K170177 · Toshibamedical Systems Corporation · JAK · Jun 30, 2017 · Radiology
Device Facts
Record ID
K170177
Device Name
Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1
Applicant
Toshibamedical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Jun 30, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Indications for Use
This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to imaqe whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc. The Aquilion ONE has the capability to provide volume sets of the entire These volume sets can be used to perform specialized studies, orqan. usinq indicated software/hardware, of the whole orqan by a trained and qualified physician. FIRST 2.1 is an iterative reconstruction alqorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.
Device Story
Aquilion ONE (TSX-305A/3) V8.3 is a multi-slice helical CT scanner; consists of gantry, couch, and console. Captures cross-sectional volume data sets of whole organs in single rotation. Incorporates FIRST 2.1 iterative reconstruction algorithm to reduce radiation exposure dose and improve high-contrast spatial resolution. Operated by trained physicians/technicians in clinical settings. Output consists of diagnostic-quality cross-sectional images. Healthcare providers use images for specialized organ studies and clinical decision-making. Benefits include reduced patient radiation dose and improved image quality compared to filtered back projection (FBP). Includes suRE Subtraction Angio and suRE Subtraction Iodine Mapping for enhanced visualization of contrast agents and blood vessels.
Clinical Evidence
Bench testing using phantoms validated spatial resolution, CT number accuracy, uniformity, and noise properties. Dose reduction study (59.2-82.4% reduction) compared FIRST 2.1 to FBP using model observer approach. Clinical evaluations conducted for suRE Subtraction Angio and Iodine Mapping using visual assessment scoring by radiologists; results showed improved visualization of contrast enhancement and abdominal organs. Diagnostic quality confirmed by American Board Certified Radiologist review of representative images.
Indicated for patients requiring cross-sectional volume imaging of the whole body, including head and whole organs (e.g., brain, heart, pancreas). Used by trained physicians for specialized studies. No specific age or gender contraindications stated.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Aquilion ONE Vision with FIRST 2.0 (CCRS-001B) V7.4 (K161009)
Submission Summary (Full Text)
{0}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 30, 2017
Toshiba Medical Systems Corporation % Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K170177
Trade/Device Name: Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1 Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: May 26, 2017 Received: May 30, 2017
Dear Mr. Tadeo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{1}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Michael D.'Hara
For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
### 510(k) Number (if known)
### K170177
### Device Name
Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1
### Indications for Use (Describe)
This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to imaqe whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire These volume sets can be used to perform specialized studies, orqan. usinq indicated software/hardware, of the whole orqan by a trained and qualified physician.
FIRST 2.1 is an iterative reconstruction alqorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.
Type of Use (Select one or both, as applicable)
ك Prescription Use (Part 21 CFR 801 Subpart D)
- Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
1.
## 510(k) SUMMARY
- SUBMITTER'S NAME: Toshiba Medical Systems Corporation 1385 Shimoishigami Otawara-Shi, Tochigi-ken, Japan 324-8550
- 2. OFFICIAL CORRESPONDENT: Akinori Hatanaka Senior Manager, Regulatory Affairs and Vigilance
- 3. ESTABLISHMENT REGISTRATION: 9614698
#### 4. CONTACT PERSON:
Orlando Tadeo, Jr. Manager, Regulatory Affairs Toshiba America Medical Systems, Inc 2441 Michelle Drive Tustin, CA 92780 (714) 669-7459
- 5. Date Prepared:
January 17, 2017
#### 6. TRADE NAME(S):
Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1
#### COMMON NAME: 7.
System, X-ray, Computed Tomography
- 8. DEVICE CLASSIFICATION (Regulatory Class, CFR Reference, Name): Class II (per 21 CFR 892.1750, Computed Tomography X-ray System)
#### 9. PRODUCT CODE / DESCRIPTION:
JAK / Computed Tomography X-Ray System
{4}------------------------------------------------
#### 10. PERFORMANCE STANDARD:
This device conforms to applicable Performance Standards for Ionizing Radiation Emitting Products [21 CFR, Subchapter J, Part 1020]
| Product | Marketed | Regulation | Regulation | Product Code | 510(k) | Clearance |
|------------------------------------------------------------------|------------------------------------------|--------------------|----------------------------------------|---------------------------------------------------|---------|------------|
| | by | Number | Name | | Number | Date |
| Aquilion<br>ONE Vision<br>with FIRST<br>2.0 (CCRS-<br>001B) V7.4 | Toshiba<br>America<br>Medical<br>Systems | 21 CFR<br>892.1750 | Computed<br>Tomography<br>X-ray System | JAK:<br>System, X-ray,<br>Tomography,<br>Computed | K161009 | 07/22/2016 |
| Aquilion<br>ONE (TSX-<br>305A/3)<br>V7.3 | Toshiba<br>America<br>Medical<br>Systems | 21 CFR<br>892.1750 | Computed<br>Tomography<br>X-ray System | JAK:<br>System, X-ray,<br>Tomography,<br>Computed | K160587 | 06/09/2016 |
#### 11. PREDICATE DEVICE:
#### 12. REASON FOR SUBMISSION:
Modification to a previously cleared device
#### 13. DEVICE DESCRIPTION:
Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1 is a whole body multi-slice helical CT scanner, consisting of a gantry, couch and a console used for data processing and display. This device captures cross sectional volume data sets used to perform specialized studies, using indicated software/hardware, by a trained and qualified physician. This system is based upon the technology and materials of previously marketed Toshiba CT systems. In addition, the subject device incorporates the latest iterative reconstruction technology, FIRST 2.1, intended to reduce exposure dose while maintaining and/or improving image quality.
#### 14. INDICATIONS FOR USE:
This device is indicated to acquire and display cross sectional volumes of the whole body, to include the head, with the capability to image whole organs in a single rotation. Whole organs include but are not limited to brain, heart, pancreas, etc.
The Aquilion ONE has the capability to provide volume sets of the entire organ. These volume sets can be used to perform specialized studies, using indicated software/hardware, of the whole organ by a trained and qualified physician.
FIRST 2.1 is an iterative reconstruction algorithm intended to reduce exposure dose and improve high contrast spatial resolution for abdomen, pelvis, chest, cardiac, extremities and head applications.
#### 15. SUBSTANTIAL EQUIVALENCE:
The Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1, is substantially equivalent to the Aquilion ONE (TSX-305A/3) V7.3, which received premarket clearance under K160587, marketed by Toshiba America Medical Systems. The changes made to the subject device include the addition
{5}------------------------------------------------
of FIRST 2.1, an iterative reconstruction algorithm that allows the exposure dose to be reduced while maintaining and/or improving image quality as seen when using FBP (filtered back projection) and improves spatial resolution over FBP. A comparison of the technological characteristics between the subject and the predicate devices is included below.
| ltem | Aquilion ONE<br>(TSX-305A/3) V8.3<br>with FIRST 2.1 | Aquilion ONE<br>(TSX-305A/3)<br>V7.3 | Aquilion ONE Vision<br>with FIRST 2.0<br>(CCRS-001B) V7.4 |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------|--------------------------------------------------------------------------------------------------|
| 510(k) Number | This submission | K160587 | K161009 |
| Exposure Dose Reduction<br>(Anatomical Region) | AIDR 3D (Whole<br>Body)<br>FIRST 2.1<br>(Abdomen, pelvis,<br>chest, cardiac,<br>extremities and<br>head) | AIDR 3D (Whole<br>Body) | AIDR 3D (Whole<br>Body)<br>FIRST 2.0<br>(Abdomen, pelvis,<br>chest, cardiac, and<br>extremities) |
| Quantitative Dose<br>Reduction Claim | Yes | None | Yes |
| Image Quality Claim | Improved Spatial<br>Resolution over<br>Filtered Back<br>Projection | No Change | Improved Spatial<br>Resolution over<br>Filtered Back<br>Projection |
| sure Subtraction Angio | Available | Not Available | New Feature |
| suRE Subtraction lodine Mapping | Available | Not Available | New Feature |
#### 16. SAFETY:
The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the following standards IEC60601-1, IEC60601-1-2, IEC60601-1-2, IEC60601-1-3, IEC60601-1-4, IEC60601-1-6, IEC60601-2-28, IEC60601-2-32, IEC60601-2-44, IEC60825-1, IEC62304, IEC62366, NEMA PS 3.1-3.18, NEMA XR-25 and NEMA XR-26. Additionally, this device complies with all applicable requirements of the radiation safety performance standards, as outlined in 21 CFR §1010 and §1020.
#### 17. TESTING
Risk analysis and verification/validation testing conducted through bench testing demonstrate that the established specifications for the device have been met. CT image quality metrics were performed, utilizing phantoms, which validated that the subject device is substantially equivalent to the predicate device with regard to spatial resolution, CT number magnitude/uniformity, noise properties and low contrast detectability/CNR performance.
## Image Quality Evaluation
CT image quality metrics were performed, utilizing phantoms, which demonstrated that the subject device is substantially equivalent to or demonstrates an improvement to the predicate
{6}------------------------------------------------
device with regard to contrast-to noise ratio, CT number accuracy, uniformity, slice sensitivity profile, modulation transfer function, low contrast detectability, line pair gauge, standard deviation of noise and noise power spectra.
### Quantitative Dose Reduction/Spatial Resolution Evaluations
A dose reduction study was conducted and based on the results, a dose reduction claim with the range 59.2 to 82.4% is supported, as well as a claim of dose reduction that demonstrated a 49.2% noise reduction with FIRST on the Aquilion ONE Genesis Edition. In addition, the testing demonstrates superior LCD performance with FIRST.
These claims were evaluated by comparing Toshiba's iterative reconstruction algorithm, FIRST version 2.1, to filtered backprojection on the Aquilion ONE Genesis Edition system. The dose reduction value was established by demonstrating comparability of low contrast detectability at full dose reconstructed with filtered backprojection (FBP), and at reduced dose reconstructed with FIRST 2.1 (IR). A model observer approach which incorporates some aspects of human vision was used for investigation and a non-inferiority analysis conducted. The noise reduction value was established by comparing the standard deviation of noise magnitude at full dose with FBP with the standard deviation of noise magnitude at variety of dose reduction levels using IR. The study was designed to use clinically realistic dose levels (i.e. acquisition techniques).
## PUREViSION Optics Quantitative LCD and Noise Improvement
The subject device demonstrated improvements in low contrast detectability and noise reduction at the same dose as evaluated via model observer studies (MITA-FDA LCD Head and MITA-FDA LCD Body phantoms).
## suRE Subtraction Angio Evaluation
A clinical evaluation was conducted to assess whether or not the visibility of contrast enhancement was improved using 80855ubtraction Angio over the predicate device. Utilizing a visual assessment scoring method by doctors and clinical case examples it was determined that the visualization of contrast enhancement, especially blood vessels, are improved when compared to the predicate device.
## suRE Subtraction Iodine Mapping Evaluation
In order to determine if Subtraction lodine Mapping functions as intended as an adjunct visualization tool to assess the enhancement of the visibility of contrast agents and the ability to provide additional information of abdominal organs, a clinical evaluation was conducted. Utilizing a visual assessment scoring method of clinical case examples it was determined that បឋម Subtraction lodine Mapping provides additional information for the assessment of abdominal organs, differential enhancement mapping images are useful visualization tools and CE Boost images provide improved CNR.
Representative diagnostic images, reviewed by an American Board Certified Radiologist, including head, chest, abdomen/pelvis, extremity and cardiac exams were also obtained using the subject device which demonstrates that the device produces images of diagnostic quality and; therefore, performs as intended.
Software Documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
{7}------------------------------------------------
Additionally, testing of the subject device was conducted in accordance with the applicable standards published by the International Electrotechnical Commission (IEC) for Medical Devices and CT Systems.
#### 18. CONCLUSION
The Aquilion ONE (TSX-305A/3) V8.3 with FIRST 2.1 performs in a manner similar to and is intended for the same use as the predicate device, as indicated in product labeling. Based upon this information, conformance to standards, successful completion of software validation, application of risk management and design controls and the performance data presented in this submission it is concluded that the subject device is substantially equivalent in safety and effectiveness to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.