus outputs is concerned. The small extensions of available current from 10 to 15 milliamperes for pulses equal to or less than 1 millisecond are still within the 20 micro-joule charge per pulse limit of the OCS-1 predicate device.
The additional available compliance voltage from 20 to 30 volts is necessary to support the increased current from 10 ma in the OCS-1 to 15 ma in the S12X. Specifications that limit the maximum pulse frequency at 1 and. 2 millisecond puse widths to 50 Hz act to keep the maximum power output within the 60 milliwatt limit.
R. Chatton, President
September 5, 2008 Date
John B. Chatten, Presid ent Chatten Associates, Inc.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines forming its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the bird symbol.
Public Health Service
JUL - 1 2009
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Chatten Associates, Inc. % Mr. John Chatten 1094 New Dehaven Avenue West Conshohocken, Pennsylvania 19428
Re: K082629
Trade Name: S12X with ESAx Option Regulation Number: 21 CFR 882.1310 Regulation Name: Cortical Electrode Regulatory Class: II Product Code: GYC Dated: June 17, 2009 Received: June 19, 2009
### Dear Mr. Chatten:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
## Page 2- Mr. John Chatten
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours
F.n. Whr. Der Dr
Mark N. Melkerson Director Division of Surgical, Orthopedic, and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## 4. Indications for Use
510(k) Number (if known):
Device Name: S12X with ESAx Option, Chatten Associates, Inc.
Indications for Use:
The S12X Cortical Stimulator is intended for Intraoperative cortical stimulation mapping to aid in cortical resections in the vicinity of essential cortex. The device is intended for use only by medically trained and qualified personnel, within a hospital or medical environment. The ESAx option is intended for use with the S12X for facilitating the switching of the patient electrodes.
Prescription Use X (Part 21 CFR 801 Subprt D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page
(Division Sign=Off) Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K082629
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.