The pen needles are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
Device Story
Sterile, single-use hypodermic needles for subcutaneous fluid delivery; used with commercially available pen injectors. Operation: needle hub attaches to pen injector via screwing or snapping motion; back end of cannula punctures drug reservoir port; protective caps removed prior to injection; pen injector automatically delivers fluid; needle unscrewed and discarded after use. Used in clinical or home settings by patients or healthcare providers. Benefits: enables precise, sterile administration of prescribed medications like insulin/exenatide. Features 6-bevel needle point geometry for improved insertion.
Clinical Evidence
Bench testing only. No clinical data presented. Verification activities included testing for needle assembly attachment, tubing dimensions, flow rate, bond strength, needle point sharpness, lubrication, dose accuracy, stiffness, resistance to breakage/corrosion, and extractable metals per ISO 11608-2, ISO 9626, and ISO 7864 standards.
Technological Characteristics
Austenitic stainless steel tubing (ISO 9626); sterile, non-pyrogenic, single-use. 6-bevel needle point geometry. Hub interface allows screwing or snapping (Clickfine) or screwing-only (Penfine Classic). Dimensions: 4-12mm lengths. Sterilization: validated cycle. Connectivity: none (mechanical).
Indications for Use
Indicated for patients requiring subcutaneous injection of fluids, including insulin and exenatide, using compatible pen injector devices.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 9, 2015
Ypsomed AG c/o Mr. Lee Leichter P/L Biomedical 10882 Stonington Avenue Fort Myers, FL 33913
Re: K150662
Trade/Device Name: Clickfine Pen Needle and Penfine Classic Pen Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: May 7, 2015 Received: May 11, 2015
Dear Mr. Leichter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150662
### Device Name
Clickfine Pen Needle and Penfine Classic Pen Needle
Indications for Use (Describe)
The pen needles are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 8. 510(k) Summary
| | Submitted By/ | | | |
|----------------------|-------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | Contact Person: | Stephan Affolter<br>Ypsomed AG<br>Brunnmattstrasse 6<br>3401 Burgdorf<br>Switzerland<br>Tel. 0041-34 424 3382<br>Fax 0041-34 424 4122 | Head of Quality System & Regulatory Affairs<br>E-mail: stephan.affolter@ypsomed.com | |
| Alternative Contact: | | Lee Leichter<br>President<br>P/L Biomedical<br>10882 Stonington Avenue<br>Fort Myers, FL 33913<br>USA<br>Tel. (239) 244-1448<br>Fax. (815) 550-0162<br>E-mail: leichter@plbiomedical.com | | |
| | Date Prepared: | May 07, 2015 | | |
| 8.1 | Trade/Proprietary Name: | | Clickfine Pen Needle<br>Penfine Classic Pen Needle | |
| 8.2 | Common/Usual Name: | | Injection Pen Needle | |
| 8.3 | Classification Name: | | Hypodermic single lumen needle | |
| 8.4 | Classification Panel. | | Classification: FDA has classified Hypodermic single lumen needles in Class II<br>according to 21 CFR 880.5570. Final Order was published in the Federal Register on<br>October 21, 1980 after review by the General Hospital and Personal Use Devices | |
| | | Panel: 80 | Procode: FMI | |
#### 8.5 Predicate Device: K131125 Clickfine Pen Needles
- Purpose of Submission: To add a second product line of needles (Penfine Classic Pen 8.6 Needles) that are attached to pens by using only a "classic" screwing or twisting action in addition to the existing Clickfine Pen Needle line that can also be "clicked" or "snapped" onto the pen. The submission also includes other minor changes to both product lines, including the addition of needle point geometry with 6-bevels (in addition to the existing 3-bevel geometry), changes to the needle hub, labeling and sterilization cycle.
{4}------------------------------------------------
### 8.7 Device Description
All of the subject device pen needles are sterile, non-pyrogenic, single use needles designed to be used with commercially available pen-injectors for the administration of prescribed fluids. Each needle is individually packaged in a sealed protective container with a peel tab. The subject device pen needles are used by peeling back the peel tab. The hub of the currently marketed Clickfine Pen Needle is snapping or screwing onto the threaded end of the pen injector. The modified Penfine Classic Pen Needle is attached to pen injectors with only a "screwing" motion. For all needles the back end of the cannula punctures the rubber injection port of the drug reservoir in the pen-injector. The outer protective cap is then removed. The inner protective cap will remain over the needle until the fluid is ready to be injected. When the injection is needed, the inner protective cap is removed and the needle is inserted into the chosen site. The pen-injector automatically delivers the fluid through the needle. The protective cap is replaced and the needle is then unscrewed for the pen injector, discarded and replaced with a new needle before the next injection.
- 8.8 Intended Use
The intended use of the modified device remains the same as the predicate device -Clickfine Pen Needles (K131125):
The pen needles are intended for use with pen injector devices for subcutaneous injection of fluids, including insulin and exenatide.
- 8.9 Technological Characteristics and Substantial Equivalence Discussion The technological characteristics have not changed. All of the materials, processes and operating principles of the modified needles are identical to the currently marketed needles. The Penfine Classic Pen Needle has a modified interface to only allow attachment to compatible cartridge holders through a twisting or screwing motion, while the current device can be snapped, or "clicked" onto the cartridge holder. The only other change is a different grinding of the needle point to produce six, rather than three surfaces. Otherwise the needles have the same design and intended use as the currently marketed product and are compatible with the same injection pens. As such, the subject needles are substantially equivalent to the predicate Clickfine Pen Needles (K131125).
### 8.10 Performance Data
As part of the Design Control Activities, Ypsomed evaluated the potential risks, hazards or hazardous situations, potential clinical impact on the patient, identified the appropriate risk control measure and acceptability criteria associated with each change in the modified devices. Then, each risk was mitigated through appropriate verification and/or validation activities which were determined to be the testing specified in the following international and internal standards and protocols. This testing confirmed that the subject devices are substantially equivalent to the predicate devices.
All Subject Device Pen Needles have met the requirements of the relevant sections of the following standards:
- . ISO 11608-2:2012 Needle based injection systems for medical use -
{5}------------------------------------------------
Requirements and test methods – Part 2: Needles
- . ISO 9626:1991/Amd.1:2001 Stainless steel needle tubing for the manufacture of medical devices
- ISO 7864:1993 Sterile hypodermic needles for single use ●
| Test | Section of Standard |
|----------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Needle assembly<br>attachment shall fit and<br>function with needle based<br>injection systems specified<br>in ISO 11608-1 | Clause 4.2.1 of ISO 11608-2<br>The subject device pen needles were tested on<br>compatibility with needle based injection system as<br>specified in ISO 11608-1 |
| Tubing dimensions | Clause 4.2.2, Table 1 of ISO 11608-2<br>Length of patient end: +/- 1.25 mm<br>Length of cartridge end: 5.7 – 7.0 mm |
| Flow rate | Clause 4.3 of ISO 11608-2:<br>Measurement of flow rate in accordance with Annex<br>A. |
| Bond between hub and<br>needle tube | Clause 9 of ISO 11608-2 and clause 13.1 of ISO<br>7864<br>The union of the hub and needle tube shall not break<br>for at least 5 sec. while a force of at least 22 N is<br>applied |
| Needle points | Clause 4.5 of ISO 11608-2<br>Visually sharp at a 2.5x magnification; designed to<br>minimize coring and fragmentation |
| Lubrication | Clause 4.7 of ISO 11608-2<br>No visible droplets on the outside surface of the<br>needle tube. |
| Dislocation of measuring<br>point at the patient-end | Clause 8 of ISO 11608-2:<br>4, 6, 8, 10mm +/- 0.9mm, 12mm +/- 1.1mm |
| Functional compatibility<br>with needle based<br>injection systems | Clause 11 of ISO 11608-2:<br>Needle assembly torque: 0.07 +/- 0.01 Nm<br>Needle hub removal: less than 0.100 Nm<br>Dose accuracy:<br>for doses ≤ 20 ml the calculated values were within ±<br>0.01 ml of the targeted dose;<br>for doses > 20 ml the calculated values were within<br>± 5% of the targeted dose. |
| Tubing characteristics | Table 1 of ISO 9626/A1:<br>The tubing is made of austenitic stainless steel acc.<br>one of the designated types given in Table 1 of ISO<br>9626/A1. |
| Tubing diameters | Table 2 of ISO 9626/A1 |
| Test | Section of Standard |
| | Tubing dimensions meet outer and inner diameter<br>requirement as defined in table 2 of ISO 9626/A1. |
| Stiffness | Clause 9 of ISO 9626:<br>When tested in accordance with annex C of ISO 9626<br>the tubing shall show a deflection not greater than the<br>relevant value given in table 3 of ISO 9626/A1. |
| Resistance to breakage | Clause 10 of ISO 9626:<br>When tested in accordance with annex D of ISO<br>9626 and Table 4 if ISO 9626/A1 the tubing shall not<br>break. |
| Limits for acidity and<br>alkalinity | Clause 6 of ISO 9626<br>Tested in accordance with annex A and extract<br>preparation of the tubing in accordance with Annex<br>B: correction for the volume of titrant required for<br>the control fluid shall not be more than 0.04 ml<br>NaOH or 0.12 ml HCl to reach the end-point<br>titration. |
| Resistance to corrosion | Clause 11 of ISO 9626:<br>When tested in accordance with annex E, the<br>immersed half of the tubing shall show no evidence<br>of corrosion resulting from the test. |
| Patency of lumen | Clause 13.2 of ISO 7864<br>A stylet, having a diameter equivalent to 80% ±2%<br>of lumen inner diameter must pass through freely |
| Freedom of defects | Clause 11.3 in ISO 7864:<br>When examined by normal or corrected-to-normal<br>vision, the needle shall appear straight and of regular<br>cross-section and wall thickness |
| Limits for extractable<br>metals | Clause 6 of ISO 7864:<br>When tested with a recognized method the content of<br>lead, tin, zinc and iron shall be less than 5mg/l. The<br>cadmium content shall be lower than 0.1 mg/l. |
| Freedom of defects | Clause 11.3 of ISO 7864:<br>When examined by normal or corrected vision, the<br>needle lube shall appear straight and of regular cross-<br>section and wall thickness |
{6}------------------------------------------------
The verifications have shown evidence that the subject device pen needles meet the acceptance criteria of these standards. Based on the results it can be concluded that the device performance is acceptable for the product.
### 8.11 Conclusion
Y psomed AG concludes based on the information presented that the modified products are substantially equivalent to the current product (Clickfine Pen Needles) legally marketed in the USA.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.