APTUS® Wrist 2.5 System is intended for use in hand and forearm fractures, osteotomies, and arthrodeses.
Device Story
APTUS® Wrist 2.5 System consists of metallic bone plates and compatible screws for orthopedic fixation. Plates are pre-contoured for left and right distal radius anatomy (Lunate Facet and Distal Radius Rim designs). Device is used by orthopedic surgeons in clinical settings to stabilize bone fractures, osteotomies, or arthrodeses. Plates are implanted via surgical procedure; screws provide locking or non-locking fixation to bone. System provides mechanical stability to bone segments during healing; benefits include anatomical alignment and rigid fixation. Device is provided non-sterile; requires end-user sterilization.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical bench testing, including engineering analysis, static mechanical testing per ASTM F382, and dynamic fatigue testing in a simulated fracture model.
Technological Characteristics
Materials: Unalloyed titanium Grade 4 (ASTM F67) for plates; Ti-6Al-4V alloy (ASTM F136) for screws. Design: Pre-contoured anatomic plates (1.6-1.8 mm thickness). Mechanical: Locking and non-locking screw fixation. Sterilization: Non-sterile, end-user sterilized. Connectivity: None (mechanical implant).
Indications for Use
Indicated for patients requiring fixation for hand and forearm fractures, osteotomies, and arthrodeses.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal are three stylized human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 8, 2017
Medartis AG % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K172170
Trade/Device Name: APTUS® Wrist 2.5 System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS Dated: July 17, 2017 Received: July 18, 2017
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K172170
Device Name
APTUS® Wrist 2.5 System
Indications for Use (Describe)
APTUS® Wrist 2.5 System is intended for use in hand and forearm fractures, osteotomies, and arthrodeses.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.qov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summary
## Medartis AG
# APTUS® Wrist 2.5 System
July 17, 2017
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | Medartis AG<br>Hochbergerstrasse 60E<br>CH-4057 Basel, Switzerland |
|-------------------|--------------------------------------------------------------------|
| | Telephone: +41 61 633 34 34<br>Fax: +41 61 633 34 00 |
| Official Contact | Andrea Schweizer<br>Head of Quality Management |
Representative/Consultant Kevin A. Thomas, PhD Floyd G. Larson, MS, MBA PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, CA 92130 Telephone: +1-858-792-1235 Fax: +1-858-792-1236 Email: kthomas@paxmed.com flarson@paxmed.com
## DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | APTUS® Wrist 2.5 System |
|----------------------------|--------------------------------------------------------------------------------|
| Common Name | Plate, fixation, bone |
| Classification Names | Single/multiple component metallic bone fixation appliances<br>and accessories |
| Classification Regulations | 21 CFR 888.3030 |
| Product Codes | HRS |
| Classification Panel | Orthopedic Products Panel |
| Reviewing Branch | Joint Fixation Devices Branch Two (JFDB2) |
#### PREDICATE DEVICE INFORMATION
The primary predicate device is K142906. The reference predicate device is K051567.
#### INDICATIONS FOR USE
APTUS® Wrist 2.5 System is intended for use in hand and forearm fractures, and arthrodeses.
#### SUBJECT DEVICE DESCRIPTION
The subject device plates have anatomic designs and are provided in two designs, each for the left and right radius: Lunate Facet Plate and Distal Radius Rim Plate. The Lunate Facet Plates have a maximum thickness of 1.6 mm, and maximum overall dimensions of approximately 19 mm x 47 mm before bending to conform to the anatomy of the distal radius. The Distal Radius Rim Plates have a maximum thickness of 1.8 mm, and maximum overall dimensions of approximately 22 mm x 53 mm before bending to
{4}------------------------------------------------
510(k) Summary Page 2 of 3
conform to the anatomy of the distal radius. The screw holes of the subject device plates are designed to accommodate appropriately sized bone screws (locking and non-locking) presently marketed as part of the APTUS System and previously cleared under K142906, K103332, and K051567. The subject device plates also are compatible with K-wires cleared under K092038. The subject device plates are made of unalloved titanium. Grade 4, conforming to ASTM F67 Standard Specification for Unalloyed Titanium for Surgical Implant Applications (UNS R50250, UNS R50400, UNS R50550, UNS R50700).
## PERFORMANCE DATA
Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence include: biocompatibility (referenced from K142906), engineering analysis, static mechanical testing (according to ASTM F382), and dynamic mechanical testing of the subject and predicate plate designs in a simulated fracture model. Clinical data were not provided in this submission.
# EQUIVALENCE TO MARKETED DEVICE
Medartis AG submits the information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K142906, APTUS® Wrist 2.5 System, Medartis AG
K051567, APTUS® Titanium System, Medartis, Inc.
A comparison of the technological characteristics of the subject device and the primary predicate device K142906 is provided in the following table.
| Comparison | Subject Device | Primary Predicate Device |
|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------|
| | APTUS Wrist 2.5 System<br>Medartis AG | K142906<br>APTUS Wrist 2.5 System<br>Medartis AG |
| Indications for Use<br>Statement | APTUS® Wrist 2.5 System is intended for use in hand<br>and forearm fractures, osteotomies, and arthrodeses. | APTUS® Wrist 2.5 System is intended for use in hand<br>and forearm fractures, osteotomies and arthrodeses. |
| Device Characteristics | | |
| Plate Designs | Anatomic plate designs (n=2) for left and right radius<br>Lunate Facet Plate and Distal Radius Rim Plate | Anatomic plate designs (multiple) for left and right<br>radius |
| Plate Thickness | Lunate Facet Plate – 1.6 mm<br>Distal Radius Rim Plate - 1.8 mm | 0.6 mm to 3.2 mm |
| Plate Overall Dimensions<br>(approximate) | Lunate Facet Plate - 19 mm width x 47 mm length<br>Distal Radius Rim Plate – 22 mm width x 53 mm<br>length | Width: 7 mm to 36 mm<br>Length: 7 mm to 184 mm |
| Plate Material | Unalloyed titanium, grade 4 (ASTM F67) | Unalloyed titanium, grade 4 (ASTM F67) |
| Information below for Subject Device is for previously cleared compatible screws | | |
| Screw Designs | Self-drilling and self-tapping;<br>Cortical (conventional non-locking) and locking | Self-drilling and self-tapping;<br>Cortical (conventional non-locking) and locking |
| Screw Diameter | 1.5 mm and 2.5 mm | 1.5 mm and 2.5 mm |
| Screw Length | 8 mm to 34 mm | 8 mm to 34 mm |
| Screw Material | Ti-6Al-4V alloy (ASTM F136) | Ti-6Al-4V alloy (ASTM F136) |
The subject device and the predicate device have the same intended use and the same Indications for Use statement.
The subject device and predicate device plates have the same technological characteristics are fabricated from the same materials and share similar design characteristics, including plate screw holes to accommodate locking and non-locking screws. The subject device and predicate device plates also are
{5}------------------------------------------------
provided in pre-contoured designs for the left and right radius, and encompass a similar range of physical dimensions appropriate to the anatomy. Any minor differences in the technological characteristics between the subject device and the predicate device do not raise new issues of safety or efficacy.
The subject device and predicate device plates are provided non-sterile and are to be sterilized by the enduser. The subject device components are packaged using the same materials and are to be sterilized by the same methods as the predicate device.
All of the subject device final finished components are manufactured in the same facilities using identical materials and identical manufacturing processes as used for the predicate device components cleared in K142906, and therefore are substantially equivalent to the predicate device regarding biocompatibility.
Performance data provided to demonstrate substantial equivalence included engineering analysis, single cvcle bending and bending fatigue testing of the subject device and predicate device plates. In addition, in a simulated fracture models, the dynamic (fatigue) performance of the subject device constructs was superior to that of the predicate device constructs.
Differences between the subject device and predicate device plates include differently shaped anatomic designs for the distal radius. The subject device and predicate device plates have a similar overall shape and similar maximum plate thickness (subject device plates 1.6 mm, predicate plates 1.6 mm and 2.0 mm). Substantial equivalence is supported by the mechanical testing of the subject device and predicate device plate and screw constructs.
# CONCLUSION
The subject device and the predicate device plates have the same intended use, have similar technological characteristics, and are made of the same material. The subject device and predicate device plates also are provided in pre-contoured anatomic designs for the left and right radius, and encompass a similar range of physical dimensions appropriate to the anatomy. The subject device and predicate device components are packaged in similar materials and are to be sterilized by the end-user using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices K142906 and K051567.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.