Discovery 24

K150498 · Brainmaster Technologies, Inc. · GWQ · Jan 26, 2016 · Neurology

Device Facts

Record IDK150498
Device NameDiscovery 24
ApplicantBrainmaster Technologies, Inc.
Product CodeGWQ · Neurology
Decision DateJan 26, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 882.1400
Device ClassClass 2
AttributesPediatric

Indications for Use

Indicated for prescription use to acquire, record, transmit, and display physiological and data for electroencephalograph (EEG) studies of patients of all ages.

Device Story

Discovery 24 is a 24-channel electrophysiological acquisition and biofeedback device. It acquires EEG signals via standard clinical sensor placement (electrocap or free sensors). The hardware module connects to a Windows-based PC via USB. BrainAvatar software on the PC processes, displays, and saves the EEG data. Used in clinical settings by trained professionals; provides real-time EEG visualization, digital filtering, and montage editing. Enables clinicians to monitor brain activity for diagnostic studies. Benefits include high-resolution data acquisition (24-bit) and improved signal quality (high common-mode rejection ratio) for clinical EEG assessment.

Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical safety (IEC 60601-1), performance requirements for electroencephalographs (IEC 60601-2-26), and software validation testing.

Technological Characteristics

24-channel EEG acquisition system. Connectivity: USB. Power: Isolated via USB. A/D resolution: 24-bit. Digitization rate: 1024 Hz. Sampling rate: 256 Hz. Input impedance: >1000GOhm. Common mode rejection ratio: >120 dB. Frequency range: 0.5-70 Hz. Software: BrainAvatar. Patient isolation: Optical and magnetic (>2500V). Compatible with standard DIN-compliant touch-proof sensors.

Indications for Use

Indicated for prescription use to acquire, record, transmit, and display physiological and data for electroencephalograph (EEG) studies of patients of all ages.

Regulatory Classification

Identification

An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 26, 2016 Brainmaster Technologies, Inc. % Ms. Maria Griffin Official Correspondent MDI Consultants, Inc. 55 Northern Blvd. Suite 200 Great Neck, New York 11021 Re: K150498 Trade/Device Name: Discovery 24 Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO Dated: December 1, 2015 Received: December 4, 2015 # Dear Ms. Griffin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR {2}------------------------------------------------ Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Michael J. Hoffmann -A for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150498 Device Name Discovery 24 Indications for Use (Describe) Indicated for prescription use to acquire, record, transmit, and data for electroencephalograph studies of patients of all ages. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. # *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {4}------------------------------------------------ # 510(K) SUMMARY The assigned 510(k) number is: - 1. Submitter's Identification: Brainmaster Technologies, Inc. 195 Willis Street, Suite 3 Bedford. OH 44146 Contact Person: Tom Collura, President Tel: 440-232-6000 ext. 205 Date Summary Prepared: January 25, 2016 ### 2. Trade Name of the Device: BrainMaster™ Discovery 24 - 3. Common or Usual Name: full-montage standard electroencephalograph ### Classification: 4. Regulation: 21 CFR 882.1400 Product Code: GWQ #### న. Predicate Device Information: The Vanguard System, k941551 ### 6. Device Description: The BrainMaster™ Discovery 24 are two-channel, four- channel, and 24-channel electrophysiological acquisition and biofeedback devices. The 24-channel (Discovery 24) digital EEG devices use industry-accepted Windowsbased PC as the control and display console. The devices consist of the following: - BrainMaster™ Discovery 24 Hardware Module 1. - 2. BrainMaster™ software - BrainMaster™ BrainAvatar Acquisition software - 3. Electrocap, which is sold separately. The system also includes a full set of DIN-compliant touch-proof sockets, either on 4. the main housing, or in an accessory cable, that allows the use of compliant sensors for use of "free sensors" without a cap. These products are designed for use with a standard IBM PC/AT compatible computers (Windows 7 and later) and associated peripheral equipment such as keyboard, mouse, and monitor(s). When assembled with a computer, the BrainMaster™ Discovery 24 system will acquire EEG and display and save it to disc. {5}------------------------------------------------ # 7. Intended Use: Indicated for prescription use to acquire, record, transmit, and display physiological and data for electroencephalograph (EEG) studies of patients of all ages. ## 8. Technological Comparison to Predicate Devices: | PRODUCT | Cleveland Clinic<br>Vanguard System | BrainMaster<br>Discovery 24 | |--------------------------------|-------------------------------------|--------------------------------| | SCROLLED LIVE EEG | yes | yes | | SYNCH VIDEO/EEG | yes | no | | ADJUSTABLE VIEW | yes | yes | | ADJUSTABLE VIEW<br>DURATION | yes | yes | | ONSCREEN MARKERS | yes | yes | | DIGITAL FILTERING | yes | yes | | MONTAGE EDITOR | yes | yes | | REMOTE SCREEN VIEW | yes | yes | | POWER | CPU (EISA) bus | Isolated Power via USB<br>port | | # OF EEG CHANNELS | 32-64 | 24 | | A/D RESOLUTION | 12 bits | 24 bit | | DIGITIZATION RATE | 200 Hz | 1024 Hz | | SAMPLING RATE | 200 Hz | 256 Hz | | IMPEDANCE CHECK | Yes | Yes | | INDICATIONS FOR USE | Clinical EEG | Clinical EEG | | SOFTWARE | EMach | BrainAvatar | | PATIENT ISOLATION | optical | optical & magnetic | | ISOLATION VOLTAGE | >2500V | >2500V | | INTERFACE | EISA bus | USB | | NOTCH FILTERING | 60/50 Hz | 60/50 Hz | | COMMON MODE<br>REJECTION RATIO | >110 dB | >120 dB | | GAIN | 1000 | 16 | | INPUT IMPEDANCE | >100MOhm | >1000GOhm | | INPUT NOISE | 1.5uV at input | 1.5uV at input | | SELF-CAL | yes | yes | | FREQUENCY RANGE | 0.5 - 70 Hz | 0.5 - 70 Hz | | Input Range | 1250 µV p-p | 100 mV p-p | {6}------------------------------------------------ # 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: The following non-clinical tests were performed to support the determination of substantial equivalence; IEC 60601-1-2 Ed. 3: 2007 Medical Electrical Equipment Part 1-2: General Requirements for Basic Safety and Essential Performance IEC 60601-1-26 Ed. 3.0 B: 2012 Medical Electrical Equipment Part 2-26: Particular Requirements for the Basic Safety and Essential Performance of Electroencephalographs Software Validation Testing per approved procedures The devices passed all testing. # 10. Discussion of Clinical Tests Performed: Not applicable. ### 11. Discussion of Substantial Equivalence The BrainMaster system has the same intended use in the same environments as the predicate device. It has the same manner of use, being a computer-based digital EEG system with basic functions, and it has a similar requirement for training and expectations of user. The safety requirements and expectations are the same. The systems have comparable performance in terms of data sampling and accuracy, and speed and resolution of displays. They use identical patient connectors and methodology, being standard clinical EEG sensor placement. ### 12. Conclusions: Based on the information provided in this submission we conclude that BrainMaster™ Discovery 24 is substantially equivalent to the predicate and are safe and effective for its intended use.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...