Piccolo Composite® Nailing System
K153536 · CarboFix Orthopedics , Ltd. · HSB · Jan 6, 2016 · Orthopedic
Device Facts
| Record ID | K153536 |
| Device Name | Piccolo Composite® Nailing System |
| Applicant | CarboFix Orthopedics , Ltd. |
| Product Code | HSB · Orthopedic |
| Decision Date | Jan 6, 2016 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 888.3020 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric, high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, malunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.
Device Story
Intramedullary nailing system for proximal femur fracture fixation; consists of carbon fiber reinforced polymer (CFRP) nails and lag screws with titanium alloy components; includes tantalum markers for radiographic visualization. Used by orthopedic surgeons in clinical/surgical settings to stabilize fractures. Device provides mechanical support to bone segments via lag and interlocking screws. Benefits include fracture stabilization and support for bone healing in trauma or pathological conditions.
Clinical Evidence
Bench testing only. Performance characteristics evaluated via static and dynamic bending of construct and lag screw torque to failure testing. Results demonstrate performance comparable to predicate devices.
Technological Characteristics
Intramedullary fixation rod system. Materials: Carbon fiber reinforced polymer (CFRP), titanium alloy, and tantalum markers. Dimensions: 11mm mid-shaft diameter, up to 17mm proximal diameter; lengths 180mm-460mm. Mechanical fixation via lag and interlocking screws.
Indications for Use
Indicated for patients with stable and unstable proximal femur fractures, including trauma, nonunions, malunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
- Piccolo Composite® Nailing System (K151010, K102369, K111056)
- Gamma3 Nail System (K043431)
Related Devices
- K151010 — Piccolo Composite Nailing System · CarboFix Orthopedics , Ltd. · Jul 28, 2015
- K173652 — Piccolo Composite Nailing System · CarboFix Orthopedics , Ltd. · Dec 28, 2017
- K221489 — Artemis Proximal Femoral Nail System · Glw, Inc. · Oct 27, 2022
- K201379 — Artemis Proximal Femoral Nail System · Glw, Inc. · Feb 19, 2021
- K100238 — HFN 125 DEGREE 9-13 MM, 180MM, HFN 130 DEGREE 9-13 MM, HFN RH 125 DEGREE 9-13MM · DePuy Orthopaedics, Inc. · May 11, 2010
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are facing to the right. The image is surrounded by a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 6, 2016
CarboFix Orthopedics, Limited Yael Rubin Director of Regulatory Affairs 11 Ha'hoshlim St. Herzeliya, 4627411 ISRAEL
Re: K153536
Trade/Device Name: Piccolo Composite® Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB Dated: December 8, 2015 Received: December 10, 2015
Dear Yael Rubin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Lori A. Wiggins -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K153536
Device Name Piccolo Composite® Nailing System
Indications for Use (Describe)
Piccolo Composite Proximal Femur Nails
The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <span></span> | |
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
## CarboFix Orthopedics Ltd.
## Piccolo Composite® Nailing System – Proximal Femur
## Applicant Name
CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzeliya 4672411, Israel
## Contact Person
Yael Rubin CarboFix Orthopedics Ltd. 11 Ha'hoshlim St., Herzliya 4672411, Israel Tel: +972 9 9511511, Fax: +972 9 9548939
## Date Prepared
December 2015
## Trade/Proprietary Name
Piccolo Composite® Nailing System
#### Common Name
Intramedullary Nailing System
#### Classification Name
Rod, Fixation, Intramedullary and Accessories (Class II, per 21 CFR §888.3020; Product Code HSB)
{4}------------------------------------------------
## Predicate Devices
- 트 Piccolo Composite® Nailing System (CarboFix Orthopedics Ltd .; K151010, K102369, K111056)
- Gamma3 Nail System (Stryker (Howmedica Osteonics Corp.); K043431 and more)
## Intended Use/Indications for Use
## Piccolo Composite Proximal Femur Nails
The Piccolo Composite Proximal Femur Nails are indicated for the treatment of stable and unstable proximal femur fractures (pertrochanteric, intertrochanteric, high subtrochanteric fractures, and combinations of these fractures), including, fractures resulting from trauma, nonunions, malunions, pathological fractures, impending pathological fractures, tumor resections, and revision procedures.
## System Description
The Piccolo Composite Nailing System includes nails, screws and a set of instruments. The Piccolo Composite nail indicated for treatment of the proximal femur is a cylindrical rod. Nail mid-shaft diameter is 11mm, with the proximal end diameter of up to 17mm. Nail lengths are 180mm, 200mm, and in the range of 300 – 460mm. The nail provides for holes at the proximal and distal sections, designed for the insertion of a lag screw and interlocking screws. The lag screw is of 10.4mm diameter, with its length being in the range of 80mm to 110mm. The nails and lag screws are made of carbon fiber reinforced polymer and incorporate small amount of titanium alloy. Tantalum markers are embedded within the carbon fiber reinforced polymer, where applicable, to enable visualization during imaging. The distal screws are made of titanium alloy.
## Substantial Equivalence
The Piccolo Composite Proximal Femur Nailing System intended use, design, materials, technological characteristics, and principles of operation are substantially equivalent to those of the predicate devices, as applicable.
{5}------------------------------------------------
Performance characteristics for the Piccolo Composite Proximal Femur Nailing System components included static and dynamic bending of construct, and lag screw torque to failure, and are comparable to those of predicate devices (as applicable), thus demonstrating that the device is safe and effective for its intended use.