Optilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls

K172868 · The Binding Site Group , Ltd. · DCN · Feb 28, 2018 · Immunology

Device Facts

Record IDK172868
Device NameOptilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein Controls
ApplicantThe Binding Site Group , Ltd.
Product CodeDCN · Immunology
Decision DateFeb 28, 2018
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5270
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K172868 · Feb 28, 2018Optilite C-Reactive Protein Reagent,Optilite C-Reactive Protein Calibrator,Optilite C-Reactive Protein ControlsThe Binding Site Group , Ltd.Clinical serum samplesClinical serum samples were used in a method comparison study to demonstrate substantial equivalence between the Optilite C-Reactive Protein Kit and the predicate device.Method comparison; Clinical samples; Substantial equivalence

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Method comparison study110 clinical samples and 83 normal donor samples; Sample Size: 193Roche Diagnostics Tina-Quant C-Reactive Protein Gen. 3 assayMethod comparison (regression statistics and clinical concordance)

Indications for Use

The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using the Binding Site Optilite analyser. Measurement of C-Reactive Protein aids in evaluation of the amount of injurv to body tissues and for evaluation of infection, tissue injurv, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. The Optilite C-Reactive Protein Calibrator is intended for the calibration of the Optilite C-Reactive Protein Reagent on the Optilite analyser. The Optilite C-Reactive Protein Controls are intended for use in quality control by monitoring accuracy and precision for the Optilite C-Reactive Protein Reagent.

Device Story

The Optilite C-Reactive Protein (CRP) system is an in vitro diagnostic test used in clinical laboratories. It utilizes serum samples to measure CRP concentration via immunoturbidimetry. The device consists of a reagent (antiserum and reaction buffer), calibrator, and controls, all designed for use on the Binding Site Optilite analyser. During operation, the sample reacts with specific anti-CRP antiserum to form insoluble complexes. The analyser passes light through the suspension and measures transmitted light via a photodiode; the intensity of transmitted light is inversely proportional to the CRP concentration. The system automatically calculates concentrations using a stored calibration curve. Healthcare providers use these results to assess tissue injury, infection, and inflammatory states in conjunction with other clinical findings. The system benefits patients by providing quantitative data to support diagnostic and monitoring decisions for inflammatory conditions.

Clinical Evidence

Bench testing only. Performance evaluated per CLSI guidelines (EP05-A3, EP06-A, EP17-A2, EP07-A2). Precision: total CV% 2.6–6.2%. Linearity: R=1.00 over 2.70–316.03 mg/L. LoQ: 5.00 mg/L. No interference observed for common endogenous substances or 14 therapeutic drugs. Method comparison (n=113) against predicate showed Passing-Bablok regression y = 1.00x + 5.56 (r=1.00). Reference interval verified against literature consensus (<5 mg/L).

Technological Characteristics

Immunoturbidimetric assay; goat anti-human CRP antiserum; reaction buffer (TRIS pH 7.5, 0.099% sodium azide). Photometric detection at 340 nm. Traceable to ERM-DA474. Automated processing on Optilite analyzer. Reagent stability: 5 months shelf-life, 3 months open-vial, 30 days on-board.

Indications for Use

Indicated for the quantitative determination of C-reactive protein (CRP) in serum to aid in the evaluation of tissue injury, infection, and inflammatory disorders in patients. Prescription use only.

Regulatory Classification

Identification

A C-reactive protein immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the C-reactive protein in serum and other body fluids. Measurement of C-reactive protein aids in evaluation of the amount of injury to body tissues.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K172868 B. Purpose for Submission: Previously cleared assay on a new instrument C. Measurand: C-Reactive Protein D. Type of Test: Quantitative immunoturbidimetry E. Applicant: The Binding Site F. Proprietary and Established Names: Optilite C-Reactive Protein Reagent Optilite C-Reactive Protein Calibrator Optilite C-Reactive Protein Controls G. Regulatory Information: 1. Regulation section: 21 CFR 866.5270, C-reactive protein immunological test system 2. Classification: Class II 3. Product code: DCN, System, Test, C-Reactive Protein {1} 4. Panel: Immunology (82) H. Intended Use: 1. Intended uses: The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using The Binding Site Optilite analyser. Measurement of C-Reactive Protein aids in evaluation of the amount of injury to body tissues and for evaluation of infection, tissue injury, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. The Optilite C-Reactive Protein Calibrator is intended for the calibration of the Optilite C-Reactive Protein Reagent on the Optilite analyser. The Optilite C-Reactive Protein Controls are intended for use in quality control by monitoring accuracy and precision for the Optilite C-Reactive Protein Reagent. 2. Indications for use: Same as Intended uses 3. Special conditions for use statement: Prescription use only 4. Special instrument requirements: The Binding Site Optilite analyzer (K110035) I. Device Description: The Optilite C-Reactive Protein Reagent is comprised of a dual wedge containing the following: - Antiserum: Supplied in stabilized liquid form. Preservatives: 0.099% sodium azide, TRIS pH 8.0. - Reaction Buffer: Containing 0.099% sodium azide, TRIS pH 7.5 as preservatives The Optilite C-Reactive Protein Calibrator is comprised of the following: - Pooled human serum, supplied in stabilized liquid form. Containing 0.099% sodium azide, as preservative {2} The Optilite C-Reactive Protein Controls are comprised of the following: - Pooled human serum, supplied in stabilized liquid form. Containing 0.099% sodium azide, as preservative J. Substantial Equivalence Information: 1. Predicate device name: Roche Diagnostics Tina-Quant C-Reactive Protein Gen. 3 2. Predicate 510(k) number: K083444 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device: Optilite C-Reactive Protein Reagent | Predicate: Tina-Quant C-Reactive Protein Gen. 3 | | Intended use | The Optilite C-Reactive Protein Reagent is intended for the quantitative in vitro determination of C-reactive protein (CRP) concentration in serum using the Binding Site Optilite analyzer. Measurement of C-Reactive Protein aids in evaluation of the amount of injury to body tissues and for evaluation of infection, tissue injury, and inflammatory disorders. This test should be used in conjunction with other laboratory and clinical findings. | Immunoturbidometric assay for the in vitro quantitative determination of CRP in human serum and plasma on Roche automated clinical chemistry analyzers | | Method | Turbidimetry | Same | | Reference Interval | <5 mg/L | Same | {3} | Differences | | | | --- | --- | --- | | Item | Device: Optilite C-Reactive Protein Reagent | Predicate: Tina-Quant C-Reactive Protein Gen. 3 | | Antibody | Anti-human CRP (goat) | Latex particles coated with anti-human CRP (mouse) | | Sample type | Serum | Serum, Li-heparin and EDTA plasma | | Measuring Range | 5–285 mg/L (neat) 25–1425 mg/L (1/5 dilution) | 0.3–350 mg/L (neat) 0.6–700 mg/L (1/2 dilution) | | Analyzer | Optilite | Hitachi 912, 917, Modular P | | Calibration | 6-points single calibrator diluted on analyzer | 5-points single calibrator diluted on analyzer | | Wavelength | 340 nm | 570 nm / 800 nm | | Traceability | ERM-DA474 | CRM470 | | Open vial stability | Three months at 2 to 8°C | Not stated | | On board stability | 30 days | 84 days opened and refrigerated on the analyzer | ## K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry - Review Criteria for Assessment of C-Reactive Protein (CRP), High Sensitivity C-Reactive Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays Guidance for Industry and FDA Staff CLSI EP17-A2: Evaluation of Detection Capability for Clinical Laboratory Measurement Procedures; Approved Guideline—Second Edition CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline—Second Edition CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures; A Statistical Approach CLSI EP05-A3: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Third Edition {4} # L. Test Principle: The determination of soluble antigen (CRP) concentration by turbidimetric methods involves the reaction with specific antiserum (anti-CRP) to form insoluble complexes. When light is passed through the suspension formed, a portion of the light is transmitted and focused onto a photodiode by an optical lens system. The amount of transmitted light is indirectly proportional to the specific protein concentration in the test sample. Concentrations are automatically calculated by reference to a calibration curve stored within the instrument. # M. Performance Characteristics: 1. Analytical performance: All results for analytical performance met the sponsor's predetermined acceptance criteria for each study. # a. Precision/Reproducibility: The studies were based on CLSI guideline EP05-A3, where four serum samples were tested in duplicate per run, two runs per day, over 21 days, using one reagent lot over three analyzers. The between-lot reproducibility was previously demonstrated in K161982. The total SD and $\mathrm{CV\%}$ calculation was based on the within-run, between-run, and between-day data. The results are shown in the table below: | Sample | Mean (mg/L) | Within-run | | Between-run | | Between-day | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | SD | CV% | SD | CV% | SD | CV% | SD | CV% | | Level 1 | 6.8 | 0.18 | 2.7 | 0.09 | 1.4 | 0.37 | 5.4 | 0.42 | 6.2 | | Level 2 | 9.5 | 0.12 | 1.2 | 0.12 | 1.3 | 0.37 | 3.9 | 0.41 | 4.3 | | Level 3 | 21.5 | 0.22 | 1.0 | 0.18 | 0.8 | 0.47 | 2.2 | 0.55 | 2.6 | | Level 4 | 65.4 | 0.49 | 0.7 | 0.52 | 0.8 | 1.90 | 2.9 | 2.03 | 3.1 | | Between-Instrument | | | --- | --- | | SD | CV% | | 0.20 | 3.0 | | 0.06 | 0.7 | | 0.17 | 0.8 | | 1.64 | 2.5 | # b. Linearity/assay reportable range: A linearity study was performed following CLSI guideline EP06-A. The linearity of this assay was confirmed using 14 serially diluted serum samples over the range of $2.70 - 316.03\mathrm{mg / L}$ . The results of weighted linear regression analysis are summarized as follows: | Dilution Range (mg/L) | Slope (95% CI) | Y-Intercept (95% CI) | Correlation R | | --- | --- | --- | --- | | 2.70–316.03 | 1.01 (0.97–1.05) | -0.06 (-0.40–0.28) | 1.00 | {5} c. Traceability, Stability, Expected values (controls, calibrators, or methods): i) Traceability: The calibration of the assay is traceable to the international reference standard ERM-DA474. ii). Kit Stability: Real-time stability - A study to establish shelf-life stability of the Optilite CRP Kit is on-going. Currently available data supports that the reagent is stable for five months from the date of manufacture when stored at recommended temperature of 2–8°C. Open-vial stability - The Optilite CRP Kit reagents can be stored opened at 2–8°C for up to three months. On-board stability - The Optilite CRP Kit reagents can be stored on-board the Optilite analyser for at least 30 days. d. Detection limit: The analytical sensitivity was determined in accordance with CLSI guideline EP17-A2. The Limit of Blank (LoB) was determined by testing four native serum samples run five times per day, over three days, using two reagent lots, to give a total of 60 results per lot. The LoB was estimated for each lot as the 95% percentile of the distribution. The Limit of Detection (LoD) was determined by testing four serum samples run five times per day, over three days, using two reagent lots to give a total of 60 results per lot. The LoD calculation followed a parametric analysis. The Limit of Quantitation (LoQ) was determined by testing four serum samples targeted to be close to the bottom of the measuring range, each tested five times per day over three days using two reagent lots. For LoB, LoD, and LoQ, the highest result obtained from the two lots tested was taken as the final result (see below): | Detection Limit | Concentration | | --- | --- | | LoB | 1.35 mg/L | | LoD | 2.66 mg/L | | LoQ | 5.00 mg/L | {6} # e. Analytical specificity: Interferences were assessed according to CLSI guideline EP07-A2 by testing serum samples with CRP concentrations falling at approximately $9\mathrm{mg / L}$ , $60\mathrm{mg / L}$ and $150\mathrm{mg / L}$ . Each sample was spiked with interfering substances and tested in replicates of three. Controls were prepared by spiking the same volume of the buffer without interfering substances into the serum samples. Percentage interference was calculated using the mean measurements of the test samples and the controls. A difference between the results within $\pm 10\%$ was considered as absence of interference. Endogenous Interference: The data demonstrated that the assay was not affected by the levels of the following substances: hemoglobin (5 g/L), bilirubin (200 mg/L), triglyceride (500 mg/dL), intralipid (250 mg/dL), and rheumatoid factor (2417 IU/mL). Drug Interference: The data demonstrated that the assay was not affected by the 14 therapeutic drugs tested at the concentrations given below. | Drug | Concentration tested | | --- | --- | | Acetaminophen | 1324 μmol/L | | Acetylsalicylic Acid | 3.63 mmol/L | | Amoxicillin | 206 μmol/L | | Ascorbic Acid | 342 μmol/L | | Caffeine | 308 μmol/L | | Cefotaxime | 673 μmol/L | | Theophylline | 222 μmol/L | | Chloramphenicol | 155 μmol/L | | Cimetidine | 79.2 μmol/L | | Digoxin | 7.8 nmol/L | | Fluconazole | 245 μmol/L | | Ibuprofen | 1212.5 μmol/L | | Penicillin | 75 mg/L | | Phenytoin | 198 μmol/L | # f. Assay cut-off: Not applicable # 2. Comparison studies: # a. Method comparison with predicate device: A total of 193 serum samples spanning the measuring range were assayed in singlicate by both the Optilite C-Reactive Protein and the Tina-Quant C-Reactive Protein Gen. 3 kits. The samples included 83 normal donors and 110 clinical samples. 80 samples (71 normal donors and 9 clinical samples) were excluded from data analysis due to results reporting as lower than the bottom of the measuring range for the Optilite C-Reactive Protein assay. A total of 113 samples were included in the {7} comparative data analysis and the result is as follows. Measurement procedure comparison between predicate (x) and test (y) device was evaluated using Passing-Bablok regression analysis. | Method Comparison (n=113) | | | | | | --- | --- | --- | --- | --- | | Regression Analysis | Line fit | 95% CI Slope | 95% CI Intercept | Pearson's r | | Passing-Bablok | y = 1.00x + 5.56 | 0.98–1.03 | 4.10–6.87 | 1.00 | b. Matrix comparison: Not applicable 3. Clinical studies: a. Clinical Sensitivity, Clinical specificity and Other clinical supportive data: Not applicable 4. Clinical cut-off: See expected values/reference range. 5. Expected values/Reference range: The reference interval was verified by testing 50 healthy adult donor samples. Forty-eight (48) of 50 samples tested had concentrations within the consensus reference interval taken from the literature (&lt;5 mg/L). Dati F, Schumann G, Thomas L et al. Consensus of a group of professional societies and diagnostic companies on guidelines for interim reference ranges for 14 proteins in serum based on the standardization against the IFCC/BCR/CAP reference material (CRM 470). Eur J. Clin Chem Clin Biochem 1996;34:517-520 N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...