Sofia Strep A+ FIA, Sofia 2 Analyzer

K171976 · Quidel Corporation · GTY · Dec 21, 2017 · Microbiology

Device Facts

Record IDK171976
Device NameSofia Strep A+ FIA, Sofia 2 Analyzer
ApplicantQuidel Corporation
Product CodeGTY · Microbiology
Decision DateDec 21, 2017
DecisionSESE
Submission TypeDual Track
Regulation21 CFR 866.3740
Device ClassClass 1

Indications for Use

The Sofia Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by either bacterial culture or an FDA-cleared molecular assay because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection. The Sofia Strep A+ FIA may be used with Sofia or Sofia 2.

Device Story

Sofia Strep A+ FIA is an immunofluorescent assay for detecting Group A Streptococcus (GAS) antigens. Patient throat swab samples are extracted in a reagent tube and applied to a test cassette. The cassette contains a test strip where GAS antigens bind to fluorescent microparticle-labeled antibodies. The device uses an ultraviolet (UV) light source to illuminate the fluorescent tag and a microprocessor-controlled analyzer (Sofia 2) to scan the strip and measure the fluorescent signal. The device operates in two modes: 'Walk Away' (automated timing) or 'Read Now' (manual timing followed by scanning). Results are displayed on a touchscreen as Positive or Invalid. Used in clinical/laboratory settings by professionals to aid in diagnosis. The output provides objective, automated interpretation of the assay, reducing subjective visual reading and assisting clinical decision-making for pharyngitis management.

Clinical Evidence

Bench testing only. Studies included Limit of Detection (LoD) verification, precision testing across multiple operators/lots/days, and method comparison using a panel of clinical samples. Reproducibility studies assessed intra- and inter-operator and laboratory performance. Results demonstrated that Sofia 2 and the predicate Sofia analyzer provide equivalent qualitative performance.

Technological Characteristics

Microprocessor-controlled analyzer; UV light source for fluorescent detection; 4-inch color LCD touchscreen; integrated barcode scanner; power via 100-240 VAC or rechargeable lithium polymer battery. Dimensions: 19.7 cm x 11.4 cm x 12.7 cm; weight ~2.5 lbs. Uses method-specific algorithms to process fluorescent signals from test cassettes.

Indications for Use

Indicated for patients with signs and symptoms of pharyngitis (e.g., sore throat) to detect Group A Streptococcal antigens. Intended for professional and laboratory use.

Regulatory Classification

Identification

Streptococcus spp. serological reagents are devices that consist of antigens and antisera (excluding streptococcal exoenzyme reagents made from enzymes secreted by streptococci) used in serological tests to identify Streptococcus spp. from cultured isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Streptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.

Predicate Devices

Submission Summary (Full Text)

{1} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K171976 B. Purpose for Submission: This submission is a Dual 510(k) and CLIA Waiver by Application (Dual Submission) tracked as K171976 and CW170009. CW170009 was submitted for CLIA Waiver of the Sofia Strep A+ FIA with the Sofia 2 analyzer. To demonstrate equivalent performance of the Sofia Strep A+ FIA on the Sofia and Sofia 2 analyzers using elements of an assay migration study approach. C. Measurand: Group A β-hemolytic Streptococcus (GAS; Streptococcus pyogenes) antigens in throat swab specimens. D. Type of Test: The Sofia Strep A+ FIA is an immunofluorescence-based lateral flow *in vitro* diagnostic test for the qualitative detection of GAS antigens isolated from throat swab specimens obtained from symptomatic patients. E. Applicant: Quidel Corporation F. Proprietary and Established Names: Sofia Strep A+ FIA and Sofia 2 G. Regulatory Information: 1. Regulation Section: 21 CFR 866.3740 - Streptococcus spp. Serological Reagents 2. Classification: Class I 3. Product Code(s): {2} GTY – Antigens, Streptococcus spp., All Groups KHO – Fluorometer, for Clinical Use 4. Panel: Microbiology (83) H. Intended Use: 1. Intended Use(s): The Sofia Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by either bacterial culture or an FDA-cleared molecular assay because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection. The Sofia Strep A+ FIA may be used with Sofia or Sofia 2. 2. Indication(s) for Use: Same as intended use. 3. Special Conditions for Use Statement(s): For prescription use only 4. Special Instrument Requirements: To be used only with Sofia or Sofia 2 analyzers I. Device Description: 1. Overview The Sofia Strep A+ FIA is a rapid lateral flow immunoassay for the qualitative detection of GAS antigens from throat swabs collected from patients with signs and symptoms of pharyngitis. The Sofia Strep A+ FIA components have not been modified; a detailed description of the lateral flow device is available in submission K141775, which describes Sofia Strep A+ FIA use with the original Sofia analyzer. The purpose of this submission is to obtain 510(k) clearance for the Sofia Strep A+ FIA for use with the newly developed Sofia 2 analyzer. The Sofia and Sofia 2 analyzers are similar in function and design. Both systems utilize the same fail-safes and failure alert 2 {3} mechanisms, the same calibration and assay-specific cartridges, and the same ultraviolet light-emitting diodes (UV LEDs) to excite the fluorophore. The primary difference between the original Sofia and Sofia 2 analyzer is the design of the optical detection system. Sofia uses a motorized optics unit to collect fluorescent signal data as it performs a series of scans across the longitudinal axis of the test strip, whereas Sofia 2 captures a still image of the entire test strip window using a complimentary color-oxide semiconductor (CMOS) camera. To emulate Sofia, the Sofia 2 analyzer converts pixels captured by the CMOS camera to fluorescent signal data, which is analyzed in an equivalent manner to Sofia to yield qualitative test results. 2. Quality Control See Decision Summary for K141775. 3. Results Interpretation There are three possible test results for the Sofia Strep A+ FIA: 1) Positive for Strep A, 2) Negative for Strep A, and 3) Invalid. If an invalid test result is reported, the Sofia Strep A+ FIA should be repeated with a new patient sample and a new test cassette. Note: The Sofia and Sofia 2 analyzers may be set to one of two operating modes: "Walk Away" or "Read Now." Time to results for the Sofia and Sofia 2 analyzers are described below. - In Walk Away Mode, the user inserts the test cassette into the analyzer immediately following addition of the specimen to the Sofia Strep A+ FIA sample port. The analyzer automatically times the test development and provides positive or negative test results after 5 minutes. - In the Read Now Mode, the user incubates the test cassette on the benchtop for 5 minutes before inserting the cassette into the analyzer. Positive and negative test results are displayed within 1 minute. No changes were made to test components in the Sofia Strep A+ kits or calibration cassette. J. Substantial Equivalence Information: 1. Predicate Device Name(s): Sofia Strep A+ FIA on Sofia analyzer 2. Predicate 510(k) Number: K141775 {4} 3. Comparison with Predicate: | Similarities and Differences | | | | --- | --- | --- | | | Predicate Device | Proposed Device | | Item | Sofia Strep A+ FIA on Sofia (K141775) | Sofia Strep A+ FIA on Sofia 2 (K171976) | | Regulation | 866.3740 | 866.3740 | | Product code | GTY | GTY | | Device class | I | I | | Technology principle of operation | Lateral flow immunoassay | Same | | Assay targets | GAS antigen | Same | | Assay results | Qualitative | Same | | Intended Use | The Sofia Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by bacterial culture because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection. | The Sofia Strep A+ FIA detects Group A Streptococcal antigens from throat swabs from patients with signs and symptoms of pharyngitis, such as sore throat. All negative test results should be confirmed by either bacterial culture or an FDA-cleared molecular assay because negative results do not preclude Group A Strep infection and should not be used as the sole basis for treatment. The test is intended for professional and laboratory use as an aid in the diagnosis of Group A Streptococcal infection. The Sofia Strep A+ FIA may be used with Sofia or Sofia 2. | | Specimen types | Throat swab | Same | | Internal assay controls | procedural control zone and reference line | Same | | External controls | Test kit contains positive and negative control swabs | Same | | Instrument | Sofia | Sofia 2 | | Dimensions | 24 cm x 16 cm x 10 cm | 19.7 cm x 11.4 cm x 12.7cm | | Weight | 3 lbs | ~2.5 lbs | | Power supply | 100-240 VAC, self-switching, or with 4 AA batteries | 100-240 VAC, self-switching, or with rechargeable lithium polymer battery | | Printer | Integrated | External | | Assay/instrument interface | Drawer | Same | {5} | Similarities and Differences | | | | --- | --- | --- | | | Predicate Device | Proposed Device | | **Item** | Sofia Strep A+ FIA on Sofia (K141775) | Sofia Strep A+ FIA on Sofia 2 (K171976) | | User interface | 3.5 inch diagonal color LCD display and numeric keypad with function specific buttons touchscreen display | 4 inch color LCD touchscreen display | | Sample ID | External hand-held barcode scanner | Integrated barcode scanner | | Cassette ID | Integrated barcode scanner | Integrated barcode scanner with custom 0.3MP camera | | Development Modes | Walk Away only | Walk Away and Read Now | | Time to obtain results | 5 minutes | Same | **K. Standard/Guidance Document Referenced (if applicable):** Not Applicable **L. Test Principle:** See decision summary for K141775. The FDA-cleared Sofia Strep A+ FIA has not been modified, and it is compatible with the newly developed Sofia 2 analyzer. **M. Performance Characteristics** Limit of Detection, Reproducibility, Assay Precision, Read Now vs Walk Away Mode, Analyzer Calibration, Transport Media/Sample Stability, and Method Comparison Studies were conducted to demonstrate equivalent performance of the Sofia Strep A+ FIA on the Sofia and Sofia 2 analyzers, where relevant. Results from each of these studies are described in sections $M(1a-1d)$ and $M(2)$ below. Additional analytical and clinical performance data for the Sofia Strep A+ FIA are available under K141775. **1. Analytical Performance** **a. Limit of Detection** The limit of detection (LoD) of the Sofia Strep A+ FIA was established on both the Sofia and Sofia 2 analyzers. The study included three GAS strains spiked into negative clinical throat swab matrix. Briefly, dose-response curves were generated using three GAS strains from which the C95 concentration for each strain was interpolated. The preliminary LoD concentration was determined for each strain with two lots of cassettes on each instrument platform and defined as the lowest dilution yielding $100\%$ positive results, where the next lower dilution returned one or more {6} negative results. The LoD was verified by testing each GAS strain to confirm that the expected proportion of positive results was obtained. An additional 60 replicates for each GAS strain were tested at the preliminary LoD on one cassette lot on each instrument platform. Testing spanned three days with at least 20 replicates run per day for each strain/instrument. The LoDs of the two instrument platforms were compared to demonstrate similar performance of the Sofia Strep A+ FIA with both analyzers. Testing was performed in Read Now Mode in accordance with the Package Insert of Sofia Strep A+ FIA and user manuals of the two analyzers. A Calibration Check of each Sofia and Sofia 2 analyzer was performed prior to use on the first day of testing and the results were recorded. In addition, the External Positive and Negative Controls were tested with each cassette lot for every day of testing and on each analyzer used in the study. Results of control testing were recorded. The LoD and Signal/Cut-off (S/CO) results are shown in Table 1 below. Table 1: Sofia Strep A+ FIA LoD Study Results on the Sofia and Sofia 2 Analyzers | Strain | Sofia | | Sofia 2 | | | --- | --- | --- | --- | --- | | | LOD (CFU/test) | S/CO | LOD (CFU/test) | S/CO | | Bruno | 3.94E+04 | 1.71 | 4.07E+04 | 1.704 | | CDC-SS-1402 | 1.31E+05 | 1.93 | 1.60E+05 | 1.851 | | CDC-SS-1460 | 8.03E+04 | 1.65 | 7.45E+04 | 1.934 | The established LoDs for the Sofia Strep A+ FIA on the Sofia and Sofia 2 analyzers with the 3 strains of GAS ranged from $3.94\mathrm{E} + 04$ to $1.60\mathrm{E} + 05$ . The study confirmed that the LoD generated for the Sofia Strep A+ FIA on Sofia 2 is similar to the LoD generated on Sofia. # b. Reproducibility Reproducibility of the Sofia Strep A+ FIA on the Sofia and Sofia 2 analyzers was evaluated at three different laboratories. A total of 270 samples were tested across 3 sites on both analyzers. Two operators at each site tested nine coded contrived samples (prepared in negative clinical matrix) from the reproducibility panel over 5 days per site. Test panels included Negative (no bacteria), Low Positive (1X LoD GAS), and Moderate Positive (2-3X LOD GAS) samples. The study demonstrated intra- and inter-laboratory reproducibility with a panel of GAS samples of various concentrations (See Table 2 and Table 3 below). When testing the low Strep A positive samples (C95), Sofia and Sofia 2 reported the samples positive $98.9\%$ (89/90) of the time. When testing the moderately positive samples, Sofia and Sofia 2 called the Strep A+ sample positive $100\%$ (90/90) of the time. Out of the 270 samples tested, there were no invalid test results obtained with either the Sofia or Sofia 2 {7} analyzers. Studies were conducted in either Walk Away (225 data points) or Read Now Mode (45 data points) on both the Sofia and Sofia 2 analyzers. Table 2. Qualitative Reproducibility Study Results Using the Sofia Analyzer | Site | Operator | Negative | Strep A Low Positive (1X LoD) | Strep A Moderate Positive (2-3X LoD) | | --- | --- | --- | --- | --- | | 1 | 1 | 15/15 | 15/15 | 15/15 | | | 2 | 15/15 | 14/15 | 15/15 | | 2 | 1 | 15/15 | 15/15 | 15/15 | | | 2 | 15/15 | 15/15 | 15/15 | | 3 | 1 | 15/15 | 15/15 | 15/15 | | | 2 | 15/15 | 15/15 | 15/15 | | Combined Total | | 90/90 | 89/90 | 90/90 | | Percent Agreement (95% CI) | | 100% (95.9-100%) | 98.9% (94.0-99.9%) | 100% (95.9-100%) | Table 3. Qualitative Reproducibility Study Results Using the Sofia 2 Analyzer | Site | Operator | Negative | Strep A Low Positive (1X LoD) | Strep A Moderate Positive (2-3X LoD) | | --- | --- | --- | --- | --- | | 1 | 1 | 15/15 | 15/15 | 15/15 | | | 2 | 15/15 | 14/15 | 15/15 | | 2 | 1 | 15/15 | 15/15 | 15/15 | | | 2 | 15/15 | 15/15 | 15/15 | | 3 | 1 | 15/15 | 15/15 | 15/15 | | | 2 | 15/15 | 15/15 | 15/15 | | Combined Total | | 90/90 | 89/90 | 90/90 | | Percent Agreement (95% CI) | | 100% (95.9-100%) | 98.9% (94.0-99.9%) | 100% (95.9-100%) | The quantitative results reported in terms of Signal/Cut-off (S/CO) for the Reproducibility Study are shown below in Table 4 for the Sofia analyzer and Table 5 for the Sofia 2 analyzer. The mean, standard deviation (SD) and the percent coefficient of variation (%CV) are provided for each of the panel members by site and overall average. {8} Table 4. S/CO-based Reproducibility Results Using the Sofia Analyzer | Site | Sofia Analyzer | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Negative | | | Strep A Low Positive | | | Strep A Moderate Positive | | | | | Mean | SD | %CV | Mean | SD | %CV | Mean | SD | %CV | | 1 | 0.238 | 0.077 | 32.3 | 2.528 | 0.749 | 29.6 | 8.424 | 3.479 | 41.3 | | 2 | 0.235 | 0.079 | 33.4 | 3.317 | 1.001 | 30.2 | 11.186 | 3.657 | 32.7 | | 3 | 0.229 | 0.101 | 44.0 | 3.047 | 0.699 | 22.9 | 10.424 | 2.397 | 23.0 | | Avg | 0.234 | 0.085 | 36.4 | 2.964 | 0.882 | 29.7 | 10.011 | 3.396 | 33.9 | Table 5. S/CO-based Reproducibility Results Using the Sofia 2 Analyzer | Site | Sofia 2 Analyzer | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Negative | | | Strep A Low Positive | | | Strep A Moderate Positive | | | | | Mean | SD | %CV | Mean | SD | %CV | Mean | SD | %CV | | 1 | 0.229 | 0.092 | 40.2 | 2.639 | 1.058 | 40.1 | 9.214 | 3.717 | 40.3 | | 2 | 0.218 | 0.114 | 52.4 | 3.635 | 1.388 | 38.2 | 12.688 | 4.319 | 34.0 | | 3 | 0.226 | 0.131 | 58.0 | 3.577 | 0.877 | 24.5 | 12.100 | 3.535 | 29.2 | | Avg | 0.224 | 0.112 | 50.1 | 3.284 | 1.205 | 36.7 | 11.334 | 4.121 | 36.4 | Any quantitative differences between sites did not impact the final qualitative results reported. c. Assay Precision To supplement the Reproducibility Study, within-laboratory precision of the Sofia and Sofia 2 analyzers was evaluated. The precision of Sofia and Sofia 2 analyzers with the Sofia Strep A+ FIA was examined using three lots of devices and two operators testing the same samples for twelve (12) days over two calibration cycles for a total of 144 data points for each analyzer and each sample type. One strain of GAS was tested at Low Positive (approximately 1X LoD) and Moderate Positive (approximately 2-3X LoD) levels. A Negative sample was also included in the test panel. One Sofia and two Sofia 2 analyzers (one for back-up) were used in this study. When an error for saving the test result happened on the Sofia 2 analyzer, the cassette was re-scanned immediately on the backup Sofia 2 analyzer. The testing was conducted in Read Now Mode. Overall percent agreement between the expected results and the precision study results for the Sofia and Sofia 2 analyzers is shown in Table 6 below. The qualitative call rates for both Sofia and Sofia 2 analyzers were the same for the Negative, Low Positive and Moderate Positive samples: $100\%$ negative for the Negative samples and $100\%$ positive for the Low and Moderate Positive samples. There were three (3) invalid tests reported in the study. {9} Table 6. Qualitative Precision Study Results Summary | Sample Type | Sofia | | Sofia 2 | | | --- | --- | --- | --- | --- | | | Overall Agreement | % Overall Agreement | Overall Agreement | % Overall Agreement | | Negative | 144/144 | 100 | 144/144 | 100 | | Low Positive (1X LoD) | 144/144 | 100 | 144/144 | 100 | | Moderate Positive (2-3X LoD) | 144/144 | 100 | 144/144 | 100 | For each of the sample types listed above, the mean S/CO, standard deviation (SD), and percent coefficient of variation (%CV) were calculated. Results for samples tested using the Sofia and Sofia 2 analyzers are shown in Table 7. Table 7. S/CO-based Precision Study Results Summary | Platform | Sample Type | Mean S/CO | SD | % CV | | --- | --- | --- | --- | --- | | Sofia | Negative | 0.25 | 0.07 | 26.6% | | | Low Positive (1X LoD) | 2.88 | 0.81 | 28.0% | | | Moderate Positive (2-3X LoD) | 13.95 | 4.38 | 31.4% | | Sofia 2 | Negative | 0.24 | 0.07 | 30.2% | | | Low Positive (1X LoD) | 2.88 | 0.81 | 27.9% | | | Moderate Positive (2-3X LoD) | 14.21 | 4.96 | 34.9% | Overall, the study results showed that Sofia and Sofia 2 analyzers generated equivalent qualitative overall agreement results when used by multiple operators and operated over multiple days. d. Read Now vs Walk-Away Mode/Temperature Study The purpose of this study was to document testing results of the Sofia Strep A+ FIA in Read Now and Walk Away Modes on the Sofia 2 analyzer. In the Read Now Mode, the cassette with loaded sample was developed for 5 minutes on the bench top before insertion into the Sofia 2 analyzer for scanning. In Walk Away Mode, the cassette was placed in the Sofia 2 analyzer immediately and developed for 5 minutes in the Sofia 2 analyzer. Additionally, testing was performed at different ambient temperatures. The entire test kit (including the Sofia 2 analyzer) was equilibrated to, and tested at, each of the following environmental conditions: 15°C, room temperature, and 30°C. The temperature range evaluated was the same as the range currently approved for the Sofia Strep A+ FIA and the existing Sofia analyzer. For sample testing, one GAS strain was tested at approximately 2-3X LoD. Contrived Negative Matrix was also tested as the negative sample. All positive and negative samples were in 100% agreement with the expected results regardless of the temperature and reading mode. This study demonstrated that the correct results can be obtained with the Sofia Strep A+ FIA when the test materials are stored and tested at temperatures from 15° to 30°C in both Read Now and Walk Away Modes on Sofia 2 analyzer. {10} e. Analyzer Calibration The purpose of this study was to assess the Sofia 2 intra-instrument variability and performance during the maximum 30-day calibration cycle. To verify the intra-instrument variability, the study was executed using five Sofia 2 analyzers. On day 1, all analyzers were calibrated. The variability in RFU for each of the four (4) lines was analyzed by each instrument to determine the percentage change in RFU over three reads per working day for 45 calendar days. The study data showed that the Sofia 2 analyzer demonstrated intra-instrument variability of less than 6% (+/-3%) over the maximum calibration interval of 30 days, and therefore, met the acceptance specification over this interval. f. Transport Media and Sample Stability The purpose of this study was to evaluate the stability of GAS spiked on - Puritan Rayon Swabs stored in a transport tube with a secured cap - BBL Culture Swabs stored in the Collection and Transport system containing Stuart's media The study included testing the contrived swab specimens spiked on Puritan Rayon swabs stored in dry transported tubes with secured caps as well as spiked on BBL Culture Swabs stored in the Collection and Transport system for up to 50 hours at two different temperatures. The seeded swabs were stored and kept at 4°C for 0, 6, 24, 48 and 50 hours and room temperature for 0, 6, 24, 48 and 50 hours. Six (6) replicates for each condition were tested. The cassettes were read in Read Now Mode on the Sofia and Sofia 2 analyzers. When the negative matrix was tested with BBL CultureSwab and Dry Puritan Rayon swabs after storage at room temperature and at 2-8°C on Sofia and Sofia 2 analyzers, all valid negative swabs were negative and yielded 100% negativity up to 50 hours. There was one (1) invalid result. When the positive samples were tested with BBL CultureSwab and Dry Puritan Rayon swabs after storage at room temperature and at 2-8°C on Sofia and Sofia 2 analyzers, all swabs were positive and yielded 100% positivity up to 50 hours. There were three (3) invalid tests. If not processed soon after collection, Quidel has stated that swabs can be held in any clean, dry plastic tube or sleeve up to 24 hours at room temperature (23°C) or refrigerated (2°C to 8°C) up to 48 hours. For the BBL CultureSwab with Liquid Stuart Media, the recommended storage is 2-8°C or ambient temperature up to 48 hours. 10 {11} # 2. Comparison Studies # a. Method Comparison A multi-center study was conducted to evaluate the performance of the Sofia Strep $\mathrm{A + }$ FIA run on the Sofia 2 analyzer vs the Sofia analyzer using a panel of known positive and negative clinical and contrived samples prepared in clinical negative matrix. One-hundred (100) Strep A positive and (100) Strep A negative samples were incorporated into each panel (tested positive or negative with the current FDA-cleared Sofia Strep $\mathrm{A + }$ FIA on Sofia). According to the test protocol, positive panel members were prepared so that approximately $60\%$ to $80\%$ of the samples had Strep A levels close to the Sofia Strep $\mathrm{A + }$ FIA cut-off with approximately one-half around the LoD and the remainder as moderate positive samples. The remaining positive samples were evenly distributed across the range of the assay. Negative panel members were prepared so that approximately $30\%$ to $40\%$ were High Negatives. Each panel member was assigned to one of the analyte concentration levels based on the signal-to-cut off ratio (S/CO) that was obtained from testing on the Sofia analyzer. Table 8 below shows the S/CO for each level. Table 8. S/CO Target Values | Panel Member | GAS S/CO | | --- | --- | | Negative | <0.5 | | High Negative | 0.55-0.99 | | Low Positive (1X LoD) | 1.1-2.4 | | Moderate Positive (2-3X LoD) | 3.0-6.0 | | High Positive (4-5X LoD) | 6.1-9 | | Very High Positive (>5X LoD) | >10.3 | The final composition of the Method Comparison Panel was reported as shown in Table 9 below. Table 9. Method Comparison Panel | Panel Member | Quantity per panel | | --- | --- | | Negative | 70 | | High Negative | 30 | | Low Positive (1X LoD) | 30 | | Moderate Positive (2-3X LoD) | 36 | | High Positive (4-5X LoD) | 19 | | Very High Positive (>5X LoD) | 15 | All samples were coded and used to prepare randomized panels, with an identical panel of specimens prepared for each of three sites. Each site received a set of two hundred (200) numbered tubes each containing 2 swabs, as well as Positive and Negative Control swabs. For all analyte levels combined, the Positive Percent Agreement (PPA) and Negative Percent Agreement (NPA) values between the Sofia {12} and the Sofia 2 results were $98.7\%$ and $96.9\%$ , respectively (See Table 10 below). Agreement between the results obtained on the Sofia and Sofia 2 analyzers for each of the individual analyte levels tested is shown in Table 11 below. There were 12 discordant results for GAS across all three sites. Ten discordant samples were categorized as "High Negative" samples, and two discordant samples were characterized as "Negative" samples. For the two discordant samples originally prepared as "Negative" samples, results were reported as positive by the Sofia analyzer but negative by the Sofia 2 analyzer. The two false positives by the Sofia analyzer appeared to be due to contamination of the samples. Discordant results were observed across all three sites and were generated by multiple operators; there was no trend indicating any instrument bias. Table 10. Method Comparison Study of GAS Results for the Sofia and Sofia 2 Analyzers | | Sofia Positive | Sofia Negative | Total | | --- | --- | --- | --- | | Sofia 2 Positive | 369 | 7b | 376 | | Sofia 2 Negative | 5a | 219 | 224 | | Total | 374 | 226 | 600 | | | PPA: 98.7% (369/374), 95% CI: (96.9%-99.4%) NPA: 96.9% (219/226), 95% CI: (93.7%-98.5%) | | | ${}^{a}$ There were 5 discordant Sofia 2 negative/Sofia positive results for Strep A,which included 3 high negative (C5) and 2 true negative (C0) specimens. The 2 true negatives (C0) showing false positive results in Sofia appeared to be due to contamination. There were 7 discordant Sofia 2 positive/Sofia negative results for Strep A, which included all high negative (C5) specimens. {13} Table 11: Agreement Stratified by GAS Level | Analyte Level | Qualitative Method Comparison Results | | | | --- | --- | --- | --- | | | | Sofia Positive | Sofia Negative | | Negative | Sofia 2 Positive | 0 | 0 | | | Sofia 2 Negative | 2* | 208 | | High Negative | Sofia 2 Positive | 69 | 7 | | | Sofia 2 Negative | 3 | 11 | | Low Positive (1X LoD) | Sofia 2 Positive | 90 | 0 | | | Sofia 2 Negative | 0 | 0 | | Moderate Positive (2-3X LoD) | Sofia 2 Positive | 108 | 0 | | | Sofia 2 Negative | 0 | 0 | | High Positive (4-5X LoD) | Sofia 2 Positive | 57 | 0 | | | Sofia 2 Negative | 0 | 0 | | Strong Positive (>5X LoD) | Sofia 2 Positive | 45 | 0 | | | Sofia 2 Negative | 0 | 0 | *Documented as operator error (potential contamination) The Method Comparison Study results demonstrated that the performance of the Sofia Strep $\mathrm{A + }$ FIA on the Sofia 2 analyzer was comparable to the Sofia analyzer (predicate device). # N. Instrumentation/System Description 1. Instrument Name: Sofia and Sofia 2 analyzers 2. System Description: The Sofia instruments are equipped with UV LED lighting and optical detection systems that are designed to collect and analyze fluorescent data emitted from the Sofia Strep A+ FIA test cassettes. The instrument software contains embedded assay-specific algorithms that are used to convert the fluorescent signal data into a qualitative test result. 3. Software: FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types. Yes $\underline{\mathbf{X}}$ or No {14} 4. **Level of Concern** Moderate 5. **Software Description** The Sofia 2 analyzer images the test strip and utilizes an embedded software algorithm to process the resulting data set into a qualitative test result. A positive test result for the analyte is determined by detection and analysis of the fluorescent signal at the test and reference lines, which are processed by the assay-specific algorithm. The algorithm employs a smoothing filter to the data and, identifies a peak maxima, minima and width, then calculates the RFU value based on peak height for the analyte test line. Results are presented on a screen and can be printed on an integrated printer. 6. **Specimen Identification** Each cassette is labeled with a barcode that is detected by the Sofia 2 internal barcode reader. The 2-dimensional barcode contains lot, expiration, and test method information. If the cassette is not in the correct orientation, the barcode cannot be read and the analyzer will display an error. 7. **Calibration** To ensure that signal drift is controlled, the Sofia 2 analyzer has a calibration algorithm that reminds the operator to check the calibration status of the instrument every 30 days. The calibration cassette uses a highly stable fluorescent reagent along with a unique barcode. The insertion of this cassette triggers the Sofia 2 analyzer to self-check its calibration. O. Other supportive Instrument Characteristics Data Not Covered in the “Performance Characteristics” Section Above N/A P. **Proposed Labeling** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. Q. **Conclusion** The submitted information in the premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...