K172586 · Vios Medical, Inc. · DRT · Jun 22, 2018 · Cardiovascular
Device Facts
Record ID
K172586
Device Name
Vios Monitoring System™ Model 2050
Applicant
Vios Medical, Inc.
Product Code
DRT · Cardiovascular
Decision Date
Jun 22, 2018
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2300
Device Class
Class 2
Indications for Use
The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-lead ECG, heart rate, respiratory rate, pulse rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure, and patient posture and activity. VMS allows for the input of body temperature, and can display data from peripheral devices. VMS can generate alerts when rate-based carthythmias are detected and when physiological vital signs fall outside of selected parameters.
Device Story
Wireless mobile platform for monitoring adult patient vitals in healthcare facilities; operated by medically qualified personnel. System components: patient-worn Chest Sensor (CS2050), pulse oximetry adapters (L2050E/F), and BSM SW (B2050) running on commercial IT platform. Chest Sensor acquires 3-channel ECG, bioimpedance, 2-channel pulse oximetry, and tri-axial accelerometer data; transmits via Bluetooth to BSM SW. BSM SW processes data to generate 7-lead ECG, respiratory rate, SpO2, pulse rate, posture, activity, and non-invasive blood pressure (from external Bluetooth cuff). System provides real-time display and generates alerts for rate-based arrhythmias or out-of-range vital signs. Enables continuous patient surveillance; facilitates timely clinical intervention; improves patient safety through automated monitoring and alerting.
Clinical Evidence
Clinical testing performed for pulse oximetry (per IEC 80601-2-61) and respiratory rate. Non-clinical testing included electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), alarm standards (IEC 60601-1-8), ECG monitoring (IEC 60601-2-27), and patient monitoring (IEC 60601-2-49). Biocompatibility per ISO 10993, usability per EN 62366, and risk management per ISO 14971.
Technological Characteristics
Wireless mobile platform; Bluetooth connectivity. Chest Sensor: rechargeable, tri-axial accelerometer, bioimpedance, ECG, pulse oximetry. Adapters: plastic, non-sterile. BSM SW: standalone software on commercial IT platform. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC 60601-2-49, EC53, ISO 10993, EN 62366, ASTM D4169-16, EN 62304, ISO 14971, IEC 80601-2-61.
Indications for Use
Indicated for physiological vital signs monitoring of adult (18+) patients in healthcare facilities.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The FDA logo is composed of two parts: the Department of Health and Human Services logo on the left and the FDA acronym and full name on the right. The Department of Health and Human Services logo is a stylized depiction of a human figure, while the FDA acronym and full name are written in blue, with the acronym in a blue square.
June 22, 2018
Vios Medical, Inc. Megan Graham Quality/Regulatory Advisor 7300 Hudson Blvd N St. Paul, Minnesota 55128
Re: K172586
Trade/Device Name: Vios Monitoring System™ Model 2050 Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: DRT, DQA, DPZ, DRG Dated: May 22, 2018 Received: May 23, 2018
Dear Megan Graham:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
# Page 2 - Megan Graham
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Dlorali Castillo
**For**
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172586
Device Name Vios Monitoring System™ Model 2050
#### Indications for Use (Describe)
The Vios Monitoring System (VMS) is intended for use by medically qualified personnel for physiological vital signs monitoring of adult (18+) patients in healthcare facilities. It is indicated for use in monitoring of 7-lead ECG, heart rate, respiratory rate, pulse rate, functional oxygen saturation of arterial hemoglobin, non-invasive blood pressure, and patient posture and activity. VMS allows for the input of body temperature, and can display data from peripheral devices. VMS can generate alerts when rate-based carthythmias are detected and when physiological vital signs fall outside of selected parameters.
Type of Use (Select one or both, as applicable)
| <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> |
|-------------------------------------------------------------------------------------------------------|
| <span> <span style="font-size: 16px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
| Submitter: | Vios Medical, Inc.<br>700 Commerce Dr.<br>Suite 190<br>Woodbury, MN 55125 |
|-------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Megan Graham<br>Director, Quality and Compliance<br>megan@viosmedical.com<br>Office: 651-764-8467<br>Fax: 651-237-7003 |
| Date Prepared: | June 22, 2018 |
| Trade Name: | Vios Monitoring System™ Model 2050 |
| Common Name: | Vitals signs monitor |
| Primary Regulation: | 21 CFR 870.2300 Cardiac Monitor including Cardiotachometer & Rate Alarm |
| Other Regulation: | 21 CFR 890.2910 Transmitters And Receivers, Physiological Signal,<br>Radiofrequency<br>21 CFR 870.2710 Ear oximeter<br>21 CFR 870.2700 Oximeter |
| Classification: | Class II |
| Review Panels: | Cardiovascular, Anesthesiology |
| Product Codes: | Primary: DRT<br>Other: DQA, DPZ, DRG |
| Predicate Devices: | VMS Model 1000 (reference K150992)<br>VisiMobile Monitoring System (reference K152341)<br>Nonin Pulse Ox Model 6100C (reference K092101). |
| Device Description: | The Vios Monitoring System (VMS) Model 2050 is a wireless mobile<br>medical device platform that allows caregivers in healthcare settings to<br>monitor patient vitals.<br><br>VMS includes Vios-proprietary monitoring software and a Vios-<br>proprietary vitals sensor with two Vios-proprietary adapters. It is<br>compatible with a medical grade, Bluetooth™-enabled NIBP cuff.<br><br>The VMS BSM SW Model B2050 is stand-alone software that can receive,<br>analyze, and display physiological vitals data from one or more patient-<br>worn sensors via standard communication protocols (Bluetooth™). It<br>runs on a commercial IT platform and is intended to be used in<br>conjunction with the Vios Chest Sensor and Vios Lead Adapters and can<br>support peripheral, medical grade, Bluetooth™-enabled devices. |
| | The VMS Chest Sensor Model CS2050 is a small, patient-worn, non-sterile,<br>multiple use, and rechargeable sensor that acquires 3-channel ECG,<br>bioimpedance, 2-channel pulse oximetry, and tri-axial accelerometer data.<br>The sensor contains signal acquisition firmware (embedded software) and<br>wirelessly communicates acquired data via standard communication<br>protocols (Bluetooth™) to the BSM SW for analysis and display. The Chest<br>Sensor has a button that, when pressed, sends a patient call alert to the<br>BSM SW.<br>VMS Chest Sensor Adapter Models L2050E (Pulse Ox Ear Adapter) and<br>L2050F (Pulse Ox Finger Adapter) are plastic, non-sterile, patient-worn,<br>multiple use pulse oxygenation sensors that connect to the Vios Chest<br>Sensor and are secured to the patient via medical grade ECG electrodes. |
| Indications for Use: | The Vios Monitoring System (VMS) is intended for use by medically<br>qualified personnel for physiological vital signs monitoring of adult (18+)<br>patients in healthcare facilities. It is indicated for use in monitoring of 7-<br>lead ECG, heart rate, respiratory rate, pulse rate, functional oxygen<br>saturation of arterial hemoglobin, non-invasive blood pressure, and<br>patient posture and activity. VMS allows for the input of body<br>temperature, and can display data from peripheral devices. VMS can<br>generate alerts when rate-based cardiac arrhythmias are detected and<br>when physiological vital signs fall outside of selected parameters. |
| Summary of<br>Technology<br>Comparison | The Vios Monitoring System Model 2050 technology is based on the Vios<br>Monitoring System Model 1000 and shares a common architecture and<br>run on a commercial IT platform. The VMS Model 2050 Chest Sensor uses<br>a standard medical grade ECG electrode and an adapter plate to collect<br>additional ECG, impedance, accelerometer, and pulse oximetry data<br>(finger or ear). The VMS Model 2050 BSM SW has been updated to<br>generate and display 7-lead ECG data, respiratory rate, SpO2, pulse rate,<br>posture, activity, and non-latched physiological alarms. It can also display<br>non-invasive blood pressure data from a medical grade, Bluetooth™-<br>enabled NIBP cuff. |
| Summary of Non-<br>Clinical, Clinical,<br>and Conformance<br>Testing | The safety, effectiveness, and substantial equivalency of the VMS Model<br>2050 have been confirmed through the following non-clinical, clinical, and<br>conformance testing:<br>Electrical safety, EMC, and vitals sign monitoring standards (IEC<br>60601-1, IEC 60601-1-2, IEC 60601-1-8, IEC 60601-2-27, IEC<br>60601-2-49, EC53)Biocompatibility standards (ISO 10993)Usability and human factors standards (EN 62366)Transportation Simulation testing (ASTM D4169-16)Software development life cycle (EN 62304)Risk Management (ISO 14971)Pulse oximetry clinical testing (IEC 80601-2-61)Respiratory Rate clinical testing |
| Conclusion: | Vios Medical's evaluation of the substantial equivalence of the Vitals<br>Monitoring System Model 2050 to the predicate devices was based on a<br>comparison of device classification, intended use, indications for use and |
| | |
| contraindications, warnings, technical characteristics, and performance | |
| characteristics. Vios Medical also performed conformance testing and | |
| clinical testing to demonstrate substantial equivalency. Based on this | |
| comparison, Vios Medical concludes that Vios Monitoring System Model | |
| 2050 is substantially equivalent to the predicate devise and does not | |
| introduce new safety or effectiveness issues. | |
{4}------------------------------------------------
{5}------------------------------------------------
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.