Optima XR240amx, AutoGrid

K173602 · Ge Medical Systems, LLC · IZL · Jan 12, 2018 · Radiology

Device Facts

Record IDK173602
Device NameOptima XR240amx, AutoGrid
ApplicantGe Medical Systems, LLC
Product CodeIZL · Radiology
Decision DateJan 12, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1720
Device ClassClass 2
AttributesPediatric

Indications for Use

The Optima XR240amx is intended to take exposures utilizing film, computed radiography (CR), or wireless detectors, which are intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR). Optima XR240amx is a self-contained; battery operated mobile radiographic imaging system designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects). The Optima XR240amx is indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room. The system is indicated for taking radiographic exposures of the skull, spinal column, chest, abdomen, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications. The Optima XR240amx incorporates AutoGrid, which is an optional image processing software installed as a part of the systems Helix image processing software. AutoGrid can be used in lieu of an anti-scatter grid to improve image contrast in general radiographic images by reducing the effects of scatter radiation.

Device Story

Optima XR240amx is a battery-operated mobile X-ray system; captures radiographic images using film, CR, or wireless digital detectors. Features AutoGrid software, an optional image processing module within the Helix software suite. AutoGrid functions as a virtual anti-scatter grid; estimates and subtracts scatter radiation from images to improve contrast. Eliminates need for physical anti-scatter grids; removes Source-to-Image Distance (SID) constraints associated with physical grids. Operated by clinicians in clinical settings (emergency, ICU, OR, etc.) for patients unable to be moved to traditional radiology rooms. Output is a diagnostic radiographic image; assists physicians in detecting orthopedic damage, tumors, pneumonia, or foreign objects. Benefits include improved workflow, reduced physical equipment handling, and consistent image contrast regardless of SID.

Clinical Evidence

Bench testing only. No clinical studies performed. Bench testing demonstrated that AutoGrid software achieves image contrast equivalent to physical anti-scatter grids. Verification and validation testing confirmed safety and effectiveness.

Technological Characteristics

Battery-operated mobile X-ray system. Includes Helix image processing software with optional AutoGrid module. Connectivity via DICOM. Complies with IEC 60601-1, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-54, and IEC 62366 standards.

Indications for Use

Indicated for adult and pediatric patients for general-purpose diagnostic radiographic examinations of skull, spinal column, chest, abdomen, and extremities. Used in clinical environments including emergency, ICU, NICU, cardiac, and OR. Contraindicated for mammography.

Regulatory Classification

Identification

A mobile x-ray system is a transportable device system intended to be used to generate and control x-ray for diagnostic procedures. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ January 12, 2018 Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue. Underneath the FDA logo is the word "ADMINISTRATION". GE Medical Systems, LLC % Mr. Chris Paulik Regulatory Affairs Program Manager 3000 N. Grandview Blvd. WAUKESHAWI 53188 Re: K173602 Trade/Device Name: Optima XR240amx with AutoGrid Regulation Number: 21 CFR 892.1720 Regulation Name: Mobile x-ray system Regulatory Class: II Product Code: IZL, MQB Dated: November 20, 2017 Received: November 21, 2017 Dear Mr. Paulik: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); U.S. Food & Drug Administration 10903 New Hampshire Avenue Silv er Spring, MD 20993 w ww.fda.gov {2}------------------------------------------------ and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Michael D. O'Hara For Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined in a stylized, cursive font. The logo is enclosed within a circular border, and there are decorative swirls or flourishes around the circle. The color of the logo and the border is blue. Section 4: Indications for Use (FORM FDA 3881) # Optima XR240amx with AutoGrid {4}------------------------------------------------ ## Indications for Use 510(k) Number (if known) #### K173602 Device Name Optima XR240amx with AutoGrid #### Indications for Use (Describe) The Optima XR240amx is intended to take exposures utilizing film, computed radiography (CR), or wireless detectors, which are intended to replace radiographic film screen systems in all general purpose diagnostic procedures, for digital radiography (DR). Optima XR240amx is a seff-contained; battery operated mobile radiographic imaging system designed to generate diagnostic radiographic images (medical x-rays) that may increase the ability to detect disease or injury early enough for a medical problem to be managed, treated, or cured. Medical x-rays are used in many types of examinations and procedures, some examples include: x-ray radiography (to find orthopedic damage, tumors, pneumonias, foreign objects). The Optima XR240amx is indicated for use on adult and pediatric patients for general-purpose diagnostic radiographic examinations and procedures. Its mobility enables general-purpose radiographic procedures throughout the clinical environment, or as needed within the emergency, intensive care, premature birth ward, cardiac and operating departments, for patients that may not be able to be moved or in cases where it is unsafe or impractical to move them to a traditional RAD room. The system is indicated for taking radiographic exposures of the skull, spinal column, extremities, and other body parts with the patient sitting, standing, or lying in the prone or supine position. This device is not intended for mammographic applications. The Optima XR240amx incorporates AutoGrid, which is an optional image processing software installed as a part of the systems Helix image processing software. AutoGrid can be used in lieu of an anti-scatter grid to improve image contrast in general radiographic images by reducing the effects of scatter radiation. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {5}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {6}------------------------------------------------ Image /page/5/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' in a stylized, cursive font, enclosed within a circle. There are water droplet-like shapes surrounding the circle. The logo is in a light blue color. # 510(k) Summary In accordance with 21 CFR 807.92 the following summary of information is provided: | Date<br>Submitted: | November 20, 2017 | |-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter: | GE Medical Systems, LLC<br>3000 N. Grandview Blvd<br>Waukesha, WI 53188, USA | | Primary<br>Contact<br>Person: | Chris Paulik<br>Regulatory Affairs Program Manager<br>GE Healthcare<br>262-548-2010<br>Christopher.A.Paulik@ge.com | | Secondary<br>Contact<br>Person: | Diane Uriell<br>Regulatory Affairs Director<br>GE Healthcare<br>262-290-8218<br>Diane.Uriell@ge.com | | Device Trade<br>Name: | Optima XR240amx with AutoGrid | | Common/Usual<br>Name: | Mobile X-ray system | | Regulation,<br>Classification,<br>and Product<br>Code: | Regulation Name: Mobile X-Ray System<br>Regulation: 21 CFR 892.1720<br>Classification: Class II<br>Product Codes: IZL, MQB | | | | | Predicate<br>Device(s): | Optima XR240amx (K162990) | | | Regulation Name: Mobile X-Ray System | | | Regulation: 21 CFR 892.1720 | | | Classification: Class II | | | Product Codes: IZL, MQB | | Device<br>Description: | The Optima XR240amx with AutoGrid is intended to take exposures,<br>using a wired or remote exposure switch, utilizing film, computed<br>radiography (CR), or cleared wireless radiographic detectors, which are<br>intended to replace radiographic film screen systems in all general<br>purpose diagnostic procedures, for digital radiography (DR). | | | Optima XR240amx with AutoGrid is a self-contained; battery operated<br>mobile radiographic imaging system designed to generate diagnostic<br>radiographic images (medical x-rays) that may increase the ability to<br>detect disease or injury early enough for a medical problem to be<br>managed, treated, or cured. Medical x-rays are used in many types of<br>examinations and procedures, some examples include: x-ray<br>radiography (to find orthopedic damage, tumors, pneumonias, foreign<br>objects). | | | The Optima XR240amx with AutoGrid system is indicated for use on<br>adult and pediatric patients for general-purpose diagnostic radiographic<br>examinations and procedures. Its mobility enables general-purpose<br>radiographic procedures throughout the clinical environment, or as<br>needed within the emergency, intensive care, premature birth ward,<br>cardiac and operating departments, for patients that may not be able to<br>be moved or in cases where it is unsafe or impractical to move them to a<br>traditional RAD room. | | | The incorporation of AutoGrid software into the Helix Image Processing<br>software on the Optima XR240amx allows the user to improve the image<br>contrast in general radiographic images by reducing the effects of<br>scatter radiation in lieu of incorporating a physical anti-scatter grid onto<br>the cleared wireless detectors. | | | The system is indicated for taking radiographic exposures of the skull,<br>spinal column, chest, abdomen, extremities, and other body parts with<br>the patient sitting, standing, or lying in the prone or supine position. | | | This device is not intended for mammographic applications. | | Intended Use: | The Optima XR240amx is intended to take exposures utilizing film,<br>computed radiography (CR), or wireless detectors, which are intended to<br>replace radiographic film screen systems in all general purpose<br>diagnostic procedures, for digital radiography (DR). | | | Optima XR240amx is a self-contained; battery operated mobile<br>radiographic imaging system designed to generate diagnostic<br>radiographic images (medical x-rays) that may increase the ability to<br>detect disease or injury early enough for a medical problem to be<br>managed, treated, or cured. Medical x-rays are used in many types of<br>examinations and procedures, some examples include: x-ray<br>radiography (to find orthopedic damage, tumors, pneumonias, foreign<br>objects). | | | The Optima XR240amx is indicated for use on adult and pediatric<br>patients for general-purpose diagnostic radiographic examinations and<br>procedures. Its mobility enables general-purpose radiographic<br>procedures throughout the clinical environment, or as needed within the<br>emergency, intensive care, premature birth ward, cardiac and operating<br>departments, for patients that may not be able to be moved or in cases<br>where it is unsafe or impractical to move them to a traditional RAD<br>room. | | | The system is indicated for taking radiographic exposures of the skull,<br>spinal column, chest, abdomen, extremities, and other body parts with<br>the patient sitting, standing, or lying in the prone or supine position. | | | This device is not intended for mammographic applications. | | | The Optima XR240amx incorporates AutoGrid, which is an optional<br>image processing software installed as a part of the systems Helix<br>image processing software. AutoGrid can be used in lieu of an anti-<br>scatter grid to improve image contrast in general radiographic images<br>by reducing the effects of scatter radiation. | | Technology: | The Optima XR240amx with AutoGrid employs the same fundamental<br>scientific technology as the predicate device. They are both battery<br>operated mobile x-ray systems that capture exposures utilizing film, CR<br>plates, or a wireless detector. The intended use is the same between the<br>Optima XR240amx with AutoGrid and the predicate device. The Optima<br>XR240amx with AutoGrid did not change the input power, battery<br>subsystem, drive subsystem, x-ray generation, and exposure control<br>from the predicate devices. The difference being introduced is a<br>modification to the Helix image processing software to incorporate an<br>optional image processing software called AutoGrid. AutoGrid can be<br>used in lieu of a physical anti-scatter grid attached to the cleared<br>wireless detectors to improve image contrast in general radiographic<br>images by reducing the effects of scatter radiation. The use of the<br>AutoGrid software can improve user workflow since there is no need to<br>physically incorporate an anti-scatter grid into the image chain. Its use<br>can also alleviate issues caused by the incorrect Source to Image<br>Distance (SID) setup with physical anti-scatter grids, since the AutoGrid<br>software does not require a specific SID to function. | | The AutoGrid software reduces the effect of scatter radiation on the<br>image by estimating the scatter radiation within the image, and then<br>subtracting this scatter estimate from the image. The AutoGrid software<br>can be configured at three global strength options (Low, Medium, and<br>High). The strength indicates the amount of scatter reduction that will<br>occur during image processing. The Low strength corresponds to the<br>amount of scatter reduction that would occur through using a physical<br>6:1 ratio grid, Medium a physical 8:1 ratio grid, and High a physical 12:1<br>ratio grid. The AutoGrid software can be used in any application where a<br>physical grid was originally indicated, including both adult and pediatric<br>imaging techniques or protocols.<br>The AutoGrid software is an optional feature that cannot be used in<br>conjunction with a physical grid. | | | Determination<br>of Substantial<br>Equivalence: | Summary of Non-Clinical Tests: | | | The modification to the Optima XR240amx to include AutoGrid and its<br>applications comply with voluntary standards: | | | ES60601-1 Medical electrical equipment - Part 1: General<br>requirements for basic safety and essential performance ; IEC 60601-1-3 Medical Elec. Equipment - P. 1: General Req. for Safety<br>3. Collateral Standard: General Req. for Radiation Protection in<br>Diagnostic XRay Equipment; IEC 60601-1-6 Medical electrical equipment - Part 1-6: General<br>requirements for safety - Collateral Standard: Usability ; IEC 60601-2-54 Medical electrical equipment - Part 2- 54: Particular<br>requirements for the basic safety and essential performance of X-ray<br>equipment for radiography and radioscopy ; IEC 62366 Medical devices - Application of usability engineering to<br>medical Devices; PS 3.1 - 3.20 Digital Imaging and Communications in Medicine<br>(DICOM) set. (Radiology). | | | The following quality assurance measures were applied to the<br>development of the system: | | | Risk Analysis Requirements Reviews Design Reviews Testing on unit level (Module verification) Integration testing (System verification) Performance testing (Verification) Safety testing (Verification) | | Simulated use testing (Validation)…
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...