Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0
K173468 · Toshibamedical Systems Corporation · JAK · Feb 23, 2018 · Radiology
Device Facts
Record ID
K173468
Device Name
Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0
Applicant
Toshibamedical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Feb 23, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, usinq indicated software/hardware, by a trained and qualified physician. FIRST 3.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.
Device Story
Ultra-high resolution multi-slice helical CT scanner; consists of gantry, couch, and console. Inputs: X-ray projection data. Processing: Iterative reconstruction (FIRST 3.0) and image reconstruction (NR, HR, SHR modes). Outputs: Cross-sectional volume sets/images. Used in clinical settings by trained physicians/technicians. FIRST 3.0 algorithm reduces exposure dose while maintaining/improving image quality. High-resolution modes (HR/SHR) improve spatial resolution for detailed anatomical assessment. Output viewed on console for diagnostic decision-making; aids in specialized studies. Benefits: Enhanced image detail, potential dose reduction, and improved size discrimination of small structures.
Clinical Evidence
Bench testing only. No clinical trials. Performance validated using phantoms (MITA-FDA LCD, Catphan) to assess CT number accuracy, contrast-to-noise ratio, uniformity, MTF, and low-contrast detectability. Model observer study demonstrated 50% improvement in size discrimination in HR mode. Cadaver and cardiac stent images used to demonstrate diagnostic quality. Radiologist review of representative images confirmed diagnostic utility.
Indicated for whole-body (including head) cross-sectional volume acquisition and display in patients requiring CT imaging. FIRST 3.0 algorithm indicated for abdomen, pelvis, chest, cardiac, extremities, and head applications to reduce exposure dose and improve high-contrast spatial resolution.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Aquilion PRIME, TSX-303A/A, /B and /F, v6.00 (K141741)
Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1 (K170177)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. The logo on the left is the Department of Health and Human Services logo. The logo on the right is the Food and Drug Administration logo. The FDA logo is in blue and contains the words "FDA U.S. FOOD & DRUG ADMINISTRATION".
Toshiba Medical Systems Corporation % Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive TUSTIN CA 92780
February 23, 2018
Re: K173468
Trade/Device Name: Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: January 10, 2018 Received: January 11, 2018
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert Ochs. Ph.D Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
# K173468
Device Name
Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0
# Indications for Use (Describe)
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head. The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, usinq indicated software/hardware, by a trained and qualified physician.
FIRST 3.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.
Type of Use (Select one or both, as applicable)
ك Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) SUMMARY
- 1. SUBMITTER'S NAME: Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
- OFFICIAL CORRESPONDENT: 2. Naofumi Watanabe Senior Manager, Regulatory Affairs and Vigilance
- 3. ESTABLISHMENT REGISTRATION: 9614698
### 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
### ട. Date Prepared:
February 20, 2018
# 6. TRADE NAME(S):
Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0
### 7. COMMON NAME:
System, X-ray, Computed Tomography
- 8. DEVICE CLASSIFICATION (Regulatory Class, CFR Reference, Name): Class II (per 21 CFR 892.1750, Computed Tomography X-ray System)
# 9. PRODUCT CODE / DESCRIPTION:
JAK / Computed Tomography X-Ray System
### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
Image /page/3/Picture/20 description: The image shows the Canon logo in red, with the words "CANON GROUP" in black below it. The words "Page 1 of 6" are located in the bottom left corner of the image. The logo and text are crisp and clear against a white background.
{5}------------------------------------------------
#### 11. PREDICATE DEVICE:
| Product | Marketed | | Regulation Regulation | Product Code | 510(k) | Clearance |
|-------------------------------------------------------------------------------------|------------------------------------------|--------------------|----------------------------------------|---------------------------------------------------|---------|------------|
| | by | Number | Name | | Number | Date |
| Aquilion PRIME,<br>TSX-303A/A, /B and /F,<br>v6.00<br>(Primary Predicate<br>Device) | Toshiba<br>America<br>Medical<br>Systems | 21 CFR<br>892.1750 | Computed<br>Tomography<br>X-ray System | JAK:<br>System, X-ray,<br>Tomography,<br>Computed | K141741 | 11/11/2014 |
| Aquilion ONE (TSX-<br>305A/3) V8.3 with<br>FIRST 2.1 | Toshiba<br>America<br>Medical<br>Systems | 21 CFR<br>892.1750 | Computed<br>Tomography<br>X-ray System | JAK:<br>System, X-ray,<br>Tomography,<br>Computed | K170177 | 06/30/2017 |
#### 12. REASON FOR SUBMISSION:
New medical device
### 13. DEVICE DESCRIPTION:
Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0 is an ultra-high resolution whole body multi-slice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. Aquilion Precision incorporates a 160-row, 0.25 mm detector, a 5.7-MHU large-capacity tube, and 0.35 s scanning, enabling wide-range scanning with short scan times to capture cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. In addition, the subject device incorporates the latest iterative reconstruction technology, FIRST 3.0, intended to reduce exposure dose while maintaining and/or improving image quality.
### 14. INDICATIONS FOR USE:
This device is indicated to acquire and display cross-sectional volumes of the whole body, to include the head.
The Aquilion Precision has the capability to provide volume sets. These volume sets can be used to perform specialized studies, using indicated software, by a trained and qualified physician.
FIRST 3.0 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.
### 15. SUBSTANTIAL EQUIVALENCE:
The Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0, is substantially equivalent to the Aquilion PRIME, TSX-303A/A, /B and /F v6.00, which received premarket clearance under K141741 and is marketed by Toshiba America Medical Systems. The intended use of the Aquilion Precision is the same as that of the predicate device. Software applications that have already received 510(k) clearance are being implemented on the subject device. The main difference between the Aquilion Precision and the predicate device is the configuration of the detector as well as the implementation of two new acquisition modes, HR (High Resolution) and
{6}------------------------------------------------
SHR (Super High Resolution), in addition to the conventional, NR (Normal Resolution) mode, which provides the user with image quality improvements. The NR mode is substantially equivalent to the acquisition mode employed by the predicate device. A comparison of the technological characteristics between the subject and the predicate device is included below.
| ltem | Aquilion Precision (TSX-304A/2)<br>V8.6 with FIRST 3.0 | Aquilion PRIME, TSX-303A/A, /B<br>and /F v6.00 | |
|---------------------------------------|--------------------------------------------------------------------|------------------------------------------------|--|
| 510(k) Number | This submission | K141741 | |
| Configuration of the detector | 0.25 mm x 160 rows | 0.5 mm x 80 rows | |
| Detector | 1792 channels × 160 rows | 896 channels × 80 rows | |
| Max. scan length/rotation | 40 mm/rotation | 40 mm/rotation | |
| Acquisition Modes | NR mode: | | |
| | 0.5mm x 80 rows / 896 channels | 0.5mm x80 rows /896 channels | |
| | HR mode: | | |
| | 0.5mm x 80 rows /1792 channels | | |
| | SHR mode: | | |
| | 0.25mm x 160 rows /1792<br>channels | | |
| Reconstruction Matrix | NR mode: 512×512 | 512 x 512 | |
| | HR & SHR mode: 512×512, 1024×1024 | | |
| FOV (field of view) | 320/500mm in diameter | 180/240/320/400/500 mm in<br>diameter | |
| Wedge filter Types | Three types | Three types | |
| | Small1:FOV M | Small :FOV SS, S | |
| | Large1:FOV L | Medium :FOV M | |
| | Large2:FOV L | Large :FOV L, LL | |
| X-ray tube capacity | 5.7 MHU | 7.5 MHU | |
| Maximum tube cooling rate | 1,026 kHU/min | 1,386 kHU/min | |
| Continuous tube cooling rate | 660 kHU/min | 1,008 kHU/min | |
| Focal spots size | Small: Two types | Small: One type | |
| | Large: Four types | Large: One type | |
| CPU memory | 256 Gbyte or more | 12 Gbyte or more | |
| Image reconstruction time | Up to 80 images/s | Up to 30 images/s (0.033 s/image) | |
| (Max. speed) | (0.0125 s/image) | Option: | |
| | | Up to 60 images/s (0.017 s/image) | |
| Magnetic disk | 9.8 TB or more | 915 GB or more | |
| | Raw data: 4,000 rotations or more | Raw data: 4000 rotations | |
| | | Image data: 500,000 images | |
| Item | Aquilion Precision (TSX-304A/2)<br>V8.6 with FIRST 3.0 | Aquilion PRIME, TSX-303A/A, /B<br>and /F v6.00 | |
| 510(k) Number | This submission | K141741 | |
| BD-RE(Blu-ray Disc) | Image data: 2,000,000 images or<br>more (1024 x 1024 pixel images) | N/A | |
| | 25 GB<br>Image data:10,000 images (1024 x<br>1024 pixel images) | | |
| Noise reduction processing | AIDR 3D<br>AIDR3D Enhanced | QDS<br>AIDR 3D<br>AIDR3D Enhanced | |
| Metal artifact reduction | SEMAR(Volume, Helical)<br>SEMAR (ECG gated) | SEMAR(Volume, Helical)<br>SEMAR (ECG gated) | |
| Gantry opening size | 780 mm in diameter | 780 mm in diameter | |
| Tilt angle | $\pm$ 30° | $\pm$ 30° | |
| Patient Couch | | | |
| Vertical movement: | 2050mm | 2390mm | |
| Couch top stroke | 1750mm | 1890mm | |
| Frame slide stroke | 330mm | N/A | |
| Scan permissible range: | 2000mm | 2000mm | |
| Conventional | 1700mm | 1500mm | |
| Scan permissible range: | 1950mm | 1950mm | |
| Helical | 1650mm | 1450mm | |
| Couch-top Speed of<br>movement (fast) | 200mm/s | 200mm/s | |
| Vertical movement stroke | 520mm | 568mm | |
| Couch-top width | 470mm | 470mm | |
| Lowest couch height | 420mm | 332mm | |
| Max. guaranteed weight | 300kg | 300kg | |
| Footswitch | Standard | Standard | |
| Lateral movement unit | Optional | Optional | |
| Movement speed | 20mm/s | 10mm/s | |
| Lateral Stroke (right and<br>left) | 170mm (85mm in the lateral<br>direction) | 84mm (42mm in the lateral<br>direction) | |
| FIRST 3.0 | Optional | N/A | |
{7}------------------------------------------------
{8}------------------------------------------------
## 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1, IEC60601-1-3, IEC60601-1-6, IEC60601-1-6, IEC60601-2-28, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25, NEMA XR-26 and NEMA XR-29. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
## 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met.
# Image Quality Evaluation
CT image quality metrics were performed, utilizing phantoms, to assess CT number accuracy, contrast-to noise ratio, uniformity, slice sensitivity profile, modulation transfer function, low contrast detectability, standard deviation of noise and noise power spectra. The Aquilion Precision in NR mode is substantially equivalent to the predicate device. The HR and SHR modes yield improved high contrast spatial resolution with some tradeoff in noise and LCD relative to NR mode. The FIRST algorithm demonstrates equivalent or improved image quality relative to FBP for all modes. A high contrast spatial resolution claim of "up to 46.1 lp/cm based upon the 2% of the MTF of images of a wire test object acquired in HR mode, under the axial scan mode and reconstructed with FC90" is supported. As demonstrated by the performance of FIRST at reduced mA values for standard deviation, MTF, and visual LCD as compared to FBP at higher mA values a qualitative claim of Dose Reduction with FIRST is supported.
# Confirmation of Spatial Resolution
The spatial resolution claim was confirmed by measuring the MTF results of seven Aquilion Precision systems. It was determined that a high contrast spatial resolution claim of up to 49.5 lp/cm, based upon the 2% of the MTF of images of a wire test object acquired in HR mode and reconstructed with FC90, is supported.
# Low Contrast Detectability Equivalence in Normal Resolution vs High Resolution Modes
A study was conducted utilizing the Head and Body MITA-FDA LCD phantoms to establish a baseline LCD value (AUC) in Normal Resolution (NR) mode using a protocol based on Toshiba's standard adult abdomen protocol and standard adult head protocol, with a fixed technique at a clinically realistic CTDI, and reconstructed with Filtered Backprojection (FBP), AIDR Enhanced, and FIRST. Results of the study demonstrated that scanning in high resolution mode (HR) and in super high resolution mode (SHR) may result in equivalent or a modest increase in Xray dose as compared to scanning in normal resolution mode (NR).
# Size Discrimination Performance
A model observer study was conducted comparing detectability of a 0.5 mm change in object radius in Normal Resolution vs High Resolution mode utilizing the Catphan 10HU contrast rods, 4-9mm in diameter, which demonstrated a 50% improvement in size discrimination with High Resolution mode compared to Normal Resolution mode at the same dose.
{9}------------------------------------------------
Alternatives to clinical images of cardiac and extremities using cardiac stent and cadaver A study was conducted to provide cadaver and cardiac stent images for certain anatomical regions in lieu of sample clinical images acquired in NR and HR mode since these examples were not available. This study demonstrated that NR mode and predicate device images were substantially equivalent and that the images acquired in HR and SHR modes showed improved spatial resolution as compared to images acquired in NR mode.
Representative diagnostic images, reviewed by an American Board Certified Radiologist, including head, chest, abdomen/pelvis, extremity and cardiac exams were also obtained using the subject device which demonstrates that the device produces images of diagnostic quality and; therefore, performs as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
Cybersecurity documentation, per the FDA cybersecurity premarket guidance document "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices " issued on October 2, 2014, is also included as part of this submission.
Additionally, testing of the subject device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
### 18. CONCLUSION
The Aquilion Precision (TSX-304A/2) V8.6 with FIRST 3.0 performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device has demonstrated substantial equivalence to the predicate device and is safe and effective for its intended use.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.