PULPDENT OBA-MCP is a light-cure orthodontic adhesive composite for the direct bonding of orthodontic brackets. For dental professional use only.
Device Story
PULPDENT OBA-MCP is a light-cure orthodontic adhesive composite used by dental professionals for direct and indirect bonding of orthodontic brackets. The device incorporates modified calcium phosphate (MCP) to release fluoride, calcium, and phosphate ions into the oral environment, supporting tooth remineralization. It is formulated without Bisphenol A (BPA), BisGMA, or BPA derivatives. The adhesive is applied by the clinician and cured using a light source (20-second set time). By providing a stable bond for brackets while releasing beneficial ions, the device assists in orthodontic treatment and potentially benefits the patient by maintaining enamel health during the treatment period.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including physical property measurements (compressive strength 353 MPa, flexural strength 112.5 MPa, diametral tensile strength 51 MPa), ion release profiles, and biocompatibility testing (ISO 10993-5, ISO 7405, ANSI/ADA Specification 41) performed on the device and similar reference materials.
Technological Characteristics
Light-cure resin composite containing aliphatic dimethacrylates, diurethane dimethacrylate, barium/strontium glass, silica, sodium fluoride, and modified calcium phosphate (MCP). Standards: ISO 4049:2009, ISO 10993-5:2009, ISO 7405:2008, ANSI/ADA Standard No. 41. Form factor: stiff, off-white paste. Energy source: light-curing (20s).
Indications for Use
Indicated for direct and indirect bonding of orthodontic brackets in patients requiring orthodontic treatment. For dental professional use only.
Regulatory Classification
Identification
A bracket adhesive resin and tooth conditioner is a device composed of an adhesive compound, such as polymethylmethacrylate, intended to cement an orthodontic bracket to a tooth surface.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
January 17, 2018
Pulpdent Corporation % Roger Mastrony President MedTek LLC 2516 Kettle Creek Court Lincolnton, North Carolina 28092
Re: K173020
Trade/Device Name: OBA-MCP, Orthodontic Bracket Adhesive with MCP Regulation Number: 21 CFR 872.3750 Regulation Name: Bracket Adhesive Resin And Tooth Conditioner Regulatory Class: Class II Product Code: DYH Dated: December 12, 2017 Received: December 18, 2017
Dear Roger Mastrony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
# Mary S. Runner -S
For Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K173020
Device Name
PULPDENT OBA-MCP - Orthodontic Bracket Adhesive with MCP
#### Indications for Use (Describe)
PULPDENT OBA-MCP is a light-cure orthodontic adhesive composite for the direct bonding of orthodontic brackets. For dental professional use only.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------|
| <span style="font-family: Arial;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
# 510(k) SUMMARY K173020
DATE OF SUBMISSION: 1/11/18
OWNER: Pulpdent Corporation 80 Oakland Street Watertown, MA 02472 USA
- CONTACT: Roger S. Mastrony 2516 Kettle Creek Court Lincolnton NC 28092 Tel: 1 203 640 5047 / Email: rogermastrony@gmail.com
TYPE OF SUBMISSION: Traditional 510(k).
## DEVICE:
Trade Name: PULPDENT OBA-MCP- Orthodontic Bracket Adhesive with MCP Device Class: II Review Panel: Dental Classification Name: Orthodontic Bracket Adhesive FDA Product Code: DYH; 21 CFR Part 872.3750
# PREDICATE DEVICE:
PULPDENT OBA-Orthodontic Bracket Adhesive, Pulpdent Corporation, K911286
# REFERENCE DEVICE:
PULPDENT Solo Flowable Composite with MCP, Pulpdent Corporation, K153249
ADDITIONAL PREDICATE DEVICES:
Comparable chemical compositions for the below referenced devices can be found in the Safety Data Sheets for the each product.
3M Unitek Transbond XT Light Cure Adhesive, 3M, K880393 (file 046 in the e-copy)
GC Fuji ORTHO LC Liquid, Fuji, K950556 (file 045 in the e-copy)
DESCRIPTION: PULPDENT OBA-MCP is a light-cure orthodontic adhesive composite containing modified calcium phosphate that releases fluoride, calcium, and phosphate ions into the oral environment. OBA-MCP is not formulated with Bisphenol A. BisGMA, or BPA derivatives.
The standards applicable to the subject device, and referenced within the submission, include: ાડવ 4049:2009 Dentistry – Polymer based filling, restorative and luting materials, ISO 10993-5:2009 Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity, ISO 7405:2008 Dentistry -Evaluation of biocompatibility of medical devices used in dentistry, ANSI/ADA Standard No. 41 Evaluation of Biocompatibility of Medical Devices Used in Dentistry - ADA41-2005D.
INDICATIONS FOR USE: PULPDENT OBA-MCP is a light cure orthodontic adhesive composite for the direct and indirect bonding of orthodontic brackets. For dental professional use only.
{4}------------------------------------------------
80 Oakland Street Watertown, MA 02472 USA
# BIOCOMPATIBILITY EVALUATION:
PULPDENT OBA-MCP is substantially equivalent to the predicate device, Pulpdent OBA, which has an established history of safe and effective use as a dental material. Pulpdent OBA-MCP is also substantially equiavalent in chemcial composition to the reference device, Pulpdent Solo Flowable Composite with MCP, which has been tested with a Cytotoxicity Test (ISO 10993-5 2009) and an Implantation Test (ANSI/ADA Specification 41 and ISO 7405) and found to be biocompatible.
# DEVICE COMPARISON WITH PREDICATE AND REFERENCE PRODUCTS
The predicate product has been found to be substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Device under CFR 872.3750, Code DYH.
| | Pulpdent OBA – MCP<br>Orthodontic Bracket<br>Adhesive with MCP | Predicate Product:<br>PULPDENT OBA, K911286,<br>Orthodontic Bracket Adhesive | Reference Product: optional<br>PULPDENT Solo Flowable<br>Composite with MCP, K153249 |
|----------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| DESCRIPTION | Light cured, adhesive<br>composite, with modified<br>calcium phosphate | Light cured, adhesive composite | Light cured, resin composite with<br>modified calcium phosphate |
| TECHNOLOGY /<br>MECHANISM OF<br>ACTION | Direct and indirect bonding of<br>orthodontic brackets | Direct and indirect bonding of<br>orthodontic brackets | Dental cavity filler and restorative |
| INDICATIONS FOR<br>USE | Light cure orthodontic adhesive<br>composite for the direct and<br>indirect bonding of orthodontic<br>brackets | Orthodontic bracket adhesive | Dental restorative, base and liner |
| COMPOSITION | Aliphatic Dimethacrylates,<br>Diurethane dimethacrylate, Bis<br>(2-(Methacryloyloxy) Ethyl)<br>Phosphate, Barium glass,<br>Strontium glass, MCP (Modified<br>calcium phosphate),<br>Silica, Sodium Fluoride,<br>Camphorquinone (photo-<br>initiators) | 1,6 Hexamethylene<br>Dimethacrylate, BisGMA (Bis [4-2<br>(Hydroxy-3-Methacryloxy Propyl)<br>Phenyl] Propane-2),<br>Dimethylaminoethylmethacrylate,<br>Silanated Barium Glass, Fumed<br>Silica, Sodium Fluoride,<br>Camphorquinone, (photo-<br>initiators) | Aliphatic Dimethacrylates,<br>Diurethane dimethacrylate, Bis (2-<br>(Methacryloyloxy) Ethyl) Phosphate,<br>Barium glass, Strontium glass, MCP<br>(Modified calcium phosphate),<br>Sodium Fluoride, Camphorquinone<br>(photo-initiators), colorants |
| PHYSICAL PROPERTIES | | | |
| Appearance | Stiff, off-white paste | Stiff, off-white paste | Tooth shade paste |
| Specific gravity | 1.9 g/ml | 1.820 g/ml | 1.810 g/ml |
| Water Absorption | 1.01% | | 1.73% |
| Fluoride Release | 422 µg/g | Yes | Yes |
| Phosphate Release | 182 µg/g | No | Yes |
| Calcium Release (pH 4) | 96 µg/g | No | Yes |
{5}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
| | <i>Pulpdent</i> OBA – MCP<br>Orthodontic Bracket<br>Adhesive with MCP | <i>Predicate Product:</i><br>PULPDENT OBA, K911286,<br>Orthodontic Bracket Adhesive | <i>Reference Product:</i> optional<br>PULPDENT Solo Flowable<br>Composite with MCP, K153249 |
|-------------------------------|-----------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| Compressive<br>Strength | 353 MPa | --- | 330 ±47.9 |
| Flexural Strength | 112.5 MPa | --- | 129 ±18.7 MPa |
| Diametral Tensile<br>Strength | 51 MPa | --- | 46.5 MPa |
| Light cure set times | 20 seconds | 20 seconds | 20 seconds |
## Similarities
Pulpdent OBA-MCP and the predicate product, OBA, both share the same technology and mechanism of action, as an orthodontic bracket adhesive. They have essentially the same physical properties and nearly identical chemical formulations.
## Differences
Pulpdent OBA-MCP differs from Pulpdent OBA with the addition of MCP (Modified Calcium Phosphate). Additionally, the BisGMA has been removed from the original OBA formulation.
#### Discussion of the Differences
MCP, or Modified Calcium Phosphate, is a methacrylate moiety, with stabilized calcium and phosphate ions, that is used as a filler in dental materials. In the natural course of the tooth, and its demineralization and remineralization processes in acidic oral environments, there is an ongoing exchange of biologically available calcium and phosphate ions. The addition of calcium and phosphate ions to dental products can be seen in many premarket approved devices. Biocompatability testing performed on products with a similar chemcial composition (including the reference device) have confirmed that a calcium and phospahte addition is biocompatible (see file 012 Biocompatabiltiy Evaluation). It has also been shown that the enamel bond strength of orthodontic materials containing a calcium phosphate moiety exhibits a comparable bond strength to commercially available cements without calcium phosphate. (Zhang, et.al., file ecopy)
#### Conclusion
Taking into account the above similarities and differences, Pulpdent OBA-MCP Orthodontic Bracket Adhesive with MCP, is substantially equivalent, in design, performance and indications for use, to the predicate product, Pulpdent OBA - Orthodontic Bracket Adhesive.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.