NuVasive X-CORE Mini Cervical Expandable VBR System
Applicant
Nu Vasive, Incorporated
Product Code
PLR · Orthopedic
Decision Date
Sep 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K151651 · Sep 25, 2015
NuVasive X-CORE Mini Cervical Expandable VBR System
Nu Vasive, Incorporated
Published clinical literature; Retrospective clinical data analysis
The sponsor utilized a review of published clinical literature and an analysis of retrospective clinical data to demonstrate the safety and effectiveness of the device for expanded indications (tumors, trauma, and degenerative disorders of the cervical spine) by comparing it to the predicate device profile.
Retrospective analysis; Clinical literature review; Expanded indications
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective clinical data analysis and clinical literature review; Retrospective analysis and literature review
Patients treated for tumors, trauma, and degenerative disorders of the cervical spine
The NuVasive® X-Core® Mini Cervical Expandable VBR System is a vertebral body replacement device indicated for use in the cervical spine (C3-C7 vertebral bodies) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The NuVasive® X-Core® Mini Cervical Expandable VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine. These implants are intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The NuVasive® X-Core® Mini Cervical Expandable VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
Device Story
Expandable vertebral body replacement (VBR) system for cervical spine; replaces diseased/damaged vertebral bodies (C3-C7) caused by tumor, trauma, or degenerative conditions. Device implanted by surgeons during spinal reconstruction/corpectomy procedures. Provides structural support to spinal column; used with supplemental fixation and bone graft (autograft/allograft) to facilitate fusion. In advanced tumor cases, provides temporary stabilization without fusion. System design remains identical to previously cleared version (K142205); modification solely expands indications for use.
Clinical Evidence
Supported by clinical literature review and retrospective clinical data analysis. Evidence indicates the device's safety and effectiveness profile for treating tumors, trauma, and degenerative cervical disorders is similar to the predicate device.
Technological Characteristics
Manufactured from Ti-6Al-4V ELI (ASTM F136, ISO 5832-3). Expandable VBR system design. No changes to design or materials from predicate K142205.
Indications for Use
Indicated for skeletally mature patients requiring vertebral body replacement in the cervical spine (C3-C7) due to tumor, fracture, osteomyelitis, or post-corpectomy reconstruction for cervical degenerative disorders. Used with supplemental fixation and bone graft.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol featuring three stylized human profiles facing to the right, overlapping to create a sense of unity and connection. The symbol is meant to represent the department's focus on health and human well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 25, 2015
NuVasive, Incorporated Ms. Olga Lewis Lead Specialist, Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121
Re: K151651
Trade/Device Name: NuVasive® X-Core® Mini Cervical Expandable VBR System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: PLR Dated: August 26, 2015 Received: August 27, 2015
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K151651
### Device Name
NuVasive® X-Core® Mini Cervical Expandable VBR System
### Indications for Use (Describe)
The NuVasive® X-Core® Mini Cervical Expandable VBR System is a vertebral body replacement device indicated for use in the cervical spine (C3-C7 vertebral bodies) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The NuVasive® X-Core® Mini Cervical Expandable VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
These implants are intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The NuVasive® X-Core® Mini Cervical Expandable VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <span style="font-size: 10px;">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 10px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Nuvasive logo. The logo consists of a purple and gray leaf-like shape on the left, followed by the word "NUVASIVE" in gray. Below the word "NUVASIVE" is the phrase "Speed of Innovation" in a smaller font, with a horizontal line extending from the "S" in "Speed" to the end of the word "Innovation."
## 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR \$807.92, the following summary of information is provided:
#### Submitted by: A.
Olga Lewis Lead Specialist, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800
Date Prepared: August 26, 2015
#### B. Device Name
| Trade or Proprietary Name: | <i>NuVasive</i> ® <i>X-Core</i> ® <i>Mini Cervical Expandable VBR System</i> |
|----------------------------|------------------------------------------------------------------------------|
| Common or Usual Name: | Spinal Vertebral Body Replacement Device |
| Classification Name: | Spinal Intervertebral Body Fixation Orthosis |
| Device Class: | Class II |
|-----------------|-------------------|
| Classification: | 21 CFR § 888.3060 |
| Product Code: | PLR |
#### C. Predicate Devices
The subject NuVasive® X-Core® Mini Cervical Expandable VBR System is substantially equivalent to the primary predicate device NuVasive® X-Core® Expandable VBR System (K142205) and the additional predicate device, NuVasive® CoRoent® System (K081611).
#### D. Device Description
The X-Core® Mini Cervical Expandable VBR System is manufactured from Ti-6A1-4V ELI conforming to ASTM F136 and ISO 5832-3. The implants are available in a variety of different shapes and sizes to suit the individual pathology and anatomical conditions of the This 510(k) is to expand the indications for use of NuVasive® X-Core® patient. Mini Expandable VBR System (K142205) to include treatment of tumors, trauma, and degenerative disorders of the cervical spine.
{4}------------------------------------------------
#### Indications for Use E.
The NuVasive® X-Core® Mini Cervical Expandable VBR System is a vertebral body replacement device indicated for use in the cervical spine (C3-C7 vertebral bodies) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. The NuVasive® X-Core® Mini Cervical Expandable VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
These implants are intended for use with autograft or allogenic bone graft comprised of cancellous and/or corticocancellous bone graft, as an adjunct to fusion. The NuVasive X-Core® Mini Cervical Expandable VBR System is also intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
#### F. Technological Characteristics
As was established in this submission, the subject NuVasive® X-Core® Mini Cervical Expandable VBR System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent based on clinical data and have the same technological characteristics to its predicate devices through comparison in areas including design, intended use, performance, material composition, and function.
#### G. Performance Data
The purpose of this 510(k) is to modify the Indications for Use for the subject NuVasive X-Core® Mini Cervical Expandable VBR System. Clinical performance data on the subject device, as well as clinical literature review and retrospective clinical data analysis was performed to support the use of the subject device for expanded indications for use. Based on the published clinical literature review and retrospective clinical data analysis, it was determined that the NuVasive® X-Core® Mini Cervical Expandable VBR System used in the treatment of tumors, trauma, and degenerative disorders of the cervical spine has a safety and effectiveness profile that is similar to the predicate device. No other changes have been made to the system design since its clearance in the NuVasive® X-Core® Expandable VBR System 510(k) K142205. Therefore, no new nonclinical testing was performed for the purpose of this submission.
#### H. Conclusions
Based on the indications for use, technological characteristics, mechanical testing, and comparison to predicate devices, the subject NuVasive® X-Core® Mini Cervical Expandable VBR System has been shown to be substantially equivalent to legally marketed predicate devices.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.