K172032 · Aesculap Implants Systems, Inc. · PLR · Nov 20, 2017 · Orthopedic
Device Facts
Record ID
K172032
Device Name
Modulift Vertebral Body Replacement (VBR) System
Applicant
Aesculap Implants Systems, Inc.
Product Code
PLR · Orthopedic
Decision Date
Nov 20, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K172032 · Nov 20, 2017
Modulift Vertebral Body Replacement (VBR) System
Aesculap Implants Systems, Inc.
Clinical literature review; Retrospective study
The sponsor utilized a clinical literature review and a retrospective study to evaluate the safety and efficacy of vertebral body replacement (VBR) devices in the cervical spine, supporting the expansion of the device's indications for use.
Cervical spine; Retrospective study; Literature review; Vertebral body replacement
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective study of VBR use in the cervical spine; Retrospective study
Patients undergoing vertebral body replacement in the cervical spine
Not applicable for this study
Risks and benefits associated with VBR use in the cervical spine
Indications for Use
The Modulift VBR System is indicated for use in the cervical spine (C3-C7 vertebral bodies for the small VBR implant) and thoracolumbar spine (T1-L5 vertebral bodies for the small/medium/large VBR implant) in skeletally mature patients for partial or total replacement of a diseased, collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders. The Modulift VBR System is intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The Modulift VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. The Modulift VBR System is intended to be used with supplemental fixation systems that have been cleared by the FDA. When used in the thoracolumbar spine, the Modulift VBR System is intended for use with supplemental spinal fixation systems such as the Aesculap MACS TL or S4 Systems. When used in the Cervical Spine at one or two levels, the Modulift Vertebral Body Replacement (VBR) System is intended to be used with supplemental fixation systems (i.e., ABC Anterior Cervical System or the Quintex Cervical System). When used at more than two levels in the cervical spine, supplemental fixation should include posterior fixation that has been cleared by FDA (i.e., Aesculap S4 Cervical System).
Device Story
Adjustable vertebral body replacement (VBR) system; implanted into vertebral body space to improve spinal stability. System comprises spacers and foot plates of varying heights/sizes/angles; axial hole allows packing with bone graft; spikes on foot plates anchor to vertebral body. Height adjusted post-implantation; column mechanically locked via grub screws. Used in OR by surgeons for spinal reconstruction. Output is structural support for spinal column; facilitates fusion or provides temporary stability in advanced tumor cases. Requires concurrent use of FDA-cleared supplemental fixation systems.
Clinical Evidence
Clinical evidence consists of a comprehensive clinical literature review and a retrospective study investigating risks and benefits of VBR use in the cervical spine. Results suggest positive benefit with minimal risk for cervical applications.
Technological Characteristics
Adjustable VBR system; components manufactured from titanium alloy (Ti6Al4V, ASTM F-136) and cobalt chrome (CoCr, ASTM F1537). Features include spacers, foot plates with lordotic/kyphotic angles, and locking grub screws. Mechanical fixation via spikes on foot plates. No software or energy source.
Indications for Use
Indicated for skeletally mature patients requiring partial or total vertebral body replacement in cervical (C3-C7) or thoracolumbar (T1-L5) spine due to tumor, osteomyelitis, trauma, or degenerative corpectomy reconstruction. Used with bone graft as adjunct to fusion or for spinal column integrity in advanced tumor patients with limited life expectancy. Requires supplemental fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
NuVasive X-CORE® Mini Cervical Expandable VBR System (K151651)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION".
November 20, 2017
Aesculap® Implant Systems, Inc. % Ms. Lisa M. Boyle Manager, Regulatory Affairs 3773 Corporate Parkway Center Valley, Pennsylvania 18034
Re: K172032
Trade/Device Name: Modulift Vertebral Body Replacement (VBR) System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: PLR, MOP Dated: October 16, 2017 Received: October 18, 2017
Dear Ms. Boyle:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820);
{1}------------------------------------------------
and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Ronald P. Jea
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172032
#### Device Name
Modulift Vertebral Body Replacement (VBR) System
#### Indications for Use (Describe)
The Modulift VBR System is indicated for use in the cervical soine (C3-C7 vertebral bodies for the small VBR implant) and thoracolumbar spine (T1-L5 vertebral bodies for the small/medium/large VBR implant) in skeletally mature patients for partial or total replacement of a diseased, collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.
The Modulift VBR System is intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The Modulift VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
The Modulift VBR System is intended to be used with supplemental fixation systems that have been cleared by the FDA. When used in the thoracolumbar spine, the Modulift VBR System is intended for use with supplemental spinal fixation systems such as the Aesculap MACS TL or S4 Systems. When used in the Cervical Spine at one or two levels, the Modulift Vertebral Body Replacement (VBR) System is intended to be used with supplemental fixation systems (i.e., ABC Anterior Cervical System or the Quintex Cervical System). When used at more than two levels in the cervical spine, supplemental fixation should include posterior fixation that has been cleared by FDA (i.e., Aesculap S4 Cervical System).
Type of Use (Select one or both, as applicable)
| <span style="text-decoration: overline;"></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="text-decoration: overline;"></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|
CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
# 4. 510(k) Summary
| Device Trade Name: | Modulift Vertebral Body Replacement (VBR) System |
|--------------------|-------------------------------------------------------------------------------------------|
| Manufacturer: | Aesculap® Implant Systems, Inc.<br>3773 Corporate Parkway<br>Center Valley, PA 18034 |
| Contact: | Lisa M. Boyle<br>Manager, Regulatory Affairs<br>800-258-1946 x 5274<br>610-791-6882 (fax) |
| Date Prepared: | November 16, 2017 |
| Classifications: | 21 CFR §888.3060, Spinal Intervertebral Body Fixation Orthosis |
| Class: | II |
| Product Codes: | PLR/MQP |
# Indications For Use:
The Modulift VBR System is indicated for use in the cervical spine (C3-C7 vertebral bodies for the small VBR implant) and thoracolumbar spine (T1-L5 vertebral bodies for the small/medium/large VBR implant) in skeletally mature patients for partial or total replacement of a diseased, collapsed, damaged, or unstable vertebral body due to tumor, osteomyelitis, trauma (i.e. fracture), or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in degenerative disorders.
The Modulift VBR System is intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The Modulift VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical, thoracic, and lumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
The Modulift VBR System is intended to be used with supplemental fixation systems that have been cleared by the FDA. When used in the thoracolumbar spine, the Modulift VBR System is intended for use with supplemental spinal fixation systems such as the Aesculap MACS TL or S4 Systems. When used in the Cervical Spine at one or two levels, the Modulift Vertebral Body Replacement (VBR) System is intended to be used with supplemental fixation systems (i.e., ABC Anterior Cervical System or the Quintex Cervical System). When used at more than two levels in the cervical spine, supplemental fixation should include posterior fixation that has been cleared by FDA (i.e., Aesculap S4 Cervical System).
{4}------------------------------------------------
# Device Description:
The Modulift VBR System is an adjustable vertebral body replacement device that is implanted into the vertebral body space to improve stability of the system is comprised of spacers and foot plates of various heights and sizes to fit the anatomical needs of a wide variety of patients. The device can be adjusted to the required height after implantation. Once it is adjusted to the desired height the column is mechanically locked in place by means of locking screws (grub screws). Each spacer has an axial hole to allow grafting material to be packed inside the device. Spikes on the end of the foot plates improve the anchoring of the implant to the vertebral body. The foot plates of the device are available in various lordotic and kyphotic angles. Components are manufactured from titanium alloy (Ti6Al4V) per ASTM F-136, and cobalt chrome (CoCr) per ASTM F1537
The purpose of the subject 510(k) is to expand the indications to include use in the cervical spine (C3-C7) of the small VBR devices.
# Primary Predicate Device:
NuVasive X-CORE® Mini Cervical Expandable VBR System (K151651).
#### Additional Predicate Device:
Aesculap Implant Systems (AIS) Modulift VBR System (K110864/K133802/K142150)
#### Comparison of Technological Characteristics:
The purpose of this 510(k) is to modify the Indications for Use for the subject Modulift Small VBR Implant. A comprehensive clinical literature review and retrospective study have been provided to investigate the risks and benefits associated with VBR use in the cervical spine. No other changes have been made to the system design since its clearance in the Modulift VBR System 510(k) K133802.
#### Performance Testing:
The Modulift Small VBR Implant and predicate NuVasive X-CORE® Mini Cervical Expandable VBR System are similar in design, material, and indicated use, and are both cleared devices. A comprehensive clinical literature review and retrospective study has been conducted to investigate the risks and benefits associated with using vertebral body replacement devices in the cervical spine. The clinical literature suggests that there is a positive benefit associated with use of VBR in the cervical spine with minimal risk.
#### Conclusion:
The Modulift VBR System has been modified to expand the indications to permit use in the cervical spine (C3-C7). Based on the indications for use, technological characteristics, mechanical testing, and comparison to the predicate device, the subject Modulift VBR System has been shown to be substantially equivalent to the legally marketed predicate device(s).
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.