K172531 · Siemens Medical Solutions USA, Inc. · OUO · Nov 14, 2017 · Radiology
Device Facts
Record ID
K172531
Device Name
Biograph mMR with mMR Angio Transfer Option
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
OUO · Radiology
Decision Date
Nov 14, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1200
Device Class
Class 2
Indications for Use
The Siemens MR-PET system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging. These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders. The MR is intended to produce transverse, sagittal, coronal and oblique crosssectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures wen performed with MR compatible devices, such as MR safe biopsy needles. The PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The combined system utilizes the MR for radiation correction maps for PET studies. The system provides inherent anatomical reference for the fissed PET and MR images due to precisely aligned MR and PET image coordinate systems.
Device Story
The Biograph mMR with mMR Angio Transfer Option is an integrated MR-PET scanner modified with a mechanical patient transfer system. It enables seamless patient movement between the Biograph mMR and Siemens Artis Q/Artis zee interventional systems. The option includes a Transferboard, Guiding Slide Way, Coil Adapter Cover, and Transfer Support Cart. The patient remains on the Transferboard during transport and imaging, eliminating manual lifting. Operated by trained healthcare professionals in clinical/interventional settings. The system maintains the diagnostic capabilities of the predicate Biograph mMR, providing fused anatomical and metabolic images. The mechanical modifications facilitate intraoperative workflows, improving patient safety and efficiency by reducing handling during transitions between imaging and interventional suites.
Clinical Evidence
Bench testing only. No clinical data provided. Performance testing included mechanical testing per IEC 60601-1/ANSI AAMI ES 60601-1 and system-level verification and validation to ensure the transfer mechanism functions as intended without compromising the safety or diagnostic performance of the MR-PET system.
Technological Characteristics
Integrated MR-PET system. Mechanical modifications: Transferboard, Guiding Slide Way, Coil Adapter Cover, and Transfer Support Cart. Standards: IEC 60601-1 (Safety), IEC 60601-1-2 (EMC), IEC 60601-2-33 (MR Safety), ISO 14971 (Risk Management), IEC 62366 (Usability), AAMI/ANSI/IEC 62304 (Software Lifecycle), NEMA MS 4/9/NU 2 (Imaging Performance), ISO 10993-1 (Biocompatibility). Connectivity: DICOM.
Indications for Use
Indicated for use by trained healthcare professionals to aid in detection, localization, and diagnosis of diseases and disorders (e.g., cardiovascular, neurological, cancer) via MR and PET imaging. Used for simultaneous or single-modality imaging, including interventional procedures with MR-compatible devices.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Biograph mMR with Software syngo MR E11P (K163234)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a caduceus, which is a traditional symbol of medicine. The text on the right is written in a clear, sans-serif font, with "FDA" in a larger font size than the rest of the text.
Siemens Medical Solutions USA, Inc. Cordell Fields Regulatory Affairs Specialist 65 Valley Stream Parkway, Mail Code 65-1A Malvern, Pennsylvania 19355
Re: K172531
Trade/Device Name: Biograph mMR with mMR Angio Transfer Option Regulation Number: 21 CFR 892.1200 Regulation Name: Emission Computed Tomography System Regulatory Class: Class II Product Code: OUO, LNH, LNI, KPS Dated: August 21, 2017 Received: August 22, 2017
Dear Cordell Fields:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
November 14, 2017
{1}------------------------------------------------
### Page 2 - Cordell Fields
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara For
Robert A. Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The background is white, providing a strong contrast that makes the text stand out. The letters are evenly spaced and appear to be part of a logo or branding element.
### Section 4: Indications for Use Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration.
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known)
K172531
Device Name Biograph mMR with mMR Angio Transfer Option
Indications for Use (Describe)
The Siemens MR-PET system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders.
The MR is intended to produce transverse, sagittal, coronal and oblique crosssectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures wen performed with MR compatible devices, such as MR safe biopsy needles.
The PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
The combined system utilizes the MR for radiation correction maps for PET studies. The system provides inherent anatomical reference for the fissed PET and MR images due to precisely aligned MR and PET image coordinate systems.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995, *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." FORM FDA 3881 (7/17) Page 1 of 1 PSC Publishing Services (301) 443-6740
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The word is presented in all capital letters and is centered in the image.
# Section 5 510(k) Summary
## Biograph mMR with mMR Angio Transfer Option
| Company Name<br>and Address | Siemens Medical Solutions USA. Inc.<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared | August 21, 2017 |
| Conformance<br>statement: | This 510(k) summary of safety and effectiveness information is<br>being submitted in accordance with the requirements of the<br>Safe Medical Device Act of 1990 and 21 CFR § 807.92. |
#### 1. General Information
| Importer / Distributor | Siemens Medical Solutions USA. Inc.<br>40 Liberty Boulevard<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Registration Number: 2240869 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Location of | Siemens Healthcare GmbH |
|---------------|---------------------------------|
| Manufacturing | Henkestrasse 127 |
| Site | Erlangen Bayern, Germany 91052 |
| | Registration Number: 3002808157 |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a teal, sans-serif font. The letters are bold and evenly spaced. The background is plain white, which makes the word stand out.
### 2. Contact Persons
| Contact Person | Cordell L. Fields, Esq.<br>Regulatory Affairs Technical Specialist<br>Siemens Medical Solutions USA, Inc.<br>65 Valley Stream Parkway<br>Mail Code 65-1A<br>Malvern, PA 19355, USA<br>Phone: (610) 448-6469<br>Cell: (610) 306-3167<br>Fax: (610) 640-4481<br>E-mail: cordell.fields@siemens-healthineers.com |
|-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Alternate Contact<br>Person | Kimberly Rendon |
| | Sr. Manager Regulatory Affairs |
| | Siemens Medical Solutions USA, Inc. |
| | 65 Valley Stream Parkway |
| | Mail Code 65-1A |
| | Malvern, PA 19355, USA |
| | Phone: (610) 448-6480 |
| | Fax: (610) 448-1787 |
| | E-mail: Kimberly.rendon@siemens-healthineers.com |
### 3. Device Name and Classification
| Trade Name: | Biograph mMR with mMR Angio Transfer Option |
|------------------------------|----------------------------------------------------------------------------------------------|
| Classification Name | Tomographic Imager Combining Emission Computed<br>Tomography with Nuclear Magnetic Resonance |
| Classification Panel: | Radiology |
| Classification<br>Regulation | 21 CFR § 892.1200<br>21 CFR § 892.1000 |
| Device Class: | II |
| Primary Product<br>Code: | OUO |
| Secondary Product<br>Codes: | LNH, LNI, KPS |
{5}------------------------------------------------
#### 4. Legally Marketed Predicate Devices
| Trade Name | Biograph mMR with Software syngo<br>MR E11P |
|------------------------------|----------------------------------------------------------------------------------------------|
| 510(k) # | K163234 cleared February 28, 2017 |
| Classification Name | Tomographic Imager Combining Emission Computed<br>Tomography with Nuclear Magnetic Resonance |
| Classification Panel: | Radiology |
| Classification<br>Regulation | 21 CFR § 892.1200<br>21 CFR § 892.1000 |
| Device Class: | II |
| Primary Product<br>Code: | OUO |
| Secondary Product<br>Codes: | LNH, LNI, KPS |
| Recall Information | "There are no Recalls for this Predicate Device." |
#### 5. Device Description
The new mMR Angio Transfer Option is being introduced for the Biograph mMR system.
The new option is intended to provide a patient transfer from the SIEMENS Artis Q and Artis Q.zen (K123529) or Artis zee/zeego SW VC21 (K141574) System to the Biograph mMR and vice versa.
To achieve this functionality a Transferboard Artis (further named as Transferboard) will be used where the patient is located and transferred. This Transferboard can be adapted on the standard table of the Artis Q and Artis Q.zen (K123529) or Artis zee/zeego SW VC21 (K141574) System. The standard table is identical for all models. By swiveling the Artis Q. Artis Q.zen or Artis zee /zeego table (standard function) the Transferboard will be positioned in front of the Biograph mMR. Via the Guiding Slide Way the Transferboard is pushed manually on the Table Top of the Biograph mMR.
The Guiding Slide Way is also included in the option and is installed on the Table Top of the Biograph mMR instead of a regular patient cushion.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The letters are a bright teal color. The background is plain white, which makes the word stand out.
The Transferboard is also used to move the patient into the bore of the Biograph mMR for further diagnostic imaging.
The basic device and its functionality remain unchanged except for the following modifications:
- . Coil Adapter Cover (CAC): The Coil Adapter Cover will replace the standard mMR coil adapter cover on the foot end of the mMR tabletop to enable the overtraveling of the Transferboard (TFB).
- Guiding Slide Wav (GSW): A Guiding Slide Way (GSW) will be adapted on the ● existing Biograph mMR tabletop. This is a foldable option and can be removed for diagnostic mMR examination in case the transfer option is not needed.
- Transferboard (TFB): The Transferboard (TFB) is intended to transfer the ● patient. The patient remains on this board during the whole intraoperative procedure.
- Patient Cushion (LGH): The Patient Cushion is laid on the Transferboard (TFB) for comfortable reclining.
- Fixation Belts: The Fixation Belts are used for preventing the patient to fall of ● the Transferboard and Patient Cushion during transport, surgery and imaging,
- Transfer Support Cart (TSC): The Transfer Support Cart is part of the option. ● It ensures that both tables have the same height and that there are no vibrations during transfer due to load changes on the patient tables
The new mMR Angio Transfer Option for the Biograph mMR system requires the most recent Device Software, syngo MR E11P software cleared via 510(k) on February 28, 2017 (K163234).
### Advantage
The new features on the Biograph mMR with mMR Angio Transfer Option enables the handover of a patient previously examined by SIEMENS Artis Q. Artis Q.zen or Artis zee/zeego System and vice versa. The advantage is that the patient can simply stay on the Transfer Board. There is no need to lift him or her from one patient support system to another.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented.
### 6. Indication for Use
The indications for use for the subject device are the same as the predicate device and are as follows:
The Siemens MR-PET system combines magnetic resonance diagnostic devices (MRDD) and Positron Emission Tomography (PET) scanners that provide reqistration and fusion of high resolution physiologic and anatomic information, acquired simultaneously and isocentrically. The combined system maintains independent functionality of the MR and PET devices, allowing for single modality MR and / or PET imaging.
These systems are intended to be utilized by appropriately trained health care professionals to aid in the detection, localization, and diagnosis of diseases and disorders.
The MR is intended to produce transverse, sagittal, coronal and oblique crosssectional MR images, spectroscopic images and/or spectra, and displays the internal structure and/or function of the human body. Other physical parameters derived from the images and/or spectra may also be produced. Depending on the region of interest, approved contrast agents may be used, as described in their labeling. This system may also be used for imaging during interventional procedures when performed with MR compatible devices, such as MR safe biopsy needles.
The PET images and measures the distribution of PET radiopharmaceuticals in humans to aid the physician in determining various metabolic (molecular) and physiologic functions within the human body for evaluation of diseases and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer.
The combined system utilizes the MR for radiation-free attenuation correction maps for PET studies. The system provides inherent anatomical reference for the fused PET and MR images due to precisely aligned MR and PET image coordinate systems.
### Substantial Equivalence 7.
The subject device, mMR Angio Transfer Option for the Biograph mMR, was developed from the previous version of the Biograph mMR (with software version svnao MR E11P).
As described above, only the hardware for the device was modified. Software is not affected.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the word "SIEMENS" in a sans-serif font. The word is written in all capital letters. The color of the text is a teal or cyan color.
### 8. Summary of Technological Characteristics of the Subject Device as Compared with the Predicate Device
## Technological Characteristics
The new mMR Angio Transfer Option for the Biograph mMR has the same technological characteristics as the predicate device Biograph, mMR with syngo MR E11P (K163234, cleared February 28, 2017).
The subject device is substantially equivalent to the predicate device. As described above, the main changes are mechanical and enable the transfer of the patient form the Artis Q. Artis Q.zen or Artis zee/zeego System to the Biograph mMR.
#### 9. Nonclinical Performance Testing
The following performance testing was conducted on the subject device for covering the mMR Angio Transfer Option for the Biograph mMR. Existing performance test for the Biograph mMR remain valid.
- Mechanical Test according to IEC 60601-1 / ANSI AAMI ES 60601-1 ●
- . System verification and validation.
The results from each set of tests demonstrate that the device performs as intended and is thus substantially equivalent to the predicate devices to which it has been compared.
mMR Angio Transfer Option for the Biograph mMR conforms to the applicable FDA recognized and international IEC, ISO and NEMA standards with regards to performance and safety as recommended by the respective MR FDA Guidance Document, "Guidance for the Submission of Premarket Notifications for Magnetic Resonance Diagnostic Devices", issued in 2016.
For the mMR Angio Transfer Option for the Biograph mMR, new testing for the standards indicated by an asterisk (*) was conducted for the proposed modifications. Standards not indicated by an asterisk are also tested for the Biograph mMR however their validity is not affected by the proposed modifications, i.e. the old evidence is still valid.
{9}------------------------------------------------
## SIEMENS
| Recognition<br>Number | Product Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
|-----------------------|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|------------------------------------------|
| * 19-4 | General | Medical electrical equipment –<br>part 1: general requirements<br>for basic safety and essential<br>performance | ES60601-<br>1:2005/(R) 2012<br>and A1:2012 | AAMI /<br>ANSI |
| 19-1 | General | Medical electrical equipment –<br>Part 1-2: General requirements<br>for basic safety and essential<br>performance – Collateral<br>standard: Electromagnetic<br>compatibility – Requirements<br>and tests | 60601-1-2 Edition<br>3:2007-03 | IEC |
| 12-271 | Radiology | Medical electrical equipment –<br>Part 2-33: Particular<br>requirements for the basic<br>safety and essential<br>performance of magnetic<br>resonance equipment for<br>medical diagnosis | 60601-2-33 Ed.<br>3.1:2013 | IEC |
| * 5-40 | General | Medical devices – Application<br>of risk management to medical<br>devices | 14971 Second<br>edition 2007-03-<br>01 | ISO |
| * 5-67 | General | Medical devices – Application<br>of usability engineering to<br>medical devices | 62366:2007 | IEC |
| 13-32 | Software | Medical device software –<br>Software life cycle processes | 62304:2006 | AAMI<br>ANSI<br>IEC |
| 12-232 | Radiology | Acoustic Noise Measurement<br>Procedure for Diagnosing<br>Magnetic Resonance Imaging<br>Devices | MS 4-2010 | NEMA |
| Recognition<br>Number | Product Area | Title of Standard | Reference<br>Number and<br>date | Standards<br>Development<br>Organization |
| 12-288 | Radiology | Characterization of Phased<br>Array Coils for Diagnostic<br>Magnetic Resonance Images<br>(MRI) | MS 9-2008 | NEMA |
| 12-300 | Radiology | Digital Imaging and<br>Communications in Medicine<br>(DICOM) Set 03/16/2012<br>Radiology | PS 3.1 – 3.20<br>(2016) | NEMA |
| 12-265 | Radiology | Performance measurements of<br>positron emission tomographs | NU 2-2012 | NEMA |
| * 2-220 | Biocompati<br>bility | Biological Evaluation Of<br>Medical Devices - Part 1:<br>Evaluation And Testing Within<br>A Risk Management Process | 10993-1 Fourth<br>Edition 2009-10-<br>15 | ISO |
{10}------------------------------------------------
Table 1: Recognized Consensus Standards for mMR Angio Transfer Option
### 10. General Safety and Effectiveness Concerns
Instructions for use are included in the device labeling, and the information provided will enable the user to operate the device in a safe and effective manner. Several safety features including visual and audible warnings are incorporated into the system design.
Also, the Biograph mMR with mMR Angio Transfer Option is continually monitored, and if an error occurs, the system functions will be blocked, and an error message will be displayed.
Risk management according to ISO 14971:2007 is ensured via a hazard analysis which is used to identify potential hazards. These potential hazards are controlled via software development, verification, and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice, and all equipment is subject to final performance testing, Furthermore, the operators are health care professionals familiar with and responsible for the mMR examinations to be performed.
{11}------------------------------------------------
## 11. Conclusion as to Substantial Equivalence
mMR Angio Transfer Option for the Biograph mMR has the same intended use and the same technological characteristics as the predicate device Biograph mMR with syngo MR E11P (K163234).
The new features on the Biograph mMR with mMR Angio Transfer Option enables the handover of a patient previous examined by SIEMENS Artis Q. Artis Q.zen or Artis zee/zeego System and vice versa. The advantage is that the patient can simply stay on the Transfer Board. There is no need to lift the patient from one patient support system to another.
These modifications improve the user's workflow to enable an intraoperative procedure together with SIEMENS Artis Q and Artis Q.zen (K123529) or Artis zee/zeego SW VC21 (K141574) System.
The differences between the subject device and the predicate device include incorporation / adaptations of cleared features from the MAGNETOM Artis Combi Suite (K140253) which do not raise new issues of safety or effectiveness.
Therefore, Siemens believes that the subject device, Biograph mMR System with mMR Angio Transfer Option is substantially equivalent to the predicate device, Biograph mMR with syngo MR E11P.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.