K172811 · Endogastric Solutions, Inc. · ODE · Oct 19, 2017 · Gastroenterology, Urology
Device Facts
Record ID
K172811
Device Name
EsophyX Z, EsophyX2 HD
Applicant
Endogastric Solutions, Inc.
Product Code
ODE · Gastroenterology, Urology
Decision Date
Oct 19, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1500
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172811 · Oct 19, 2017
EsophyX Z, EsophyX2 HD
Endogastric Solutions, Inc.
Published clinical literature
The sponsor utilized data from clinical literature to support a specification change for the tissue mold clamping pressure, confirming that the change does not impact the safety or effectiveness of the device.
Literature review; Clamping pressure specification; Design change support
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Literature review regarding tissue clamping pressure; Literature review
—
Not applicable for this study
Safety and effectiveness of increased tissue mold clamping pressure
Indications for Use
The EndoGastric Solutions EsophyX Device models (EsophyX2 HD and EsophyX Z) with SerosaFuse Fasteners and Accessories are indicated for use in transoral tissue approximation, full thickness plication and ligation in the GI tract and are indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal junction and reduce hiatal hernia ≤ 2cm in size in patients with symptomatic chronic gastroesophageal reflux disease. Patients with hiatal hernias larger than 2cm may be included, when a laparoscopic hiatal hernia repair reduces the hernia to 2cm or less.
Device Story
EsophyX2 HD and EsophyX Z are disposable, single-use, all-mechanical, flexible fastener delivery systems used in Transoral Incisionless Fundoplication (TIF) procedures. Operated by physicians in an OR environment; a separate endoscope inserted through the device shaft provides continuous visualization. The operator uses proximal mechanical controls to capture and position tissue at the gastroesophageal junction. The device deploys polypropylene SerosaFuse fasteners to create a permanent surgical partial fundoplication. EsophyX2 HD allows individual fastener deployment; EsophyX Z uses an automated mechanical trigger to deploy two fasteners simultaneously. The device temporarily clamps tissue during fastener deployment. By restoring the valve at the gastroesophageal junction, the device treats GERD symptoms, potentially reducing reliance on pharmacological therapy.
Clinical Evidence
Bench testing only. No clinical data presented. Substantial equivalence is based on identical design, materials, and principles of operation to the predicate devices, with literature supporting the minor specification change for clamping pressure.
Indicated for patients with symptomatic chronic gastroesophageal reflux disease (GERD) who require and respond to pharmacological therapy. Used for transoral tissue approximation, full-thickness plication, and ligation in the GI tract to narrow the gastroesophageal junction and reduce hiatal hernias ≤ 2cm. Patients with hernias > 2cm are eligible if reduced to ≤ 2cm via laparoscopic repair.
Regulatory Classification
Identification
An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Special Controls
*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.
In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:
1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left, and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
October 19, 2017
EndoGastric Solutions, Inc. Steve J. Hoffman Corporate Compliance Officer 18109 NE 76th Street. Suite 100 Redmond, WA 98052
Re: K172811
Trade/Device Name: EsophyX Z Device with SerosaFuse Fasteners and Accessories EsophyX2 HD Device with SerosaFuse Fasteners and Accessories Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: ODE Dated: September 26, 2017 Received: October 4, 2017
Dear Steve J. Hoffman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
OMB No. Form Approved: 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page
510(k) Number (if known) K172811 Device Name
EsophyX Z Device with SerosaFuse Fasteners and Accessories
EsophyX2 HD Device with SerosaFuse Fasteners and Accessories
Indications for Use (Describe)
EsophyX2 HD Device with SerosaFuse Fasteners and Accessories: The EndoGastric Solutions EsophyX, HD Device with SerosaFuse Fasteners and Accessories is indicated for use in transoral tissue approximation, full thickness plication in the Gl tract and is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal junction and reduce hiatal hernia ≤ 2cm in size in patients with symptomatic chronic gastroesophageal reflux disease. Patients with hiatal hernias larger than 2cm may be included, when a laparoscopic hiatal hernia repair reduces the hernia to 2cm or less,
EsophyX Z Device with SerosaFuse Fasteners and Accessories: The EndoGastric Solutions EsophyX Z Device with SerosaFuse Fasteners and Accessories is indicated for use in transoral tissue approximation, full thickness plication in the GI tract and is indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal iunction and reduce hiatal hernia ≤ 2cm in size in patients with symptomatic chronic gastroesophageal reflux disease. Patients with hiatal hernias larger than 2cm may be included, when a laparoscopic hiatal hernia repair reduces the hernia to 2cm or less.
Type of Use (Select one or both, as applicable)
7 Prescription Use (Part 21 CFR 801 Subpart D) _ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{3}------------------------------------------------
# 510(k) Summary
EsophyX®2 HD and EsophyX Z Fastener Delivery Devices
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
#### 510(k) Number: _____________________________
#### Applicant Information [807.92(a)(1)]:
| Date Prepared: | August 26, 2017 |
|-------------------|---------------------------------------------------------------------------------|
| Name:<br>Address: | EndoGastric Solutions, Inc.<br>18109 NE 76th St. Suite 100<br>Redmond, WA 98052 |
| Phone: | 425-307-9200 |
| Fax: | 425-307-9201 |
| Contact Person: | Steven J Hoffman |
| Phone Number: | 425.307.9226 |
| Office: | 425.307.9200 |
| Facsimile Number: | 425.307.9201 |
#### Device Information [807.92(a)(2)]:
| Device Trade Name: | EsophyX® Z Device with SerosaFuse Fasteners and Accessories |
|---------------------------|---------------------------------------------------------------|
| | EsophyX₂ HD Device with SerosaFuse Fasteners and Accessories |
| Common Name: | Endoscopic Clip Applier, Implantable Fastener and Accessories |
| Classification Name(s): | Endoscope and Accessories |
| Product Code/ Regulation: | ODE / 21 CFR 876.1500 |
| Classification: | Class II |
## Predicate Device(s) [807.92(a)(3)]:
EndoGastric Solutions, EsophyX2 HD, EndoGastric Solutions, EsophyX Z, K171307
#### Device Description: [807.92(a)(4)]:
The subject EsophyX devices are unchanged from the currently cleared devices. The principles of operation for the devices are the same and also remain unchanged from the prior clearance. The indications for use for both models are identical.
The EndoGastric Solutions EsophyX Device models (EsophyX2 HD and EsophyX Z, K171307) with SerosaFuse Fasteners and Accessories are a prescription use only, disposable, single use systems consisting of an all mechanical, flexible fastener delivery device with user controls outside the patient's body. The devices are used in an operating room environment, using the Transoral Incisionless Fundoplication (TIF) procedure.
The device is limited contact and materials include acrylic/polycarbonate plastics, machined
{4}------------------------------------------------
### 510(k) Summary (Continued):
aluminum, stainless steel, nitinol wire, UV cured adhesives, pellethane and nylon.
The permanent contact, implantable tissue fasteners are constructed of injection molded polypropylene.
These transoral devices and tissue fasteners are provided sterile (EO). A separate, commercial endoscope operating independently down the center of the devices' flexible shaft lumen provides visualization of the procedure at all times, from device insertion through extraction. Two fasteners are loaded from a replaceable fastener cartridge containing ten fasteners for each channel. The loaded fasteners are pushed from the proximal device end, to the distal end down two separate channels via lumens in a flexible shaft where they are then in position ready for deployment into tissue. The operator, controlling the distal end of the devices through the mechanical controls at the proximal end of the devices, captures and positions a fold of tissue at the gastroesophageal junction. The fastener controls deploy both fasteners simultaneously in the EsophyX Z model, or individually in the EsophyX2 HD model, at the captured tissue position. This creates a permanent surgical partial fundoplication. Additional fasteners are used as needed to complete the valve restoration.
## Intended Use / Indications for Use [807.92(a)(5)]:
Both EsophyX2 HD and Z Fastener Delivery Devices with SerosaFuse Fasteners and accessories are intended for transoral tissue approximation. plication and fastening of tissue in the GI tract, for the endoluminal treatment of gastroesophageal reflux disease.
The Indications for Use for both products is as follows:
The EndoGastric Solutions EsophyX Device models (EsophyX2 HD and EsophyX Z) with SerosaFuse Fasteners and Accessories are indicated for use in transoral tissue approximation, full thickness plication and ligation in the GI tract and are indicated for the treatment of symptomatic chronic gastroesophageal reflux disease in patients who require and respond to pharmacological therapy. It is also indicated to narrow the gastroesophageal junction and reduce hiatal hernia ≤ 2cm in size in patients with symptomatic chronic qastroesophaqeal reflux disease. Patients with hiatal hernias larger than 2cm may be included, when a laparoscopic hiatal hernia repair reduces the hernia to 2cm or less.
#### Summary of the Technological Characteristics compared to the Predicate Device [807.92(a)(6)]:
Both EsophyX models are cleared for the same indications for use with the subject devices unchanged from the corresponding predicates (please reference K171307). Both devices use the same method of loading fasteners at the proximal end of the device and moving them down to the distal end in preparation of deployment. Tissue is retracted, secured in position, plicated and fastened in the same manner. Both use the TIF procedure.
As with the identical predicate device, the EsophyX Z differs in design in the operator controls on the proximal end when compared to the EsophyX2HD. In the EsophyX2 HD, posterior and anterior fasteners can be deployed using individual controls. In the EsophyX Z device, these controls have been automated with a mechanical trigger, which deploys both fasteners at the same time.
{5}------------------------------------------------
# 510(k) Summary (Continued):
Further, the EsophyX Z device/EsophyX2 HD devices (K171307) were used as the predicate without additional in vivo data needed for substantial equivalence as there are no changes made to the devices for this specification change proposal. Therefore, the devices have the same indications for use and are identical technologically.
No technological differences exist between the subject and predicate devices.
#### Performance Data on which Substantial Equivalence is Based [807.92(b)(1) and (2)]:
As both submission devices are identical to the predicates devices, the performance characteristics remain the same as those cleared previously in submission K171307.
As previously cleared in the referenced submission, both products are purely mechanical, have no device specific guidance documents, consensus standards are referenced, verification and validation testing remains unchanged as well as biocompatibility testing. In all instances, both devices function as intended and operation observed continues as designed.
Because there are no device changes, the safety and effectiveness profiles for both devices remain identical to the predicate devices.
The engineering specification change to increase the upper, design specification limit for the clamping pressure of the current devices' operational feature which temporarily clamps apposed tissue until a fastener is deployed to hold tissue together is proposed to provide a moderate increase in margin due to variations in the manufacturing process. This could cause the clamping pressure to exceed the currently cleared specification due to minor variations in the control system drag within each device. This specification change is supported by literature.
Based on the data presented in the literature, we attest the product specification change for the EsophyX2 HD and EsophyX Z devices does not change either device's operation, safety or effectiveness.
#### Conclusions Drawn from Performance Data [807.92(b)(3)]:
With the only change compared to the predicate EsophyX devices' clearance is a change to the product specification allowing a moderate increase in the upper limit of the temporary Tissue Mold clamping pressure occurring during each fastener deployment, no change is seen in the performance of either subject device.
As the subject devices and the predicate devices are the EsophyX2 HD and EsophyX Z are as safe and effective as the predicate EsophyX2 HD and EsophyX Z devices. Both devices have the same intended uses and indications, technological characteristics, and principles of operation as their predicate devices. The specification change does not alter the intended therapeutic use of the devices and does not affect safety and effectiveness when used as labeled. Device performance data remains the same and is therefore as safe and effective as the predicate devices. Thus, the EsophyX2 HD and EsophyX Z are substantially equivalent to their predicates.
{6}------------------------------------------------
## Additional Information [807.92(d)]:
None.
### Concluding Statements:
As the proposed and predicate devices remain the same in design and use, the proposed EsophyX2 HD and EsophyX Z are as safe and effective as the predicate devices. All devices have the same intended use and indications, technological characteristics, and principles of operation.
The specification change to increase the margin for the upper clamping pressure limit does not alter the intended therapeutic use of the device and does not affect its safety and effectiveness when used as labeled. Device performance data, i.e., tissue clamping performance, remains the same and is therefore as safe and effective as the predicate devices. Thus, the proposed devices are substantially equivalent to the predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.