The Magic3 Go® Intermittent Urinary Catheter is intended for urological use only. It is intended for use by adult and pediatric patients of all ages for bladder management including urine drainage, collection, and measurement. The device is passed to the urinary bladder via the urethra.
Device Story
Ready-to-use, single-use, sterile silicone intermittent urinary catheter; features hydrophilic coating with self-hydrating layer requiring no water activation; includes polyethylene Sure-Grip™ insertion sleeve for non-touch handling; available in French sizes 6–20Fr, lengths 10" and 16", and straight or coudé tip designs. Used by patients or clinicians for bladder management; device inserted via urethra to bladder to facilitate urine drainage, collection, and measurement. Benefits include ease of use, reduced contamination risk via non-touch sleeve, and elimination of pre-activation steps.
Clinical Evidence
No clinical data. Substantial equivalence supported by biocompatibility testing (ISO 10993-5, ISO 10993-10) and non-clinical functional performance testing per BS EN 1616:1997 + A1:1999. Lubricity performance established via equivalence to predicate K150345.
Technological Characteristics
All-silicone single lumen catheter; polyethylene Sure-Grip™ insertion sleeve; hydrophilic coating with self-hydrating layer. Latex, DEHP, and phthalate-free. Sterile, single-use. Complies with BS EN 1616:1997 + A1:1999. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for adult and pediatric patients of all ages requiring bladder management, including urine drainage, collection, and measurement via urethral catheterization.
Regulatory Classification
Identification
A urological catheter and accessories is a flexible tubular device that is inserted through the urethra and used to pass fluids to or from the urinary tract. This generic type of device includes radiopaque urological catheters, ureteral catheters, urethral catheters, coudé catheters, balloon retention type catheters, straight catheters, upper urinary tract catheters, double lumen female urethrographic catheters, disposable ureteral catheters, male urethrographic catheters, and urological catheter accessories including ureteral catheter stylets, ureteral catheter adapters, ureteral catheter holders, ureteral catheter stylets, ureteral catheterization trays, and the gastro-urological irrigation tray (for urological use).
{1}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 20, 2017
C.R. Bard, Inc. Nancy Underwood Regulatory Affairs Specialist 8195 Industrial Blvd Covington, GA 30014
K172247 Re:
Trade/Device Name: Magic3 Go® Intermittent Urinary Catheter Regulation Number: 21 CFR§ 876.5130 Regulation Name: Urological Catheter and Accessories Regulatory Class: II Product Code: EZD Dated: July 25, 2017 Received: July 26, 2017
Dear Nancy Underwood:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{2}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Charles Viviano -S
Benjamin R. Fisher, Ph.D. For Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
510(k) Number (if known)
#### K172247
#### Device Name
Magic3 Go® Intermittent Urinary Catheter
Indications for Use (Describe)
The Magic3 Go® Intermittent Urinary Catheter is intended for urological use only. It is intended for use by adult and pediatric patients of all ages for bladder management including urine drainage, collection, and measurement. The device is passed to the urinary bladder via the urethra.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------|----------------------------------------------------------------------|
| <input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (7/17)
Page 1 of 1
Trade Secret | Confidential Information
Notify Bard Medical Division
{4}------------------------------------------------
# Magic3 Go® Intermittent Urinary Catheter 510(k) Summary 21 CFR 807.92
As required by the Safe Medical Devices Act of 1990, coded under Section 513, Part (I)(3)(A) of the Food, Drug and Cosmetic Act, a 510(k) summary upon which substantial equivalence determination is based is as follows:
### Submitter Information:
Bard Medical Division C. R. Bard, Inc. 8195 Industrial Boulevard Covington GA 30014 Phone: 770-784-6284 Fax: 770-385-4706 Contact Person: Nancy Underwood, Regulatory Affairs Specialist Date of Submission: July 25, 2017
## Subject Device Name:
| Name of Device: | Magic3 Go® Intermittent Urinary Catheter |
|-----------------------------------------|------------------------------------------------------|
| Common or Usual<br>Name: | Urological Catheter |
| Classification Name: | Urological catheter and accessories |
| FDA Pro Code: | EZD |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR §876.5130 |
| Predicate Devices:<br>Primary Predicate | |
| Name of Device: | Magic3 Go® Intermittent Urinary Catheter,<br>K150345 |
| Common or Usual<br>Name: | Urological Catheter |
| Classification Name: | Urological catheter and accessories |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR §876.5130 |
| Name of Device: | Hydrophilic Personal Catheter®, K000723 |
| Common or Usual<br>Name: | Urological Catheter |
| Classification Name: | Urological catheter and accessories |
| Regulatory Class: | II |
| Regulation Number: | 21 CFR §876.5130 |
{5}------------------------------------------------
### Device Description:
The Magic3 Go® Intermittent Urinary Catheter is a ready-to-use silicone intermittent catheter with a self-hydrating over a hydrophilic coating. The device is provided sterile and is for single use.
The device consists of an all silicone single lumen catheter with a polyethylene insertion sleeve (Sure-Grip™) which is not made from natural rubber latex, DEHP and/or other phthalates. The catheter has four drainage eyes located in the proximal tip and a tapered funnel located at the distal end. The catheter will be offered in multiple French sizes (6 – 20Fr), multiple lengths (10" & 16") and two tip designs (straight and coudé).
The outer surface of the catheter has two coatings. The first coating is a hydrophilic coating. Over the hydrophilic coating, a self-hydrating coating is applied. The additional coating does not require activation with water prior to use. The Sure-Grip™ insertion sleeve provides the user with an ergonomically designed area for a secure grip and notouch area with which to insert the catheter.
## Indications for Use of Device:
The Magic3 Go® Intermittent Urinary Catheter is intended for urological use only. It is intended for use by adult and pediatric patients of all ages for bladder management including urine drainage, collection, and measurement. The device is passed to the urinary bladder via the urethra.
## Technological Comparison to Predicate Devices:
The Magic3 Go® Intermittent Urinary Catheter has similar technological characteristics as the predicate device, Magic3 Go® Intermittent Urinary Catheter previously cleared via K150345 and the Hydrophilic Personal Catheter® previously cleared via K000723. The subject and predicate devices are based on the following technological elements:
- Same intended use .
- . Similar indications for use
- Same drainage eye position
{6}------------------------------------------------
- . Similar tip configurations
- Provided sterile for single-use .
- . Composed of biocompatible materials
- Same catheter materials
- . Same self-hydrating coating
- . Similar design features
- Similar product offerings
#### Performance Data:
The following performance data were provided in support of the substantial equivalence determination:
#### Biocompatibility
The biocompatibility evaluation of the subject device was conducted in accordance with Use of International Standard ISO 10993-1, Biological evaluation of medical devices and testing within a risk management process – July 21, 2016. The subject device is considered a mucosal contacting surface device with limited exposure. The battery of testing included the following:
- . Cytotoxicity (ISO 10993-5)
- . Sensitization (ISO 10993-10)
- . Vaginal Mucosal Irritation (ISO 10993-10)
#### Non-clinical functional performance testing
Non-clinical functional performance testing of the subject device was performed in accordance with BS EN 1616:1997 + A1:1999, Sterile urethral catheters for single use.
Coefficient of friction (CoF) testing was previously conducted on the cleared predicate device K150345. Since the subject device is identical to the predicate device K150345, in terms of catheter material and coatings, it has been determined that lubricity testing was not necessary on the subject device. The lubricity of the subject device is determined to be equivalent to the predicate device, K150345.
{7}------------------------------------------------
## Conclusions:
The Magic3 Go® Intermittent Urinary Catheter is substantially equivalent to the legally marketed predicate devices as demonstrated by the same intended use, similar indications for use, similar technological characteristics and performance data, and does not raise different questions of safety and effectiveness
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.