Pulpdent Solo Flowable Composite with MCP is a fluoride, calcium and phosphate-releasing flowable composite used by dental professionals as a restorative, base and liner.
Device Story
Pulpdent Solo Flowable Composite with MCP is a light-cured, radiopaque, glass-filled resin restorative material. It incorporates modified calcium phosphate (MCP) filler to release calcium and phosphate ions into the oral environment, with optional fluoride release. Used by dental professionals in clinical settings as a restorative, base, or liner for dental cavities. The material is applied to the cavity and light-cured for 20 seconds. The release of ions is intended to support the dental restoration process. It is formulated without Bisphenol A, BisGMA, or BPA derivatives.
Clinical Evidence
Bench testing only. Biocompatibility confirmed via Cytotoxicity Test (ISO 10993-5:2009) and Implantation Test (ANSI/ADA Specification 41 and ISO 7405). Physical property testing included flexural strength (129 ± 18.7 MPa), compressive strength (330 ± 47.9 MPa), diametral tensile strength (46.5 MPa), polymerization shrinkage (2.1%), water absorption (1.73%), and water solubility (0.09%).
Technological Characteristics
Light-cured, glass-filled resin composite. Composition: Diurethane dimethacrylate, Bis (2-(Methacryloyloxy) Ethyl) Phosphate, Barium glass, Strontium glass, Camphorquinone, Sodium fluoride (optional), MCP (Modified calcium phosphate), Colorants. Form factor: Tooth-shade paste. Curing: Light-cure (20 seconds). Biocompatibility: ISO 10993-5, ISO 7405, ANSI/ADA Specification 41.
Indications for Use
Indicated for use by dental professionals as a restorative, base, and liner for dental cavities.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are black against a white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 1, 2016
PULPDENT CORPORATION c/o Roger Mastrony MedTek LLC. 2516 Kettle Creek Court Lincolnton, North Carolina 28092
Re: K153249
Trade/Device Name: Pulpdent Solo Flowable Composite with MCP Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth-shade resin material Regulatory Class: Class II Product Code: EBF Dated: March 25, 2016 Received: March 28, 2016
Dear Mr. Roger Mastrony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{1}------------------------------------------------
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Pulpdent Solo Flowable Composite with MCP
Indications for Use:
Pulpdent Solo Flowable Composite with MCP is a fluoride, calcium and phosphate-releasing flowable composite used by dental professionals as a restorative, base and liner.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{3}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
### 510(k) SUMMARY
DATE OF SUBMISSION: April 1, 2016
OWNER: Pulpdent Corporation 80 Oakland Street Watertown, MA 02472 USA
CONTACT: Roger S. Mastrony 2516 Kettle Creek Court Lincolnton NC 28092 Tel: 1 203 640 5047 Email: rogermastrony@gmail.com
TYPE OF SUBMISSION: Traditional 510(k).
#### DEVICE:
Trade Name: Pulpdent Solo Flowable Composite with MCP Device Class: II Classification Name: Tooth-shade resin material FDA Product Code: EBF
#### PREDICATE DEVICE:
Pulpdent ACTIVA BioACTIVE-RESTORATIVE, K130223 (Pulpdent RMGI Fill)
NOTE: Pulpdent RMGI Fill was the working name for ACTIVA BioACTIVE-RESTORATIVE that was used during the design and development of the product before a marketing name was decided upon. It is the same exact formula.
#### REFERENCE DEVICE:
Bosworth AEGIS V with ACP, K103585
#### DESCRIPTION:
Pulpdent Solo Flowable Composite with MCP is a light-cured, radiopaque, glass-filled resin with MCP (modified calcium phosphate filler) that releases calcium and phosphate ions into the oral environment and which may be formulated to release fluoride ions. Solo Flowable Composite with MCP contains no Bisphenol A, no BisGMA and no BPA derivatives. It is available in tooth shades, has low solubility, low water sorption and high physical properties.
#### INTENDED USE:
Pulpdent Solo Flowable Composite with MCP is a fluoride, calcium and phosphate-releasing flowable composite used by dental professionals as a restorative, base and liner.
{4}------------------------------------------------
80 Oakland Street Watertown, MA 02472 USA
# BIOCOMPATIBILITY EVALUATION:
Pulpdent Solo Flowable Composite with MCP was tested with a Cytotoxicity Test (ISO 10993-5.2009) and an Implantation Test (ANSI/ADA Specification 41 and ISO 7405) and found to be biocompatible.
### COMPARISON WITH PREDICATE AND REFERENCE PRODUCTS:
| | Pulpdent Solo Flowable<br>Composite<br>K153249 / EBF | Predicate Product:<br>Pulpdent ACTIVA BioACTIVE<br>RESTORATIVE (RMGI Fill)<br>K130223 / EMA | Reference Product:<br>Bosworth AEGIS V with ACP<br>K103585 / EBF, EMA | |
|-------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| DESCRIPTION /<br>TECHNOLOGY | Light-cured, glass-filled resin<br>with modified calcium<br>phosphate | Dual-cure, resin-modified glass ionomer<br>with ion releasing resins and glass | Light-cured, glass-filled resin<br>with amorphous calcium<br>phosphate | |
| INTENDED USE | Dental restorative, base and<br>liner | Dental restorative* | Dental restorative and liner | |
| COMPOSITION | Diurethane dimethacrylate<br>Bis (2-(Methacryloyloxy) Ethyl)<br>Phosphate<br>Barium glass<br>Strontium glass<br>Camphorquinone<br>Sodium fluoride (optional)<br>MCP (Modified calcium<br>phosphate)<br>Colorants | Diurethane dimethacrylate<br>Bis (2-(Methacryloyloxy) Ethyl)<br>Phosphate<br>Barium glass<br>lonomer glass<br>Polyacrylic acid/maleic acid copolymer<br>Dual-cure chemistry<br>Sodium fluoride<br>Colorants. | BisGMA (Bis [4-2 (hydroxy-3-<br>methacryloxy propyl) phenyl]<br>propane-2)<br>Barium glass<br>Camphorquinone<br>ACP (Amorphous Calcium<br>phosphate) | |
| PHYSICAL PROPERTIES | | | | |
| Light cure set time | 20 seconds | 20 seconds | 20 seconds | |
| Depth of cure | 2.84 mm | 2.80 mm | | |
| Appearance | Tooth-shade paste | Tooth-shade paste | Paste | |
| Odor | Mild, characteristic | Mild, characteristic | Mild, characteristic | |
| Specific gravity | 1.810 ± 0.03 g /ml | Part A: 1.580 ± 0.02 g/mL<br>Part B: 1.620 ± 0.02 g/mL | 1.46 g/ml | |
| Flexural strength (MPa) | 129 ± 18.7 | 108.4 ± 14.7 | | |
| Flexural modulus GPa | 8.0 | ---- | | |
| Compressive strength<br>(MPa) | 330 ± 47.9 | 270 ± 14 | | |
| Diametral tensile<br>strength | 46.5 MPa | 44.4 ± 2.7 | | |
| | Pulpdent Solo Flowable<br>Composite<br>K153249 / EBF | Predicate Product:<br>Pulpdent ACTIVA BioACTIVE<br>RESTORATIVE (RMGI Fill)<br>K130223 / EMA | Reference Product:<br>Bosworth AEGIS V with ACP<br>K103585 / EBF, EMA | |
| Polymerization<br>shrinkage | 2.1 % | 1.7 % | | |
| Water absorption | 1.73% | 1.77% | | |
| Water solubility | 0.09 % | 0.25 % | | |
| Fluoride releasing | Yes (optional) | Yes | No | |
| Calcium releasing | Yes | Yes | Yes | |
| Phosphate releasing | Yes | Yes | Yes | |
{5}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
* Note: Although the intended use of Solo Flowable and ACTIVA RESTORATIVE are not identical, in dentistry, products are often used for multiple purposes. The inclusion of the indications for use does not change the intended use of the device because placing a "base' is one step in the process of cavity restoration. The device is placed in the same location for the purpose of filling a dental cavity and restoring function. The difference between a 'base' and a 'restorative' is the amount of the device material used.
## Similarities
Pulpdent Solo Flowable Composite, ACTIVA BioACTIVE-RESTORATIVE and Bosworth Aeqis V with ACP are all glass-filled resin composites that release ions into the oral environment. All the devices have the same basic structure, i.e., glass-filled resin.
Solo Flowable Composite, ACTIVA BioACTIVE-RESTORATIVE and Bosworth Aegis V with ACP are all intended to be used as dental restoratives; Solo Flowable Composite and AEGIS may also be used as a base and/or liner under cavity restorations.
Solo Flowable Composite, ACTIVA and AEGIS all release calcium and phosphate ions. Solo Flowable may be formulated to release fluoride ions like ACTIVA does.
Solo Flowable Composite and ACTIVA contain the same monomers, i.e., diurethane dimethacrylate and Bis (2-(methacryloyloxy) ethyl) phosphate.
Solo Flowable Composite and AEGIS are one-part light-cured materials. Solo Flowable Composite, ACTIVA and AEGIS all light cure in 20 seconds.
The predicate and reference medical devices have been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3690 and 872.3275.
## Differences
Solo Flowable Composite and ACTIVA are formulated with urethane dimethacrylate and Bis(2-(Methacryloyloxy) Ethyl) Phosphate. Bosworth Aegis V with ACP is formulated with BisGMA (Bis [4-2 (hydroxy-3-methacryloxy propyl) phenyl] propane-2). The difference is not significant because both monomers have an established history of use in dental materials and no substantial history of adverse reactions.
ACTIVA is a two-part, resin-modified glass ionomer composite that has three setting mechanisms: light cure, self-cure resin chemistry, and acid-base glass ionomer reaction. Solo Flowable Composite and AEGIS are one-part, light-cured materials. This difference in curing chemistry is not significant because they are both used extensively in dental composites, have no substantial history of adverse events, and give the dentists a choice of curing method based on the clinical situation.
{6}------------------------------------------------
80 Oakland Street Watertown, MA 02472 USA
# Conclusion
Taking into account the above similarities and differences, Pulpdent Solo Flowable Composite with MCP is substantially equivalent in design, composition, performance and intended use to ACTIVA BioACTIVE-RESTORATIVE and Bosworth Aegis V with ACP.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.