K052235 · 3M Company · EMA · Sep 27, 2005 · Dental
Device Facts
Record ID
K052235
Device Name
LCR
Applicant
3M Company
Product Code
EMA · Dental
Decision Date
Sep 27, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 872.3275
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Primary teeth restorations Small Class I restorations Class III and Class V restorations Transitional restorations Filling defects and undercut areas in crown preparations Core buildup where at least half the coronal tooth structure is remaining to provide structural support for the crown Laminate sandwich technique
Device Story
LCR is a light-curing, radiopaque, fluoride-releasing, resin-modified glass ionomer restorative system. It consists of a base and catalyst paste and a primer. The system is delivered via a multi-dose Clicker dispensing system or unit dose, replacing traditional powder/liquid systems to ensure consistent mixing ratios. Used by dental professionals in clinical settings for tooth restoration and core buildup. The device provides improved wear, polishability, and esthetics compared to previous powder/liquid systems. It is applied to the tooth structure to restore form and function, benefiting patients through durable, esthetic, and fluoride-releasing restorative material.
Clinical Evidence
No clinical data provided. Evidence consists of chemical composition comparison to the predicate and biocompatibility testing conducted by independent research institutes. Performance characteristics were compared to predicate devices (Vitremer, Photac-Fil, Fujii II LC) via bench testing.
Technological Characteristics
Resin-modified glass ionomer cement. Components: base and catalyst paste, primer. Features: light-curing, radiopaque, fluoride-releasing. Delivery: multi-dose Clicker system or unit dose. Materials are chemically similar to predicate Vitremer system.
Indications for Use
Indicated for primary teeth restorations, small Class I restorations, Class III and V restorations, transitional restorations, filling defects/undercuts in crown preps, core buildup (with >=50% coronal tooth structure remaining), and laminate sandwich technique.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
Predicate Devices
Vitremer Core Buildup/Restorative System (K925032)
{0}------------------------------------------------
SEP 2 7 2005
#### 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS IV.
## Submitter
| Company: | 3M ESPE |
|------------------------------------|-------------------------------------------------|
| Street: | 3M Center Bldg 260-2B-12 |
| City, State, Zip-code: | St. Paul, MN 55144 |
| Country: | U.S.A |
| Establishment Registration Number: | 2110898 |
| Contact: | Karen O'Malley<br>Regulatory Affairs Specialist |
| Phone: | 651 736-7326 |
| Fax: | 651 736-0990 |
| E-mail: | kdomalley@mmm.com |
| Date of Submission: | August 9, 2005 |
Name of Device
| Proprietary Name: | LCR |
|----------------------|---------------------------|
| Classification Name: | Dental Cement |
| Common Name: | Glass Ionomer Restorative |
## Predicate Devices:
| Vitremer Core Buildup/Restorative System....... | K925032 |
|-------------------------------------------------------------|---------|
| Photac-Fil Aplicap .................................... | K925027 |
| Fujii II LC................................................ | K993973 |
LCR is classified as a Dental Cement (21 C.F.R. § 872.3275) because it is a resin modified glass ionomer base cement used as a temporary tooth filling.
3M ESPE is submitting this Special 510(k) for modifications to Vitremer Core Buildup/Restorative System (K 925032). The modifications of Vitremer Core Buildup/Restorative System are minor changes of the chemical composition, while the basic chemical design remains the same. The change results in primarily change to the physical form of the delivered product and the means to dispense the product for placement in the dental procedure. A summary of the design change features is:
- Paste/ paste bulk and unit dose delivery system for ease of mixing and delivery as . compared to existing powder/ liquid systems.
- Consistent cement properties because of exact mixing ratio ●
- Less preparation and clean-up steps ◆
- Improved wear and polishability of the restoration .
- Improved esthetics of the restoration .
Image /page/0/Picture/16 description: The image shows the logo for 3M ESPE. Above the logo, the text "Special 510(k) LCR" is printed in a smaller font size. The 3M ESPE logo is printed in a bold, sans-serif font.
{1}------------------------------------------------
LCR is designed to retain the excellent physical, mechanical and clinical properties of existing Vitremer Core Buildup/Restorative System. Like Vitremer Core Buildup/Restorative System, LCR has the same basic chemical composition and material characteristics for patient application. LCR is a light curing, radiopaque, fluoride-releasing, resin-modified glass ionomer restorative system. The restorative system consists of a base and catalyst paste and a primer. LCR is available in 8 shades. The restorative paste will be available in both the multi-dose 3M ESPE Clicker dispensing system or a unit dose option.
The modified light-curing glass ionomer restorative system, LCR, has the following similarities to the currently marketed Vitremer Core Buildup/Restorative System:
- . Both restoratives have the same intended use.
- . Both are resin modified glass ionomer cements incorporating the same major chemical components.
- . Both restoratives have the same shelf life and storage conditions.
To provide evidence for safety, the chemical composition of LCR was compared to Vitremer Core Buildup/Restorative System. Additionally, independent research institutes carried out biocompatibility testing. The results show that LCR is a safe device when used as directed. To demonstrate the effectiveness of LCR, the performance characteristics of LCR were compared to 3M ESPE Vitremer Core Buildup/Restorative System, 3M ESPE Photac-Fil Aplicap and GC Fujill LC.
In summary, the modified dental restorative system, LCR, described in this Special 510(k) pre-market notification submission is substantially equivalent to the predicate devices. This is our position with regard to intended use, major chemical components, shelf life, storage conditions, safety and effectiveness.
Image /page/1/Picture/7 description: The image shows the text "Special 510(k) LCR" on the top line. Below that is the text "3M ESPE" in a larger, bold font. The text is black and the background is white. The image appears to be a logo or branding for a product or company.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the circumference of the circle. Inside the circle is a stylized emblem that resembles an eagle or bird-like figure with three curved lines representing its wings or body.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP 2 7 2005
Ms. Karen O'Malley Regulatory Affairs Specialist 3M Company 3M Center, Building 260-2B-17 St. Paul. Minnesota 55144-1000
Re: K052235
Trade/Device Name: LCR, Light-Curing Glass Ionomer Restorative System Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: September 9, 2005 Received: September 12, 2005
Dear Ms. O'Malley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. O'Malley
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and inome (s as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Ches
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{4}------------------------------------------------
# Indications for Use
| 510(k) Number: | K052235 |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name: | LCR, Light-curing glass ionomer restorative system |
| Indications for Use: | Primary teeth restorations<br>Small Class I restorations<br>Class III and Class V restorations<br>Transitional restorations<br>Filling defects and undercut areas in crown preparations<br>Core buildup where at least half the coronal tooth<br>structure is remaining to provide structural support for the<br>crown<br>Laminate sandwich technique |
Prescription Use × (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Michelle for MSR
(Division Sign-Off) (Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
Page _1_ of _1__
1052235 510(k) Number __
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.