K130223 · Pulpdent Corporation · EMA · Mar 29, 2013 · Dental
Device Facts
Record ID
K130223
Device Name
RMGI FILL
Applicant
Pulpdent Corporation
Product Code
EMA · Dental
Decision Date
Mar 29, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Indications for Use
Pulpdent RMG/ Fill is a resin-modified glass ionomer preparation used by dental professionals as a restorative filling material.
Device Story
Pulpdent RMGI Fill is a resin-modified glass ionomer restorative material. It features a bioactive resin matrix and bioactive glass fillers that release beneficial ions into the oral environment. The device is supplied as a two-paste formulation in an automix syringe. It utilizes three setting mechanisms: light cure, self-cure resin chemistry, and an acid-base glass ionomer reaction. It is intended for use by dental professionals in clinical settings for dental restorations. The material is free of Bisphenol A, BisGMA, and BPA derivatives. Clinical benefits include its use as a restorative filling material, with performance characteristics such as flexural strength, compressive strength, and radiopacity comparable to existing dental cements.
Clinical Evidence
Bench testing only. Performance metrics reported: flexural strength (102.0 MPa), flexural modulus (4.3 GPa), compressive strength (280.0 MPa), diametral tensile strength (42.0 MPa), light cure set time (20 seconds), self-cure set time (2 min 20 sec intraoral), shear bond strength (28 MPa), radiopacity (2 mm aluminum equivalent), and polymerization shrinkage (1.7%). No clinical trial data provided.
Technological Characteristics
Resin-modified glass ionomer (RMGI) paste-paste formulation. Contains diurethane dimethacrylate, methacrylate-based monomers/oligomers, polyacrylic/maleic acid copolymer, water, barium borosilicate glass, silica, and aluminofluorosilicate ionomer glass. Setting mechanisms: light cure, self-cure, and acid-base reaction. Radiopaque. Supplied in automix syringe. No Bisphenol A, BisGMA, or BPA derivatives.
Indications for Use
Indicated for use by dental professionals as a restorative filling material in dental restorations.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
KVJo-223
MAR 2 9 2013
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
## 510(k) SUMMARY
DATE OF SUBMISSION: January 25, 2013
#### OWNER: Pulpdent Corporation
#### CONTACT:
Kenneth J. Berk 80 Oakland Street PO Box 780 Watertown, MA 02472 USA
Telephone: Fax: Email:
617-926-6666 617-926-6262 ken@pulpdent.com
#### DEVICE:
Trade Name: Pulpdent RMGI Fill Classification Name: Dental cement FDA Product Code: EMA, 21 CFR Part 872.3275
#### PREDICATE DEVICES:
Pulpdent RMGI Low Viscosity
- 3M ESPE Ketac Nano Glass Ionomer Restorative
GC Fuji Filling LC
#### INTENDED USE:
Pulpdent RMG/ Fill is a resin-modified glass ionomer preparation used by dental professionals as a restorative filling material.
## DESCRIPTION:
Pulpdent RMG/ Fill is a resin-modified glass ionomer preparation with both a bioactive resin matrix and bioactive glass fillers. In this context 'bioactive' refers to the release of beneficial ions from the resin and glass fillers into the oral environment. As a resin-modified glass ionomer, Pulpdent RMG/ Fill has three setting mechanisms: light cure, self-cure resin chemistry, and acid-base glass ionomer reaction. It contains no Bisphenol A, no BisGMA and no BPA derivatives. Pulpdent RMG/ Fill is a paste-paste formulation provided in an automix syringe that is used by dental professionals as filling material in dental restorations.
#### SUMMARY OF PERFORMANCE TESTING - BENCH
| Flexural strength | 102.0 MPa |
|---------------------------------|-----------------------------------------------------|
| Flexural modulus | 4.3 GPa |
| Compressive strength | 280.0 MPa |
| Diametral tensile strength | 42.0 MPa |
| Light cure set time | 20 seconds |
| Self-cure set time (intraoral) | 2 minutes, 20 seconds at 37°C |
| Self-cure set time (extra-oral) | 3 minutes, 30 seconds from beginning of mix at 20°C |
| Shear bond strength | 28 MPa (RMGI Fill to composite) |
| Radiopacity | Equivalent to 2 mm aluminum |
| Polymerization shrinkage | 1.7% |
1 / 5
{1}------------------------------------------------
# PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
# COMPARISON WITH PREDICATE PRODUCTS:
Pulpdent RMGI Fill is substantially equivalent in design, composition, performance and intended use to the predicate products:
| Product | Description | Intended Use | Composition |
|---------------------------------------------------------------|---------------------------------------------------------------|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pulpdent RMGI Fill | Resin-modified glass<br>ionomer in two pastes | Dental restorative | Part A: Diurethane dimethacrylate and other<br>methacrylate- based monomers and oligomers;<br>polyacrylic acid/maleic acid copolymer, water, barium<br>borosilicate glass, silica, reducing agents,<br>photoinitiator, colorants.<br><br>Part B: Diurethane dimethacrylate and other<br>methacrylate- based monomers and oligomers;<br>aluminoflurosilicate ionomer glass, silica, oxidizing<br>agents. |
| Pulpdent<br>RMGI Low Viscosity<br>K123265 | Resin-modified glass<br>ionomer in two pastes | Dental liner, base or<br>luting agent. | Part A: Diurethane dimethacrylate and other<br>methacrylate- based monomers and oligomers,<br>polyacrylic acid/maleic acid copolymer, water, barium<br>borosilicate glass, silica, reducing agents,<br>photoinitiator.<br><br>Part B: Diurethane dimethacrylate and other<br>methacrylate- based monomers and oligomers,<br>aluminoflurosilicate ionomer glass, silica, oxidizing<br>agents. |
| 3M Espe<br>Ketac Nano Glass<br>Ionomer Restorative<br>K052235 | Light cure, resin-<br>modified glass<br>ionomer in two pastes | Dental restorative | Part A: Silane-treated glass, zirconia and silica;<br>ionomer glass; BisGMA; PEGDMA; HEMA; TEGDMA.<br><br>Part B: Silane-treated ceramic; co-polymer of acrylic<br>and itaconic acids; HEMA; water. |
| GC<br>Fuji Filling LC<br>K051427 | Light cure, resin-<br>modified glass<br>ionomer in two pastes | Dental restorative | Part A: Aluminoflurosilicate ionomer glass; urethane<br>dimethacrylate; HEMA.<br><br>Part B: Polyacrylic acid; water; amorphous, fumed<br>silica; urethane dimethacrylate. |
{2}------------------------------------------------
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
| | Compressive<br>Strength<br>MPa | Flexural Strength<br>MPa | Flexural Modulus<br>GPa | Diametral<br>Tensile strength<br>(DTS) MPa |
|--------------------------------|--------------------------------|--------------------------|-------------------------|--------------------------------------------|
| Pulpdent RMGI Fill | 280 | 102 | 4.3 | 42.0 |
| Pulpdent RMGI<br>Low Viscosity | 239 | 88 | | 36 |
| 3M ESPE<br>Ketac Nano | 236 | 35 | 5.4 | 32 |
| GC<br>Fuji Filling LC | 165 | 28 | 12.3 | 15 |
From the above comparisons and the bench testing, it can be concluded that Pulpdent RMG/ Fill is substantially equivalent in design, composition, performance and intended use to the products listed above. The predicate products have been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3275 and have been used by dental professionals for more than 20 years. A search of the relevant scientific literature shows that resinmodified glass ionomers, used as a restorative filling material, have been extensively studied. See References below.
#### REFERENCES
Specific to Pulpdent RMGI Fill and RMGI Low Viscosity (K123265)
- Pameijer CH, et al. Biocompatibility of four experimental formulations [including Pulpdent 1. RMGI Fill] in subcutaneous connective tissue of rats. Pulpdent Corporation. 2011 Nov.
- 2. Pameijer CH. Report on the retention of Embrace WetBond cement and a RMGI cement (Pulpdent). Pulpdent Corporation 2012 Aug.
- 3. Pameijer CH et al. Marginal leakage with RMGI filled restorations and castings cemented with a RMGI luting cement. University of El Salvador/Argentine Dental Association, Buenos Aires, Argentina. 2012 Jun.
- 4. Pameijer CH. Microleakage of four experimental resin-modified glass ionomer restorative materials. Pulpdent Corporation. 2011 Apr.
- ട്. Pameiier CH, Zmener O. Histopathological evaluation of RMGI cement, auto- and lightcured, used as a luting agent. A subhuman primate study. Pulpdent Corporation. March 2011.
#### General
- Booth SE. Surface properties and the bioactivity of glass-ionomer dental cements and દ. related materials. PhD Thesis, University of Greenwich. 2010.
{3}------------------------------------------------
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
- 7. Comisi JC. Using bioactive materials to achieve proactive dental care. Oral Health. 2011 Dec. Dec;34-46.
- 8. Dülgergil CT. et al. Atraumatic restorative treatment with resin-modified glass ionomer material: short term results of a pilot study. Med Princ Pract. 2005 Jul-Aug;14(4):277-80.
- Daou MH, et al. Clinical evaluation of four different dental restorative materials. one year റു. results. Schweiz Monatsschr Zahnmed. 2008;118(4):290-5.
- 10. Donly KJ, et al. Clinical performance and caries inhibition of resin-modified glass ionomer cement and amalgam restorations. J Am Dent Assoc. 1999 Oct;130(10):1459-66.
- 11. Ermis RB. Two-vear clinical evaluation of four polyacid-modified resin composites and a resin-modified glass-ionomer cement in Class V restorations. Quintessence Int. 2002 Jul-Aug;33(7):542-8.
- 12. Gandolfo MG. et al. Biomimetic Remineralization of human dentin using promising innovative calcium silicate hybrid "smart" materials. Dent Mater. 2011 Nov:27(11):1055-69.
- 13. Goldstep F. Proactive intervention dentistry: a model for oral care through life. Compend Contin Educ Dent. 2012 Jun;33(6):394-6, 398-402; quiz 404, 416.
- Gomes SC, et al. Clinical and histologic evaluation of the periodontal response to 14. restorative procedures in the dog. Int J Periodontics Restorative Dent. 2005 Feb;25(1):39-47.
- Hübel S. Meiàre I. Conventional versus resin-modified glass-ionomer cement for Class II 15. restorations in primary molars. A 3-year clinical study. Int J Paediatr Dent. 2003 Jan;13(1):2-8.
- 16. Kotsanos N. Arizos S. Evaluation of a resin modified glass ionomer serving both as indirect pulp therapy and as restorative material for primary molars. Eur Arch Paediatr Dent. 2011 Jun;12(3):170-5.
- 17. Loof J et al. A comparative study of the bioactivity of three materials for dental applications. Dent Mater. 2008 May;24(5):653-9. Epub 2007 Aug 28.
- 18. Loguercio AD, et al. Five-year double-blind randomized clinical evaluation of a resinmodifed glass ionomer and a polyacid-modified resin in non-carious cervical lesions. J Adhes Dent. 2003 Winter;5(4):323-32.
- Matsuya S, et al. Structure of bioactive glass and its application to glass ionomer cement. 19. Dent Mater J. 1999 Jun;18(2):155-66.
- 20. Mitra SB, et al. Fluoride release and recharge behavior of a nano-filled resin-modified glass ionomer compared with that of other fluoride releasing materials. Am J Dent. 2011 Dec;24(6):372-8.
- 21. Mousavinasab SM, et al. Flexural strength and morphological characteristics of resinmodified glass-ionomer containing bioactive glass. J Contemp Dent Pract. 2011 Jan 1;12(1):41-6.
- Murray PE, et al. Restorative pulpal and repair responses. JADA 2001 Apr,132:482-91. 22.
{4}------------------------------------------------
## PULPDENT CORPORATION
80 Oakland Street Watertown, MA 02472 USA
Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com
- Nicholson JW, Czarnecka B. The biocompatibility of resin-modified glass-ionomer cements 23. for dentistry. Dent Mater. 2008 Dec;24(12):1702-8. :
- Pereira AF, et al. Influence of restorative techniques on marginal adaptation and dye 24. penetration around Class V restorations. Gen Dent. 2012 Jan-Feb;60(1):e17-21.
- Roberts HW, et al. Hardness of three resin-modified glass-ionomer restorative materials as 25. a function of depth and time. J Esthet Restor Dent. 2009;21(4);262-72.
- 26. Salata LA et al. Recent advances in the use of glass ionomers: bone substitutes. Rev Odontol Univ Sao Paulo. 1999 Apr/June;13(2):203-7.
- Sampaio PC, et al. Effect of conventional and resin-modified glass-ionomer liner on dentin 27. adhesive interface of Class I cavity walls after thermocycling. Oper Dent. Jul-Aug;36(4):403-12.
- 28. Stanley HR. Local and systemic responses to dental composites and glass ionomers. Adv Dent Res. 1992 Sep:6:55-64.
- 29. Tarim B. et al. Pulpal response to a resin-modified glass-ionomer material on non-exposed and exposed monkey pulps. Quintessence Int. 1998 Aug;29(8):535-42.
- 30. Yli-Urpo H, et al. Compound changes and tooth mineralization effects of glass ionomer cements containing bioactive glass (S53P4), an in vivo study. Biomaterials. 2005 Oct;26(30):5934-41.
- 31. Zhou SL, et al. In vitro study of the effects of fluoride-releasing dental materials on Remineralization in an enamel erosion model. J Dent. 2012 Mar;40(3):255-63.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the department's mission to protect the health of all Americans and provide essential human services. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 29, 2013
Mr. Kenneth J. Berk Director of Research Pulpdent Corporation 80 Oakland Street PO Box 780 WATERTOWN MA 02472 USA
Re: K130223
Trade/Device Name: Pulpdent RMGI FILL Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: January 25, 2013 Received: February 6, 2013
Dear Mr. Berk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kwame O. Ulmer -S for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Pulpdent RMGI FILL
Indications for Use:
Pulpdent RMG/ Fill is a resin-modified glass ionomer preparation used by dental professionals as filling material in dental restorations.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
~ ---------------------------------------------------------------------------------------------------------------------------------------------------------------------------
・・・・・・・・
Andrew l. Steen [ _ Steen ] _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _ _
(Division Sign-Off) Division of Anesthesiology, General Hospital nfection Control, Dental Devices
Page 1 of 1
..
K130223 510(k) Number:
.
- -
. ... . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
・・・・・・・・ :
2-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.