K171984 · Ypsomed AG · FMI · Oct 11, 2017 · General Hospital
Device Facts
Record ID
K171984
Device Name
Clickfine Pen Needle, Penfine Classic Pen Needle
Applicant
Ypsomed AG
Product Code
FMI · General Hospital
Decision Date
Oct 11, 2017
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Clickfine and Penfine Classic Pen Needle are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
Device Story
Sterile, single-use pen needles for subcutaneous fluid delivery (e.g., insulin, exenatide) via pen-injectors; siliconized stainless-steel needle in polypropylene hub with protective cap; back end punctures drug reservoir port; needle inserted into injection site for fluid delivery; used in clinical or home settings by patients or healthcare providers; modification involves increased inner diameter (extra thin-walled) for 31G and 32G needles to improve flow; performance verified via ISO 9626:2016 testing for breakage and stiffness.
Clinical Evidence
Bench testing only. No clinical data. Verification testing performed per ISO 11608-2:2012, ISO 9626:2016, and ISO 7864:2016 confirmed that needles with increased inner diameters met all acceptance criteria for stiffness and resistance to breakage.
Technological Characteristics
Siliconized stainless-steel needle; polypropylene hub and protective cap; radiation sterilization; extra thin-walled tubing (ISO 9626:2016); 31G and 32G gauges; 4-12 mm lengths; mechanical assembly for pen-injector interface.
Indications for Use
Indicated for patients requiring subcutaneous injection of fluids, including insulin and exenatide, using pen injector devices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
Clickfine Pen Needle and Penfine Classic Pen Needle (K150662)
Submission Summary (Full Text)
{0}------------------------------------------------
## 510(k) Summary - File # K171984
### 8.1. Date prepared September 19, 2017
### 8.2. Submitter's information
| Company Name: | Ypsomed AG |
|------------------|-----------------------------------------------------------------|
| Company Address: | Brunnmattstrasse 6<br>CH - 3401 Burgdorf, Switzerland |
| Contact Person: | Stephan Affolter<br>Head of Quality System & Regulatory Affairs |
| Phone number: | 0041-34 424 3382 |
| Fax number: | 0041-34 424 4122 |
| Email: | stephan.affolter@ypsomed.com |
### Alternative contact
| Company Name: | P/L Biomedical |
|------------------|------------------------------------------------------|
| Company Address: | 10882 Stonington Avenue<br>Fort Myers, FL 33913, USA |
| Contact Person: | Lee Leichter |
| Phone number: | (239) 244-1448 |
| Fax number: | (815) 550-0162 |
| Email: | leichter@plbiomedical.com |
#### 8.3. Trade Name, Common Name, Classification
| Trade Name: | Clickfine Pen Needle and<br>Penfine Classic Pen Needle |
|----------------------|--------------------------------------------------------|
| Common Name: | Injection Pen Needle |
| Classification Name: | Hypodermic single lumen needle |
| Regulation Number: | 21 CFR 880.5570 |
| Regulatory Class: | Class II |
| Product Code | FMI |
### 8.4. Identification of predicate Device(s)
| The identified predicate device within this submission are as follows: | |
|------------------------------------------------------------------------|--------------------------------------------------------|
| Manufacturer: | Ypsomed AG |
| Trade Name: | Clickfine Pen Needle and<br>Penfine Classic Pen Needle |
| Common Name: | Injection Pen Needle |
| Classification Name: | Hypodermic single lumen needle |
| Regulation Number: | 21 CFR 880.5570 |
| Regulatory Class: | Class II |
| Product Code | FMI |
| FDA 510(k) #: | K150662 |
{1}------------------------------------------------
### 8.5. Comparison with Predicate Devices:
The subject device has the same materials, fundamental scientific technology and device performance as the predicate device (K150662). The only difference is that the subject devices of the same gauge (outer diameter) will now have a larger inner diameter, i.e. extra thin walled needles according to ISO 9626:2016.
The table below provides a side by side comparison of the subject device compared to its predicate.
| Characteristic | Predicate device | Subject device | Comments |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------|
| Indications for<br>use | The Clickfine and Penfine<br>Classic Pen Needle are<br>intended for use with pen<br>injector devices for<br>subcutaneous injection of<br>fluids, including insulin and<br>exenatide. | The Clickfine and Penfine<br>Classic Pen Needle are<br>intended for use with pen<br>injector devices for<br>subcutaneous injection of<br>fluids, including insulin and<br>exenatide. | Unchanged |
| Needle length | 4, 4.5, 5, 6, 8, 10, and 12 mm | 4, 4.5, 5, 6, 8, 10, and 12 mm | Unchanged |
| Needle gauge | 32, 31, 30 and 29 gauge | 32, 31, 30 and 29 gauge | Unchanged |
| Needle outer<br>diameter | 0.23 mm, 0.25, 0.3 and 0.33 mm | 0.23 mm, 0.25, 0.3 and 0.33 mm | Unchanged |
| Needle inner<br>diameter | Normal walled needle:<br>≥ 0.114 mm (31 gauge)<br>≥ 0.133 mm (30 gauge)<br>≥ 0.133 mm (29 gauge)<br><br>Thin walled needle:<br>≥ 0.105 mm (32 gauge)<br>≥ 0.125 mm (31 gauge)<br>≥ 0.165 mm (30 gauge)<br>≥ 0.190 mm (29 gauge) | Normal walled needle:<br>≥ 0.114 mm (31 gauge)<br>≥ 0.133 mm (30 gauge)<br>≥ 0.133 mm (29 gauge)<br><br>Thin walled needle:<br>≥ 0.105 mm (32 gauge)<br>≥ 0.125 mm (31 gauge)<br>≥ 0.165 mm (30 gauge)<br>≥ 0.190 mm (29 gauge) | Unchanged |
| | n/a | Extra thin walled needle:<br>≥ 0.146 mm (31 gauge),<br>≥ 0.125 mm (32 gauge) | The needles have<br>been modified to<br>have Extra thin<br>wall tubing |
### 8.6. Description of the Device
The subject devices are sterile, non-pyrogenic, single use pen needles designed to be used with commercially available pen-injectors. The back end of the needle punctures the rubber injection port of the drug reservoir in the pen. When the injection is needed, the needle is inserted into the chosen site and the pen-injector delivers the fluid through the needle.
The materials of construction, (siliconized stainless-steel needle encased in a polypropylene hub with a polypropylene protective cap) as well as the sterile packaging (a polypropylene outer protective container and the peel tab) which provides the sterile barrier and the manufacturing (which includes radiation sterilization) are unchanged from the currently marketed devices. They subject devices will be marketed in the same
{2}------------------------------------------------
sizes as the predicate devices - 32 and 31 gauges and 4, 4.5, 5, 6, 8, 10, and 12 mm lengths.
#### 8.7. Indication for Use
The pen needles are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
#### 8.8. Technological Characteristics
The predicate device Clickfine and the Penfine Classic pen needle are comprised of a siliconized stainless steel needle encased in a needle hub. An outer protective cap snaps onto the hub over the needle. This assembly fits into a larger outer protective container, which provides the sterile barrier along with the peel foil.
These characteristics remain the same for the proposed devices.
### 8.9. Substantial Equivalence
Performance testing:
All Subject Device Pen Needles have met the requirements of the relevant sections of the following standards recognized by FDA:
- a) ISO 11608-2:2012 Needle based injection systems for medical use Requirements and test methods - Part 2: Needles
- b) ISO 9626:2016 Stainless steel needle tubing for the manufacture of medical devices - Requirements and test methods
- c) ISO 7864:2016 Sterile hypodermic needles for single use Requirements and test methods
The following risks were adequately addressed through design controls:
| Device<br>Modification | Risks | Verification<br>Activity | Acceptance<br>Criteria | Results of<br>Verification |
|--------------------------------------------------------------------------------|--------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------|
| Larger inner diameter of needle on<br>31 Gauge and 32 Gauge needles | Breakage of needle | Testing for<br>resistance to<br>breakage was<br>conducted acc. ISO<br>9626:2016 | When tested acc. Annex C<br>of ISO 9626:2016, the<br>tubing shall not show<br>visible breakage when<br>examined by normal or<br>corrected vision after<br>testing. | All samples tested met the acceptance<br>criteria for resistance to breakage acc. ISO<br>9626:2016. |
| | Bending of needle | Testing for stiffness<br>was conducted acc.<br>ISO 9626:2016 | The tubing shall show a<br>maximum defined<br>deflection when tested in<br>accordance with annex B<br>of ISO 9626:2016:<br>max. deflection 0.22 mm<br>for 32 G and max. 0.21<br>mm for 31 G | All samples tested met the acceptance<br>criteria for stiffness acc. ISO 9626:2016. |
| Change in labeling to add<br>information regarding increased<br>inner diameter | None | None Required.<br>Changes do not<br>impact the safe or<br>effective use of the<br>product | N/A | N/A |
### 8.10. Conclusion
The verifications have shown evidence that the subject device pen needles meet the acceptance criteria of these standards. Based on the results it can be concluded that the subject device is substantially equivalent to the predicate device.
{3}------------------------------------------------
# Please wait...
If this message is not eventually replaced by the proper contents of the document, your PDF viewer may not be able to display this type of document.
You can upgrade to the latest version of Adobe Reader for Windows®, Mac, or Linux® by visiting http://www.adobe.com/go/reader_download.
For more assistance with Adobe Reader visit http://www.adobe.com/go/acrreader.
Windows is either a registered trademark of Microsoft Corporation in the United States and/or other countries. Mac is a trademark of Apple Inc., registered in the United States and other countries. Linux is the registered trademark of Linus Torvalds in the U.S. and other countries.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo in blue, followed by the words "U.S. FOOD & DRUG" in a larger font and "ADMINISTRATION" in a smaller font below it.
October 11, 2017
Ypsomed AG % Lee Leichter President P/L Biomedical 10882 Stonington Avenue Fort Myers, Florida 33913
Re: K171984
Trade/Device Name: Clickfine Pen Needle, Penfine Classic Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: Class II Product Code: FMI Dated: September 8, 2017 Received: September 13, 2017
Dear Lee Leichter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good
{5}------------------------------------------------
manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
for
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.