K140932 · Amsel Medical Corporation · FZP · Jan 23, 2015 · General, Plastic Surgery
Device Facts
Record ID
K140932
Device Name
AMSEL OCCLUDER DEVICE
Applicant
Amsel Medical Corporation
Product Code
FZP · General, Plastic Surgery
Decision Date
Jan 23, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Amsel Occluder Device is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated and within the size range of 2.0mm to 7.0mm diameter.
Device Story
Amsel Occluder Device is a mechanical occlusion implant for open general surgery. Input: target vessel structure (2.0mm–7.0mm). Operation: hand-held, syringe-like delivery device penetrates/transfixes vessel with a needle; deploys a metallic clip construct consisting of two compression components and an interlocking titanium rod; components are compressed and locked on either side of the vessel to achieve permanent occlusion. Output: permanently occluded vessel. Used in hospitals/surgery centers by trained surgeons. Delivery system detaches from the implant after successful locking. Benefits: provides secure, permanent vessel ligation via transfixing mechanical compression. Device is single-patient use and disposable.
Clinical Evidence
No human clinical data. Evidence consists of bench testing and animal studies. Bench testing included clip closure distance/repeatability, strain analysis, corrosion testing (ASTM F2129-08, ASTM F3044-14), detachment force, holding pressure (>700mmHg), and release mechanism function. Animal study evaluated in-vivo safety, deployment, usability, and occlusion performance in arteries and veins (2.0mm–7.0mm) with histology. MR imaging testing performed to support MR conditional labeling.
Technological Characteristics
Materials: Nitinol and titanium. Energy: Manual (mechanical). Form factor: Hand-held syringe-like delivery system with implantable clip construct. Connectivity: None. Sterilization: EtO (single-use). Standards: ASTM F2129-08, ASTM F3044-14. Software: None.
Indications for Use
Indicated for open general surgery procedures on tubular structures or vessels requiring metal ligating clips, specifically for structures 2.0mm to 7.0mm in diameter.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
Predicate Devices
Neurosel Medical Ltd. Twister Superelastic Ligating Clip (K032238)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of a human face in profile, composed of three overlapping silhouettes. The silhouettes are arranged in a way that suggests a sense of community and collaboration. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 23, 2015
Amsel Medical Corporation % Leo Basta NorthStar Biomedical Associates 93 Benefit Street Providence, Rhode Island 02904
Re: K140932
Trade/Device Name: Amsel Occluder Device Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable clip Regulatory Class: Class II Product Code: FZP Dated: November 28, 2014 Received: December 1, 2014
Dear Leo Basta:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# INDICATIONS FOR USE STATEMENT
510(k) Number (if known): K140932
Device Name: Amsel Occluder Device
## Indications for Use:
The Amsel Occluder Device is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated and within the size range of 2.0mm to 7.0mm diameter.
Prescription Use √
AND/OR
Over-The-Counter Use
(Part 21 C.F.R. 801 Subpart D)
(21 C.F.R. 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Amsel Medical Corp. Rev A
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Amsel Medical. The logo features a stylized letter "A" in blue and gray, with a curved line running through it. To the right of the "A" is the text "AMSEL" in gray, with the word "MEDICAL" in blue underneath. The logo has a clean and professional design.
Traditional Premarket Notification Submission - 510(k)
510(k) SUMMARY Amsel Occluder Device 510(k) Number K140932 Prepared January 22, 2015
## 5.1 Company Name
Amsel Medical Corp. 171 Hamilton Street Cambridge, MA 02139 Tel: (617) 395-8825 Fax: (617) 608-9080
### 5.2 Contact Person
Leo Basta NorthStar Biomedical Associates for Amsel Medical Corporation 93 Benefit Street Providence, RI 02904 617.834.9866 (phone) 401.454.1733 lbasta@northstarbiomedical.com
And/or
Orly Maor 25 A Sirkin Street Kfar-Saba 4442156, Israel Tel: 972-7453607 Fax: 972-153-9-7453607 oram.ma@gmail.com
## 5.3 Trade/Proprietary Name
Amsel Occluder Device
## Classification Name 5.4 Clip Implantable
## ર.5 Product Code/Regulation No. Product code: FZP, Regulation No. 878.4300.
Amsel Medical Corp. Rev A
{5}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Amsel Medical. The logo features a stylized letter "A" on the left, with the word "AMSEL" in a larger font size to the right of the "A". Below "AMSEL" is the word "MEDICAL" in a smaller font size. The colors used in the logo are shades of blue and gray.
## 5.6 Device Classification Class II
- 5.7 Panel General & Plastic Surgery
### 5.8 Predicate Devices
- 1. Neurosel Medical Ltd. Twister Superelastic Ligating Clip cleared under K032238, (product code FZP, Regulation No. 878.4300).
- 2. Ethicon Inc. LigaClip, cleared under K834267 (product code FZP, Regulation No. 878.4300).
### 5.9 Intended Use
The Amsel Occluder Device is intended for use in open general surgery procedures on tubular structures or vessels wherever a metal ligating clip is indicated and within the size range of 2.0mm to 7.0mm diameter.
# 5.10 Device Description
The Amsel Occluder Device involves the implantation of a mechanical occlusion device that penetrates into and through the target vessel structure and when compressed and locked, clamps the vessel shut.
The device consists of a metallic clip construct comprised of two compression components and an interlocking implant rod that when compressed together as a unit and locked on either side of a vessel results in occlusion. The delivery device consists of a hand held syringe-like device loaded with the components of the implant construct. The device is for use in general surgical procedures. Once positioned correctly, the occlusion construct is actuated through a series of semiautomatic steps resulting in the two components becoming compressed on either side of the vessel and locked together to occlude the vessel. Once occlusion is achieved, the delivery system, which is mechanically locked to the implant rod, is detached from the implant and the needle is withdrawn leaving the occlusion construct implanted, permanently occluding the vessel.
### 5.11 Substantial Equivalence
The Amsel Occluder Device is substantially equivalent to the legally marketed predicate devices for the following reasons:
- The Amsel Occluder Device has the same intended use and similar ● indications for use as the predicate devices,
{6}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Amsel Medical. The logo features a stylized letter "A" in blue and gray, with a curved line running through it. To the right of the "A" is the text "AMSEL" in gray, with the word "MEDICAL" in blue below it. The logo has a clean and professional design.
- All devices use similar technology to deliver a clip to mechanically . compress the tissue,
- . The Amsel Occluder employs similar or the same materials as its predicates,
- The Amsel Occluder and the predicate devices use a delivery applicator that ● offers smooth, efficient deployment of the clip/s, and
- All three devices share similar dimensional attributes. .
The following table contains a summary of the descriptive and technological characteristics of the Amsel Occluder Device compared to the predicate devices.
{7}------------------------------------------------
| | AMSEL OCCLUDER DEVICE | TWISTER SUPERELASTIC LIGATING<br>CLIP | LIGACLIP | SE |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | Not yet known | K032238 | K834267 | Same |
| Manufacturer | Amsel Medical Corp. | Neurosel Medical Ltd. | Ethicon Inc. | Same |
| Product Code | FZP | FZP | FZP | Same |
| CFR | 878.4300 | 878.4300 | 878.4300 | Same |
| Intended Use &<br>Indications for Use | The Amsel Occluder Device is<br>intended for use in open general<br>surgery procedures on tubular<br>structures or vessels wherever a<br>metal ligating clip is indicated and<br>within the size range of 2.0mm to<br>7.0mm diameter. | The Twister Superelastic Ligating Clip,<br>Size M is intended for the permanent<br>occlusion or ligation of blood vessels<br>and other tubular body structures,<br>wherever a metal ligating clip is<br>indicated, and within the size range of<br>2.0 to 3.5mm diameter. | The LIGACLIP MULTIPLE CLIP<br>APPLIER is intended for use on<br>tubular structures or vessels wherever<br>a metal ligating clip is indicated. The<br>tissue being ligated should be<br>consistent with the size of the clip. | Similar. Size range for the<br>Amsel Occluder was<br>demonstrated through<br>animal testing performed<br>using the device. |
| Environments of<br>Use | Hospitals and surgery center. | Hospitals and surgery center. | Hospitals and surgery center. | Same |
| Pre Loaded \<br>Loaded in OR | Pre-loaded | Pre-loaded | Loaded in the OR | Same as Neurosel |
| | AMSEL OCCLUDER DEVICE | TWISTER SUPERELASTIC LIGATING<br>CLIP | LIGACLIP | SE |
| Clip Size | One size | S,M,L | M,M-L, L | Clip size range tested in<br>animal and laboratory<br>bench testing demonstrated<br>single clip performs as<br>intended through upper<br>limit labeled vessel size. |
| Vessel Size | 2.0mm-7.0mm | 2.0mm-3.5mm (M) | 2.0mm-6.0mm | Clip size range tested in<br>animal and laboratory<br>bench testing demonstrated<br>single clip performs as<br>intended over labeled vessel<br>size. |
| Length of clip | 5.5mm | unknown | 5.5mm | Similar. Testing<br>demonstrated that Amsel<br>clip length is sufficient for<br>safe and effective use. |
| Secured to vessel or<br>tubular structure | Mechanical, by compressing the<br>clip components (sliding and<br>locking two nitinol elements one<br>against the other on either side of<br>the vessel and locked together with<br>a titanium rod) | Mechanical compression | Mechanical compression | Same |
| Closure method | Penetrating through (transfixing)<br>the vessel using a needle and then<br>deploying locking the vessel<br>externally | External vessel compression | External vessel compression | Similar. All occlude via<br>external compression of the<br>vessel. Amsel closure<br>technology demonstrated as<br>safe and effective in animal<br>and bench testing. |
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Amsel Medical. The logo features a stylized letter "A" in blue and gray, with the words "AMSEL MEDICAL" in gray next to it. The word "AMSEL" is above the word "MEDICAL". There is a reflection of the logo below it.
{9}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for Amsel Medical. The logo consists of a stylized letter A in blue and gray, followed by the words "AMSEL" in gray and "MEDICAL" in blue. The logo has a reflection effect below it.
| | AMSEL OCCLUDER DEVICE | TWISTER SUPERELASTIC LIGATING<br>CLIP | LIGACLIP | SE |
|-----------------------------------|------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clip Material | Nitinol and titanium | Nitinol | Titanium | Similar. Amsel uses the<br>same and additional<br>material. All materials<br>tested for biocompatibility<br>and functionality and any<br>differences do not adversely<br>affect safety or<br>effectiveness. |
| Single Patient Use,<br>Disposable | Yes | Yes | Yes | Same |
| Sterilization | Sterile for single use EtO | Sterile for single use<br>EtO | Sterile for single use<br>EtO | Same |
| Prescription Use | Yes. The device should be used<br>only by trained surgeon. | Yes. The device should be used only by<br>trained surgeon. | Yes. The device should be used only<br>by trained surgeon. | Same |
{10}------------------------------------------------
As demonstrated in the completed battery of preclinical tests that were conducted by the company minor technological differences between the Amsel Occluder Device and the predicate devices do not raise new questions of safety and effectiveness. Any minor differences in technological characteristics have been tested and reported on in this notification and demonstrate that such differences do not adversely affect the safety, effectiveness, or intended performance of the device. Therefore, the Amsel Occluder Device is substantially equivalent to the legally marketed predicate devices. Testing in support of the SE determination included laboratory bench testing, biocompatibility testing, shelf life testing and animal testing.
# 5.12 Performance Characteristics of the Amsel Occluder Device
The Amsel Occluder Device underwent a full battery of bench testing and animal testing to demonstrate its safe and effective performance in occluding vessels. It was concluded that the device is safe and effective for its intended use. Tests were performed in comparison to the predicate device. Specifically, testing included:
- Clip Closure Distance to verify Clip closure distance to ensure it is within the designed specification Clip Closure Repeatability and Reliability ● to verify the repeatable compression and locking of the occluder clip ● Strain Analysis the purpose was to analyze the maximal strain of the clips during occlusion Corrosion Test ● Verification that the device does not develop corrosion post implantation. Testing was performed per ASTM 2129-08 Standard Test Method for Conducting Cyclic Potentiodynamic Polarization Measurements to Determine the Corrosion Susceptibility of Small Implant Devices and ASTM F3044-14 Test Method for Standard Test Method for Evaluating the Potential for Galvanic Corrosion for Medical Implants Performance Evaluation ● to verify the performance of the Amsel Occluder device in a simulated model ● Device to Implant Detachment Force to verify that the force required to detach the clip from the delivery system is sufficiently robust to prevent inadvertent detachment of the delivery system from the clip ● Holding Pressure Evaluation verification of occlusion performance by pressurizing occluded structure through pressures over 700mmHg ● Release Mechanism Function Testing to verify proper detachment of the delivery system from the implant during operation
{11}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the logo for Amsel Medical. The logo features a stylized letter "A" in blue and gray, with the word "AMSEL" in gray and "MEDICAL" in blue to the right of the letter. The logo is clean and modern, with a professional look.
## Σ Deployment Force Testing
- The force required to deploy the Amsel Occluder Device """"""""""""""""""""""""""was tested in comparison to a predicate. Results were that """""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""""" """""""""""""" the device
- Biocompatibility Testing ●
- to verify that all body contact materials are biocompatible for the contact and duration of contact
- Animal study including holding pressure and histology
- preclinical in-vivo use of the Amsel device compared to a predicate device to evaluate the safety and effectiveness of the Amsel device in terms of introduction, deployment, usability, and occlusion performance in arteries and veins 2.0mm to 7.0mm
- . MR Imaging Testing
-
to support the MR conditional claim in the device's labeling.
All testing performed demonstrated that the Amsel Occluder is a safe and effective device for vessel and tubular structure occlusion and is considered substantially equivalent to its predicate devices.
# 5.13 Conclusion
Amsel Medical believes that, based on the descriptive information and testing provided in this submission, the Amsel Occluder Device is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.