PMT Posterior Fixation System

K172279 · Providence Medical Technology, Inc. · NKG · Sep 20, 2017 · Orthopedic

Device Facts

Record IDK172279
Device NamePMT Posterior Fixation System
ApplicantProvidence Medical Technology, Inc.
Product CodeNKG · Orthopedic
Decision DateSep 20, 2017
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3075
Device ClassClass 2
AttributesTherapeutic

Intended Use

The PMT Posterior Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine from T1 to T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PMT Posterior Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

Device Story

PMT Posterior Fixation System; bone screw and rod construct for spinal stabilization. Components: polyaxial screws (3.5mm/4.0mm diameter; 8-24mm length), fixation rods (3.5mm/3.8mm diameter; 15-240mm length), and locking set screws. Surgeon implants screws into pedicle or posterior lateral mass of cervical/thoracic vertebrae; rods connect screws; set screws lock rod orientation. Used in OR by surgeons to immobilize spinal segments as adjunct to fusion or for palliative stabilization in advanced tumor cases. Benefits: spinal column immobilization/restoration of integrity.

Clinical Evidence

Bench testing only. Static compression bending, static torsion, and dynamic compression bending performed per ASTM F1717.

Technological Characteristics

Materials: 6Al-4V Titanium Alloy (ASTM F136). Design: Polyaxial screw and rod construct. Dimensions: Screws 3.5-4.0mm diameter, 8-24mm length; Rods 3.5-3.8mm diameter, 15-240mm length. Sterilization: Supplied sterile, single-use. No software or electronic components.

Indications for Use

Indicated for patients with acute/chronic instabilities of cervical (C1-C7) and thoracic (T1-T3) spine, including traumatic fractures/dislocations, instability/deformity, pseudarthrosis, tumors, and degenerative disease (radiculopathy, myelopathy, discogenic pain, facet instability). Also indicated for spinal column integrity restoration in patients with advanced cervical tumors and limited life expectancy where fusion is not achievable.

Regulatory Classification

Identification

Posterior cervical screw systems are comprised of multiple, interconnecting components, made from a variety of materials that allow an implant system to be built from the occiput to the upper thoracic spine to fit the patient's anatomical and physiological requirements, as determined by preoperative cross-sectional imaging. Such a spinal assembly consists of a combination of bone anchors via screws (i.e., occipital screws, cervical lateral mass screws, cervical pedicle screws, C2 pars screws, C2 translaminar screws, C2 transarticular screws), longitudinal members (e.g., plates, rods, including dual diameter rods, plate/rod combinations), transverse or cross connectors, interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors), and closure mechanisms (e.g., set screws, nuts). Posterior cervical screw systems are rigidly fixed devices that do not contain dynamic features, including but not limited to: non-uniform longitudinal elements or features that allow more motion or flexibility compared to rigid systems.Posterior cervical screw systems are intended to provide immobilization and stabilization of spinal segments in patients as an adjunct to fusion for acute and chronic instabilities of the cervical spine and/or craniocervical junction and/or cervicothoracic junction such as: (1) Traumatic spinal fractures and/or traumatic dislocations; (2) deformities; (3) instabilities; (4) failed previous fusions (e.g., pseudarthrosis); (5) tumors; (6) inflammatory disorders; (7) spinal degeneration, including neck and/or arm pain of discogenic origin as confirmed by imaging studies (radiographs, CT, MRI); (8) degeneration of the facets with instability; and (9) reconstruction following decompression to treat radiculopathy and/or myelopathy. These systems are also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion.

Special Controls

*Classification.* Class II (special controls). The special controls for posterior cervical screw systems are:(1) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (2) Nonclinical performance testing must demonstrate the mechanical function and durability of the implant. (3) Device components must be demonstrated to be biocompatible. (4) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (5) Labeling must include the following: (i) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (ii) Intended use and indications for use including levels of fixation; (iii) Device specific warnings, precautions, and contraindications that include the following statements: (A) “Precaution: Preoperative planning prior to implantation of posterior cervical screw systems should include review of cross-sectional imaging studies ( *e.g.,* CT and/or MRI) to evaluate the patient's cervical anatomy including the transverse foramen, neurologic structures, and the course of the vertebral arteries. If any findings would compromise the placement of these screws, other surgical methods should be considered. In addition, use of intraoperative imaging should be considered to guide and/or verify device placement, as necessary.”(B) “Precaution: Use of posterior cervical pedicle screw fixation at the C3 through C6 spinal levels requires careful consideration and planning beyond that required for lateral mass screws placed at these spinal levels, given the proximity of the vertebral arteries and neurologic structures in relation to the cervical pedicles at these levels.” (iv) Identification of magnetic resonance (MR) compatibility status; (v) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user, and; (vi) Detailed instructions of each surgical step, including device removal.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 September 20, 2017 Providence Medical Technology, Inc. Mr. Edward Liou Chief Operating Officer 1331 North California Boulevard. Suite 320 Walnut Creek, California 94596 Re: K172279 Trade/Device Name: PMT Posterior Fixation System Regulatory Class: Unclassified Product Code: NKG Dated: July 26, 2017 Received: July 28, 2017 Dear Mr. Liou: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely, Ronald P. Jean - S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K172279 Device Name PMT Posterior Fixation System #### Indications for Use (Describe) The PMT Posterior Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine from T1 to T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PMT Posterior Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. | Type of Use (Select one or both, as applicable) | | |-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------| | <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # Section 6. 510(k) Summary | 510(k) Owner: | Providence Medical Technology, Inc.<br>1311 N. California Blvd, Suite 320<br>Walnut Creek, CA 94596<br>T: 415-923-9376<br>F: 415-923-9377 | |-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Edward Liou<br>Email: ed@providencemt.com<br>T: 415-923-9376 | | Date Summary Prepared: | September 19, 2017 | | Trade Name: | PMT Posterior Fixation System | | Common Name: | Orthosis, Cervical Pedicle Screw Spine Fixation | | Device Classification Regulation: | Unclassified | | Device Product Code & Panel: | NKG<br>Orthopaedic and Rehabilitation Devices Panel | | Primary Predicate Device: | Zimmer Spine, Inc. Virage OCT Spinal Fixation System (K151031) | #### Device Information ## A. Device Description The PMT Posterior Fixation System is a bone screw and rod construct that will be supplied sterile and single use only. It consists of polyaxial screws, fixation rods, and locking set screws. The polyaxial screws are designed to be utilized for pedicle fixation or anchored to the posterior lateral mass of the cervical (C1 to C7) and thoracic (T1 to T3) vertebra. A set screw is used to connect and anchor the rod to each screw head while locking the orientation of the polyaxial screw. The PMT Posterior Fixation System is manufactured from 6Al-4V Titanium Alloy, conforming to ASTM F136. The polyaxial screw, available in diameters of 3.5mm and 4.0mm, is self-tapping with a range of screw lengths between 8mm to 24mm. The rod is offered in two diameters, 3.5mm and 3.8mm, with a range in lengths of 15mm to 240mm. The screw head that is assembled to the polyaxial screw accepts the 3.5mm or 3.8mm rod. The set screw comes in one size and has a matching thread profile to the screw head. {4}------------------------------------------------ #### B. Intended Use and Indications for Use The PMT Posterior Fixation System is intended to provide immobilization of spinal segments as an adjunct to fusion for the following acute and chronic instabilities of the cervical spine (C1 to C7) and the thoracic spine from T1 to T3: traumatic spinal fractures and/or traumatic dislocations; instability or deformity; failed previous fusions (e.g., pseudarthrosis); tumors involving the cervical spine; and degenerative disease, including intractable radiculopathy and/or myelopathy, neck and/or arm pain of discogenic origin as confirmed by radiographic studies, and degenerative disease of the facets with instability. The PMT Posterior Fixation System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion. ## C. Summary of Technological Characteristics The fundamental operational principles, surgical approach, design, materials, and performance of the PMT Posterior Fixation System and the predicate device are essentially the same. Therefore, the technological characteristics of the PMT Posterior Fixation System do not raise any new or different safety and effectiveness questions not addressed in the predicate device. ### D. Summary of Performance Testing Non-clinical testing conducted per ASTM F1717 (static compression bending, static torsion, and dynamic compression bending) has demonstrated that the PMT Posterior Fixation System is substantially equivalent to the predicate device. ## E. Basis of Substantial Equivalence The PMT Posterior Fixation System is substantially equivalent compared to the cleared predicate device, Virage OCT Spinal Fixation System (K151031), on the basis the devices have the same intended use and indications for use, same technological characteristics, similar design features, and same materials and principles of operation. Performance data demonstrate that the PMT Posterior Fixation System is as safe and effective as the predicate device.
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...