Amsel Occluder Device

K172479 · Amsel Medical · FZP · Sep 14, 2017 · General, Plastic Surgery

Device Facts

Record IDK172479
Device NameAmsel Occluder Device
ApplicantAmsel Medical
Product CodeFZP · General, Plastic Surgery
Decision DateSep 14, 2017
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4300
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Amsel Occluder Device is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated and within the size range of 2.0mm to 7.0mm diameter.

Device Story

Amsel Occluder Device is a mechanical implantable clip used in open general surgery. It is designed to occlude tubular structures or vessels ranging from 2.0mm to 7.0mm in diameter. The device functions as a standard metal ligating clip, providing mechanical closure of vessels or structures to achieve hemostasis or ligation. It is intended for use by surgeons in a clinical or operating room setting. The device provides a permanent mechanical barrier to flow within the targeted structure, aiding in surgical management of vessels.

Clinical Evidence

No clinical data provided; bench testing only.

Technological Characteristics

Implantable metal ligating clip for tubular structures/vessels. Dimensions accommodate 2.0mm to 7.0mm diameters. Mechanical occlusion principle. Class II device.

Indications for Use

Indicated for use in open general surgery procedures on tubular structures or vessels requiring metal ligating clips, specifically for structures with diameters between 2.0mm and 7.0mm.

Regulatory Classification

Identification

An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 14, 2017 Amsel Medical % Mr. Leo Basta Northstar Biomedical Associates 93 Benefit Street Providence, Rhode Island 02904 Re: K172479 Trade/Device Name: Amsel Occluder Device Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: August 15, 2017 Received: August 16, 2017 Dear Mr. Basta: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device- Image /page/0/Picture/11 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. {1}------------------------------------------------ related adverse events) (21 CFR 803): good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm_for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Jennifer R. Stevenson -S3 For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K172479 Device Name Amsel Occluder Device #### Indications for Use (Describe) The Amsel Occluder Device is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated and within the size range of 2.0mm to 7.0mm diameter. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | |----------------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | # CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%