K172651 · Encore Medical L.P. · OQG · Dec 13, 2017 · Orthopedic
Device Facts
Record ID
K172651
Device Name
FMP Extended Liners
Applicant
Encore Medical L.P.
Product Code
OQG · Orthopedic
Decision Date
Dec 13, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3358
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Joint replacement is indicated for patients suffering from disability due to: - noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head; - rheumatoid arthritis; - correction of functional deformity; - femoral fracture. This device may also be indicated in the salvage of previously failed surgical attempts. This device is to be used for uncemented applications.
Device Story
FMP Extended Liners are acetabular liners designed as a line extension to the X-alt Highly Cross Linked Acetabular Liner with Vitamin E system. The device expands femoral head compatibility by allowing larger femoral heads to be used with smaller acetabular cups. Used in orthopedic hip replacement surgery; implanted by surgeons in a clinical/hospital setting. The device functions as a bearing surface within a hip prosthesis system. It is intended for uncemented applications. The device is comparable to predicate liners in design, material, and surgical technique, with modifications to collar height and liner thickness to accommodate larger femoral heads. Benefits include improved surgical flexibility for surgeons to address patient-specific anatomical needs during hip arthroplasty.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing: wear (ISO 14242-2/3), impingement (ASTM F2582-14), push-out, lever-out, torsion (ASTM F1820-13), and range of motion (ISO 21535:2009). Bacterial endotoxin testing met USP <161> limits.
Technological Characteristics
Acetabular liner for hip prosthesis. Materials and design are consistent with predicate X-alt liners. Testing standards: ISO 14242-2/3 (wear), ASTM F2582-14 (impingement), ASTM F1820-13 (torsion/push-out/lever-out), ISO 21535:2009 (range of motion). Bacterial endotoxin levels monitored per USP <161>.
Indications for Use
Indicated for patients with disability due to noninflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or femoral fracture; also for salvage of failed surgical attempts. For uncemented applications.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
Predicate Devices
X-alt Highly Cross Linked Acetabular Liner with Vitamin E (K130365)
Stryker Trident Large Diameter Acetabular Insert (K062419)
Reference Devices
Foundation® Porous Coated Hemispherical Acetabular Cup (K974093)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 13, 2017
Encore Medical, L.P. Teffany Hutto Manager, Regulatory Affairs 9800 Metric Blvd. Austin, Texas 78758
Re: K172651
Trade/Device Name: FMPTM Extended Liners Regulation Number: 21 CFR 888.3358 Regulation Name: Hip Joint Metal/Polymer/Metal Semi-Constrained Porous-Coated Uncemented Prosthesis Regulatory Class: Class II Product Code: OQG, LPH Dated: November 28, 2017 Received: November 29, 2017
Dear Teffany Hutto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Teffany Hutto
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| 510(k) Number (if known) | K172651 |
|--------------------------|-----------------------|
| Device Name | FMPT™ Extended Liners |
**Indications for Use (Describe)**
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular necrosis of the natural femoral head;
- rheumatoid arthritis;
- correction of functional deformity;
- femoral fracture.
This device may also be indicated in the salvage of previously failed surgical attempts.
This device is to be used for uncemented applications.
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: underline;">Prescription Use</span> (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
|-----------------------------------------------------------------------------------------------|---------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
င္ခင္မွ ခ်စ္ေ႕
{3}------------------------------------------------
# 510(k) Summary
Date: September 1 , 2017
Manufacturer: DJO Surgical (Legal Name: Encore Medical, L.P.) 9800 Metric Blvd Austin, TX 78758
Contact Person: Teffany Hutto Manager, Regulatory Affairs Phone: (512) 834-6255 Fax: (512) 834-6313 Email: teffany.hutto@djoglobal.com
| Product | Classification | Product Code |
|------------------------|----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| FMP Extended<br>Liners | Class II | LPH - Hip joint metal/polymer/metal semi-constrained porous-<br>coated uncemented prosthesis per 21 CFR 888.3358<br><br>OQG - Hip prosthesis, semi-constrained, cemented,<br>metal/polymer, + additive, porous, uncemented per 21 CFR<br>888.3358 |
#### Description:
This Traditional 510(k) is a line extension to the X-alt Highly Cross Linked Acetabular Liner with Vitamin E product line to expand femoral head compatibility within the system by making larger femoral heads compatible with smaller acetabular cups.
#### Indications for Use:
Joint replacement is indicated for patients suffering from disability due to:
- noninflammatory degenerative joint disease including osteoarthritis and avascular ● necrosis of the natural femoral head;
- rheumatoid arthritis;
- correction of functional deformity; ●
- . femoral fracture.
This device may also be indicated in the salvage of previously failed surgical attempts.
This device is to be used for uncemented applications.
#### Predicate Devices:
| Device | Manufacturer | 510(k) Number | 510(k) Type |
|--------------------------------------------------------------|----------------------|---------------|-------------|
| X-alt Highly Cross Linked<br>Acetabular Liner with Vitamin E | Encore Medical, L.P. | K130365 | Primary |
| Stryker Trident Large Diameter<br>Acetabular Insert | Stryker | K062419 | Additional |
| Foundation® Porous Coated<br>Hemispherical Acetabular Cup | Encore Medical, L.P. | K974093 | Reference |
{4}------------------------------------------------
# Comparable Features to Predicate Device(s):
These components are comparable to the predicate devices in indications, material, design features, manufacturing methods, surgical implantation technique, intended use, packaging, and sterilization.
### Key Differences in Subject Device to Predicate:
Collar height and liner thickness compared to K130365.
Bacterial endotoxin testing conducted based on DJO Surgical's Endotoxin Assessment: alternative to batch testing plan met the bacterial endotoxin limits defined in USP <161> as evidenced through ongoing monitoring conducted concurrently with dose audits. Testing conducted with the most recent dose audit using the worst case product demonstrated an acceptable BET level. The subject FMP Extended Liners require fewer manufacturing operations and less handling than the worst case product (FMP Shell). Refer to NDPCAW-00085 for new product review.
Non-Clinical Testing: Testing performed includes wear per ISO 14242-2/3, impingement per ASTM F2582-14, push-out, lever-out, torsion per ASTM F1820-13, and range of motion per ISO 21535:2009. All evaluations determined that the devices are substantially equivalent to the applicable predicate devices.
Clinical Testing: Clinical testing is not required.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.