The uMI 780 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images. This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Device Story
uMI 780 PET/CT system combines PET and CT modalities for diagnostic imaging. PET component measures quantitative distribution of radiopharmaceuticals; CT component provides anatomical tomographic images and photon attenuation maps. System performs spatial alignment and image registration/fusion to provide anatomical reference for PET findings. Operated by qualified healthcare professionals in clinical settings. Workstation software manages patient data, scan control, image reconstruction, and reading. Output consists of fused PET/CT images used by physicians for disease assessment, staging, and treatment planning. Benefits include improved diagnostic accuracy through combined metabolic and anatomical data.
Clinical Evidence
Bench testing only. No clinical study included. Performance verified via NEMA NU 2:2012 testing and compliance with electrical/radiation safety standards (IEC 60601 series, AAMI ANSI ES60601-1).
Indicated for diagnostic imaging to assist in detection, localization, diagnosis, staging, treatment planning, and treatment response evaluation for diseases and disorders, including but not limited to oncology, cardiology, and neurology. Intended for use by qualified healthcare professionals.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
Predicate Devices
Biograph mCT and mCT Flow PET/CT Scanners (K151486)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Shanghai United Imaging Healthcare Co., Ltd. % Shumei Wang QM&RA VP NO.2258 Chengbei Road,Jiading District Shanghai. 201807 CHINA
Re: K172143
Trade/Device Name: uMI 780 PET/CT System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: II Product Code: KPS, JAK, Dated: March 10, 2018 Received: March 13, 2018
Dear Shumei Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
April 13, 2018
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and
Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D'Hara For
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized symbol that resembles a shield or crest. The symbol is divided into sections by a white line, creating a modern and abstract design.
## Indications for Use
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drup Administration Indications for Use
Form Approved:OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below
510(k) Number( if known)
K172143
Device Name
uMI 780 PET/CT System
#### Indications for Use (Describe):
The uMI 780 PET/CT is a diagnostic imaging system that combines two existing imaging modalities -PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration of PET and CT images provides anatomical reference for the findings in the PET images.
This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
Type of use:
> Precription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (Part 21 CFR 801 Subpart D)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
PSC Publishing Services (301) 443-6740 FF
### SECTION 77 - 1 of 1
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for United Imaging. The text "UNITED IMAGING" is displayed in bold, sans-serif font. To the right of the text is a stylized logo consisting of a dark gray shield shape with a white vertical line running down the center and a horizontal line intersecting it near the top, forming a stylized "U" shape.
# SECTION 16
# 510(k) Summary
SECTION 16 - Page 1 of 7
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the words is a stylized "U" shape, which is made up of two vertical lines and a horizontal line in the middle. The logo is simple and modern, and the colors are muted.
## 510(k) SUMMARY
#### 1. Date of Preparation: March 9, 2018
#### 2. Sponsor Identification
| Client Name: | Shanghai United Imaging Healthcare Co.,Ltd. |
|---------------------------------------|---------------------------------------------------------|
| Client Address: | NO.2258 Chengbei Road, Jiading District, Shanghai,China |
| Establishment<br>Registration Number: | Not yet registered |
| Contact Person: | Wang Shumei |
| Position: | QM&RA VP |
| Tel: | +86-21-67076888-6776 |
| Fax: | +86-21-67076889 |
| Email: | shumei.wang@united-imaging.com |
#### 3. Identification of Proposed Device
| Model(s): | uMI 780 |
|----------------------|-----------------------------------------------------------------------------------------------------------------|
| Trade Name: | uMI 780 |
| Common Name: | Positron Emission Tomography and<br>Computed Tomography System |
| Classification Name: | Emission Computed Tomography System per 21 CFR 892.1200<br>Computed Tomography X-Ray System per 21 CFR 892.1750 |
| Regulation Number: | 21 CFR 892.1200 |
| Product Code: | KPS and JAK |
| Review Panel: | Radiology |
| Classification: | Class II |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for United Imaging. The words "UNITED" and "IMAGING" are stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" symbol, which is also in a bold, sans-serif font. The logo is simple and modern, and the use of a bold font gives it a strong and confident look.
#### Proposed Device Description 4.
The uMI 780 PET/CT system is a combined multi-slice X-Ray Computed Tomography and Positron Emission Tomography scanner. This system is intended to be operated by qualified healthcare professionals for performing diagnostic imaging examinations. The spatial alignment and precise image registration between PET and CT ensure the PET and CT images of the same region can be fused accurately for reading. PET measures the distribution of PET radiopharmaceuticals inside the human body quantitatively. CT produces the anatomical information of the same scanned region, and provides accurate localization for the findings in the PET images. The attenuation information contained in the CT images can be utilized in the PET image reconstruction to ensure quantitation accuracy. The PET system has time-of-flight capability with a timing resolution of 520ps. It has a 300mm-long axial field of view (FoV) and a system sensitivity higher than 15cps/kBq.
The uMI 780 PET/CT system also includes a patient table, a workstation with associated software installed. The software is used for patient management, data management, scan control, image reconstruction and image reading. All patient images produced by the system conform to the DICOM 3.0 standard.
### 5. Intended Use Statement of Proposed Device
The uMI 780 PET/CT is a diagnostic imaging system that combines two existing imaging modalities - PET and CT. The quantitative distribution information of PET radiopharmaceuticals within the patient body measured by PET can assist healthcare providers in assessing the metabolic and physiological functions. CT provides diagnostic tomographic anatomical information as well as photon attenuation information for the scanned region. The accurate registration and fusion of PET and CT images provides anatomical reference for the findings in the PET images.
This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, treatment planning and
## SECTION 16 - Page 3 of 7
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the words "UNITED IMAGING" stacked on top of each other in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is formed by two vertical lines and a horizontal line connecting them at the top. The logo is simple and modern, and the use of bold text and a clean design makes it easily recognizable.
treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.
#### Identification of Predicate Device(s) 6.
| 510(k) Number: | K151486 |
|----------------|------------------------------------------------------------------------------|
| Product Name: | Positron Emission Tomography (PET) System<br>Computed Tomography (CT) System |
| Model Name: | Biograph mCT and mCT Flow PET/CT Scanners |
### 7. Non-clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- > Recognition Number 19-4:AAMI ANSI ES60601-1: 2005/(R)2012 and A1:2012
- > Recognition Number 12-269:IEC 60601-1-3: Edition 2.1
- > Recognition Number 12-257:IEC 60601-2-44: Edition 3.0
- > Recognition Number 19-1:IEC 60601-1-2: Edition 3
- > Recognition Number 12-273:IEC 60825-1:Edition 2.0
- > Recognition Number 12-270:NEMA 61223-3-5: First edition
- > Recognition Number 12-265:NEMA NU 2:2012
- > Recognition Number 2-174:ISO 10993-10:Third Edition
- > Recognition Number 2-245:ISO 10993-5:Third edition
- > Recognition Number 12-225:NEMA XR 25: 2010
- > Recognition Number 12-287:NEMA XR 28:2013
- > NEMA XR 29:2013
#### 8. Clinical Test Conclusion
No clinical study is included in this submission.
## SECTION 16 - Page 4 of 7
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image contains the logo for United Imaging. The logo consists of the word "UNITED" stacked on top of the word "IMAGING", both in a bold, sans-serif font. To the right of the text is a stylized "U" shape, which is also in a bold font. The color of the logo is a dark gray.
#### 9. Substantial Equivalence Comparison
| Table 1 Comparison of Technology Characteristics | | | | | | |
|--------------------------------------------------|--|--|--|--|--|--|
|--------------------------------------------------|--|--|--|--|--|--|
| Item | Proposed Device(s) | Predicate Device(s) |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | KPS,JAK | KPS,JAK |
| Regulation No. | 21 CFR 892.1200 | 21 CFR 892.1200 |
| Class | II | II |
| Intended Use | The uMI 780 PET/CT is a<br>diagnostic imaging system that<br>combines two existing imaging<br>modalities - PET and CT. The<br>quantitative distribution<br>information of PET<br>radiopharmaceuticals within the<br>patient body measured by PET can<br>assist healthcare providers in<br>assessing the metabolic and<br>physiological functions. CT<br>provides diagnostic tomographic<br>anatomical information as well as<br>photon attenuation information for<br>the scanned region. The accurate<br>registration and fusion of PET and<br>CT images provides anatomical<br>reference for the findings in the<br>PET images.<br>This system is intended to be<br>operated by qualified healthcare<br>professionals to assist in the<br>detection, localization, diagnosis,<br>staging, restaging, treatment<br>planning and treatment response<br>evaluation for diseases and<br>disorders in, but not limit to,<br>oncology, cardiology and<br>neurology. | The Siemens Biograph mCT systems<br>are combined X-Ray Computed<br>Tomography (CT) and Positron<br>Emission<br>Tomography (PET) scanners that<br>provide registration and fusion of<br>high resolution physiologic and<br>anatomic information.<br>The CT component produces cross-<br>sectional images of the body by<br>computer reconstruction of X-Ray<br>transmission data from either the<br>same axial plane taken at different<br>angles or spiral planes taken at<br>different angles. The PET subsystem<br>images and measures the distribution<br>of PET radiopharmaceuticals in<br>humans for the purpose of<br>determining various metabolic<br>(molecular) and physiologic<br>functions within the human body and<br>utilizes the CT for fast attenuation<br>correction maps for PET studies and<br>precise anatomical reference for the<br>fused PET and CT images.<br>The system maintains independent<br>functionality of the CT and PET<br>devices, allowing for single modality<br>CT and / or PET diagnostic imaging.<br>These systems are intended to be<br>utilized by appropriately trained<br>health care professionals to aid in<br>detecting, localizing, diagnosing,<br>staging and restaging of lesions,<br>tumors, disease and organ function<br>for the evaluation of diseases and<br>disorders such as, but not limited to,<br>cardiovascular disease, neurological<br>disorders and cancer. The images<br>produced by the system can also be<br>used by the physician to aid in<br>radiotherapy treatment planning and<br>interventional radiology procedures. |
| Item | Proposed Device(s) | Predicate Device(s) |
| PET Specification | | |
| Sensitivity | >/=15cps/kBq | >/=9.4cps/kBq |
| NECR Peak Value | >/=165kcps@16kBq/cc | >/=165 kcps@28kBq/cc |
| Peak True Count Rate | >/=500kcps@30kBq/cc | >/=575kcps@40kBq/cc |
| PET Scatter Fraction | =0.44</td =40%</td | |
| Count Rate Bias | = ±5%</td = ±5%</td | |
| Axial FWHM@1cm | =3.5mm</td =6.0mm</td | |
| Transaxail FWHM@1cm | =3.5mm</td =7.3mm</td | |
| Axial FWHM@10cm | =4.0mm</td =6.6mm</td | |
| Transaxial FWHM@10cm | =4.0mm</td =7.5mm</td | |
| CT Specification | | |
| Scan Regime | Continuous Rotation | Continuous Rotation |
| Scan Modes | Topo<br>Axial Scan<br>Helical Scan | Topo<br>Axial Scan<br>Helical Scan |
| Z-plane coverage | 40mm | 38.4mm |
| Number of detector row | 80 | 64 |
| Minimum slice thickness | 0.5mm | 0.6mm |
| Rotation speed | Up to 0.3 sec for 360 °rotation | Up to 0.3 sec for 360 °rotation |
| Table Maximum table load | 250kg | 227kg |
| Safety | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.