Piccolo Composite Nailing System

K173652 · CarboFix Orthopedics , Ltd. · HSB · Dec 28, 2017 · Orthopedic

Device Facts

Record IDK173652
Device NamePiccolo Composite Nailing System
ApplicantCarboFix Orthopedics , Ltd.
Product CodeHSB · Orthopedic
Decision DateDec 28, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3020
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric, high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.

Device Story

Intramedullary nailing system for proximal femur fracture fixation; consists of carbon fiber reinforced polymer (CFRP) nails, titanium alloy screws, and surgical instruments. Nails feature cylindrical geometry (11mm mid-shaft, up to 17mm proximal diameter) with lengths 180-460mm. Tantalum markers embedded in CFRP enable radiographic visualization. Used by orthopedic surgeons in clinical/OR settings to stabilize fractures via internal fixation. Device provides mechanical support to bone segments during healing. Benefits include radiolucency of CFRP material and structural stability comparable to metallic predicate systems.

Clinical Evidence

Bench testing only. Performance characteristics evaluated via static and dynamic bending tests of implant constructs. Bacterial endotoxin testing performed. No clinical data provided.

Technological Characteristics

Materials: Carbon fiber reinforced polymer (CFRP), titanium/titanium alloy, tantalum markers. Design: Cylindrical intramedullary rod with proximal/distal interlocking screw holes. Dimensions: 11mm mid-shaft diameter, 180-460mm lengths. Principle: Mechanical internal fixation. Sterilization: Not specified.

Indications for Use

Indicated for patients with stable or unstable proximal femur fractures, including pertrochanteric, intertrochanteric, and high subtrochanteric fractures. Applicable for trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.

Regulatory Classification

Identification

An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the U.S. Food & Drug Administration. The Department of Health & Human Services logo is on the left, and the U.S. Food & Drug Administration logo is on the right. The FDA logo is blue and white and includes the words "FDA U.S. FOOD & DRUG ADMINISTRATION". December 28, 2017 CarboFix Orthopedics Ltd. Yael Rubin Director of Regulatory Affairs 11 Ha'hoshlim Street Herzeliya, Israel 4672411 Re: K173652 Trade/Device Name: Piccolo Composite® Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: November 26, 2017 Received: November 28, 2017 Dear Yael Rubin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); {1}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. 510(k) Number (if known) K173652 Device Name Piccolo Composite® Nailing System Indications for Use (Describe) Piccolo Composite Proximal Femur Nails The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(K) Summary # CarboFix Orthopedics Ltd. # Piccolo Composite® Proximal Femur Nailing System # Applicant Name CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel # Contact Person Yael Rubin CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939 # Date Prepared December 2017 # Trade/Proprietary Name Piccolo Composite® Nailing System # Common Name Intramedullary Nailing System # Classification Name Rod, Fixation, Intramedullary and Accessories (21 CFR §888.3020; Product Code HSB). {4}------------------------------------------------ # Predicate Devices # Primary Predicate: - Piccolo Composite® Proximal Femur Nailing System (CarboFix Orthopedics Ltd.; K153536, K151010) Additional Predicate: - Gamma3® Nail System (Stryker (Howmedica Osteonics Corp.); K043431 and ■ more) # Intended Use/Indications for Use The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric, high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures. # System Description The Piccolo Composite Nailing System includes nails, screws and a set of instruments. The Piccolo Composite nail indicated for treatment of the proximal femur is a cylindrical rod. Nail mid-shaft diameter is 11mm, with the proximal end diameter of up to 17mm. Nail lengths are 180mm, 200mm, and in the range of 300 – 460mm. The nail provides for holes at the proximal and distal sections, designed for the insertion of a lag screw and interlocking screws. The lag screw is of 10.4mm diameter, with its length being in the range of 80mm to 110mm. The nails (and, optionally, the lag screws) are made of carbon fiber reinforced polymer and incorporate small amount of titanium/titanium alloy. Tantalum markers are embedded within the carbon fiber reinforced polymer, where applicable, to enable visualization during imaging. The distal screws, and, optionally, the lag screws, are made of titanium alloy. {5}------------------------------------------------ # Substantial Equivalence The Piccolo Composite Proximal Femur Nailing System intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable. Performance characteristics for the Piccolo Composite Proximal Femur Nailing System components included static and dynamic bending of implants construct, and are comparable to those of predicate devices (as applicable), thus demonstrating that the device is safe and effective for its intended use. In addition, bacterial endotoxin testing was conducted.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%