K173460 · Memory MD, Inc. · GWQ · Feb 16, 2018 · Neurology
Device Facts
Record ID
K173460
Device Name
NeuroEEG
Applicant
Memory MD, Inc.
Product Code
GWQ · Neurology
Decision Date
Feb 16, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.1400
Device Class
Class 2
Attributes
Pediatric
Indications for Use
NeuroEEG is intended for prescription use to acquire, record, transmit, and display electrical brain activity of patients of all ages.
Device Story
NeuroEEG is a 16-lead portable, non-invasive electroencephalograph (EEG) system. It utilizes electrodes placed according to the international 10-20 standard to capture electrical brain activity. The device transmits signals via a wireless Bluetooth channel to a stationary PC or laptop running MemoryMD software. It is intended for use in medical practices, rehabilitation institutions, diagnostic centers, neurosurgical clinics, OR, ICU, ER, and clinical research institutes under the direction of a licensed medical professional. The device does not provide diagnostic conclusions; it serves as an assessment aid for healthcare providers to view and interpret EEG data. The system operates on an internal battery and maintains patient isolation by having no connection to power supplies or PCs during registration. It benefits patients by enabling the acquisition and display of physiological brain activity for clinical assessment.
Clinical Evidence
Bench testing only. The device demonstrated compliance with electrical safety and electromagnetic compatibility standards, including AAMI/ANSI ES60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, and IEC 60601-2-26. Software verification and validation were conducted per FDA guidance and IEC 62304, confirming that all functions, performance requirements, and hazard mitigations met predetermined acceptance criteria.
Technological Characteristics
16-lead EEG system; 24-bit A/D resolution; 500 Hz sampling rate; 2.4 GHz wireless Bluetooth connectivity; internal battery power; 10-20 electrode system compatibility; input impedance min 1GOhms; frequency band 0-80 Hz; notch filtering 100/60/50 Hz. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-1-11, IEC 60601-2-26, and FCC Part 15. Software lifecycle processes follow IEC 62304.
Indications for Use
Indicated for prescription use to acquire, record, transmit, and display electrical brain activity of patients of all ages.
Regulatory Classification
Identification
An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2018
MemoryMD Inc. Abdus-Salaam Muwwakkil Chief Quality Officer 205 East 42nd St 14th FL New York, New York 10017
Re: K173460
Trade/Device Name: NeuroEEG Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWQ Dated: November 6, 2017 Received: November 21, 2017
Dear Abdus-Salaam Muwwakkil:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Michael J. Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K173460
Device Name NeuroEEG
Indications for Use (Describe)
NeuroEEG is intended for prescription use to acquire, record, transmit, and display electrical brain activity of patients of all ages.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K173460
## 510(k) Summary
| Identification of Submitter | |
|--------------------------------------|---------------------------------------------------------------------------|
| Primary Contact Name | Abdus-Salaam Muwwakkil,<br>Chief Quality Officer<br>MemoryMD Inc. |
| Address: | 205 E 42nd Street, 14th Floor<br>New York, New York 10017,<br>USA |
| Telephone: | 215-341-6373 |
| Fax: | 646-304-1064 |
| Email: | abdus@memorymd.com |
| Name / Address of<br>Manufacturer: | MemoryMD Inc.<br>67 35th street, c250<br>Brooklyn, New York 11232,<br>USA |
| Date Prepared | 11/06/2017 |
| Identification of the subject device | |
| Device Proprietary Name: | NeuroEEG |
| Common Name: | Full-Montage Standard<br>Electroencephalograph |
| Regulation: | 21 CFR 882.1400 |
| Review Panel: | Neurology |
| Product Code: | GWQ |
| Classification Panel: | Neurology |
| Device Class: | Class II |
| Predicate Device | | | | |
|------------------|--------------------------------|-------------------------|-----------------------------------------------------------------------------------------|--|
| 510(k) Number | Classification<br>Product Code | Trade of<br>Proprietary | Manufacturer | |
| K150498 | GWQ | Discovery 24 | Brainmaster<br>Technologies, Inc.<br>195 Willis Street,<br>Suite 3 Bedford, OH<br>44146 | |
{4}------------------------------------------------
#### Device Description
"NeuroEEG" is a 16-lead electroencephalograph. Electrodes correspond to the international 10-20 standard. It is a portable, non-sterile, non-invasive, non-radiation emitting electroencephalogram (EEG) device that works with a stationary PC with uninterruptible power supply (UPS) or a laptop with an internal battery.
Signal transfer occurs between NeuroEEG and PC via wireless Bluetooth channel. The accompanying MemoryMD software runs on the Windows operating system (OS) Windows 8.1 and later.
This device will be used "By Prescription" pursuant to 21 CFR 801 Subpart D. The medical use of data acquired by "NeuroEEG" is to be performed under the direction and interpretation of a licensed medical professional. This device does not provide any diagnostic conclusion about a subject's condition.
The generated data serves as an assessment aid at medical practices, rehabilitation institutions, diagnostic centers, neurosurgical clinics, OR, ICU, ER, and clinical research institutes.
The device consists of the following components:
- Registration Module: ●
- Bluetooth Adapter
- Charger and USB Cable
- Connection cable of electrode system NeuroCap ●
- Connection cable of electrode system StatNet
- Cuff;
- Flash Card;
- Flash Card Reading Device; ●
- Software Flash Disk with Software Description
- . User's Guide
#### Intended Use of Device
NeuroEEG is intended for prescription use to acquire, record, transmit, and display electrical brain activity of patients of all ages.
#### Test Summary
Electrical safety and electromagnetic compatibility (EMC) tests have been performed by the accredited laboratories and show full compliance with the standards below. The device under consideration has passed the tests according to (FDA recognition number - standard designation number):
- 19-4 AAMI / ANSI ES60601-1:2005/(R)2012 And A1:2012, C1:2009/(R)2012 And . A2:2010/(R)2012 (Consolidated Text) Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
{5}------------------------------------------------
- . 19-12 - IEC 60601-1-2 Edition 4.0 2014-02, Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests
- 5-89 IEC 60601-1-6 Edition 3.1 2013-10, Medical Electrical Equipment Part 1-6: . General Requirements For Basic Safety And Essential Performance - Collateral Standard: Usabilitv
- 19-16 IEC 60601-1-11 Edition 2.0 2015-01, Medical Electrical Equipment Part 1-11: . General Requirements For Basic Safety And Essential Performance - Collateral Standard: Requirements For Medical Electrical Equipment And Medical Electrical Systems Used In The Home Healthcare Environment
- IEC 60601-2-26:2012 (Third Edition): Particular requirements for the basic safety and ● essential performance of electroencephalographs
- FCC Part 15: 2015 Subpart B .
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff. "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "General Principles of Software Validation". Verification and Validation activities have been successfully performed on the software package, including assurance that functions work as designed, performance requirements and specifications have been met, and that all hazard mitigations have been fully implemented. All testing has met the predetermined acceptance values. Traceability of the requirements specified in the requirement specifications and functional specifications is ensured during component integration, software validation and system testing.
The device under consideration has passed the test according to (Recognition Number standard):
- . 13-79: IEC 62304 Edition 1.1 2015-06, Medical Device Software - Software Life Cycle Processes.
### Comparison to predicate device and conclusion
The technological characteristics. features, specifications, materials, mode of operation, and intended use of the proposed NeuroEEG is substantially equivalent to the predicate device quoted below. The differences between the subject device and predicate device does not raise new issues of safety or effectiveness.
| | Discovery 24E<br><b>K150498</b> | "NeuroEEG" by<br>MemoryMD<br><b>K173460</b> |
|------------|---------------------------------|---------------------------------------------|
| MODALITIES | EEG | EEG |
{6}------------------------------------------------
| INDICATION FOR USE | Indicated for prescription<br>use to acquire, record and<br>display physiological and<br>data for<br>electroencephalograph<br>studies of patients of all<br>ages | NeuroEEG is intended<br>for prescription use to<br>acquire, record,<br>transmit, and display<br>electrical brain activity of<br>patients of all ages. |
|------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| SYSTEM<br>COMPONENTS | Hardware Module; Brain<br>Avatar Acquisition<br>software; Electrode cap<br>(sold separately); USB<br>Cable w/chokes; Touch-<br>Proof sockets that allow<br>use of compliant "free<br>sensors' without a cap | Registration Module,<br>Bluetooth adapter,<br>Charger and USB Cable,<br>Connection cable of<br>electrode system Neuro<br>Cap connection,<br>Connection cable<br>electrode system StatNet,<br>Cuff, Flash card, Flash<br>Card Reading Device,<br>Software Flash Disk with<br>Software description,<br>User's Guide |
| # OF EEG CHANNELS | 24 | 16 |
| A/D RESOLUTION | 24 bit | 24 bit |
| DIGITIZATION RATE | 1024 Hz | 500 Hz |
| SAMPLING RATE | 256 Hz | 500 Hz |
| IMPEDANCE CHECK | Yes | Yes |
| BATTERY CHARGE<br>CURRENT | n/a | 400 mA |
| WIRELESS<br>CONNECTION<br>CHANNEL<br>FREQUENCY | n/a | 2.4 GHz |
| ENVIRONMENT FOR<br>USE | Clinical EEG | Clinical and experimental<br>EEG; laboratory research;<br>Rehabilitation centers;<br>OR, ICU, ER |
| SOFTWARE | BrainAvatar | MemoryMD |
| PATIENT ISOLATION | optical & magnetic | No connection of the<br>product to the power<br>supply or PC during the<br>registration. |
| ISOLATION<br>VOLTAGE | >2500V | No connection of the<br>product to the power<br>supply or PC during the<br>registration. |
| Connection to PC | USB, optically and<br>magnetically isolated | Bluetooth |
| NOTCH FILTERING | 60/50 Hz | 100/60/50 Hz |
| COMMON MODE<br>REJECTION RATIO | >120 dB | >110 dB |
| GAIN | 16 | 6 |
| INPUT IMPEDANCE | > 1000GOhms | min 1GOhms |
| INPUT NOISE | 1.5uV at input | Less than 2 uV |
| SELF-Calibration | Yes | Yes |
| FREQUENCY BAND | 0.5 - 70 Hz | 0 - 80 Hz |
| Input Range | 100 mV p-p | $\pm$ 0.8V |
| Amplifier bandwidth | 0.000 - 100 Hz | 0 - 125 Hz |
| A/D accuracy | 24 bits; resolution 0.01<br>microvolts EEG, 0.4<br>microvolts DC | 24 bits; resolution 0.1<br>microvolts EEG, 0.1<br>microvolts DC |
| Power | Isolated Power via USB<br>port | Internal battery |
| Shelf Life | n/a | 5 years |
{7}------------------------------------------------
The "NeuroEEG" device is similar in intended use, operating principals and fundamental scientific technology to the following predicate device: "Discovery 24" (K150498: Date of Premarket Notification Approval: January 26, 2016).
As shown in the table below, both systems have the same manner of use, being computer-based digital EEG systems with basic functions. They have similar requirements for training and expectations of user. The safety requirements and expectations are the same. The systems have comparable performance in terms of data sampling, accuracy, speed and resolution of displays. They use identical patient connectors and methodology, being standard 10-20 EEG sensor placement.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.