The Destiknee™ Total Knee System is indicated for patients with severe knee pain and disability due to: · Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, traumatic arthritis, polyarthritis. · Correction of functional deformities. · Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy. · Moderate valgus, varus, or flexion trauma. · Knee fractures untreatable by other methods. The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.
Device Story
Destiknee™ Total Knee System replaces femoral, tibial, and patellar articular surfaces of knee joint. System comprises metallic and non-metallic components in various styles/sizes. Intended for cemented fixation in surgical setting. Device restores joint function, corrects deformities, and alleviates pain in patients with severe knee pathology. Surgeon implants components during total knee arthroplasty. Benefits include improved mobility and pain relief.
Clinical Evidence
Bench testing only. Fatigue testing of the Tibial Base Plate was performed per ASTM F1800-12 and ISO 14879-1:2000. Sterility and endotoxin testing (USP) were conducted to ensure manufacturing safety.
Technological Characteristics
Semi-constrained cemented knee prosthesis. Components manufactured from metallic and non-metallic materials. Fatigue testing performed per ASTM F1800-12 and ISO 14879-1:2000. Sterilization via validated methods. System is a modular implant set.
Indications for Use
Indicated for patients with severe knee pain and disability resulting from rheumatoid arthritis, osteoarthritis, traumatic arthritis, polyarthritis, functional deformities, post-traumatic loss of knee joint contour (patellofemoral erosion/dysfunction/prior patellectomy), moderate valgus/varus/flexion trauma, or knee fractures untreatable by other methods. Intended for cemented, single-use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
November 30, 2017
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
Meril Healthcare Pvt. Ltd. % Linda Braddon President & Chief Executive Officer Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, Georgia 30188
Re: K172936
Trade/Device Name: Destiknee™ Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: September 27, 2017 Received: October 04, 2017
Dear Linda Braddon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{1}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known)
### K172936
Device Name DestikneeTM Total Knee System
Indications for Use (Describe)
The DestikneeTM Total Knee System is indicated for patients with severe knee pain and disability due to:
· Severe knee joint pain, loss of mobility, and disability due to: rheumatoid arthritis, traumatic arthritis, polyarthritis.
· Correction of functional deformities.
· Post-traumatic loss of knee joint contour, particularly when there is patellofemoral erosion, dysfunction, or prior patellectomy.
· Moderate valgus, varus, or flexion trauma.
· Knee fractures untreatable by other methods.
The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 5. 510(k) SUMMARY
## 510(k) Summary
This 510(k) Summary is submitted in accordance with the requirements of 21 CFR Part 807. Section 92.
#### 5.1 Applicant:
Meril Healthcare Private Limited Survey No. 135/2/B & 174/2, First Floor, H1-H3, Meril Park Muktanand Marg, Chala, Vapi - 396 191, Gujarat, INDIA
#### 5.2 Regulatory Correspondent:
Linda Braddon, Ph.D President and CEO Secure BioMed Evaluations 7828 Hickory Flat Highway, Suite 120 Woodstock, GA, 30188 Email: lgb@securebme.com Phone: +-770-837-2681
### Meril Healthcare Contact Information:
Bhavik Gondaliya Manager- regulatory Affairs Meril Healthcare Private Limited Survey No. 135/2/B & 174/2, First Floor, H1-H3, Meril Park, Muktanand Marg, Chala, Vapi - 396 191, Gujarat, INDIA Mobile: +91 – 9687604713 Fax: +91 260-3052125 E mail: Bhavik.gondaliya@merillife.com
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the word "Meril" in a sans-serif font. The word is in a light blue color, and there is a small yellow circle above the "i" in "Meril". The background is also light blue.
| 5.3 | Date prepared: | 27, September 2017 |
|-----|----------------------|---------------------------------------------------------------------------------------------------------------------|
| 5.4 | Device information: | |
| | Proprietary Name: | DestikneeTM Total Knee System |
| | Common / Usual Name: | Total Knee Replacement |
| | Regulation name: | Knee joint patellofemorotibial<br>polymer/metal/polymer semi-constrained cemented<br>prosthesis per 21 CFR 888.3560 |
| | Product Code: | JWH |
| | Device Class: | Class II |
#### 5.5 Predicate Devices
| Component | Predicate<br>Device | Manufacturer | Submission<br>Number |
|-----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------------------------|------------------------------|
| Femoral Component,<br>Tibial Component (Tibial Base<br>Plate and Tibial Articular<br>Surface), All Poly Tibial<br>Component, All Poly Patella | DestikneeTM<br>Total Knee<br>System | Meril Healthcare<br>Pvt. Ltd. | 510(k)<br>Number:<br>K160771 |
### 5.6 Device Description:
The Destiknee™ Total Knee System is a system of components intended to replace the Femoral, Tibial and Patella articular surfaces of the knee joint. The components are
{5}------------------------------------------------
available in many styles and sizes and are manufactured from metallic and non-metallic materials.
The Destiknee TM Total Knee System was originally cleared under 510(k): K160771. The purpose of current submission is to add Tibial Base Plate (size 7 and size 8) and All Poly Patella (size 37mm and 40mm) in Destiknee™ Total Knee System.
#### 5.7 Indications for Use:
The Destiknee™ Total Knee System is indicated for patients with severe knee pain and disability due to:
- Severe knee joint pain, loss of mobility, and disability due to: rheumatoid . arthritis, osteoarthritis, traumatic arthritis, polyarthritis.
- Correction of functional deformities. .
- Post-traumatic loss of knee joint contour, particularly when there is . patellofemoral erosion, dysfunction, or prior patellectomy.
- Moderate valgus, varus, or flexion trauma. .
- . Knee fractures untreatable by other methods.
The Destiknee™ Total Knee System is intended for cemented use only. This device is for single use only.
#### 5.8 Comparison of Technological characteristics:
The Destiknee™ Total Knee system is substantially equivalent to the previously cleared predicate device based on similarities in intended use, device design/technological characteristics, materials and sterilization method.
#### Non clinical Performance data: 5.9
The components of Destiknee™ Total Knee System were subjected to following nonclinical performance testing to evaluate device function/ mechanical performance.
- Fatigue test of Tibial Base Plate (ASTM F1800-12; ISO 14879-1:2000) ●
{6}------------------------------------------------
Endotoxin testing has demonstrated that the manufacturing process does not introduce Endotoxin as a bi-product of the manufacturing and cleaning process.
Destiknee™ Total Knee System is subjected to 14 day USP sterility testing and USP Endotoxin testing prior to release.
Meril Healthcare Pvt. Ltd. Confirms that the Bacterial Endotoxin Test [(BET)/LAL test] will be conducted on final sterilized device of every batch prior to releasing to US market.
### 5.10 Conclusion
Based on performance testing results and similarities in intended use, device design/technological characteristics, materials and sterilization method, the Destiknee™ Total Knee System is considered substantially equivalent to the previously cleared predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.